How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.

André Euler1, Natalia Saltybaeva2, Hatem Alkadhi2.   

Abstract

OBJECTIVES: To assess the impact of patient off-centering on organ dose and image noise for head and thoracoabdominal CT in a pediatric phantom.
METHODS: An anthropomorphic phantom simulating a 5-year-old child was used. Semiconductor dosimeters were placed in various cranial and thoracoabdominal organs. Head and thoracoabdominal CT were performed using automatic tube current modulation (ATCM) and default bowtie filters. The phantom was imaged repeatedly at vertical table positions ranging from - 6 to + 6 cm from the 0-position. Tube current time products (TCTP), organ doses, and image noise were recorded. Scatter radiation was measured in the thyroid for head CT. The effect of ATCM and bowtie filters was assessed.
RESULTS: Depending on patient position, organ doses differed up to 22% for the supratentorial brain, 34% for the infratentorial brain, 19% for the eyes, 28% for the lungs, 25% for the stomach, and 22% for the liver compared with those in the 0-position. The relation between position and dose was linear and mainly affected by the bowtie filter in head CT, while it was quadratic and affected by ATCM and bowtie filter in thoracoabdominal CT. It further depended on the relative position of each organ to the isocenter. An inverse relation was found between position and image noise. Scatter radiation was not significantly related to patient positioning (p = 0.21).
CONCLUSIONS: In pediatric CT, vertical patient positioning had a substantial impact on radiation dose with differences of up to 34%, depending on the body region and location of each individual organ. KEY POINTS: • Patient off-centering has a substantial impact on organ radiation dose and image noise in pediatric CT. • Impact of patient off-centering on radiation dose and noise differs between head and thoracoabdominal CT. • Differences are caused by both ATCM and bowtie filter in thoracoabdominal CT, but mainly by bowtie filter in head CT.

Entities:  

Keywords:  Child; Multidetector computed tomography; Phantoms, imaging; Radiation dosimeters; Radiation exposure

Mesh:

Year:  2019        PMID: 31278575     DOI: 10.1007/s00330-019-06330-5

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  7 in total

1.  Effect of patient centering on patient dose and image noise in chest CT.

Authors:  Touko Kaasalainen; Kirsi Palmu; Vappu Reijonen; Mika Kortesniemi
Journal:  AJR Am J Roentgenol       Date:  2014-07       Impact factor: 3.959

2.  Investigating the CT localizer radiograph: acquisition parameters, patient centring and their combined influence on radiation dose.

Authors:  J W Lambert; S Kumar; J S Chen; Z J Wang; R G Gould; B M Yeh
Journal:  Br J Radiol       Date:  2015-01-22       Impact factor: 3.039

3.  Prevalence and Severity of Off-Centering During Diagnostic CT: Observations From 57,621 CT scans of the Chest, Abdomen, and/or Pelvis.

Authors:  Oladunni O Akin-Akintayo; Lauren F Alexander; Rebecca Neill; Elizabeth A Krupinksi; Xiangyang Tang; Pardeep K Mittal; William C Small; Courtney C Moreno
Journal:  Curr Probl Diagn Radiol       Date:  2018-02-23

4.  Vertical off-centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantoms.

Authors:  Natalia Saltybaeva; Hatem Alkadhi
Journal:  Med Phys       Date:  2017-09-25       Impact factor: 4.071

5.  Automatic patient centering for MDCT: effect on radiation dose.

Authors:  Jianhai Li; Unni K Udayasankar; Thomas L Toth; John Seamans; William C Small; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2007-02       Impact factor: 3.959

6.  The effects of patient positioning when interpreting CT dose metrics: A phantom study.

Authors:  Rebecca M Marsh; Michael S Silosky
Journal:  Med Phys       Date:  2017-03-17       Impact factor: 4.071

7.  Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies.

Authors:  M A Habibzadeh; M R Ay; A R Kamali Asl; H Ghadiri; H Zaidi
Journal:  Phys Med       Date:  2011-07-08       Impact factor: 2.685

  7 in total
  1 in total

1.  Ultra-high resolution computed tomography of joints: practical recommendations for acquisition protocol optimization.

Authors:  Pedro Augusto Gondim Teixeira; Nicolas Villani; Malik Ait Idir; Edouard Germain; Charles Lombard; Romain Gillet; Alain Blum
Journal:  Quant Imaging Med Surg       Date:  2021-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.