Daiki Eto1, Ryota Tanaka1, Yosuke Suzuki1, Yuhki Sato1, Hiroki Itoh1.
Abstract
BACKGROUND: Recently, the Nanopia® TDM Zonisamide reagent using the latex particle-enhanced turbidimetric immunoassay (LTIA) method was developed. The aim of this study was to compare the differences in serum zonisamide (ZNS) concentrations quantified by the high-performance liquid chromatography (HPLC) method and the LTIA method using a TBA-25FR analyzer.Entities:
Keywords: antiepileptic; high-performance liquid chromatography; latex-enhanced turbidimetric immunoassay; therapeutic drug monitoring; zonisamide
Mesh:
Substances:
Year: 2019 PMID: 31218755 PMCID: PMC6757127 DOI: 10.1002/jcla.22940
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Figure 1Chromatograms of (A) blank serum; (B) serum spiked with LLOQ at concentration of 1.0 µg/mL; (C) serum spiked with high QC at concentration of 75 µg/mL; (D) serum sample of a patient. LLOQ, lower limit of quantitation; QC, quality control
Validation results for measuring zonisamide concentrations in human serum by the HPLC method
| Zonisamide concentrations measured by HPLC (µg/mL) | ||||
|---|---|---|---|---|
| LLOQ | QC A | QC B | QC C | |
| 1 | 3 | 30 | 75 | |
| Intra‐day | ||||
| Mean (µg/mL) | 1.04 | 3.02 | 30.18 | 73.17 |
| Accuracy (%) | 3.73 | 0.59 | 0.59 | −2.44 |
| Precision (%CV) | 1.17 | 1.33 | 1.86 | 1.57 |
| Inter‐day | ||||
| Mean (µg/mL) | 0.94 | 2.93 | 28.81 | 70.25 |
| Accuracy (%) | −5.78 | −2.45 | −3.98 | −6.33 |
| Precision (%CV) | 10.27 | 4.95 | 6.39 | 9.00 |
Abbreviations: CV, coefficient of variation; HPLC, high‐performance liquid chromatography; LLOQ, lower limit of quantitation; QC, quality control.
Validation results for measuring zonisamide concentrations in human serum by the LTIA method
| Zonisamide concentrations measured by LTIA (µg/mL) | ||||
|---|---|---|---|---|
| LLOQ | QC A | QC B | QC C | |
| 5 | 8 | 25 | 50 | |
| Intra‐day | ||||
| Mean (µg/mL) | 5.23 | 6.70 | 22.60 | 45.20 |
| Accuracy (%) | 4.67 | −15.80 | −9.50 | −9.50 |
| Precision (%CV) | 3.56 | 1.80 | 2.50 | 2.40 |
| Inter‐day | ||||
| Mean (µg/mL) | 5.33 | 7.20 | 23.60 | 46.20 |
| Accuracy (%) | 6.67 | −10.60 | −5.70 | −7.60 |
| Precision (%CV) | 2.48 | 5.20 | 3.70 | 2.20 |
Abbreviations: CV, coefficient of variation; LLOQ, lower limit of quantitation; LTIA, latex‐enhanced turbidimetric immunoassay; QC, quality control.
Figure 2Relationship between serum zonisamide concentrations measured by HPLC and LTIA methods (A), and a Bland‐Altman plot for serum zonisamide concentrations measured by HPLC and LTIA methods (B). HPLC, high‐performance liquid chromatography; LTIA, latex‐enhanced turbidimetric immunoassay