| Literature DB >> 31183150 |
Lan-Gui Xie1, Xia Zhao1, Si-Hong Dou1, Long Tang1, Hui-Min Sun1.
Abstract
A novel reference film was characterized to improve the oxygen gas transmission measurement accuracy of plastic materials for pharmaceutical packaging. The material processing, homogeneity, stability, jointly determined value and uncertainty evaluation were discussed. The film is the first reference film characterized by multiple laboratories using both manometric and coulometric methods. The oxygen transmission rate of the reference film was 20.53 with the expanded uncertainty of 1.36. The newly characterized reference film can be used in the calibration and self-calibration of oxygen transmission measurement equipment and analytical method verification to improve the measurement accuracy and achieve traceable data.Entities:
Keywords: certified reference material; coulometric method; manometric method; oxygen gas transmission; polyethylene terephthalate; uncertainty evaluation
Year: 2019 PMID: 31183150 PMCID: PMC6502387 DOI: 10.1098/rsos.190142
Source DB: PubMed Journal: R Soc Open Sci ISSN: 2054-5703 Impact factor: 2.963
Oxygen transmission rates of the same plastic film obtained with different instruments and different methods from 22 Chinese laboratories in 2014.
| laboratory number | oxygen gas transmission rates (ml m−2 d−1) |
|---|---|
| L08 | 79.4 |
| L32 | 85.5 |
| L11 | 87.2 |
| L18 | 87.2 |
| L14 | 87.4 |
| L12 | 90.4 |
| L01 | 102.7 |
| L31 | 104.3 |
| L24 | 108.0 |
| L07 | 108.7 |
| L05 | 114.7 |
| L02 | 117.1 |
| L26 | 118.7 |
| L04 | 120.1 |
| L03 | 120.9 |
| L06 | 121.9 |
| L25 | 122.0 |
| L09 | 122.3 |
| L16 | 122.4 |
| L27 | 122.7 |
| L15 | 123.0 |
| L30 | 123.8 |
| average value | 108.0 |
| RSD | 14.4%a |
aRSD 14.4% suggests the significant dispersion of the oxygen transmission rates obtained by different laboratories using different methods. In general, the RSD of oxygen transmission rates for a homogeneous plastic film should not be higher than 10% and the data should be normally distributed.
Figure 1.Statistical dispersion of oxygen transmission rates shown in table 1. The horizontal and vertical coordinates represent the oxygen transmission rates reported by 22 laboratories and the referred laboratory number of the corresponding oxygen transmission rate, respectively. The statistic histogram shows that measurement data from 22 laboratories are not a normal distribution.
Figure 2Representative sample selection on the film of six units. The dark squares are the sampling units and each sample unit includes three square pieces.
Thickness homogeneity of film CCRM195017.
| thickness (μm) | |||
|---|---|---|---|
| group no. | test no. 1 | test no. 2 | test no. 3 |
| 1 | 82.24 | 82.74 | 81.56 |
| 2 | 81.77 | 82.35 | 82.77 |
| 3 | 83.14 | 82.68 | 82.47 |
| 4 | 82.89 | 82.23 | 82.20 |
| 5 | 82.36 | 82.44 | 81.22 |
| 6 | 82.57 | 81.92 | 82.09 |
| 7 | 81.67 | 82.70 | 82.61 |
| 8 | 81.79 | 82.31 | 81.25 |
| 9 | 82.57 | 82.49 | 82.70 |
| 10 | 81.34 | 82.93 | 81.94 |
| 11 | 81.94 | 81.55 | 81.87 |
| 12 | 82.34 | 82.76 | 82.25 |
| 13 | 81.56 | 82.68 | 82.63 |
| 14 | 82.40 | 81.41 | 82.22 |
| 15 | 82.55 | 83.02 | 82.83 |
| average value | 82.27 | ||
| RSD (%) | 0.61 | ||
Oxygen gas transmission homogeneity of film CCRM195017.
| oxygen gas transmission rates (ml m−2 (0.1 MPa d)−1) | |||
|---|---|---|---|
| group no. | test no. 1 | test no. 2 | test no. 3 |
| 1 | 18.719 | 19.576 | 20.078 |
| 2 | 20.793 | 19.821 | 20.019 |
| 3 | 20.345 | 20.017 | 19.021 |
| 4 | 20.363 | 20.218 | 19.023 |
| 5 | 20.488 | 20.526 | 21.153 |
| 6 | 19.894 | 19.875 | 20.309 |
| 7 | 19.812 | 18.744 | 19.579 |
| 8 | 20.675 | 19.857 | 20.705 |
| 9 | 19.760 | 19.773 | 19.992 |
| 10 | 20.686 | 19.966 | 19.579 |
| 11 | 19.981 | 20.897 | 20.007 |
| 12 | 19.962 | 20.609 | 19.658 |
| 13 | 20.495 | 19.335 | 19.586 |
| 14 | 21.014 | 20.710 | 20.609 |
| 15 | 19.685 | 19.956 | 20.367 |
| average value | 20.050 | ||
| RSD (%) | 2.85 | ||
Short-term stability of film CCRM195017 at 60°C.
| oxygen gas transmission rates (ml m−2 (0.1 MPa d)−1) | ||||
|---|---|---|---|---|
| storage time (day) | test no. 1 | test no. 2 | test no. 3 | average value |
| 0 | 20.251 | 21.166 | 20.582 | 20.666 |
| 3 | 19.854 | 19.794 | 19.579 | 19.742 |
| 7 | 20.095 | 20.245 | 19.592 | 19.977 |
| 11 | 19.806 | 19.811 | 19.793 | 19.804 |
| 14 | 20.078 | 20.295 | 19.972 | 20.115 |
Short-term stability of film CCRM195017 at −20°C.
| oxygen gas transmission rates (ml m−2 (0.1 MPa d)−1) | ||||
|---|---|---|---|---|
| storage time (day) | test no. 1 | test no. 1 | test no. 1 | average value |
| 0 | 18.964 | 19.957 | 19.517 | 19.480 |
| 3 | 20.059 | 20.467 | 19.959 | 20.162 |
| 7 | 20.907 | 20.581 | 20.301 | 20.596 |
| 11 | 20.039 | 20.089 | 20.187 | 20.105 |
| 14 | 19.589 | 19.920 | 20.360 | 19.957 |
Long-term stability of film CCRM195017 at 25°C.
| oxygen gas transmission rates (ml m−2 (0.1 MPa d)−1) | ||||
|---|---|---|---|---|
| test no. | 0 month | the 3rd month | the 6th month | the 18th month |
| 1 | 20.74 | 20.83 | 21.34 | 20.32 |
| 2 | 20.79 | 20.32 | 21.18 | 20.64 |
| 3 | 20.33 | 19.41 | 20.78 | 20.42 |
| 4 | 20.83 | 19.77 | 20.86 | 20.76 |
| 5 | 20.77 | 20.79 | 21.00 | 21.00 |
| 6 | 20.54 | 20.76 | 21.26 | 20.65 |
| 7 | 20.89 | 20.54 | / | / |
| 8 | 20.84 | 20.84 | / | / |
| 9 | 20.75 | 20.75 | / | / |
| 10 | 20.89 | 20.89 | / | / |
| average | 20.74 | 20.49 | 21.07 | 20.63 |
Oxygen gas transmission rates of film CCRM195017 measured by 4 CNAS laboratories using the manometric method or the coulometric method.
| oxygen gas transmission rates measured by the manometric method (ml m−2 (0.1 MPa d)−1) | oxygen gas transmission rates measured by the coulometric method (ml m−2 d−1) | |||||||
|---|---|---|---|---|---|---|---|---|
| test no. | M1 lab | M2 lab | M3 lab | M4 lab | M1 lab | M2 lab | M3 lab | M4 lab |
| 1 | 21.3 | 20.2 | 20.66 | 19.738 | 20.92 | 20.02 | 20.17 | 20.743 |
| 2 | 21.7 | 20.6 | 19.90 | 19.676 | 20.01 | 20.16 | 20.20 | 20.790 |
| 3 | 22.3 | 20.4 | 20.35 | 20.241 | 20.40 | 20.64 | 20.60 | 20.326 |
| 4 | 20.4 | 19.8 | 21.16 | 20.166 | 20.48 | 20.76 | 20.40 | 20.834 |
| 5 | 21.2 | 20.5 | 20.27 | 19.654 | 21.02 | 20.32 | 20.35 | 20.765 |
| 6 | 22.1 | 20.4 | 20.62 | 19.803 | 20.68 | 20.84 | 20.26 | 20.540 |
| 7 | 22.0 | 19.5 | 19.63 | 20.993 | 20.16 | 20.88 | 20.22 | 20.892 |
| 8 | 21.0 | 20.4 | 21.04 | 19.911 | 20.43 | 21.05 | 20.15 | 20.844 |
| 9 | 22.3 | 20.9 | 20.51 | 20.922 | 19.89 | 20.87 | 20.22 | 20.754 |
| 10 | 22.2 | 20.7 | 19.86 | 19.538 | 19.73 | 20.97 | 20.26 | 20.892 |
| mean | 21.65 | 20.34 | 20.40 | 20.06 | 20.37 | 20.65 | 20.28 | 20.74 |
| s.d. | 0.65 | 0.42 | 0.50 | 0.52 | 0.43 | 0.36 | 0.14 | 0.18 |
| assigned value | 20.53 | |||||||
Variations of the components affecting the manometric test.
| component | description | value | standard derivation | relative standard derivation |
|---|---|---|---|---|
| Δ | pressure change in the low-pressure chamber | 25.87 Pa | 0.2582 Pa | 0.010 |
| Δ | time interval | 30 min | 0 min | 0 |
| volume of the bottom chamber | 5.30 cm3 | 0.0053 cm3 | 0.001 | |
| transmission area | 50.24 cm2 | 0.01 cm2 | 0.0002 | |
| chamber temperature | 23°C | 0.23°C | 0.011 | |
| pressure difference between two sides of the film | 101.02 KPa | 124.275 Pa | 0.001 |
Variations of the components affecting the coulometric test.
| component | description | value | standard derivation | relative standard derivation |
|---|---|---|---|---|
| accuracy of the coulometric sensor | / | / | 0.00577 | |
| transmission area | 50 cm2 | 0 cm2 | 0 | |
| chamber temperature | 23°C | 0.2898°C | 0.0126 |