Improved Real-World Glycemic Control With Continuous Glucose Monitoring System Predictive Alerts.

Sinu Bessy Abraham1, Siddharth Arunachalam1, Alex Zhong1, Pratik Agrawal1, Ohad Cohen2, Chantal M McMahon1.   

Abstract

BACKGROUND: Most standalone real-time continuous glucose monitoring (RT-CGM) systems provide predictive low and high sensor glucose (SG) threshold alerts. The durations and risk of low and high SG excursions following Guardian™ Connect CGM system predictive threshold alerts were evaluated.
METHODS: Continuous glucose monitoring system data uploaded between January 2, 2017 and May 22, 2018 by 3133 individuals using multiple daily injections (MDIs) or continuous subcutaneous insulin infusion (CSII) therapy were deidentified and retrospectively analyzed. Glucose excursions were defined as SG values that went beyond a preset low or high SG threshold for ≥15 minutes. For a control group, thresholds were based on the median of the low SG threshold limit (70 mg/dL) and the high SG threshold limit (210 mg/dL) preset by all system users. During periods when alerts were not enabled, timestamps were identified when a predictive alert would have been triggered. The time before low horizon was 17.5 minutes and the time before high horizon was 15 minutes, of all users who enabled alerts. Excursions occurring after a low SG or high SG predictive alert were segmented into prevented, ≤20, 20-60, and >60 minutes.
RESULTS: Excursions were prevented after 59% and 39% of low and high SG predictive alerts, respectively. The risk of a low or high excursion occurring was 1.9 (P < 0.001, 95% CI, 1.88-1.93) and 3.3 (P < 0.001, 95% CI, 3.20-3.30) times greater, respectively, when alerts were not enabled.
CONCLUSIONS: The predictive alerts of the RT-CGM system under study can help individuals living with diabetes prevent some real-world low and high SG excursions. This can be especially important for those unable to reach or maintain glycemic control with basic RT-CGM or CSII therapy.

Entities:  

Keywords:  continuous glucose monitoring; hyperglycemia; hypoglycemia; predictive threshold alerts; type 1 diabetes; type 2 diabetes

Year:  2019        PMID: 31272204      PMCID: PMC7783013          DOI: 10.1177/1932296819859334

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  48 in total

1.  Responses to continuous glucose monitoring in subjects with type 1 diabetes using continuous subcutaneous insulin infusion or multiple daily injections.

Authors:  David Rodbard; Lois Jovanovic; Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2009-12       Impact factor: 6.118

2.  Predictors of fear of hypoglycemia in adolescents with type 1 diabetes and their parents.

Authors:  Linda A Gonder-Frederick; Craig D Fisher; Lee M Ritterband; Daniel J Cox; Likun Hou; Aditi A DasGupta; William L Clarke
Journal:  Pediatr Diabetes       Date:  2006-08       Impact factor: 4.866

Review 3.  Impact of CGM on the Management of Hypoglycemia Problems: Overview and Secondary Analysis of the HypoDE Study.

Authors:  Norbert Hermanns; Lutz Heinemann; Guido Freckmann; Delia Waldenmaier; Dominic Ehrmann
Journal:  J Diabetes Sci Technol       Date:  2019-03-06

4.  Real-time continuous glucose monitoring in type 1 diabetes: a qualitative framework analysis of patient narratives.

Authors:  John C Pickup; Melissa Ford Holloway; Kritika Samsi
Journal:  Diabetes Care       Date:  2014-12-31       Impact factor: 19.112

5.  The Impact of Continuous Glucose Monitoring on Markers of Quality of Life in Adults With Type 1 Diabetes: Further Findings From the DIAMOND Randomized Clinical Trial.

Authors:  William H Polonsky; Danielle Hessler; Katrina J Ruedy; Roy W Beck
Journal:  Diabetes Care       Date:  2017-04-07       Impact factor: 19.112

6.  Continuous Glucose Monitoring vs Conventional Therapy for Glycemic Control in Adults With Type 1 Diabetes Treated With Multiple Daily Insulin Injections: The GOLD Randomized Clinical Trial.

Authors:  Marcus Lind; William Polonsky; Irl B Hirsch; Tim Heise; Jan Bolinder; Sofia Dahlqvist; Erik Schwarz; Arndís Finna Ólafsdóttir; Anders Frid; Hans Wedel; Elsa Ahlén; Thomas Nyström; Jarl Hellman
Journal:  JAMA       Date:  2017-01-24       Impact factor: 56.272

7.  Comparison of continuous blood glucose measurement with conventional documentation of hypoglycemia in patients with Type 2 diabetes on multiple daily insulin injection therapy.

Authors:  Reinhard Zick; Bettina Petersen; Mario Richter; Cornelia Haug
Journal:  Diabetes Technol Ther       Date:  2007-12       Impact factor: 6.118

8.  Frequency of severe hypoglycemia in patients with type I diabetes with impaired awareness of hypoglycemia.

Authors:  A E Gold; K M MacLeod; B M Frier
Journal:  Diabetes Care       Date:  1994-07       Impact factor: 19.112

9.  The burden of severe hypoglycemia in type 1 diabetes.

Authors:  Jieruo Liu; Rosa Wang; Michael L Ganz; Yurek Paprocki; Doron Schneider; James Weatherall
Journal:  Curr Med Res Opin       Date:  2017-11-10       Impact factor: 2.580

10.  Effect of continuous glucose monitoring on hypoglycemia in type 1 diabetes.

Authors:  Tadej Battelino; Moshe Phillip; Natasa Bratina; Revital Nimri; Per Oskarsson; Jan Bolinder
Journal:  Diabetes Care       Date:  2011-02-19       Impact factor: 19.112

View more
  9 in total

1.  Flash glucose monitoring and glycemic control in type 1 diabetes with subcutaneous insulin infusion.

Authors:  Sara Lomelino Pinheiro; Margarida Bastos; Luísa Barros; Miguel Melo; Isabel Paiva
Journal:  Acta Diabetol       Date:  2021-11-16       Impact factor: 4.280

Review 2.  Artificial intelligence perspective in the future of endocrine diseases.

Authors:  Mandana Hasanzad; Bagher Larijani; Hamid Reza Aghaei Meybodi; Negar Sarhangi
Journal:  J Diabetes Metab Disord       Date:  2022-01-11

3.  The state of artificial intelligence-based FDA-approved medical devices and algorithms: an online database.

Authors:  Stan Benjamens; Pranavsingh Dhunnoo; Bertalan Meskó
Journal:  NPJ Digit Med       Date:  2020-09-11

4.  Mitigation of Rebound Hyperglycemia With Real-Time Continuous Glucose Monitoring Data and Predictive Alerts.

Authors:  Giada Acciaroli; John B Welsh; Halis Kaan Akturk
Journal:  J Diabetes Sci Technol       Date:  2021-01-05

Review 5.  Advanced Diabetes Management Using Artificial Intelligence and Continuous Glucose Monitoring Sensors.

Authors:  Martina Vettoretti; Giacomo Cappon; Andrea Facchinetti; Giovanni Sparacino
Journal:  Sensors (Basel)       Date:  2020-07-10       Impact factor: 3.576

6.  Practical aspects of diabetes technology use: Continuous glucose monitors, insulin pumps, and automated insulin delivery systems.

Authors:  Brynn E Marks; Kristen M Williams; Jordan S Sherwood; Melissa S Putman
Journal:  J Clin Transl Endocrinol       Date:  2021-12-04

Review 7.  Artificial Intelligence in Perioperative Medicine: A Proposed Common Language With Applications to FDA-Approved Devices.

Authors:  Ryan L Melvin; Matthew G Broyles; Elizabeth W Duggan; Sonia John; Andrew D Smith; Dan E Berkowitz
Journal:  Front Digit Health       Date:  2022-04-25

Review 8.  Lessons from Use of Continuous Glucose Monitoring Systems in Digital Healthcare.

Authors:  Hun-Sung Kim; Kun-Ho Yoon
Journal:  Endocrinol Metab (Seoul)       Date:  2020-09-22

Review 9.  Hypoglycaemia detection and prediction techniques: A systematic review on the latest developments.

Authors:  Omar Diouri; Monika Cigler; Martina Vettoretti; Julia K Mader; Pratik Choudhary; Eric Renard
Journal:  Diabetes Metab Res Rev       Date:  2021-03-24       Impact factor: 4.876

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.