Independent recalculation outperforms traditional measurement-based IMRT QA methods in detecting unacceptable plans.

Stephen F Kry1,2, Mallory C Glenn1,2, Christine B Peterson2,3, Daniela Branco1,2, Hunter Mehrens1, Angela Steinmann1,2, David S Followill1,2.   

Abstract

PURPOSE: To evaluate the performance of an independent recalculation and compare it against current measurement-based patient specific intensity-modulated radiation therapy (IMRT) quality assurance (QA) in predicting unacceptable phantom results as measured by the Imaging and Radiation Oncology Core (IROC).
METHODS: When institutions irradiate the IROC head and neck IMRT phantom, they are also asked to submit their internal IMRT QA results. Separately from this, IROC has previously created reference beam models on the Mobius3D platform to independently recalculate phantom results based on the institution's DICOM plan data. The ability of the institutions' IMRT QA to predict the IROC phantom result was compared against the independent recalculation for 339 phantom results collected since 2012. This was done to determine the ability of these systems to detect failing phantom results (i.e., large errors) as well as poor phantom results (i.e., modest errors). Sensitivity and specificity were evaluated using common clinical thresholds, and receiver operator characteristic (ROC) curves were used to compare across different thresholds.
RESULTS: Overall, based on common clinical criteria, the independent recalculation was 12 times more sensitive at detecting unacceptable (failing) IROC phantom results than clinical measurement-based IMRT QA. The recalculation was superior, in head-to-head comparison, to the EPID, ArcCheck, and MapCheck devices. The superiority of the recalculation vs these array-based measurements persisted under ROC analysis as the recalculation curve had a greater area under it and was always above that for these measurement devices. For detecting modest errors (poor phantom results rather than failing phantom results), neither the recalculation nor measurement-based IMRT QA performed well.
CONCLUSIONS: A simple recalculation outperformed current measurement-based IMRT QA methods at detecting unacceptable plans. These findings highlight the value of an independent recalculation, and raise further questions about the current standard of measurement-based IMRT QA.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  IMRT QA; Mobius ; patient-specific QA; recalculation

Year:  2019        PMID: 31152568      PMCID: PMC6692191          DOI: 10.1002/mp.13638

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Beam data modeling of linear accelerators (linacs) through machine learning and its potential applications in fast and robust linac commissioning and quality assurance.

Authors:  Wei Zhao; Ishan Patil; Bin Han; Yong Yang; Lei Xing; Emil Schüler
Journal:  Radiother Oncol       Date:  2020-10-08       Impact factor: 6.280

2.  Fiducial detection and registration for 3D IMRT QA with organ-specific dose information.

Authors:  Yi-Fang Wang; Olga Dona; Yuanguang Xu; John Adamovics; Cheng-Shie Wuu
Journal:  J Appl Clin Med Phys       Date:  2021-03-31       Impact factor: 2.102

3.  Survey results of 3D-CRT and IMRT quality assurance practice.

Authors:  Hunter Mehrens; Paige Taylor; David S Followill; Stephen F Kry
Journal:  J Appl Clin Med Phys       Date:  2020-04-30       Impact factor: 2.102

4.  What is the optimal input information for deep learning-based pre-treatment error identification in radiotherapy?

Authors:  Cecile J A Wolfs; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2022-08-27

5.  AAPM MEDICAL PHYSICS PRACTICE GUIDELINE 5.b: Commissioning and QA of treatment planning dose calculations-Megavoltage photon and electron beams.

Authors:  Mark W Geurts; Dustin J Jacqmin; Lindsay E Jones; Stephen F Kry; Dimitris N Mihailidis; Jared D Ohrt; Timothy Ritter; Jennifer B Smilowitz; Nicholai E Wingreen
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

6.  Evaluation of 4-Hz log files and secondary Monte Carlo dose calculation as patient-specific quality assurance for VMAT prostate plans.

Authors:  Philipp Szeverinski; Matthias Kowatsch; Thomas Künzler; Marco Meinschad; Patrick Clemens; Alexander F DeVries
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

7.  Independent calculation-based verification of volumetric-modulated arc therapy-stereotactic body radiotherapy plans for lung cancer.

Authors:  Tomohiro Ono; Takamasa Mitsuyoshi; Takashi Shintani; Yusuke Tsuruta; Hiraku Iramina; Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Yukinori Matsuo; Takashi Mizowaki
Journal:  J Appl Clin Med Phys       Date:  2020-05-11       Impact factor: 2.102

8.  Correlation between the γ passing rates of IMRT plans and the volumes of air cavities and bony structures in head and neck cancer.

Authors:  Zhengwen Shen; Xia Tan; Shi Li; Xiumei Tian; Huanli Luo; Ying Wang; Fu Jin
Journal:  Radiat Oncol       Date:  2021-07-21       Impact factor: 3.481

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.