| Literature DB >> 31197608 |
Andrew G Kalina1, Paul H Kalina2, Morgan M Brown3.
Abstract
INTRODUCTION: The purpose of this study was to evaluate intraocular pressure (IOP) lowering and safety of XEN® stent in medically refractory, progressive, open-angle glaucoma (OAG).Entities:
Keywords: Glaucoma, MIGS; Surgery; XEN® stent
Year: 2019 PMID: 31197608 PMCID: PMC6692422 DOI: 10.1007/s40123-019-0192-8
Source DB: PubMed Journal: Ophthalmol Ther
Demographics and baseline characteristics of the total population (n = 42)
| Parameter | Total: 47 eyes of 42 patients |
|---|---|
| Age (years) | |
| Mean ± SD | 78.15 ± 8.55 |
| Range | 54–91 |
| Sex, | |
| Female | 28 (66.7%) |
| Male | 14 (33.3%) |
| Study eye, | |
| Right | 20 (42.6) |
| Left | 27 (57.4) |
| Ethnicity, | |
| White | 41 (97.6%) |
| Asian | 1 (2.4%) |
| Hispanic or Latino | 0 |
| Black/African American | 0 |
| Glaucoma type, | |
| POAG | 32 (68.1) |
| Pseudoexfoliative | 11 (23.4) |
| Low tension | 4 (8.5) |
| Previous glaucoma procedures | |
| Selective laser trabeculoplasty, | 20 (42.6) |
| iStent, | 1 (2.1) |
| Endocyclophotocoagulation, | 1 (2.1) |
| Filtration or tube surgery, | 0 |
| IOP at baseline | |
| Mean ± SD | 22.34 (7.34) |
| Range | 11.00–36.00 |
| Mean medications at baseline, | 2.96 (1.20) |
| Total number, | |
| None | 1 (2.1) |
| One | 3 (6.4) |
| Two | 10 (21.3) |
| Three | 22 (46.8) |
| Four | 7 (14.9) |
| Five or more | 4 (8.5) |
| XEN® stent and cataract surgery, | 27 (57.4) |
| XEN® stent as standalone, | 20 (42.6) |
| Phakic, | 7 (35) |
Pseudophakic, n (% of 20) 13 (65)
Fig. 1Mean intraocular pressure (IOP) change and mean medication reduction from baseline for the overall population at each visit after XEN® stent placement. +Percentage change in mean intraocular pressure (IOP) from baseline shown in parentheses
Fig. 2Scatterplot of preoperative and postoperative intraocular pressure (IOP) at 6 months. Line shows best fit
Fig. 3Scatterplot of preoperative and postoperative intraocular pressure (IOP) at 12 months. Line shows best fit
Mean IOP in eyes receiving XEN® stent as a standalone procedure (n = 20) versus XEN® stent with cataract surgery (n = 27)
| IOP (mmHg) | ||||
|---|---|---|---|---|
| XEN® stent standalone | 95% confidence interval | XEN® stent with cataract surgery | 95% confidence interval | |
| Baseline (mean ± SD) | 24.18 ± 8.18 ( | (20.59, 27.76) | 20.98 ± 6.47 ( | (18.54, 23.42) |
| 1 week post-op (mean ± SD) | 8.83 ± 4.13 ( | (7.01, 10.64) | 12.79 ± 5.39 ( | (10.72, 14.86) |
| 1 month post-op (mean ± SD) | 11.83 ± 3.65 ( | (10.23, 13.42) | 13.72 ± 4.48 ( | (12.03, 15.41) |
| 3 months post-op (mean ± SD) | 13.37 ± 5.11 ( | (11.07, 15.67) | 12.61 ± 3.70 ( | (11.09, 14.12) |
| 6 months post-op (mean ± SD) | 13.53 ± 4.89 ( | (11.33, 15.73) | 13.46 ± 3.14 ( | (12.28, 14.65) |
| 1 year post-op (mean ± SD) | 13.04 ± 4.50 ( | (10.68, 15.39) | 13.61 ± 2.90 ( | (12.27, 14.95) |