Literature DB >> 31260115

Evaluation of tissue deformation during radiofrequency and microwave ablation procedures: Influence of output energy delivery.

Dong Liu1, Christopher L Brace1.   

Abstract

PURPOSE: The purpose of this study was to quantitatively analyze tissue deformation during radiofrequency (RF) and microwave ablation for varying output energy levels.
METHODS: A total of 46 fiducial markers which were classified into outer, middle, and inner lines were positioned into a single plane around an RF or microwave ablation applicator in each ex vivo bovine liver sample (8 cm × 6 cm × 4 cm, n = 18). Radiofrequency (500 kHz; ~35 W average) or microwave (2.4 GHz; 50-100 W output, ~35-70 W delivered) ablation was performed for 10 min (n = 4-6 each setting). CT images were acquired over the entire liver volume every 15 s. Principle strain magnitude and direction were determined from fiducial marker displacement. Normal and shear strain were then calculated such that negative strain denoted contraction and positive strain denoted expansion. Temporal variations, the final magnitudes, and angles of the strain were compared across energy delivery settings, using one-way ANOVA with post hoc Tukey's tests.
RESULTS: On average, tissue strain rates peak at around 1 min and decayed exponentially over time. No evidence of tissue expansion was observed. The tissue strains from RF and 50 W, 75 W, and 100 W microwave ablation at 10 min were -8.5%, -38.9%, -54.4%, and -65.7%, respectively, from the inner region and -3.6%, -23.7%, -41.8%, and -44.3%, respectively, from the outer region. Negative strain magnitude was positively correlated to energy delivery in the inner region (Spearman's ρ  = -0.99). Microwaves at higher powers (75-100 W) induced significantly more strain than at lower power (50 W) or after RF ablation (P < 0.01). Principal strain angles ranged from 0.8° to -8.1°, indicating that tissue deformed more in the direction transverse to the applicator than along the direction of the applicator.
CONCLUSIONS: The influence of output energy on tissue deformation during RF and microwave ablation was analyzed. Microwave ablation created significantly greater contraction than RF ablation with similar energy delivery. During microwave ablation, more contraction was noted at higher power levels and in proximity to the antenna. Contraction primarily transverse to the antenna produces ablation zones that are more elongated than the original tissue volume.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  microwave ablation; radiofrequency ablation; thermal ablation; tissue contraction; tissue deformation

Mesh:

Year:  2019        PMID: 31260115      PMCID: PMC6739156          DOI: 10.1002/mp.13688

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Gel Phantom Models for Radiofrequency and Microwave Ablation of the Liver.

Authors:  Willa J Chen; Qi Wang; Charles Y Kim
Journal:  Dig Dis Interv       Date:  2020-09-28

2.  Safety and efficacy of microwave versus radiofrequency ablation for large hepatic hemangioma: a multicenter retrospective study with propensity score matching.

Authors:  Jian Kong; Ruize Gao; Shilun Wu; Yaoping Shi; Tao Yin; Shigang Guo; Zonghai Xin; Aolei Li; Xinliang Kong; Demin Ma; Bo Zhai; Wenbing Sun; Jun Gao
Journal:  Eur Radiol       Date:  2022-01-29       Impact factor: 5.315

3.  Small single perivascular hepatocellular carcinoma: comparisons of radiofrequency ablation and microwave ablation by using propensity score analysis.

Authors:  Chao An; Wang-Zhong Li; Zhi-Mei Huang; Xiao-Ling Yu; Yu-Zhi Han; Fang-Yi Liu; Song-Song Wu; Jie Yu; Ping Liang; Jinhua Huang
Journal:  Eur Radiol       Date:  2021-01-05       Impact factor: 5.315

4.  Microwave ablation of lung tumors: A probabilistic approach for simulation-based treatment planning.

Authors:  Jan Sebek; Pinyo Taeprasartsit; Henky Wibowo; Warren L Beard; Radoslav Bortel; Punit Prakash
Journal:  Med Phys       Date:  2021-05-27       Impact factor: 4.506

5.  Direction of Tissue Contraction after Microwave Ablation: A Comparative Experimental Study in Ex Vivo Bovine Liver.

Authors:  Junhyok Lee; Hyunchul Rhim; Min Woo Lee; Tae Wook Kang; Kyoung Doo Song; Jeong Kyong Lee
Journal:  Korean J Radiol       Date:  2022-01       Impact factor: 3.500

6.  Fat Quantification Imaging and Biophysical Modeling for Patient-Specific Forecasting of Microwave Ablation Therapy.

Authors:  Frankangel Servin; Jarrod A Collins; Jon S Heiselman; Katherine C Frederick-Dyer; Virginia B Planz; Sunil K Geevarghese; Daniel B Brown; Michael I Miga
Journal:  Front Physiol       Date:  2022-02-03       Impact factor: 4.566

7.  2450-MHz microwave ablation of liver metastases under 3.0 T wide-bore magnetic resonance guidance: a pilot study.

Authors:  Kaihao Xu; Zhaonan Li; Yiming Liu; Zaoqu Liu; Chaoyan Wang; Dechao Jiao; Xinwei Han
Journal:  Sci Rep       Date:  2022-07-25       Impact factor: 4.996