Literature DB >> 31305068

Experimentally Validating Open Tubular Liquid Chromatography for a Peak Capacity of 2000 in 3 h.

Piliang Xiang1, Yu Yang1, Zhitao Zhao1, Apeng Chen1, Shaorong Liu1,2.   

Abstract

The advancements in life science research mandate effective tools capable of analyzing large numbers of samples with low quantities and high complexities. As an essential analytical tool for this research, liquid chromatography (LC) encounters an ever-increasing demand for enhanced resolving power, accelerated analysis speed, and reduced limit of detection. Although theoretical studies have indicated that open tubular (OT) columns can produce superior resolving power under comparable elution pressures and analysis times, ultrahigh-resolution and ultrahigh-speed open tubular liquid chromatography (OTLC) separations have never been reported. Here we present experimental results to demonstrate the predicted potential of this technique. We use a 2 μm i.d. × 75 cm long OT column coated with trimethoxy(octadecyl)silane for separating pepsin/trypsin digested E. coli lysates and routinely produce exceptionally high peak capacities (e.g., 1900-2000 in 3-5 h). We reduce the column length to 2.7 cm and exhibit the capability of OTLC for ultrafast separations. Under an elution pressure of 227.5 bar, we complete the separation of six amino acids in ∼800 ms and resolve these compounds within ∼400 ms. In addition, we show that OTLC has low attomole limits of detection (LOD) and each separation requires samples of only a few picoliters. Importantly, no ultrahigh elution pressures are required. With the ultrahigh resolution, ultrahigh speed, low LOD, and low sample volume requirement, OTLC can potentially be a powerful tool for biotech research, especially single cell analysis.

Entities:  

Year:  2019        PMID: 31305068     DOI: 10.1021/acs.analchem.9b01465

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  Picoflow Liquid Chromatography-Mass Spectrometry for Ultrasensitive Bottom-Up Proteomics Using 2-μm-i.d. Open Tubular Columns.

Authors:  Piliang Xiang; Ying Zhu; Yu Yang; Zhitao Zhao; Sarah M Williams; Ronald J Moore; Ryan T Kelly; Richard D Smith; Shaorong Liu
Journal:  Anal Chem       Date:  2020-03-27       Impact factor: 6.986

2.  Estimation of low-level components lost through chromatographic separations with finite detection limits.

Authors:  Nicole M Devitt; Joe M Davis; Mark R Schure
Journal:  J Chromatogr A       Date:  2020-05-31       Impact factor: 4.759

3.  A comparison of 2 micron inner diameter open tubular column liquid chromatography with pressure-driven isocratic, slip-flow, and electrochromatographic modes of operation: a theoretical study.

Authors:  Mark R Schure; Matthew D Beauchamp
Journal:  J Chromatogr A       Date:  2020-12-24       Impact factor: 4.759