| Literature DB >> 31295087 |
Ekaterina Royba1,2, Mikhail Repin1, Sergey Pampou2, Charles Karan2, David J Brenner1, Guy Garty1.
Abstract
We developed a fully-automated dicentric chromosome assay (DCA) in multiwell plates. All operations, from sample loading to chromosome scoring, are performed, without human intervention, by the second-generation Rapid Automated Biodosimetry Tool II (RABiT-II) robotic system, a plate imager and custom software, FluorQuantDic. The system requires small volumes of blood (30 µl per individual) to determine radiation dose received as a result of a radiation accident or terrorist attack. To visualize dicentrics in multiwell plates, we implemented a non-classical protocol for centromere FISH staining at 37°C. The RABiT-II performs rapid analysis of chromosomes after extracting them from metaphase cells. With the use of multiwell plates, many samples can be screened at the same time. Thus, the RABiT-II DCA provides an advantage during triage when risk-based stratification and medical management are required for a large population exposed to unknown levels of ionizing radiation.Entities:
Year: 2019 PMID: 31295087 DOI: 10.1667/RR15266.1
Source DB: PubMed Journal: Radiat Res ISSN: 0033-7587 Impact factor: 2.841