| Literature DB >> 31179187 |
Patorn Piromchai1, Charoiboon Puvatanond1, Virat Kirtsreesakul2, Saisawat Chaiyasate3, Sanguansak Thanaviratananich1.
Abstract
BACKGROUND: Nasal irrigation is widely used as an adjunctive treatment for nasal diseases. There is little evidence regarding the efficacy of the devices used in this procedure. The objective of this survey was to evaluate the effectiveness of nasal irrigation devices based on the experiences of patients and physicians.Entities:
Keywords: Allergic rhinitis; Cold; Devices; Nasal irrigation; Sinusitis
Year: 2019 PMID: 31179187 PMCID: PMC6542345 DOI: 10.7717/peerj.7000
Source DB: PubMed Journal: PeerJ ISSN: 2167-8359 Impact factor: 2.984
Symptom scores for high-pressure devices.
| 1. Improve overall symptom | 7.45 ± 2.12 | 8.46 ± 1.60 | 1.02 | 0.008 | 0.27 to 1.7 |
| 2. Improve nasal congestion | 7.33 ± 2.32 | 8.28 ± 1.75 | 0.95 | 0.025 | 0.12 to 1.78 |
| 3. Decrease runny nose | 7.23 ± 2.27 | 8.06 ± 1.88 | 0.82 | 0.061 | −0.04 to 1.69 |
| 4. Decrease blowing nose | 7.23 ± 2.24 | 7.97 ± 2.17 | 0.74 | 0.123 | −0.20 to 1.68 |
| 5. Decrease viscosity | 7.50 ± 1.84 | 8.32 ± 1.89 | 0.82 | 0.051 | −0.01 to 1.64 |
| 6. Improve sinus pain/headache | 6.26 ± 2.66 | 7.81 ± 2.07 | 1.55 | 0.010 | 0.37 to 2.72 |
| 7. Decrease post-nasal drip | 6.13 ± 2.70 | 7.89 ± 2.23 | 1.76 | 0.003 | 0.63 to 2.90 |
| 8. Improve taste and smell | 6.35 ± 2.66 | 7.79 ± 2.60 | 1.44 | 0.038 | 0.08 to 2.79 |
| 9. Decrease sneezing | 6.83 ± 2.71 | 7.56 ± 2.31 | 0.74 | 0.224 | −0.46 to 1.93 |
| 10. Decrease cough | 6.50 ± 2.79 | 7.17 ± 2.70 | 0.68 | 0.412 | −0.96 to 2.31 |
| 11. Clear the secretion | 6.30 ± 2.74 | 8.36 ± 1.95 | 2.05 | <0.001 | 1.03 to 3.08 |
| 12. Improve sleep quality | 7.1 ± 2.49 | 8.35 ± 1.88 | 1.25 | 0.020 | 0.20 to 2.30 |
Notes.
p < 0.05.
confidence interval
Ease of use, learning curve, and satisfaction scores of high-pressure devices.
| 1. Ease of use | 8.61 ± 1.99 | 8.83 ± 1.74 | 0.23 | 0.567 | −0.57 to 1.03 |
| 2. Learning curve | 8.71 ± 1.84 | 8.99 ± 1.49 | 0.27 | 0.444 | −0.43 to 0.97 |
| 3. Satisfaction score | 8.19 ± 1.96 | 8.94 ± 1.85 | 0.76 | 0.075 | −0.08 to 1.60 |
| 4. Would recommend the device to others | 7.86 ± 2.58 | 9 ± 1.75 | 0.45 | 0.012 | 0.26 to 2.03 |
Notes.
p < 0.05.
confidence interval
Adverse events of high-pressure devices.
| 1. Pain/discomfort | 2 (6.4%) | 1 (1.2%) | −5.5% | 0.140 | −14.16 to 3.86 |
| 2. Retained fluid in sinuses | 4 (12.9%) | 23 (29.9%) | 16.97% | 0.066 | 1.35 to 32.8 |
| 3. Epistaxis | 0 (0%) | 1 (1.3%) | 1.30% | 0.524 | −1.23 to 3.83 |
| 4. Headache | 0 (0%) | 1 (1.3%) | 1.30% | 0.524 | −1.23 to 3.83 |
| 5. Aspiration | 0 (0%) | 5 (6.5%) | 6.49% | 0.146 | 0.9 to 11.00 |
| 6. Ear pain/hearing loss | 1 (3.2%) | 8 (10.4%) | 7.16% | 0.223 | −2.06 to 16.39 |
| 7. Salty taste | 2 (6.4%) | 11 (14.3%) | 7.83% | 0.258 | −3.82 to 19.49 |
Notes.
p < 0.05.
confidence interval
Symptom scores for large-volume devices.
| 1. Improve overall symptom | 7.84 ± 1.88 | 8.46 ± 1.60 | 0.62 | 0.011 | 0.15 to 1.10 |
| 2. Improve nasal congestion | 7.69 ± 1.97 | 8.28 ± 1.76 | 0.59 | 0.024 | 0.08 to 1.11 |
| 3. Decrease runny nose | 7.51 ± 2.07 | 8.06 ± 1.88 | 0.55 | 0.050 | −0.00 to 1.10 |
| 4. Decrease blowing nose | 7.60 ± 2.21 | 7.97 ± 2.17 | 0.38 | 0.217 | −0.22 to 0.98 |
| 5. Decrease viscosity | 7.89 ± 2.06 | 8.32 ± 1.89 | 0.43 | 0.119 | −0.11 to 0.97 |
| 6. Improve sinus pain/headache | 6.72 ± 2.75 | 7.81 ± 2.07 | 1.09 | 0.007 | 0.30 to 1.88 |
| 7. Decrease post-nasal drip | 7.30 ± 2.23 | 7.89 ± 2.23 | 0.59 | 0.067 | −0.04 to 1.23 |
| 8. Improve taste and smell | 6.90 ± 2.86 | 7.79 ± 2.60 | 0.89 | 0.044 | 0.02 to 1.75 |
| 9. Decrease sneezing | 7.46 ± 2.34 | 7.56 ± 2.31 | 0.10 | 0.778 | −0.61 to 0.81 |
| 10. Decrease cough | 7.13 ± 2.48 | 7.17 ± 2.70 | 0.04 | 0.916 | −0.77 to 0.86 |
| 11. Clear the secretion | 7.94 ± 1.96 | 8.36 ± 1.95 | 0.41 | 0.126 | −0.12 to 0.95 |
| 12. Improve sleep quality | 8.01 ± 2.13 | 8.35 ± 1.88 | 0.34 | 0.268 | −0.27 to 0.95 |
Notes.
p < 0.05.
confidence interval
Ease of use, learning curve, and satisfaction scores of large-volume devices.
| 1. Ease of use | 8.40 ± 2.07 | 8.84 ± 1.74 | 0.44 | 0.104 | −0.09 to 0.97 |
| 2. Learning curve | 8.43 ± 2.03 | 8.98 ± 1.49 | 0.56 | 0.031 | 0.05 to 1.07 |
| 3. Satisfaction score | 8.56 ± 2.06 | 8.95 ± 1.85 | 0.39 | 0.155 | −0.15 to 0.92 |
| 4. Would recommend the device to others | 8.33 ± 2.30 | 9 ± 1.75 | 0.67 | 0.024 | 0.09 to 1.25 |
Notes.
p < 0.05.
confidence interval
Adverse events of large-volume devices.
| 1. Pain/discomfort | 16 (7.2%) | 1 (1.2%) | −5.88% | 0.055 | −10.10 to −1.65 |
| 2. Retained fluid in sinuses | 29 (13.0%) | 23 (29.9%) | 16.87% | <0.001 | 5.73 to 28.00 |
| 3. Epistaxis | 6 (2.7%) | 1 (1.3%) | −1.39% | 0.486 | −4.69 to 1.91 |
| 4. Headache | 14 (6.3%) | 1 (1.3%) | −4.98% | 0.084 | −9.05to -0.91 |
| 5. Aspiration | 25 (11.2%) | 5 (6.5%) | −4.71% | 0.234 | −11.60 to 2.17 |
| 6. Ear pain/hearing loss | 23 (10.3%) | 8 (10.4%) | 0.08% | 0.985 | −7.82 to 7.97 |
| 7. Salty taste | 34 (15.2%) | 11 (14.3%) | −0.96% | 0.839 | −10.09 to 8.17 |
Notes.
p < 0.05.
confidence interval