| Literature DB >> 31254457 |
Takanori Hara1, Shinji Niwa1, Atsushi Urikura2, Kosuke Matsubara3, Takashi Hoshino4, Eiji Nishimaru5, Takuya Taniguchi6.
Abstract
BACKGROUND: Because the x-ray property of patient longitudinal axis in area detector computed tomography (ADCT) depends on a heel effect, radiation dose and beam quality are not uniform along the long axis of the patient.Entities:
Keywords: 160-mm x-ray beam; area detector computed tomography (ADCT); half-value layer (HVL); heel effect; radiation dose; volume-acquisition mode
Year: 2019 PMID: 31254457 PMCID: PMC6698757 DOI: 10.1002/acm2.12670
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Photograph of the experimental setup for dose profile measurements. The small sensor embedded in the computed tomography dose profiler probe was moved at a constant speed of 83 mm/s by the RTI Mover during continuous use of the 160‐mm non‐helical volume‐acquisition mode.
Figure 2Photograph of the experimental setup for half‐value layer measurements. The slant angles of the detection area plane in the RTI Piranha dosimeter at the respective locations were adjusted by using digital clinometer.
Figure 3(a) Photograph of the phantom used for longitudinal contrast uniformity assessment. (b) Cross‐section image of the phantom. Iodinated solution of cylinder phantom was adjusted to (A) 7.3 and (B) 2.4 mgI/ml.
Figure 4Dose profile results for (a) 256‐ and (b) 320‐row area detector computed tomography (ADCT) scanner. The heel effect in the respective ADCT scanners longitudinally differed along the beam.
Accuracy of half‐value layer with two different measurement methods.
| Tube voltage (kVp) | Half‐value layer (mm Al) | Percent‐difference (%) | |
|---|---|---|---|
| RTI Piranha | Conventional | ||
| 80 | 4.11 ± 0.00 | 4.17 ± 0.01 | −1.4 |
| 100 | 5.18 ± 0.00 | 5.34 ± 0.01 | −3.0 |
| 120 | 6.11 ± 0.00 | 6.33 ± 0.00 | −3.5 |
Data express mean ± standard deviation.
Effective energy difference within 160‐mm x‐ray beam.
| Tube voltage (kVp) | Position (mm) | Effective energy (keV) | |
|---|---|---|---|
| 256‐row ADCT | 320‐row ADCT | ||
| 80 | −75 | 37.4 | 41.6 |
| −60 | 37.5 | 40.4 | |
| −30 | 37.8 | 38.9 | |
| 0 | 38.3 | 38.1 | |
| 30 | 38.9 | 37.7 | |
| 60 | 40.0 | 37.4 | |
| 75 | 40.8 | 37.4 | |
| 120 | −75 | 50.1 | 53.1 |
| −60 | 50.2 | 51.8 | |
| −30 | 50.5 | 50.2 | |
| 0 | 51.0 | 49.3 | |
| 30 | 51.6 | 48.8 | |
| 60 | 52.7 | 48.5 | |
| 75 | 53.5 | 48.5 | |
Figure 5Influence of longitudinal beam locations on the computed tomography (CT) number of diluted contrast material. Mean CT numbers at the isocenter position (0 mm) were approximately as follows: for 256‐row area detector computed tomography (ADCT): 120 HU at 2.4 mgI/ml and 310 HU at 7.3 mgI/ml; for 320‐row ADCT: 130 HU at 2.4 mgI/ml and 360 HU at 7.3 mgI/ml.