Literature DB >> 31179411

Rapid Identification of Novel Psychoactive and Other Controlled Substances Using Low-Field 1H NMR Spectroscopy.

Lysbeth H Antonides1, Rachel M Brignall2, Andrew Costello3,4, Jamie Ellison1,3,4, Samuel E Firth1, Nicolas Gilbert1,4, Bethany J Groom1, Samuel J Hudson1, Matthew C Hulme1,4, Jack Marron1,4, Zoe A Pullen1, Thomas B R Robertson1, Christopher J Schofield3,4, David C Williamson2, E Kate Kemsley5, Oliver B Sutcliffe1,4, Ryan E Mewis1.   

Abstract

An automated approach to the collection of 1H NMR (nuclear magnetic resonance) spectra using a benchtop NMR spectrometer and the subsequent analysis, processing, and elucidation of components present in seized drug samples are reported. An algorithm is developed to compare spectral data to a reference library of over 300 1H NMR spectra, ranking matches by a correlation-based score. A threshold for identification was set at 0.838, below which identification of the component present was deemed unreliable. Using this system, 432 samples were surveyed and validated against contemporaneously acquired GC-MS (gas chromatography-mass spectrometry) data. Following removal of samples which possessed no peaks in the GC-MS trace or in both the 1H NMR spectrum and GC-MS trace, the remaining 416 samples matched in 93% of cases. Thirteen of these samples were binary mixtures. A partial match (one component not identified) was obtained for 6% of samples surveyed whilst only 1% of samples did not match at all.

Entities:  

Year:  2019        PMID: 31179411      PMCID: PMC6547625          DOI: 10.1021/acsomega.9b00302

Source DB:  PubMed          Journal:  ACS Omega        ISSN: 2470-1343


Introduction

The detection, and identification, of drugs within a seized sample is a desirable means of evidence and intelligence gathering.[1] Given the growing issue of novel psychoactive substances (NPS) and herbal products mixed with synthetic cannabinoids, known as spice,[2,3] a technique is required for the detection of the active pharmaceutical ingredient (API) within a seized sample, if indeed it is present. Such a technique is urgently needed, given that the number of NPS reported to the European Monitoring Centre for Drugs and Drug Addiction (EMCDDA) has risen from 13 in 2008 to 101 in 2014[4] and that in 2016, a survey of 625 UK inmates reported 33% had used spice in the last month.[5] The established analytical tools of GC–MS (gas chromatography–mass spectrometry) and HPLC–MS (high performance liquid chromatography–MS), which are employed by forensic laboratories (and by harm reduction point-of-care services[6]) for the identification and quantification of drugs of abuse, struggle with the high-throughput that is required to test the number of seized samples. Additionally, in some instances, the reference standards are not available for comparison. Recently, a number of studies have been reported that focus on the use of various spectroscopic techniques to achieve high sample throughput, whilst also detecting the API present in seized samples. Studies have utilized IR (infra-red) spectroscopy,[7−10] Raman spectroscopy,[11−14] NMR (nuclear magnetic resonance),[9,10,15−17] HPLC,[9,10,18−21] and GC–MS.[9,10,22−25] IR and Raman spectroscopies have been used for the analysis of more than 200 seized NPS samples, resulting in the identification of a single component or a mixture in 76% of the samples surveyed.[26] 17% of samples were partially identified, with a further 7% that were unidentifiable. One of the limitations of this study was the lack of reference spectra; the reference library needed to be extended through the independent identification of NPS via NMR and MS. A recent development concerning Raman spectroscopy is the use of a portable hand-held technology for mobile detection. A study employing 785 and 1064 nm excitation sources highlighted that the latter wavelength was more suitable, giving fewer false positives and false negatives for the NPS screened whilst also significantly lowering fluorescence.[27] In addition, IR has been used in conjunction with chemometric methods (partial least squares (PLS) discriminant analysis) to distinguish four classes of drugs with an average reliability rate of 98.6% for the 932 samples surveyed.[8] A further use of PLS-based analysis in conjunction with Raman and IR spectroscopies has been exemplified for the quantification of cocaine in samples consisting of cocaine, sodium carbonate, and either caffeine or lidocaine.[28] A further study utilized liquid chromatography coupled to high-resolution MS for the detection of NPS in wastewater from a festival in Amsterdam.[29] Implementation of an algorithm to extract accurate masses which were then compared to an in-house database of 2000 entries of NPS and transformation products facilitated the detection of eight NPS compounds. The combination of dopant-assisted positive photoionization ion mobility spectrometry (DAPP-IMS) and direct analysis in real time time-of-flight mass spectrometry (DART-TOF MS) has been employed successfully to screen drug samples with minimal sample preparation.[30] Although the method provided a fast response and high-throughput, the method was limited due to its inability to identify all the components in a mixture or unambiguously identify similar compounds, such as epimers. A method utilizing GC–MS for the analysis of cathinones, amphetamine-like stimulants, and other NPS has been developed that incorporates an almost simultaneous derivatization and extraction step, which is achieved through the use of hexyl chloroformate and dispersive liquid–liquid microextraction.[22] The chromatographic method allowed the separation of 26 stimulants, which were all resolved in under 15 min. The development of a compound database and HRMS (high-resolution mass spectrometry) spectral library, consisting of data for 875 and 252 NPS from multiple classes, respectively, has been utilized for the analysis of 13 serum samples that had been spiked with up to five NPS per sample.[31] 32 of the 35 compounds present were detected using the method developed. Furthermore, HRMS-based screening of four seized samples provided results consistent with official forensic laboratory findings. NMR, despite being a powerful analytical technique, has been underutilized for the detection of drugs within samples. This is largely because of the perceived complexity of the measurement but also the inherent cost of requiring a superconducting magnet and the associated technical costs. However, the recent development of bench-top NMR spectrometers has made this technique a much more attractive and affordable prospect. This approach is exemplified through the identification of drugs in real case samples, carried out for two samples, using a limited library of twelve spectra on a benchtop 80 MHz spectrometer and compared by visual inspection with data collected on a 600 MHz spectrometer.[32] Another report has investigated the differentiation of 65 fentanyl analogues, by eye, using 1H NMR spectra obtained on a low-field spectrometer (62 MHz).[33] Quantum mechanical spin system analysis was employed in this report, and this highlighted that field strength-independent 1H NMR libraries could be useful for drug analysis. Furthermore, the quantification of components in over-the-counter painkiller drugs has been utilized in an undergraduate practical, again using a visual inspection to identify the various components.[16] Screening of spice samples using low-field NMR has been explored by Gilard and co-workers.[34] In their study, nine synthetic cannabinoids were detected and quantified. Although high-field NMR spectroscopy, sometimes in conjunction with mass spectrometry, is required for unambiguous structural identification, low-field NMR provided useful insights, in the form of diagnostic signals, as to their chemical structure. However, the greater spectral band overlap compared to data acquired from a superconducting system did present a data analysis challenge. In this paper, we detail a fully automated NMR system that acquires the 1H NMR spectrum of a sample, processes it, and then using a developed pattern recognition algorithm, compares the spectrum to a reference library of over 300 1H NMR spectra, before finally reporting the drug(s) present (if there are any to detect). The simplistic nature of the experimental measurement overcomes the need for dedicated technical staff, and the robustness of the algorithm for spectral analysis means that it can be readily utilized in custodial suites.

Results and Discussion

Reference Library and Algorithm Development

1H NMR spectra of all reference compounds were collected using the Pulsar instrument. The reference library thus consisted of 302 spectra altogether. The large number and types of materials represented in the reference library enable a wide range of seized materials to be identified and also allow for various tests of the performance of the pattern recognition algorithm. The 1H NMR spectra were truncated into two discrete sections: the “class” region (0.46–1.54 ppm) and a fingerprint region (3.90–12.50 ppm). “Classes” of compounds possess similar backbone structures, and hence spectral features, but it was imperative that the algorithm developed could also differentiate compounds within a class. Both sections were used in the pattern recognition process. Further, to account for some variations in the dimethyl sulfoxide (DMSO) signal that was used as an internal standard, the algorithm accommodated a tolerance of ±0.06 ppm on the chemical shift of any peak. Before commencing any analysis of seized samples, the algorithm’s performance in terms of correct identifications was tested by using it to the validation sample spectra. This allowed evaluation of technical repeatability, as well as the ability of the method to detect and identify previously unseen mixtures and an estimation of the rate of false-positives. Of the 130 controlled substance replicates, the class was correctly identified in all but one of the samples (>99% success rate) and the compound in all but six (>95% success rate). Notably, all of these were from the diphenidine class, of which 5 were misidentified as other closely related diphenidines. Note that the diphenidines were somewhat over-represented amongst the technical replicates due to the ready availability of these compounds. The algorithm produces a “match score” as an output for each sample: this is the largest of the Pearson’s correlations between the sample spectrum and each of the library spectra. The median score from correct identifications across all classes was 0.993. In contrast, the median score for the incorrect identifications was 0.96. A summary figure expanding on these values is given in the Supporting Information. It is worth noting that the median score for the amphetamine class was 0.993 and that this was also the individual score for the technical replicate of MDMA; this was the most prevalent compound amongst the seized samples discussed below. For the 32 validation mixtures, there were 21 correct identifications of both compounds present in the sample. The median match score for these was 0.923. This is substantially lower than the typical match scores for single compounds, reflecting the increased difficulty of identifying unknown binary mixtures of unknown concentration ratios. A further 10 were identified as single compounds instead of mixtures, with one of the two compounds being correctly identified. The final sample was found to be a mixture, but one of the individual compounds was incorrectly identified as a different cutting agent; importantly, therefore, this was not a false positive with regards to controlled substances. From consideration of the outcomes from the validation samples, a threshold was set empirically at 0.838: match scores greater than this are indicative of a correct match, whereas identifications with lower scores are considered unreliable.

Analysis of Seized Samples

432 seized samples were analyzed as part of this work. These samples represent a seized material within the Greater Manchester region and tested by the MANchester DRug Analysis & Knowledge Exchange (MANDRAKE) partnership in accordance with the UK Misuse of Drugs Act (1971) and Misuse of Drugs Regulations (2001). All samples were analyzed using 1H NMR spectroscopy (ca. 5–10 mg of sample dissolved in 0.6 mL of deuterated DMSO) and GC–MS (see Supporting Information for method and sample preparation). The results from the 1H NMR analysis were cross-referenced with GC–MS data acquired from the same sample. The GC–MS validation approach is outlined in the Supporting Information. Compounds identified through the work detailed herein are shown in Figure .
Figure 1

Compounds detected in the seized samples as part of this work.

Compounds detected in the seized samples as part of this work. Of the 432 samples analyzed, thirteen (3%) possessed no peaks in the GC–MS chromatogram. Of these, 1H NMR analysis identified that adulterants such as lactose were present, which would require derivatization before they could be detected by GC–MS. Two samples were identified as containing lactose only (median match score 0.964) and another one as sucrose only (match score 0.968). Two further samples were indicated to be comprised of sodium gamma-hydroxybutyric acid with a median match score of 0.851. The other eight samples possessed match scores below the identification threshold (i.e. lower than 0.838), and thus their identifications were considered unreliable. Three samples returned a null result in both the 1H NMR and GC–MS, that is, they consisted of no material that possessed a 1H NMR signature, and no peak was observed in the GC–MS trace. The other 416 samples resulted in partial or complete agreement between the 1H NMR and the GC–MS outcomes. Of these 416 samples, the result from the 1H NMR analysis matched the GC–MS result exactly in 387 cases (93%), thirteen of which were identified as being binary mixtures by both GC–MS and 1H NMR. Partial agreement was obtained for 25 (6%) of samples surveyed, whilst the outcomes from just 4 samples (1%) did not match at all. A partial match here means that either the 1H NMR analysis or the GC–MS trace were missing one or more components than the other method detected. By far the most common drug identified was MDMA, accounting for 171 instances (40% of samples surveyed). Cocaine and ketamine (85 (20%) and 74 (17%) instances respectively) were the next most common. Samples found to contain mephedrone, paracetamol, benzocaine, ethylone, heroin, caffeine, diazepam, or cannabis accounted for a further 23 samples (5% of samples surveyed). Figure shows the percentage breakdown by class of single and binary component samples. A recent report from a UK festival onsite drug checking service reported that the three most common drugs identified were MDMA (57%), ketamine (13.5%), and cocaine (10%).[35] The results presented here reaffirm these drugs as the most commonly encountered and with similar frequencies.
Figure 2

Pie-chart showing the relative numbers of samples of different compositions, as identified by 1H NMR analysis and subsequently validated by GC–MS, for (a) single component samples and (b) binary mixtures.

Pie-chart showing the relative numbers of samples of different compositions, as identified by 1H NMR analysis and subsequently validated by GC–MS, for (a) single component samples and (b) binary mixtures. The median match score for the seized MDMA samples was found to be 0.986 (range 0.858–0.997). This implies that some of the samples analyzed were almost of reference standard quality. In theory, an exact match with a reference compound will have a score of 1; in practice, sources of variance such as instrumental drift, spectral noise, and concentration differences mean that this is rarely achieved. The value obtained for the technical replicate of the reference MDMA discussed above was 0.993. Figure shows the 1H NMR spectral variation of ten seized samples containing MDMA in comparison to the 1H NMR reference spectrum of MDMA.
Figure 3

Series of 1H NMR spectra of MDMA. Spectrum 1 is the reference spectrum that is included in the reference library, whereas spectra 2–11 are seized samples (sample IDs 123, 207, 244, 309, 377, 405, 561B, 582, 598A, 598B, and 613).

Series of 1H NMR spectra of MDMA. Spectrum 1 is the reference spectrum that is included in the reference library, whereas spectra 2–11 are seized samples (sample IDs 123, 207, 244, 309, 377, 405, 561B, 582, 598A, 598B, and 613). Similarly, ketamine and cocaine samples had median match scores of 0.980 and 0.973 respectively. For the cocaine samples, the 1H NMR analysis was able to differentiate between the free-base form and the hydrochloride salt. Four instances of free-base cocaine were found, for which the median match score was 0.184 higher than that of cocaine hydrochloride (median match-score for free-base cocaine was found to be 0.987). The difference between the two match scores is important, as the match score for cocaine hydrochloride is below the validation threshold in some instances. Although the 1H NMR spectroscopy analysis reveals this difference, GC–MS only reports the presence of cocaine. It is noteworthy that MDMA and ketamine both possess amine protons and the 1H NMR chemical shift is concentration-dependent (see Figure for the variation of the amine protons of MDMA at ca. δ 9.1). Thus, the algorithm developed was able to successfully deal with this and still match the 1H NMR spectrum to that of the reference spectrum. The use of deuterated DMSO also prevents this amine proton from being deuterated, and thus the molecule is not needed to be chemically changed to facilitate its successful detection. The median match scores for the other single component samples are shown in Table .
Table 1

Median Match Scores for the Detection of Single Component Mixtures (Inclusive of Null Responses)a

compound detectedmedian match score
MDMA0.986 (171)
6-Br-MDMA0.896 (2)
ketamine0.980 (74)
cocaine0.973 (85)
benzocaine0.960 (4)
mephedrone0.909 (4)
paracetamol0.939 (4)
ethylone0.939 (2)
heroin0.858 (3)
caffeine0.855 (1)
diazepam0.959 (2)
cannabis0.937 (2)
5F-ADB0.899 (20)

The number of samples identified of each type are indicated in parentheses.

The number of samples identified of each type are indicated in parentheses. The detection of 6-Br-MDMA in eight samples (two as the sole component and six as a binary mixture with magnesium stearate) highlights a way in which the method can be used to identify a drug class, despite the material not being in the reference library. Initial analysis of these eight samples indicated that the material was a possible unknown—all the match scores were below the validation threshold as they ranged from 0.755 to 0.644. Five samples reported the sample consisting of an amphetamine (MDMA), whilst the other three were indicated to be N-ethylpentylone hydrochloride, or a precursor to its production, piperonal. Despite the range of materials suggested, there is commonality, in that all of the compounds possess a 1,2-methylenedioxybenzene ring, with the benzene ring possessing a substituted aliphatic chain or functional group. Analysis of the GC–MS traces for these eight compounds identified a single component that had a retention time of 6.7 min. Subsequent analysis of the samples using high-field 1H and 13C NMR spectroscopy and MS in conjunction with previously reported[36] MS data revealed the structure to be 6-Br-MDMA. A sample was synthesized and the reference library updated accordingly with its 1H NMR spectrum. This enabled two of the samples to be identified as 6-Br-MDMA with a median match-score of 0.896, although a further reference sample needed to be created to correctly identify the other six, due to the presence of magnesium stearate. These samples are discussed further in the mixtures section. The detection of diazepam in two samples required a modification of the general procedure to enable its analysis by 1H NMR spectroscopy. In general benzodiazepines such as diazepam are produced as tablets which contain a large concentration of lactose. When part of the tablet is dissolved in DMSO both the API and excipient are soluble. The lactose is in such a high concentration that it swamps the signals from the diazepam. The subsequent result from matching the fingerprint is usually lactose. By contrast, the benzodiazepines are soluble in acetonitrile while the lactose is practically insoluble. Repeating the sample prep using deuterated acetonitrile in place of DMSO-d6 allowing the 1H NMR data acquisition and analysis to be performed under the same conditions as the other samples analyzed. The corresponding median match score for these two samples was 0.959. There are other misused pharmaceuticals which may present a similar issue with the analysis, for example, buprenorphine. The procedure using deuterated acetonitrile is also applicable in these cases. Of the 416 samples exhibiting GC–MS peaks, 24 samples were identified as synthetic cannabinoids, of which one was a mixture. The synthetic cannabinoids needed extracting from the plant material and were not seized as crystalline materials. All of the samples were typically less than 1.5% API by weight (see Supporting Information for details on the experimental protocol used to determine this) with only three samples possessing more than this threshold (highest 3.8%). Twenty samples were found to comprise 5F-ADB as the sole API. We note here that some of match scores for the samples were very similar to other materials in the database. For example, four samples had match scores that differed by only 0.001–0.005 compared to that of buprenorphine hydrochloride whilst another two samples were 0.003–0.008 different to MMB-CHMINACA. A further three samples were indicated to be 5F-ADB by GC–MS but gave a null response in terms of the 1H NMR analysis. The GC–MS traces for these three samples indicated that the API was 0.3–0.7%. It is notable that of the four samples that did not produce a match between the GC–MS result and that of the 1H NMR analysis, three of them were synthetic cannabinoid samples.

Mixtures

The algorithm developed is capable of detecting binary mixtures, which is commonplace for drugs (e.g. cocaine) as they are “cut” with another material. Of the samples analyzed, thirteen samples were identified as being a binary mixture by both GC–MS and 1H NMR, with a further 26 samples being identified as being a partial match, that is, one component not identified in the 1H NMR spectrum when two or more are present in the GC–MS trace, or vice versa. Of the thirteen binary mixtures correctly identified, 69% of these were cocaine based, the other 31% were amphetamine and caffeine (two instances) and benzocaine and ketamine (two instances). The cocaine samples were cut with caffeine, levamisole, benzocaine, or phenacetin. Four samples were identified as being a mixture of cocaine and levamisole by GC–MS (retention times 8.22 and 7.50 min respectively) and also by 1H NMR analysis (median match score 0.885). Following analysis of the gas chromatogram, the amount of levamisole in these samples was observed to range from 3.6 to 17.6%. The success of the 1H NMR analysis is particularly noteworthy here, given that levamisole and cocaine have similar aromatic moieties and therefore overlap in the 1H NMR spectrum. The class region is key to distinguishing between the two different components in these cases, whilst also facilitating the identification of both components present. One instance of a cocaine and phenacetin mixture was identified by 1H NMR (match score 0.837) and validated by the GC–MS data. Analysis of the gas chromatogram revealed that cocaine was prevalent in this mixture (79.4% cocaine and 20.6% phenacetin). The relative amount of cocaine in the single instance of cocaine mixed with caffeine was 90.2%. The match score for this sample was 0.922. A similar median match-score (0.849) was obtained for the two cocaine and benzocaine samples despite the composition of cocaine varying from 24.7 to 80.9%. Two samples were identified as amphetamine and caffeine (1H NMR median match score 0.890). From GC analysis, the relative amount of amphetamine present ranged from 23.3 to 36.7%. Similarly, the two samples that returned ketamine and benzocaine by both 1H NMR and GC–MS analysis had a 1H NMR median match score of 0.919.

Partial Matches

26 samples were identified as partial matches. Six of these that contained 6-Br-MDMA, in its hydrochloride form, were found to also consist of magnesium stearate. In the absence of 6-Br-MDMA, magnesium stearate does not dissolve in DMSO, as evidenced by the lack of peaks in the 1H NMR spectrum. However, it was observed that if magnesium stearate was added to a sample of 6-Br-MDMA hydrochloride in DMSO, a new peak was observed at 1.21 ppm in the 1H NMR spectrum with concomitant loss of the amine peak at 9.23 ppm (see Supporting Information for 1H NMR spectra). It is believed that a disproportionation reaction had occurred, leading to the exchange of the chloride for stearate which resulted in magnesium chloride being formed. The practice of increasing the water solubility of drugs using disproportionation is well-documented.[37−40] Some samples were more heavily cut with magnesium stearate than others, and thus a reference 1H NMR spectrum of 6-Br-MDMA with magnesium stearate was required. A suitable sample was prepared, a 1H NMR spectrum acquired, and the reference library was updated accordingly. Subsequently, the median match score for the six seized samples was 0.926. These were recorded as partial matches because magnesium stearate cannot be observed under the GC–MS conditions employed and so could not be validated. Ten cocaine binary mixtures were recorded as partial matches, with a further three that were found to be tertiary mixtures and one that comprised six components. One of the tertiary mixtures by GC–MS proved to comprise benzocaine, paracetamol, and cocaine in a ratio of 6.6:15.8:77.6. The 1H NMR analysis, which is limited to reporting on binary mixtures only, returned paracetamol and cocaine as the match, with a match score of 0.926. Benzocaine was not reported, most likely due to the low concentration present in the sample. Similarly, another tertiary mixture was identified to be cocaine, benzocaine, and levamisole by GC–MS and returned cocaine and benzocaine as the match in the 1H NMR analysis with a match score of 0.869. As levamisole is only present at a level of 0.4%, it is not surprising that the 1H NMR analysis identified the two most prevalent components. The final tertiary mixture was found to consist of cocaine, ketamine, and benzocaine in a 35.6:35.7:28.7 ratio. The match scores for benzocaine and cocaine, and benzocaine and ketamine, differ by only 0.007 for this sample being 0.856 and 0.849, respectively. Thus, despite the algorithm being limited to detecting two components only, appraisal of the match scores can provide evidence for the presence of three components in a mixture. The six-component mixture consisted of cocaine, levamisole, hexadecanoic acid, tolycaine, heptan-3-ol, and phenacetin with cocaine (87.9%) and phenacetin (7.4%) being the most prevalent. The 1H NMR analysis reported cocaine and phenacetin with a match-score of 0.862. It is noteworthy that the other components were all present at a level of 3% or lower, and this will have hindered their detection in the 1H NMR spectrum. Of the ten cocaine binary mixtures, three samples did not return cocaine as the primary identified API. All of these samples comprised largely of benzocaine (>79%), and hence, the 1H NMR spectrum was dominated by the signals of benzocaine. Furthermore, the lower molecular weight of benzocaine compared to cocaine results in lower signal intensity of the latter in the 1H NMR spectrum, also hindering its detection. A further six samples were cocaine and levamisole binary mixtures. For five of these samples, levamisole comprised <2% of the sample according to GC–MS analysis (with one exception where the levamisole was present at the 5% level), and this might have prevented its detection in the 1H NMR analysis. The other sample indicated levamisole as an additional component in the 1H NMR analysis. Although this represents a false-positive hit, the important aspect is that cocaine is detected in the sample as this is the main API. The final cocaine containing sample was shown to be cut with caffeine but only at the level of 1%. Again, this low amount is not conducive to its detection by 1H NMR spectroscopy. The remaining six samples were indicated by 1H NMR to have one too few or one too many components compared to the GC–MS trace. One of these samples returned only AMB-FUBINACA by 1H NMR whilst the GC–MS showed it comprised the former as well as 5F-ADB. Both components on a weight-by-weight basis were <2.5%, with AMB-FUBINACA being present at almost five times the amount of 5F-ADB. Another sample was shown to comprise alpha-d-lactose and caffeine whereas GC–MS reported only caffeine; this is as expected, as lactose would need to be derivatized in order to be detectable under the GC conditions employed. Another sample of note here is one that comprised methiopropamine and 2-aminoindane by GC–MS but was determined to be methiopropamine and amphetamine sulfate following 1H NMR analysis. Despite the difference in compositions, it must be noted that 2-aminoindane is the rigid analogue of amphetamine and hence has similar 1H NMR resonances; thus there is a scientific rationale for its detection following 1H NMR analysis.

Conclusions

There is a need for a high-throughput analytical technique with facile sample preparation to deal with increasing numbers of drug seizures made by law agencies. This paper presents an approach comprising an automated NMR benchtop spectrometer, which acquires, processes, and interprets the spectral data with minimum user knowledge. The findings of the analysis are reported in around 5 min. The approach makes use of an algorithm that analyses the acquired 1H NMR spectrum by matching it to a library of over 300 reference spectra from known substances. 432 seized samples were tested using this system, and the outcomes compared with GC–MS data, a currently accepted reference method (see Table ). The GC–MS trace in thirteen cases (3% of the samples surveyed) were devoid of any peaks, so the 1H NMR analysis could not be validated for these samples. A further three samples were found to lack any API or cutting agent after both 1H NMR and GC–MS analysis.
Table 2

Analysis Breakdown for the 432 Samples Surveyed as Part of This Worka

no. of samples432
% of samples that could not be validated3.0 (13)
% samples found to contain no API or cutting agent0.7 (3)
samples that do not match0.9 (4)
% correctly matched (single component mixtures)86.6 (374)
% correctly matched (two-component samples)3.0 (13)
% partially matched (two or more component samples)5.8 (25)

The number in parentheses indicate the number of samples of that type.

The number in parentheses indicate the number of samples of that type. For the remaining 416 samples, the outcomes by GC–MS and 1H NMR matched in 93% of cases. Of these, thirteen samples were identified as being a binary mixture by both GC–MS and 1H NMR. A partial match (one component not identified) was obtained for 6% of samples surveyed, whilst only 1% of samples resulted in contradictory outcomes. The analytical predictive power of the algorithm was exemplified in the detection of 6-Br-MDMA which identified the methylenedioxy ring structure and the class of compound. Further characterization of the samples, and the synthesis of a reference sample, led to the complete elucidation of the chemical structure. Note that the disproportionation of chloride for stearate from magnesium stearate present in the majority of seized samples of this type required an additional reference 1H NMR spectrum to be acquired and the spectral library updated accordingly. Eight samples were found to possess 6-Br-MDMA as the API, with six of these being heavily cut with magnesium stearate, as indicated by the match scores. The protocol was modified for the detection of benzodiazepines, for example, diazepam. Deuterated acetonitrile was used instead, but the methodology for detection, acquisition, processing, and reporting remained consistent with that of the samples collected in deuterated DMSO. Having demonstrated the validity of the approach, future work will be directed toward a quantification method to widen the applicability of the qualitative approach employed here, and to improve the ability to handle multicomponent mixtures.

Methodology

Materials

302 reference compounds were synthesized (see below) or obtained either from Fluorochem (UK), Sigma Aldrich (UK), BRC Fine Chemicals (https://www.brc-finechemicals.com/—prior to the UK Psychoactive Substances Act (2016), which came in to effect on the 26th May 2016 and made it an offence to manufacture, supply, possess, and import and export psychoactive substances named in this Act) or Carbosynth (UK). The following groups of chemicals were synthesized according to, or adapted from, the literature references cited: 4-methoxyamphetamines,[41] 2-, 3- and 4-fluoroamphetamines,[42] thiopropamine,[43] methiopropamine,[44] dimethyldiphenidines,[23] 3- and 4-chloroethcathinone,[45] trifluoromethylmethcathinones,[45] 3- and 4-chloro N,N-diethylcathinones,[46] 2-, 3- and 4-methcathinones,[46] 2-, 3- and 4-fluoromethcathinones,[47] diphenidine and ephenidine derivatives,[23] fentanyl derivatives,[48] fluorolintanes,[23] 25I-,25B-, 25C-, and 25F-NBOMe,[49] MMB-CHMINACA, MMB-CHMICA, 5F-ADB, AMB-FUBINACA, and AB-FUBINACA,[50] 2C-C, 2C-B, 2C-I, 2C-F, and 2C-H.[51] All synthesized reference standards were fully characterized and exhibited physical and spectral data consistent with the literature and determined to be >99% pure by high-field NMR. These consisted of a range of narcotics, NPS, and controlled substances, as well as commonly encountered non-controlled substances and adulterants. In total, there were 21 classes of compound: aminoindanes, amphetamines, atypical diphenidines, benzodiazepines, cathinones, diphenidines, ephenidines, fentanyls, fluorolintanes, methylenedioxycathinones, NBOMe, narcotics, pesticides, pharmaceuticals, phenethylamines, piperazines (including MT-45 and its derivatives), precursors, steroids, and synthetic cannabinoids. Two further “classes” of non-controlled substances were also included: a range of “cutting agents” and “background residues” (i.e., only DMSO). 432 seized samples of unknown identity were obtained from Greater Manchester Police via the MANchester DRug Analysis & Knowledge Ex-change (MANDRAKE) partnership and were stored and analyzed in accordance with the UK Misuse of Drugs Act (1971) and Misuse of Drugs Regulations (2001). The samples were collected over the period January 2017–August 2018. All of these samples were supplied in their solid, bulk forms. For the seized materials, a micro-spatula tip of the material (ca. 5–10 mg) was dissolved in 0.6 mL of deuterated DMSO. If diazepam was suspected, then 0.6 mL of deuterated acetonitrile was used instead. For spice samples, approximately 50 mg of material was weighed out and 0.6 mL of deuterated DMSO added. The spice material was heavily agitated to ensure that the API became dissolved in the DMSO. All samples were filtered through a 0.45 μm polyvinylidene difluoride syringe filter (Whatman) directly into a NMR tube. Samples were analyzed by 1H NMR and validated against contemporaneously acquired GC–MS data (GC–MS method outlined in Supporting Information). The reference samples were prepared for NMR analysis by dissolving 10 mg of compound per 1 mL of deuterated DMSO, or in the case of diazepam, deuterated MeCN. 130 compounds from the reference collection were available in sufficient quantity to prepare a replicate sample. All but three of the controlled substance classes (NBOMe, 2C-derivatives and piperazines) were represented in this validation collection. A further 32 samples of non-controlled substances in the reference collection were prepared in pairwise combinations, to assist with development of the mixture identification method.

NMR Data Acquisition

1H NMR spectra were acquired of all samples using a Pulsar benchtop NMR spectrometer (Oxford Instruments, Abingdon, UK) operating at a frequency of 59.7 MHz. The temperature of the probe was calculated to be 308.5 K by measuring the separation (in Hz, Δδ) between the CH2 and OH signals of neat ethylene glycol and implementing the equation T [K] = 466.5 – 102.00Δδ.[52] After the sample had been inserted, an automated procedure began whereby the instrument would lock on to the deuterated signature of DMSO (thus used as a chemical shift reference) before acquiring 16 scans (for the spice samples, 64 scans were acquired). Following acquisition, the data were processed in MNova (Mestrelab Research, Santiago de Compostela, Spain) using an automated script file. The processed FID was then analyzed by the pattern recognition algorithm, developed in-house using Matlab (The Mathworks Inc, Cambridge, UK). The algorithm employs a minimum distance classifier. The multivariate distance between the sample spectrum and each of the reference spectra is calculated. The sample is identified as the nearest reference compound, provided the “match score” (equal to one minus the distance) exceeds an (empirically determined) threshold; if it does not, then the outcome is considered to be tentative, unreliable, or unknown. Binary mixtures are accommodated by extending the pattern search with synthetically generated mixture spectra of pairwise combinations of the reference library. The collection of NMR data and the subsequent analysis takes approximately 5 min. If signal-to-noise was found to be poor, more scans were obtained. This was only required in less than 1% of the samples being analyzed, and this was due to the amount of sample available being low, for example, the residue material within plastic packets used to transport illicit materials.
  35 in total

1.  Synthesis, full chemical characterisation and development of validated methods for the quantification of the components found in the evolved "legal high" NRG-2.

Authors:  Osama I G Khreit; Craig Irving; Elke Schmidt; John A Parkinson; Niamh Nic Daeid; Oliver B Sutcliffe
Journal:  J Pharm Biomed Anal       Date:  2011-11-11       Impact factor: 3.935

2.  Test purchase, synthesis, and characterization of 2-methoxydiphenidine (MXP) and differentiation from its meta- and para-substituted isomers.

Authors:  Gavin McLaughlin; Noreen Morris; Pierce V Kavanagh; John D Power; John O'Brien; Brian Talbot; Simon P Elliott; Jason Wallach; Khoa Hoang; Hamilton Morris; Simon D Brandt
Journal:  Drug Test Anal       Date:  2015-04-15       Impact factor: 3.345

3.  Synthesis, full chemical characterisation and development of validated methods for the quantification of (±)-4'-methylmethcathinone (mephedrone): a new "legal high".

Authors:  Eman Y Santali; Anna-Karina Cadogan; Niamh Nic Daeid; Kathleen A Savage; Oliver B Sutcliffe
Journal:  J Pharm Biomed Anal       Date:  2011-05-23       Impact factor: 3.935

4.  An analysis of legal highs: do they contain what it says on the tin?

Authors:  Mark Baron; Mathieu Elie; Leonie Elie
Journal:  Drug Test Anal       Date:  2011-05-19       Impact factor: 3.345

5.  Synthesis of (R)-(-)- and (S)-(+)-4-fluorodeprenyl and (R)-(-)- and (S)-(+)-[N-11C-methyl]-4-fluorodeprenyl and positron emission tomography studies in baboon brain.

Authors:  A Plenevaux; S L Dewey; J S Fowler; M Guillaume; A P Wolf
Journal:  J Med Chem       Date:  1990-07       Impact factor: 7.446

6.  Keeping pace with NPS releases: fast GC-MS screening of legal high products.

Authors:  Mathieu P Elie; Leonie E Elie; Mark G Baron
Journal:  Drug Test Anal       Date:  2013-01-07       Impact factor: 3.345

7.  Synthesis and structure-activity relationships of N-benzyl phenethylamines as 5-HT2A/2C agonists.

Authors:  Martin Hansen; Karina Phonekeo; James S Paine; Sebastian Leth-Petersen; Mikael Begtrup; Hans Bräuner-Osborne; Jesper L Kristensen
Journal:  ACS Chem Neurosci       Date:  2014-01-15       Impact factor: 4.418

8.  Formulating weakly basic HCl salts: relative ability of common excipients to induce disproportionation and the unique deleterious effects of magnesium stearate.

Authors:  Christopher T John; Wei Xu; Lisa K Lupton; Paul A Harmon
Journal:  Pharm Res       Date:  2013-03-20       Impact factor: 4.200

9.  Fluoromethcathinone, a new substance of abuse.

Authors:  R P Archer
Journal:  Forensic Sci Int       Date:  2009-02-04       Impact factor: 2.395

10.  An efficient, optimized synthesis of fentanyl and related analogs.

Authors:  Carlos A Valdez; Roald N Leif; Brian P Mayer
Journal:  PLoS One       Date:  2014-09-18       Impact factor: 3.240

View more
  5 in total

1.  City checking: Piloting the UK's first community-based drug safety testing (drug checking) service in 2 city centres.

Authors:  Fiona Measham
Journal:  Br J Clin Pharmacol       Date:  2020-02-17       Impact factor: 4.335

Review 2.  Interpol review of controlled substances 2016-2019.

Authors:  Nicole S Jones; Jeffrey H Comparin
Journal:  Forensic Sci Int Synerg       Date:  2020-05-24

3.  Synthesis, characterization, and evaluation of selective molecularly imprinted polymers for the fast determination of synthetic cathinones.

Authors:  Hong Chen; Fangsheng Wu; Yibing Xu; Yuan Liu; Lun Song; Xiujuan Chen; Qun He; Wei Liu; Qiaoying Han; Zihua Zhang; Yun Zou; Wenbin Liu
Journal:  RSC Adv       Date:  2021-09-06       Impact factor: 4.036

4.  Characterization of new psychoactive substances by integrating benchtop NMR to multi-technique databases.

Authors:  Thomas Castaing-Cordier; Alejandra Benavides Restrepo; Damien Dubois; Virginie Ladroue; Fabrice Besacier; Audrey Buleté; Céline Charvoz; Anais Goupille; Denis Jacquemin; Patrick Giraudeau; Jonathan Farjon
Journal:  Drug Test Anal       Date:  2022-06-26       Impact factor: 3.234

5.  High-throughput screening of argan oil composition and authenticity using benchtop 1 H NMR.

Authors:  Yvonne Gunning; Alex J Jackson; Joshua Colmer; Fouad Taous; Mark Philo; Rachel M Brignall; Tibari El Ghali; Marianne Defernez; E Kate Kemsley
Journal:  Magn Reson Chem       Date:  2020-04-06       Impact factor: 2.447

  5 in total