Experimental and Numerical Studies on a Method to Mitigate Screening Current-Induced Field for No-Insulation REBCO Coils.

Jiho Lee1, Dongkeun Park1, Yi Li1, Yoon Hyuck Choi1, Philip C Michael1, Juan Bascuñân1, Yukikazu Iwasa1.   

Abstract

We present experimental and numerical studies on a method to mitigate screening current-induced field (SCF) for NI REBCO coil. The SCF is the major field error to incorporate a REBCO insert for a high field LTS/HTS magnet. The field-shaking technique is going to be used to mitigate the SCF of 800-MHz REBCO insert magnet (H800) for MIT 1.3-GHz LTS/HTS NMR magnet (1.3 G). The field-shaking using 500-MHz LTS background magnet generates the SCF in H800, due to huge self and mutual inductances of them. In this paper, we tested the effect of the induced current in the NI REBCO coil on the field-shaking technique to mitigate the SCF. The amount of the induced current was decided by the NI REBCO coil status; the open- or closed-loop coil. We performed the three cases of experimental tests and analyzed them. From the test results, we may conclude that we need to limit the ramp rate of L500 during the field-shaking, to minimize the induced current in the HTS insert which consists of the NI REBCO coil.

Entities:  

Keywords:  Field-shaking; HTS insert; nuclear magnetic resonance (NMR) magnet; screening current-induced field (SCF)

Year:  2019        PMID: 31178651      PMCID: PMC6550491          DOI: 10.1109/TASC.2019.2906221

Source DB:  PubMed          Journal:  IEEE Trans Appl Supercond


  3 in total

1.  Why an ac magnetic field shifts the irreversibility line in type-II superconductors.

Authors:  Ernst Helmut Brandt; Grigorii P Mikitik
Journal:  Phys Rev Lett       Date:  2002-06-20       Impact factor: 9.161

2.  A Field-Shaking System to Reduce the Screening Current-Induced Field in the 800-MHz HTS Insert of the MIT 1.3-GHz LTS/HTS NMR Magnet: A Small-Model Study.

Authors:  Jiho Lee; Dongkeun Park; Philip C Michael; So Noguchi; Juan Bascuñán; Yukikazu Iwasa
Journal:  IEEE Trans Appl Supercond       Date:  2018-02-08

3.  Construction and Test Results of Coils 2 and 3 of a 3-Nested-Coil 800-MHz REBCO Insert for the MIT 1.3-GHz LTS/HTS NMR Magnet.

Authors:  Dongkeun Park; Juan Bascuñán; Philip C Michael; Jiho Lee; Seungyong Hahn; Yukikazu Iwasa
Journal:  IEEE Trans Appl Supercond       Date:  2017-12-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.