Accuracy of an intraoral digital scanner in tooth color determination.

Vygandas Rutkūnas1, Julius Dirsė2, Vytautas Bilius2.   

Abstract

STATEMENT OF PROBLEM: Whether intraoral digital scanners with an integrated shade-taking function can substitute for colorimeters, spectrophotometers, or the visual method to reduce working time is unclear.
PURPOSE: The purpose of this clinical study was to evaluate the accuracy of the measurement of tooth shade obtained with an intraoral digital scanner in vivo.
MATERIAL AND METHODS: Shades of 120 maxillary anterior teeth were evaluated by using a SpectroShade spectrophotometer (SS) and a TRIOS 3 intraoral digital scanner (T3) on 20 participants. The matching of shade readings between the T3 and SS was used to estimate the accuracy of the T3. The percentage of readings when a difference between the shades obtained by both devices was visually perceptible (ΔE>3.7) was calculated. Each of the 120 teeth was measured 5 times to assess repeatability.
RESULTS: The accuracy of the T3 was 53.3% when the color was recorded as a Vita 3D-Master (VM) shade and 27.5% for the Vita Classical (VC) shade guide when the SS was taken as a reference. A visually perceptible color difference was found in 25% (VM) and 50.8% (VC) of situations when the shade was determined with the SS and 48.3% (VM) and 78.3% (VC) with the T3. Repeatability was 92% (VM) and 93.5% (VC) for the SS, and 90.33% (VM) and 87.17% (VC) for the T3.
CONCLUSIONS: The findings of this study revealed that the tooth color determined by the T3 does not exactly match that obtained by the SS that additional methods of measuring tooth color are recommended. The accuracy of the T3 was higher when the color was recorded as VM values rather than VC values.
Copyright © 2019 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 31227232     DOI: 10.1016/j.prosdent.2018.12.020

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  4 in total

1.  Evaluation of accuracy and characteristics of tooth-color matching by intraoral scanners based on Munsell color system: an in vivo study.

Authors:  Mingming Huang; Hongqiang Ye; Hu Chen; Yongsheng Zhou; Yunsong Liu; Yong Wang; Yuchun Sun
Journal:  Odontology       Date:  2022-02-25       Impact factor: 2.885

2.  Digital intraoral scanner devices: a validation study based on common evaluation criteria.

Authors:  Ivett Róth; Alexandra Czigola; Dóra Fehér; Viktória Vitai; Gellért Levente Joós-Kovács; Péter Hermann; Judit Borbély; Bálint Vecsei
Journal:  BMC Oral Health       Date:  2022-04-26       Impact factor: 3.747

3.  Translucency of a Dental Porcelain Mixed by Two Ceramic Slurry Methods: A Bayesian Comparison.

Authors:  Catalina Serna-Meneses; Gabriel Ocampo-Parra; Santiago Arango-Santander; Claudia Garcia-Garcia; Luis Felipe Restrepo-Tamayo; Johnatan Cardona-Jimenez; Alexander Ossa; Alejandro Pelaez-Vargas
Journal:  Int J Dent       Date:  2022-06-06

4.  Occlusion in the digital era: a report on 3 cases.

Authors:  Smaranda Buduru; Anca Mesaros; Daniel Talmaceanu; Oana Baru; Raul Ghiurca; Raluca Cosgarea
Journal:  Med Pharm Rep       Date:  2019-12-15