Prevalence and Severity of Off-Centering During Diagnostic CT: Observations From 57,621 CT scans of the Chest, Abdomen, and/or Pelvis.

Oladunni O Akin-Akintayo1, Lauren F Alexander1, Rebecca Neill1, Elizabeth A Krupinksi1, Xiangyang Tang1, Pardeep K Mittal1, William C Small1, Courtney C Moreno2.   

Abstract

PURPOSE: To determine distances between patient centroid and gantry isocenter during CT imaging of the chest, abdomen, and/or pelvis, and to evaluate differences based on patient gender, scan region, patient position, and gantry aperture.
MATERIALS AND METHODS: A water phantom and an anthropomorphic phantom were imaged in the centered position in the CT gantry and at several off-centered positions. Additionally, data from 57,621 adult chest, abdomen, and/or pelvic CT acquisitions were evaluated. Data were analyzed with an analysis of variance using the centroid-to-isocenter data as the dependent variable and the other parameters as independent variables.
RESULTS: The majority of patient acquisitions (83.7% (48271/57621)) were performed with the patient's centroid positioned below isocenter (mean 1.7 cm below isocenter (SD 1.8 cm); range 12.1 cm below to 7.8 cm above isocenter). Off-centering in the x-axis was less severe (mean 0.01 cm left of isocenter (SD 1.6 cm)). Distance between centroid and isocenter in the y-axis did not differ as a function of sex but did differ based on scan region, patient position, and gantry aperture.
CONCLUSION: Off-centering is common during CT imaging and has been previously demonstrated to impact dose and image quality.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29576415     DOI: 10.1067/j.cpradiol.2018.02.007

Source DB:  PubMed          Journal:  Curr Probl Diagn Radiol        ISSN: 0363-0188


  3 in total

1.  How patient off-centering impacts organ dose and image noise in pediatric head and thoracoabdominal CT.

Authors:  André Euler; Natalia Saltybaeva; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2019-07-05       Impact factor: 5.315

2.  The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.

Authors:  Touko Kaasalainen; Teemu Mäkelä; Mika Kortesniemi
Journal:  Eur J Radiol Open       Date:  2018-12-17

3.  Effect of technologist and patient attributes on centering for body CT examinations: Influence of cultural and ethnic factors.

Authors:  Antar Aly; Shadi Ebrahimian; Mohammed H Kharita; Mahmoud Heidous; Mohammad Zaya Ashruf; Davendra Kumar; Mannudeep K Kalra; Huda Mohd Al Naemi
Journal:  PLoS One       Date:  2022-08-19       Impact factor: 3.752

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.