A performance comparison of convolutional neural network-based image denoising methods: The effect of loss functions on low-dose CT images.

Byeongjoon Kim1, Minah Han1, Hyunjung Shim1, Jongduk Baek1.   

Abstract

PURPOSE: Convolutional neural network (CNN)-based image denoising techniques have shown promising results in low-dose CT denoising. However, CNN often introduces blurring in denoised images when trained with a widely used pixel-level loss function. Perceptual loss and adversarial loss have been proposed recently to further improve the image denoising performance. In this paper, we investigate the effect of different loss functions on image denoising performance using task-based image quality assessment methods for various signals and dose levels.
METHODS: We used a modified version of U-net that was effective at reducing the correlated noise in CT images. The loss functions used for comparison were two pixel-level losses (i.e., the mean-squared error and the mean absolute error), Visual Geometry Group network-based perceptual loss (VGG loss), adversarial loss used to train the Wasserstein generative adversarial network with gradient penalty (WGAN-GP), and their weighted summation. Each image denoising method was applied to reconstructed images and sinogram images independently and validated using the extended cardiac-torso (XCAT) simulation and Mayo Clinic datasets. In the XCAT simulation, we generated fan-beam CT datasets with four different dose levels (25%, 50%, 75%, and 100% of a normal-dose level) using 10 XCAT phantoms and inserted signals in a test set. The signals had two different shapes (spherical and spiculated), sizes (4 and 12 mm), and contrast levels (60 and 160 HU). To evaluate signal detectability, we used a detection task SNR (tSNR) calculated from a non-prewhitening model observer with an eye filter. We also measured the noise power spectrum (NPS) and modulation transfer function (MTF) to compare the noise and signal transfer properties.
RESULTS: Compared to CNNs without VGG loss, VGG-loss-based CNNs achieved a more similar tSNR to that of the normal-dose CT for all signals at different dose levels except for a small signal at the 25% dose level. For a low-contrast signal at 25% or 50% dose, adding other losses to the VGG loss showed more improved performance than only using VGG loss. The NPS shapes from VGG-loss-based CNN closely matched that of normal-dose CT images while CNN without VGG loss overly reduced the mid-high-frequency noise power at all dose levels. MTF also showed VGG-loss-based CNN with better-preserved high resolution for all dose and contrast levels. It is also observed that additional WGAN-GP loss helps improve the noise and signal transfer properties of VGG-loss-based CNN.
CONCLUSIONS: The evaluation results using tSNR, NPS, and MTF indicate that VGG-loss-based CNNs are more effective than those without VGG loss for natural denoising of low-dose images and WGAN-GP loss improves the denoising performance of VGG-loss-based CNNs, which corresponds with the qualitative evaluation.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  adversarial loss; deep learning; feature-level loss; image denoising; low-dose CT; mathematical observer; modulation transfer function; noise power spectrum

Year:  2019        PMID: 31306488     DOI: 10.1002/mp.13713

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Magician's Corner: 7. Using Convolutional Neural Networks to Reduce Noise in Medical Images.

Authors:  Nathan Robert Huber; Andrew D Missert; Bradley J Erickson
Journal:  Radiol Artif Intell       Date:  2020-09-30

2.  The synthesis of high-energy CT images from low-energy CT images using an improved cycle generative adversarial network.

Authors:  Haojie Zhou; Xinfeng Liu; Haiyan Wang; Qihang Chen; Rongpin Wang; Zhi-Feng Pang; Yong Zhang; Zhanli Hu
Journal:  Quant Imaging Med Surg       Date:  2022-01

3.  A minimum SNR criterion for computed tomography object detection in the projection domain.

Authors:  Scott S Hsieh; Shuai Leng; Lifeng Yu; Nathan R Huber; Cynthia H McCollough
Journal:  Med Phys       Date:  2022-07-10       Impact factor: 4.506

4.  ADAPTIVE-NET: deep computed tomography reconstruction network with analytical domain transformation knowledge.

Authors:  Yongshuai Ge; Ting Su; Jiongtao Zhu; Xiaolei Deng; Qiyang Zhang; Jianwei Chen; Zhanli Hu; Hairong Zheng; Dong Liang
Journal:  Quant Imaging Med Surg       Date:  2020-02

5.  Effect Evaluation of Platelet-Rich Plasma Combined with Vacuum Sealing Drainage on Serum Inflammatory Factors in Patients with Pressure Ulcer by Intelligent Algorithm-Based CT Image.

Authors:  Jingzhe Yang; Changshuan Xiao; Hailing Wen; Kui Sun; Xiaoming Wu; Xinshu Feng
Journal:  Comput Math Methods Med       Date:  2022-03-02       Impact factor: 2.809

6.  Wavelet subband-specific learning for low-dose computed tomography denoising.

Authors:  Wonjin Kim; Jaayeon Lee; Mihyun Kang; Jin Sung Kim; Jang-Hwan Choi
Journal:  PLoS One       Date:  2022-09-09       Impact factor: 3.752

7.  Low-dose CT image and projection dataset.

Authors:  Taylor R Moen; Baiyu Chen; David R Holmes; Xinhui Duan; Zhicong Yu; Lifeng Yu; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2020-12-16       Impact factor: 4.071

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.