| Literature DB >> 31168608 |
László Gellér1, Zoltán Salló1, Levente Molnár1, Tamás Tahin2, Emin Evren Özcan3, Valentina Kutyifa1,4, István Osztheimer1, Szabolcs Szilágyi1, Nándor Szegedi1, Pál Ábrahám1, Astrid Apor1, Klaudia Vivien Nagy1, Annamária Kosztin1, Dávid Becker1, Szilvia Herczeg1, Endre Zima1, Béla Merkely1.
Abstract
AIMS: The aim of our study was to investigate the long-term efficacy and safety of transseptal endocardial left ventricular lead implantation (TELVLI). METHODS ANDEntities:
Keywords: Cardiac resynchronization therapy; Endocardial; Heart failure; Lead; Transseptal
Year: 2019 PMID: 31168608 PMCID: PMC6680368 DOI: 10.1093/europace/euz116
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Patient baseline characteristics
| Patient characteristics | All patients ( |
|---|---|
| General | |
| Age at enrolment (years) | 69 ± 10 |
| Gender (male), | 44 (81) |
| Failed implant, | 25 (46) |
| Suboptimal CS anatomy, | 20 (37) |
| CRT non-responders, | 9 (17) |
| Cardiovascular history, | |
| Ischaemic cardiomyopathy | 21 (39) |
| Non-ischaemic cardiomyopathy | 33 (61) |
| Atrial fibrillation | 20 (37) |
| Baseline assessment, | |
| NYHA class | |
| Class III | 50 (93) |
| Class IV | 4 (7) |
| QRS duration (ms) | 158 ± 28 |
| LVEF (%) | 29 ± 6 |
| LVESD (mm) | 57 ± 10 |
| LVEDD (mm) | 68 ± 8 |
| Mitral regurgitation (moderate/severe), | 33 (61) |
Data are shown as mean ± SD, or as numbers and percentages.
CRT, cardiac resynchronization therapy; CS, coronary sinus; LVEF, left ventricular ejection fraction; LVESD, left ventricular end-systolic diameter; LVEDD, left ventricular end-diastolic diameter; NYHA, New York Heart Association.
Implantation characteristics
| Device type, | |
| CRT-P | 19 (35) |
| CRT-D | 35 (65) |
| Lead type, | |
| Medtronic 4076 | 10 (19) |
| Medtronic 5076 | 20 (37) |
| Medtronic 3830 | 18 (33) |
| St. Jude Medical 1888 | 6 (11) |
| Electrical parameters at implantation | |
| Sensing (mV) | 14 ± 7 |
| Threshold (mV) | 0, 9 ± 0, 4 |
| Impedance (Ω) | 713 ± 256 |
| Lead implantation positions, | |
| Posterolateral | 40 (74) |
| Lateral | 14 (26) |
Data are shown as mean ± SD, or as numbers and percentages.
CRT-P, cardiac resynchronization therapy pacemaker; CRT-D, cardiac resynchronization therapy defibrillator.
Thromboembolic complications
| Pt. no. | Sex | Age | Rhythm | Time to TE event (months) | Type of TE event | Lead type |
|---|---|---|---|---|---|---|
| 1 | M | 69 | Sinus | 46 | stroke | Medtronic 5076 |
| 2 | M | 63 | Sinus | 7 | TIA | St. Jude Medical 1888 |
| 3 | M | 61 | Sinus | 46 | stroke | Medtronic 5076 |
| 4 | M | 69 | AFib | 33 | TIA | Medtronic 4076 |
Age is given at TE event.
F, female; M, male; TE, thromboembolic; TIA, transient ischaemic attack.