Image quality and semi-quantitative measurements of the Siemens Biograph Vision PET/CT: Initial experiences and comparison with Siemens Biograph mCT PET/CT.

Joyce van Sluis1, Ronald Boellaard1, Ananthi Somasundaram1, Paul van Snick1, Ronald Borra1, Rudi Dierckx1, Gilles Stormezand1, Andor Glaudemans1, Walter Noordzij1.   

Abstract

In May 2018 the new Biograph Vision PET/CT (Siemens Healthineers, Knoxville, USA) was introduced at the University Medical Center Groningen. This study evaluated the initial experiences with this new PET/CT system in terms of perceived image quality and semi-quantitative analysis in comparison to the Biograph mCT (Siemens Healthineers, Knoxville, USA) as a reference.
Methods: A total of 20 oncological patients were enrolled and received a single 18F-FDG injected dose of 3 MBq/kg followed by a dual-imaging PET protocol. Ten patients were scanned on the Biograph mCT first, while the other ten patients were scanned on the Biograph Vision first. The locally preferred clinical reconstructions were blindly reviewed by three nuclear medicine physicians and scored 1-5 on tumor lesion delineation, overall image quality, and image noise. In addition, these clinical reconstructions were used for semi-quantitative analysis; measurement of standardized uptake values (SUVs) in tumor lesions. Reconstructions conform the European Association of Nuclear Medicine Research Ltd (EARL) specifications were also used for measurements of SUVs in tumor lesions and healthy tissues for comparison between systems.
Results: 20 oncological patients (14 men, 6 women; age 36-84, mean ± SD 61 ± 16 years) received a single 18F-FDG injected dose (range 145-405 MBq, mean ± SD 268 ± 59.3). Images acquired on the Biograph Vision were scored significantly higher on tumor lesion delineation, overall image quality, and image noise in comparison to images acquired on the Biograph mCT (P < 0.001). The overall inter-reader agreement showed a Fleiss kappa of 0.61 (95% CI 0.53 - 0.70). Furthermore, the measured SUVs in tumor lesions and healthy tissues showed a good agreement (within 95%) between PET/CT systems, in particular when using EARL compliant reconstructions on both systems.
Conclusion: In this initial study it was found that the Biograph Vision PET/CT showed improved image quality compared with the Biograph mCT in terms of lesion demarcation, overall image quality, and visually assessed signal-to-noise ratio. The two systems are comparable in semi-quantitatively assessed image biomarkers in both healthy tissues and tumor lesions. Improved quantitative performance may, however, be feasible using the clinically optimized reconstruction settings.
Copyright © 2019 by the Society of Nuclear Medicine and Molecular Imaging, Inc.

Entities:  

Keywords:  Instrumentation; Oncology: General; PET/CT; digital detectors; image quality; silicon photomultipliers; standardized uptake value

Year:  2019        PMID: 31253742     DOI: 10.2967/jnumed.119.227801

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  11 in total

1.  Comparison of Image Quality and Semi-quantitative Measurements with Digital PET/CT and Standard PET/CT from Different Vendors.

Authors:  Sung Hoon Kim; Bong-Il Song; Hae Won Kim; Kyoung Sook Won
Journal:  Nucl Med Mol Imaging       Date:  2020-08-13

2.  New standards for phantom image quality and SUV harmonization range for multicenter oncology PET studies.

Authors:  Go Akamatsu; Naoki Shimada; Keiichi Matsumoto; Hiromitsu Daisaki; Kazufumi Suzuki; Hiroshi Watabe; Keiichi Oda; Michio Senda; Takashi Terauchi; Ukihide Tateishi
Journal:  Ann Nucl Med       Date:  2022-01-14       Impact factor: 2.668

3.  Total-body PET/CT using half-dose FDG and compared with conventional PET/CT using full-dose FDG in lung cancer.

Authors:  Hui Tan; Xiuli Sui; Hongyan Yin; Haojun Yu; Yusen Gu; Shuguang Chen; Pengcheng Hu; Wujian Mao; Hongcheng Shi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2020-11-27       Impact factor: 9.236

4.  Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer.

Authors:  David Kersting; Walter Jentzen; Miriam Sraieb; Pedro Fragoso Costa; Maurizio Conti; Lale Umutlu; Gerald Antoch; Michael Nader; Ken Herrmann; Wolfgang Peter Fendler; Christoph Rischpler; Manuel Weber
Journal:  EJNMMI Phys       Date:  2021-02-15

5.  Evaluation of 18F-FDG PET/CT images acquired with a reduced scan time duration in lymphoma patients using the digital biograph vision.

Authors:  Manuel Weber; Walter Jentzen; Regina Hofferber; Ken Herrmann; Wolfgang Peter Fendler; Christoph Rischpler; Lale Umutlu; Maurizio Conti; Pedro Fragoso Costa; Miriam Sraieb; David Kersting
Journal:  BMC Cancer       Date:  2021-01-14       Impact factor: 4.430

6.  Evaluation of [68Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision.

Authors:  Manuel Weber; Walter Jentzen; Regina Hofferber; Ken Herrmann; Wolfgang Peter Fendler; Maurizio Conti; Axel Wetter; David Kersting; Christoph Rischpler; Pedro Fragoso Costa
Journal:  EJNMMI Res       Date:  2021-02-28       Impact factor: 3.138

7.  EANM procedure guidelines for brain PET imaging using [18F]FDG, version 3.

Authors:  Eric Guedj; Andrea Varrone; Ronald Boellaard; Nathalie L Albert; Henryk Barthel; Bart van Berckel; Matthias Brendel; Diego Cecchin; Ozgul Ekmekcioglu; Valentina Garibotto; Adriaan A Lammertsma; Ian Law; Iván Peñuelas; Franck Semah; Tatjana Traub-Weidinger; Elsmarieke van de Giessen; Donatienne Van Weehaeghe; Silvia Morbelli
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-12-09       Impact factor: 10.057

Review 8.  Shining Damaged Hearts: Immunotherapy-Related Cardiotoxicity in the Spotlight of Nuclear Cardiology.

Authors:  David Kersting; Stephan Settelmeier; Ilektra-Antonia Mavroeidi; Ken Herrmann; Robert Seifert; Christoph Rischpler
Journal:  Int J Mol Sci       Date:  2022-03-30       Impact factor: 5.923

9.  Brain partial volume correction with point spreading function reconstruction in high-resolution digital PET: comparison with an MR-based method in FDG imaging.

Authors:  Masanobu Ibaraki; Keisuke Matsubara; Yuki Shinohara; Miho Shidahara; Kaoru Sato; Hiroyuki Yamamoto; Toshibumi Kinoshita
Journal:  Ann Nucl Med       Date:  2022-05-26       Impact factor: 2.258

10.  Value of SiPM PET in myocardial perfusion imaging using Rubidium-82.

Authors:  S S Koenders; J A van Dalen; P L Jager; S Knollema; J R Timmer; M Mouden; C H Slump; J D van Dijk
Journal:  J Nucl Cardiol       Date:  2020-05-14       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.