Shuto Osaki1,2, Takuya Kintoki3,4, Takayo Moriuchi-Kawakami5, Kenichi Kitamura6, Shin-Ichi Wakida7,8.
Abstract
We have investigated human-stress monitoring by making use of salivary nitrate, which can be a candidate for stress markers, with ion-selective field-effect transistors (ISFETs). ISFETs are suitable for on-site single-drop analysis of salivary nitrate within 10 s. However, when ISFETs are used for salivary nitrate, ISFETs have a problem that is called the initial drift. The initial drift makes accurate nitrate monitoring difficult. Thus, the purpose of this study is to prevent the initial drift and to search for a new, simple polymer to possess a better performance of sensor responses than conventional matrix membranes, such as PVC. In this research, we investigated ISFETs using specific matrix membranes, for example KP-13, Pellethane®--, and P7281-PU. The initial drift was evaluated from the fluctuations of the response values generated by the ISFETs when immersed in saliva or aqueous solution. As a result, P7281-PU showed a prevention effect on the initial drift, both in the whole saliva and in various solutions. Furthermore, the cause of drift may be H+ diffusion, and the drift prevention effect of P7281-PU may be affected by urethane bond capturing H+ in the ion-selective membrane. This result suggests that a continuous nitrate monitoring is feasible and may be applied to wearable sensors.Entities:
Keywords: ISE; ISFETs; drift; ion-selective membrane; polyurethane; salivary nitrate; stress
Mesh:
Substances:
Year: 2019 PMID: 31212895 PMCID: PMC6630929 DOI: 10.3390/s19122713
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1pH-sensitive ion-selective field-effect transistors (pH-ISFETs) and accessory cap.
Figure A1Initial drift of NO3-ISFETs in whole saliva. (a) Same data as Figure 3f; (b) for reproducibility check.
Figure 2Calibration curves of NO3-ISFETs. (a) Only KNO3 solution. (b) In artificial saliva.
Selectivity coefficients.
| Polymer Matrix |
| ||||
|---|---|---|---|---|---|
| I | SCN | NO2 | Cl | SO4 | |
| PVC | 0.82 | 0.95 | −1.53 | −2.55 | −4.10 |
| KP-13 | 0.56 | 0.95 | −1.53 | −2.59 | −4.03 |
| Pellethane® | 0.78 | 1.09 | −1.50 | −2.71 | −4.11 |
| P7281-PU | 0.69 | 1.07 | −1.40 | −2.30 | −4.13 |
Figure 3Initial drift of NO3-ISFETs (a) in standard solution; (b) in standard solution with 0.5 wt % mucin; (c) in standard solution with CO2 bubbling; (d) in artificial saliva; (e) in saliva via Salivette; (f) in whole saliva (non-pretreatment).
Figure 4Long-term stability of P7281-PU-based ISFETs. This measurement is done in only KNO3 solution.
Figure 5The hypothesis of urethane bond capture of H+ and OH- by hydrogen bonds (in whole saliva).