Characterization and uncertainty evaluation of a new certified reference material of baicalein.

Ningbo Gong1, Xiaoying Wang1, Baoxi Zhang1, Zhaolin Gao2, Guanhua Du3, Yang Lv4.   

Abstract

A new certified reference material (CRM) of baicalein was developed aiming to control the quality of baicalein and relative pharmaceuticals. Sample preparation, homogeneity, stability, value assignment, and uncertainty of a new certified reference material (CRM) of baicalein were presented in this paper. Characterization of the material relied on three different methods, which were differential scanning calorimetry (DSC), coulometric titration method (CT) and mass balance method. DSC and CT were used for purity determination of baicalein firstly. The certified value of baicalein CRM is 99.72% with an expanded uncertainty of 0.25% (k=2). The new CRM of baicalein can be used to validate analytical methods, improve the accuracy of measurement data as well as establish meteorological traceability of analytical results.
Copyright © 2013 Elsevier B.V. All rights reserved.

Keywords:  Baicalein; Certified reference material; Coulometric titration; Differential scanning calorimetry; Mass balance method

Mesh:

Substances:

Year:  2013        PMID: 24280738     DOI: 10.1016/j.jpba.2013.10.015