Sarah Parackal1, Donghui Zou1, Robert Day1, Michael Black1, Parry Guilford1.
Abstract
OBJECTIVES: To compare the Roche Cell-Free DNA Collection Tubes® against the Streck Cell-Free DNA BCT®s for sample stability using Cell Free DNA (cfDNA) from healthy volunteers (n = 20). DESIGN &Entities:
Keywords: Blood-based biomarker; Cell free DNA; Genomic DNA contamination; Plasma DNA; Roche cell-free DNA collection tubes
Year: 2019 PMID: 31289732 PMCID: PMC6593214 DOI: 10.1016/j.plabm.2019.e00125
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Fig. 1Concentration of eluted DNA of samples drawn from 20 healthy volunteers and stored in either Streck or Roche cfDNA blood collection tubes. Samples were drawn as per the manufacturer's recommendations and stored at 22 °C until processing at five time points (Day 0, 4, 7, 10 and 14). Concentrations are reported as ng/mL of plasma.
Fig. 2Concentration of plasma DNA in Roche and Streck blood collection tubes over 14 days post collection. Whole blood was drawn from healthy volunteers (n = 20) into both Roche and Streck tubes, stored and processed at five time points (Day 0, 4, 7, 10 and 14).
Wilcoxon signed rank test results for paired sample analysis within and between streck cell-free DNA BCT®s and roche cell free DNA collection Tubes®.
| Streck Sample Pair | V | Adjusted p-value |
|---|---|---|
| Day 0 and 4 | 29 | 0.36 |
| Day 0 and 7 | 34 | 0.096 |
| Day 0 and 10 | 65 | 1.12 |
| Day 0 and 14 | 70 | 1.6 |
| Roche Sample Pair | ||
| Day 0 and 4 | 27 | 0.163 |
| Day 0 and 7 | 19 | 0.01 |
| Day 0 and 10 | 33 | 0.046 |
| Day 0 and 14 | 14 | 0.0016 |
| Streck and Roche Pair | ||
| Day 0 and 4 | 62 | 7.44 |
| Day 0 and 7 | 91 | 6.64 |
| Day 0 and 10 | 149 | 0.88 |
| Day 0 and 14 | 160 | 0.32 |
Fig. 3Electropherograms from healthy volunteers' samples stored in Streck, Roche and K3EDTA blood collection tubes. DNA peaks are quantified between two markers, lower (35 bp) and upper (10, 380 bp). cfDNA peak for each volunteer was identified between 150 and 200 bp. Genomic DNA peaks clustered around the upper marker. Graphs are shown for three volunteers A. Streck Day 0 sample, volunteer had a significant cfDNA, two smaller distinguishable peaks at 375bp and 500bp and further peaks near the upper marker characterized as gDNA. B. Streck Day 14 sample, cfDNA size peak had decreased in fluorescence, however the two previously observed peaks were still apparent. gDNA near the upper marker was at the same fluorescence intensity as the day 0 sample. C. Roche Day 0 sample, a significant cfDNA peak was apparent. No gDNA was observed near the upper marker. D. Roche Day 14 sample, fluorescence of cfDNA has diminished, increase in “jaggedness” of baseline however no gDNA were detected. E. K3EDTA Day 0 sample, an obvious cfDNA peak was present between 150bp and 200bp. F. K3EDTA Day 14 sample, significant gDNA contamination with very high fluorescence intensity partially obscuring the cfDNA in the sample.