OBJECTIVE: Recording of local field potentials (LFPs) during deep brain stimulation (DBS) is necessary to investigate the instantaneous brain response to stimulation, minimize time delays for closed-loop neurostimulation and maximise the available neural data. To our knowledge, existing recording systems lack the ability to provide artefact-free high-frequency (>100 Hz) LFP recordings during DBS in real time primarily because of the contamination of the neural signals of interest by the stimulation artefacts. APPROACH: To solve this problem, we designed and developed a novel, low-noise and versatile analog front-end (AFE) that uses a high-order (8th) analog Chebyshev notch filter to suppress the artefacts originating from the stimulation frequency. After defining the system requirements for concurrent LFP recording and DBS artefact suppression, we assessed the performance of the realised AFE by conducting both in vitro and in vivo experiments using unipolar and bipolar DBS (monophasic pulses, amplitude ranging from 3 to 6 V peak-to-peak, frequency 140 Hz and pulse width 100 µs). A full performance comparison between the proposed AFE and an identical AFE, equipped with an 8th order analog Bessel notch filter, was also conducted. MAIN RESULTS: A high-performance, 4 nV ([Formula: see text])-1 AFE that is capable of recording nV-scale signals was designed in accordance with the imposed specifications. Under both in vitro and in vivo experimental conditions, the proposed AFE provided real-time, low-noise and artefact-free LFP recordings (in the frequency range 0.5-250 Hz) during stimulation. Its sensing and stimulation artefact suppression capabilities outperformed the capabilities of the AFE equipped with the Bessel notch filter. SIGNIFICANCE: The designed AFE can precisely record LFP signals, in and without the presence of either unipolar or bipolar DBS, which renders it as a functional and practical AFE architecture to be utilised in a wide range of applications and environments. This work paves the way for the development of externalized research tools for closed-loop neuromodulation that use low- and higher-frequency LFPs as control signals.
OBJECTIVE: Recording of local field potentials (LFPs) during deep brain stimulation (DBS) is necessary to investigate the instantaneous brain response to stimulation, minimize time delays for closed-loop neurostimulation and maximise the available neural data. To our knowledge, existing recording systems lack the ability to provide artefact-free high-frequency (>100 Hz) LFP recordings during DBS in real time primarily because of the contamination of the neural signals of interest by the stimulation artefacts. APPROACH: To solve this problem, we designed and developed a novel, low-noise and versatile analog front-end (AFE) that uses a high-order (8th) analog Chebyshev notch filter to suppress the artefacts originating from the stimulation frequency. After defining the system requirements for concurrent LFP recording and DBS artefact suppression, we assessed the performance of the realised AFE by conducting both in vitro and in vivo experiments using unipolar and bipolar DBS (monophasic pulses, amplitude ranging from 3 to 6 V peak-to-peak, frequency 140 Hz and pulse width 100 µs). A full performance comparison between the proposed AFE and an identical AFE, equipped with an 8th order analog Bessel notch filter, was also conducted. MAIN RESULTS: A high-performance, 4 nV ([Formula: see text])-1 AFE that is capable of recording nV-scale signals was designed in accordance with the imposed specifications. Under both in vitro and in vivo experimental conditions, the proposed AFE provided real-time, low-noise and artefact-free LFP recordings (in the frequency range 0.5-250 Hz) during stimulation. Its sensing and stimulation artefact suppression capabilities outperformed the capabilities of the AFE equipped with the Bessel notch filter. SIGNIFICANCE: The designed AFE can precisely record LFP signals, in and without the presence of either unipolar or bipolar DBS, which renders it as a functional and practical AFE architecture to be utilised in a wide range of applications and environments. This work paves the way for the development of externalized research tools for closed-loop neuromodulation that use low- and higher-frequency LFPs as control signals.
Increasing evidence suggests that local field potential (LFP) oscillations in
the beta frequency band (13–30 Hz) can be consistently picked up in the
subthalamic nucleus (STN) of patients with Parkinson’s disease (PD) and that
their strength correlates with the severity of the disease and the efficacy of
therapy (Stanslaski , Little ). However, the last decade of LFP analysis also focused on spectral
power extraction from higher frequency bands, such as high gamma (60–80 Hz)
and 300 Hz (270–330 Hz) (Arlotti ). The power of these oscillations also correlates
with PD motor symptoms and clinical conditions, thus being eligible as a biomarker
(Priori ).As a result, activity in the aforementioned frequency bands during
stimulation could potentially be used to monitor disease progression, assess the
effects of therapy and direct patient treatment towards more effective therapeutical
strategies. Moreover, maintaining sensing during stimulation, rather than
eliminating the available neural data by simply blanking the signal chain during
stimulation, might also be vital for closed-loop neurostimulation systems.
Simultaneous neural recording and stimulation could help to maximise treatment
effectiveness for patients suffering from epilepsy. In an episodic disorder, such as
epilepsy, maintaining sensing during stimulation helps minimizing the temporal delay
between seizure detection and adaptation of the stimulation to achieve the most
effective therapy (Stanslaski ). Finally, observations during stimulation could also reveal novel
neural activity patterns that are not present in neural tissue in the absence of
stimulation. This could uncover new biomarkers of serious neurological disorders
previously masked by stimulation.However, the large difference between the amplitude of the stimulation pulses
and the relevant underlying neural activity leads to the appearance of stimulation
artefacts, which impede the accurate recording of neural signals and the processing
of potential biomarkers. More specifically, the normal amplitude of LFP signals can
range from a few microvolts (e.g. in the basal ganglia) (Goldberg 2004) to hundreds of microvolts in the cortex. Hence,
it is clear that the magnitude of LFPs is approximately 100–120 dB (five to
six orders of magnitude) smaller than that of the stimulation pulses. Therefore, the
design of an AFE that can record weak neural signals (in μV
range) in the presence of strong stimulation artefacts (in Volts range) without
being saturated, is, perhaps, the most difficult challenge associated with the
strategy of concurrent sensing and stimulation.To alleviate this problem, Rossi et al designed an
artefact-free recording system for acquisition of LFPs from the DBS lead positioned
in the STN (Rossi ). The stimulation artefact at 130 Hz and the higher harmonics were
separated from the neural signals of interest in the frequency domain using a 10th
order analog low-pass filter at 40 Hz. This high-order filter was formed by
cascading five 2nd order Sallen-Key low-pass filters, designed using Butterworth
coefficients. The advantages of this system are its high gain of 100 dB and its high
common mode rejection ratio (CMRR) of 130 dB. However, that front-end suppresses the
stimulation interference by significantly restricting the bandwidth of the recorded
LFPs and it requires a ± 15 V supply to operate. Another method to remove
stimulation artefacts is post-filtering (Al-ani
). In this case, an artefact-free
biomarker is produced by subtracting the template of the stimulation signal from the
recorded signal. However, this method degrades the signal quality (Parastarfeizabadi and Kouzani 2017).
Furthermore, it may not operate correctly in a closed-loop DBS setting where the
stimulation rate may fluctuate (Parastarfeizabadi
and Kouzani 2017).Stanslaski et al designed an implantable, chronic, adaptive
DBS device that benefits from an LFP/ECoG sensor (Stanslaski ). This device, which was
successfully validated in an ovine model of epilepsy by measuring hippocampus
seizure activity during and after stimulation, has been chronically implanted in
humans (Swann ). A
support vector machine (SVM) classification algorithm with spectral fluctuation
processing capabilities was used to separate the biomarker from the stimulation
artefact. The suggested device fits in a 39 cm3 volume, employing
front-end band-pass filtering which ensures that the instrumentation amplifier (INA)
operates within its normal range. However, an analog third-order low-pass filter at
100 Hz is used to filter chopping clock interference and stimulation interference,
thus limiting the available bandwidth for LFP recording. Moreover, the authors found
that interactions of stimulation artefact and sampling clock can create an aliased
signal in the measurement band.Finally, Pinnell et al introduced a miniature wireless
system weighing 8.5 g (including battery) for rodent use that combined multichannel
DBS and LFP recordings (Pinnell ). Its performance was verified in a working memory
task that involved 4-channel fronto-hippocampal LFP recording and bilateral
constant-current fimbria-fornix DBS. The wireless system was capable of simultaneous
recording and stimulation for a signal bandwidth between 1.5 and 100 Hz. However,
the activation of DBS resulted in prominent stimulation artefacts on the raw LFP
trace consisting of both harmonic repetitions of the stimulus frequency, and
aliasing artefacts (Pinnell ). The proposed way to alleviate this problem was to introduce
relatively lower-intensity stimulation parameters and apply a low-pass filter below
80 Hz on the recorded signals (Pinnell ).All in all, despite the advances in concurrent neural sensing and
stimulation, the issue of stimulation artefact in the recorded LFP signals has not
been fully addressed yet in the existing DBS systems (Parastarfeizabadi and Kouzani 2017). This paper focuses on the
interface between the neural tissue and the analog front-end (AFE) (which amplifies
the neural signals of interest and suppresses stimulation artefacts) prior to
digitization and presents the design and testing of a novel AFE architecture, which
enables the reliable recording of low- and higher-frequency LFP signals during
either unipolar or bipolar DBS.
Methods
System requirements, design and implementation of the AFE
As already stated in the introduction, the power of the oscillations at
physiologically significant bands, such as theta (4–7 Hz), alpha
(8–11), low beta (12–20 Hz), high beta (20–35 Hz), high
gamma (60–80 Hz), and 300 Hz (270–330 Hz) correlates with PD motor
symptoms and clinical conditions (Priori
). Based on these findings, our aim
was to design and assess a versatile AFE that provides the passband needed to
investigate the existence of possible biomarkers in these frequency bands.
Moreover, since the LFP spectral content varies among patients (Arlotti ), the
extended passband offered by the proposed AFE could lead to a more in-depth
analysis of the spectral content recorded from each patient, facilitating the
personalization of treatment for patients suffering from PD.The proposed AFE is specifically designed to acquire LFPs from DBS
electrodes placed in the STN. Post-operative LFPs are usually differentially
recorded from two DBS electrode contacts and referred to an electrode placed on
the scalp (Rossi ). Differential LFP recording offers the advantage of limiting
volume conduction (Gabriel ) and leveraging the CMRR of the front-end
amplifier to reduce the artefacts originating from DBS. The requirements for
signal acquisition are summarized in table
1. It is clear that a high gain, CMRR and dynamic range along with
low noise levels (Zhou ) were required in order to ensure high-performance LFP recording
during or without the presence of stimulation. Furthermore, according to the
literature (Denison ), when platinum–iridium (PtIr) DBS electrodes are used
for recording LFP signals (our case), adequate rejection of differential dc
offset voltages that are in the order of tens of millivolts is required.
Regarding stimulation artefact suppression, the requirement was to extend the
available bandwidth for LFP recording during stimulation beyond the limit of 100
Hz, which is the bandwidth offered by the existing DBS devices for recording
neural signals during stimulation.
Table 1
Key AFE requirements for reliable acquisition of LFPs during deep brain
stimulation (DBS).
Property
Value
Units/comments
Gain
⩾60
dB
Noise power spectral density estimate
⩽100
nV(Hz)−1
Integrated noise
⩽100
nV rms (0.5–500 Hz)
CMRR
⩾100
dB (DC to 60 Hz)
Differential DC offset to tolerate
Tens of
mV
Hours of continuous operation
⩾24
h
Input dynamic range
⩾ ±200
μV
High-pass corner
0.5
Hz
Low-pass corner
500
Hz
To meet the requirements for data acquisition, we designed and
implemented an AFE consisting of four main stages (figure 1): (i) a differential pre-amplification stage with high-pass
characteristics, which suppresses the common mode artefact voltage (CMAV); (ii)
an 8th order analog notch filter that suppresses the main frequency of the
differential mode artefact voltage (DMAV); (iii) a 2nd order analog low-pass
filter that suppresses the high-frequency harmonics of the DMAV and defines the
passband of the system; and (iv) a final amplification stage that uses a
programmable gain INA to achieve the required gain.
Figure 1
Architecture of AFE design for artefact-free LFP recording during DBS. The AFE
consists of (a) a differential pre-amplification stage with high-pass
characteristics, which suppresses the CMAV, (b) an 8th order analog notch filter
that suppresses the main frequency of the DMAV, (c) a 2nd order analog low-pass
filter that suppresses the high-frequency harmonics of the DMAV and (d) a final
amplification stage that uses a programmable gain instrumentation amplifier to
achieve the required gain. Two AFEs, based on the architecture presented above,
have been designed. They only differ in their second stage, where the first AFE
(Chebyshev notch channel) employs an 8th order Chebyshev notch filter, whereas
the second AFE (Bessel notch channel) employs an 8th order Bessel notch
filter.
The pre-amplification stage consists of an ultralow noise INA (model
AD8429, Analog Devices, USA). Taking into consideration that the high pulse
amplitudes of 2–3.5 V in a typical stimulation therapy are up to six
orders of magnitude larger than the neural signals of interest, which typically
are in the order of 1–10 μV when measured from
DBS electrodes (Stanslaski ), an artefact suppression strategy has to be
employed. The strategy followed in this design was to initially achieve a
significant suppression of the CMAV by exploiting the high CMRR offered by the
front-end INA and thus avoiding the use of input protective diodes, or passive
high-pass filters that would increase the noise and decrease the CMRR of the AFE
(Casas ).
Therefore, the gain of the front-end INA was set to 40 dB in order to provide a
high CMRR value and thus satisfied the imposed requirement on the CMRR of the
system.Another challenge that has to be taken into account in the design of the
AFE is that the placement of a metallic electrode in the tissue results in
charge redistribution, creating a capacitive double layer that can lead to
significant polarization voltages (Merrill
). These offsets can easily saturate
the high-gain front-end INA and must be adequately rejected (Spinelli , Denison ). The
strategy followed in this design was to introduce an active feedback integrator
(Nikola )
(figure 2(a)), implemented with a
single operational amplifier (OPA) (model ADA4522, Analog Devices, USA). The
addition of this OPA leads to the formation of the following transfer function:
Figure 2
Graphical representation of the four blocks constituting the AFE architecture.
The resistors and capacitors included in these blocks are characterized by a
tolerance of 0.1% and 10%, respectively. (a) The signals coming from two
contacts of the DBS electrode are subtracted and amplified with a gain of 40 dB
by an INA with high-pass characteristics (the high-pass knee frequency was set
at 0.5 Hz). (b) An eight-pole Bainter 140 Hz notch filter is used to suppress
the main frequency of the stimulation artefacts. (c) A two-pole 500 Hz
Sallen-Key low-pass filter is used to suppress the high-frequency harmonics of
the stimulation artefacts and define the passband of the system. (d) An INA
provides either 20 dB or 40 dB amplification, which is digitally determined via
a multiplexer.
Where Vout is the output voltage of the
active feedback integrator, Vdiff is the
differential input voltage, DG is the differential gain, K is
the integrator slope and s (=jω, where
ω is the angular frequency in rad/sec) is the
complex variable representing frequencies. Clearly, as ω
decreases, the magnitude of the TF decreases (for ω
→ 0 rad/sec, the TF magnitude approaches 0), whereas when
ω increases, the magnitude of TF increases (for
ω → ∞rad/sec, the TF magnitude
approximates the value of DG), which is in accordance with the function of a
high-pass filter. The zero introduced by the integrator was set at 0.5 Hz.The analog filtering stages (notch and low-pass) were introduced between
the pre-amplification and the final amplification stages in order to suppress
the DMAV. More specifically, an eight-pole Bainter notch filter (figure 2(b)) was designed and introduced in
the signal chain to suppress the main frequency of the DMAV. Since 140 Hz DBS
has proven to improve limb bradykinesia (Blumenfeld ) and continuous high
frequency stimulation (130–180 Hz) of subcortical motor nuclei has proven
to be highly effective in suppressing PD motor symptoms, and tremor observed in
essential and dystonic tremor (Cagnan ), the center frequency of the aforedescribed
analog notch filter was chosen to be equal to 140 Hz. Moreover, the stopband of
the notch filter was tuned between 125 and 155 Hz. It is important to note here
that, according to (Arlotti ), no biomarkers for PD were found in this
frequency band though research continues (Shimamoto , De Hemptinne ).Two different versions of this notch filter were designed and tested in
order to assess which of those two implementations is the most suitable for
being placed in the second stage of the final AFE architecture. The first
version was a 0.5 dB Chebyshev approximation while the second version was a
Bessel approximation. Chebyshev and Bessel approximations were chosen because
our aim was to thoroughly investigate the tradeoff between a steep filter
roll-off (Chebyshev filters) and an excellent transient response to a step/pulse
input thanks to a linear phase response (Bessel filters) (Karki 2002). Both steep roll-off and good transient and
phase response are required in neuromodulation. The former is explained by the
fact that the proximity of the sensing to the stimulation and the low magnitude
of the neural signal relative to the stimulation require the design of filters
that can sufficiently suppress the stimulation artefacts, while the latter is
required in order to achieve minimally distorted recording of neural
signals.The low-pass filtering stage (figure
2(c)) includes a two-pole classic Sallen-Key low-pass filter designed
using Bessel coefficients to ensure an excellent transient response. The role of
this filter is to suppress the high-frequency harmonics of the DMAV and define
the passband of the system. The objective of this effort was to reliably extract
LFP signals during DBS by applying techniques in the analog domain to avoid
saturation. As a result, having ensured that the LFP signals will reach the
analog-to-digital-converter (ADC), further a posteriori
low-pass filtering in the digital domain can be applied to completely remove
higher harmonics coming from stimulation. Taking into account the aforementioned
objective and the fact that a higher order low-pass filter would add extra
components that would further increase complexity and possibly power consumption
of the AFE and would occupy more space on the final printed circuit board (PCB),
a low-order filter was introduced at this stage.The final amplification stage (figure
2(d)) includes a single-ended amplification with a gain of either 20
dB or 40 dB. An INA (model AD8422, Analog Devices, USA) with its negative input
grounded was used to amplify the signals coming from the low-pass filter of the
previous stage. The gain is digitally programmable and is determined by a
multiplexer (model ADG1404, Analog Devices, USA). The first three stages
(differential pre-amplification, notch and low-pass filtering) are supplied with
± 5 V to ensure that an adequate headroom is provided to eliminate the
risk of saturation coming from electrode dc offsets and stimulation artefacts.
However, the fourth (last) stage is supplied with ± 2.5 V to be able to
interface with high-performance and low-power commercial ADC chips (e.g. model
ADS1298, Texas Instruments, USA). The fundamental building block for the design
of the filtering stages is the OPA ADA4522 by Analog Devices. Finally, the
resistors and capacitors included in this four-stage architecture are
characterized by a tolerance of 0.1% and 10%, respectively.The two designed channels (Chebyshev and Bessel notch channel) are
powered by a medical DC/DC converter (model THM 10-0521WI by Traco power), which
provides a reinforced isolation system for 5000 VACrms isolation and a very low
leakage current of less than 2 μA. On the isolated side
of the PCB hosting the designed AFEs, a low dropout voltage regulator (model
TPS7A7001DDA from Texas Instruments) is used to convert the +5 V originating
from the positive (isolated) output of the DC/DC converter into +2.5 V, while a
linear voltage regulator (model LM337IMP/NOPB from Texas Instruments) is used to
convert the −5 V originating from the negative (isolated) output of the
DC/DC converter into −2.5 V.
In vitro experimental setup for artefact suppression
testing
An in vitro experimental setup for unipolar (figure 3(a)) and bipolar (figure 3(b)) stimulation was prepared to
reproduce the stimulation and recording conditions of a typical post-operative
LFP recording session. The DBS electrode used in the experiments (electrode A in
figures 3(a) and (b), model DB-2201,
Boston Scientific Neuromodulation) is a directional eight-contact segmented DBS
lead (Rossi ).
We placed the DBS electrode in a glass container filled with tyrode solution
(128.2 mM of NaCl, 1.3 mM of CaCl2, 4.7 mM of KCl, 1.05 mM of
MgCl2, 1.19 mM of NaH2PO4, 20 mM of
NaHCO3 and 11.1 mM of glucose) at room temperature. The segmented
DBS electrode has eight contacts in total, two contacts at the two sides of the
electrode (which are contacts 0 and 3 of electrode A in figures 3(a) and (b)) and another six contacts (1a, 1b, 1c,
2a, 2b and 2c).
Figure 3
The in vitro experimental setup for unipolar (a) and bipolar (b)
stimulation. A DBS electrode (electrode A in (a) and (b), model DB-2201, Boston
Scientific Neuromodulation) was placed in a glass container filled with tyrode
solution at room temperature. The monophasic stimulation pulses (3 V
peak-to-peak amplitude, 140 Hz frequency and 100 μs
pulse width) were delivered by means of a commercial stimulator (Grass,
Astromed, Inc., USA) and the LFP signals (representing the LPF signals recorded
from the human neural tissue in a typical post-operative LFP recording session)
were injected to the solution by an Agilent 33220A waveform generator. The LFP
signals were injected to the solution as a differential signal through a second
electrode (electrode B in (a) and (b), model 401261, St. Jude Medical). One of
the four contacts of electrode B was connected to the ground of the recording
system. In both unipolar and bipolar settings the stimulation ground was
electrically isolated from the mains by using a commercial isolator (SIU5
stimulus isolation point, Grass, Astromed, Inc., USA). The LFP signals recorded
by the proposed AFE were digitized at a sampling frequency of 20 kSPS (samples
per second) and depicted on a computer by the Powerlab data acquisition system
(ADInstruments). (a) In the unipolar stimulation setting, we sense
differentially and symmetrically in space about the unipolar stimulation contact
1a of electrode A by sensing across the two nearest, equi-distant to contact 1a,
neighbour contacts (contacts 0 and 2a). However, since the surface areas of
contacts 0 and 2a differ, the sensing is not completely symmetrical and thus
some differential-mode interference from stimulation is expected to appear and
be suppressed by the analog notch filter of the proposed AFE. The anode (ground)
of the stimulator was connected to one of the contacts of a third electrode
(electrode C), which is the 8-contact Vercise DBS lead (Boston Scientific).
Electrode C was placed approximately 4 cm away from the stimulation site and
represents the case of the IPG, which acts as an anode in the unipolar
stimulation setting. (b) In the bipolar stimulation setting, two contacts of
electrode A (0 and 1a) were used for stimulation (anode and cathode of the
stimulator) and another two for recording (2a and 3).
The monophasic stimulation pulses (3 V peak-to-peak amplitude, 140 Hz
frequency and 100 μs pulse width) were delivered by a
commercial voltage-mode stimulator (Grass, Astromed, Inc., USA) and the LFP
signals (representing the LPF signals recorded from the human neural tissue in a
typical post-operative LFP recording session) were injected in the solution by
an Agilent 33220A waveform generator. The LFP signals were injected in the
solution as a differential signal through a second electrode (electrode B in
figures 3(a) and (b), model 401261, St.
Jude Medical). One of the four contacts of electrode B was connected to the
ground of the recording system. In both unipolar and bipolar settings the
stimulation ground was electrically isolated from the mains using a commercial
isolator (SIU5 stimulus isolation point, Grass, Astromed, Inc., USA). The output
impedance of the SIU5 isolator equals 1 kΩ. The LFP signals recorded by
the proposed AFE were digitized and depicted on a computer by means of the
Powerlab data acquisition system (Powerlab 16/35, ADInstruments).In a unipolar configuration one contact on the electrode is set to
cathode and the case of the implantable pulse generator (IPG) acts as an anode
(Amon and Alesch 2017). In our
unipolar stimulation setting (figure 3(a)),
we sense differentially and symmetrically in space about the unipolar
stimulation contact 1a of electrode A by sensing across the nearest (bilateral
to contact 1a) neighbour contacts, i.e. across contacts 0 and 2a. As a result, a
significant part of the interference appears as a common-mode signal at the
differential sensing pre-amplifier and is rejected by its high CMRR. However,
since the surface areas of contacts 0 and 2a differ, the sensing will not be
perfectly symmetrical and thus some differential-mode interference caused by the
stimulation is expected to appear and be suppressed by the analog notch filter
which follows in the AFE’s chain. The anode (ground) of the stimulator
was connected to one of the contacts of a third electrode (electrode C), which
is the 8-contact Vercise DBS lead (Boston Scientific). Electrode C was placed
approximately 4 cm away from the stimulation site and represents the case of the
IPG, which acts as an anode in the unipolar stimulation setting.Finally, in a bipolar configuration one electrode contact is used as the
anode and another electrode contact as the cathode, while the case of the IPG is
neutral (Schmidt and Van Rienen 2012,
Amon and Alesch 2017). In our setup
(figure 3(b)), two contacts of
electrode A (0 and 1a) were used for stimulation (as the anode and cathode of
the stimulator, respectively) and another two for recording (2a and 3). Since
contact 2a is closer to the stimulation site in comparison with contact 3, the
differential sensing of the contaminating pulses by the front-end INA is
asymmetric and thus more differential mode artefacts enter the signal chain.
Hence, in the bipolar stimulation setup shown in figure 3(b) the high CMRR of the front-end INA cannot be fully
exploited.
Results
AFE characterization—measured results
As described in section 2.1, two
versions of the 8th order Bainter notch filter were designed and introduced in
the fundamental AFE architecture shown in figure
1. The aim of this effort was to compare the achieved performance of
the two AFE architectures (Chebyshev notch channel versus Bessel notch channel)
and decide for the one that is the most suitable for neuromodulation based on
the specifications summarized in table 1.
In this section, a number of strict tests, which are typically performed on
analog electronics to assess their performance in terms of noise, linearity and
temporal response are presented and analysed.
Impulse/step response
The impulse function is defined as an infinitely high, infinitely
narrow pulse, with an area of unity (Zumbahlen 2008). In practice, when the impulse width is much
less than the rise time of the filter, the resulting response of the filter
will give a reasonable approximation to the actual impulse response of the
filter (Zumbahlen 2008). Rise time is
typically defined as the time between 10% response to 90% response of the
final value (steady state output) (Ardizzoni
2007). Since the impulse response of the total channel (Chebyshev
versus Bessel channel), rather than the sole impulse response of each notch
filter, is examined here, the rise time is determined by the front-end
first-order high-pass filter and for both channels this is equal to
where R = 1 MΩ and
C = 270 nF (see figure
2(a)).In figure 4(a), the width of
the input impulse was set at 100 μs (which is
identical to the DBS pulse duration used in later in vitro
and in vivo experiments) and the amplitude was 2 mV, which
is close to the maximum peak amplitude that can be handled by the designed
channels (=2.3 mV). Chebyshev and Bessel notch channels exhibit
approximately the same settling time (figure
4(d)). Another important test for evaluating the temporal
response of the designed AFEs is to supply them with a biphasic input pulse.
The responses of the Chebyshev and Bessel notch channels to a biphasic input
pulse are shown in figure 4(b). The
input pulse was approximately equal to 2 mV for 100
μs and −2 mV for another 100
μs. As anticipated, the responses of both
channels to a biphasic input pulse (figure
4(e)) exhibit a faster settling in comparison to their impulse
responses (figure 4(d)).
Figure 4
(a) Impulse response, (b) response to a biphasic pulse and (c) step response of
the Chebyshev (red line) and Bessel (blue line) notch channels. (d) Chebyshev
and Bessel notch channels exhibit approximately the same settling time. (e) The
response of both channels to a biphasic pulse exhibits a faster settling in
comparison to their corresponding impulse responses. (f) The step response of
the Chebyshev notch channel shows a slightly bigger overshoot and ringing in
comparison to the Bessel notch channel. As in the case of the impulse response,
the differences are not significant.
The step response of a filter, which is the integral of the impulse
response, is useful in determining the envelope distortion of a modulated
signal (Zumbahlen 2008). The two most
important features of a filter’s step response are the overshoot and
ringing. Overshoot should be minimal for good pulse response and ringing
should decay as fast as possible, so as not to interfere with subsequent
pulses. Transient response curves cannot provide a completely accurate
estimation of the output since, in practice, signals typically are not made
up of impulse pulses or steps (Zumbahlen
2008). However, these curves constitute a convenient figure of
merit so that transient responses of various filter types can be compared on
an equal footing (Zumbahlen 2008).
The step response of the Chebyshev notch channel shows a slightly bigger
overshoot and ringing in comparison to the Bessel notch channel (figure 4(f)). As in the case of the
impulse response, the differences, which are in accordance with the nature
of the two notch filters, are not significant. Finally, figure 4(c) reveals the accoupling characteristics of
the designed channels. Although the input voltage remains at 2.1 mV, the
output voltage returns back to 0 V after a settling time (time needed for
the response to reach and stay within 2% of its final value) of:
Bode magnitude plot/noise
From the Bode magnitude plot shown in figure 5(a), it is clear that both channels provide a passband
between 0.5 and 500 Hz and achieve the desired gain of 60 dB. The roll-off
of the high- and the low-pass filters equals + 20 dB/decade and −40
dB/decade, respectively, for both topologies. However, the Chebyshev notch
channel achieves a sharper transition between the passband and the stopband
at 140 Hz, compared to the transition of the Bessel notch channel. Moreover,
the Chebyshev notch channel provides stronger attenuation at the central
frequency of the notch, which is equal to 140 Hz, compared to the Bessel
notch channel. Although the most serious drawback of the Chebyshev
approximation is that it allows ripple in the frequency response in order to
achieve a faster roll-off (Karki
2002), the proposed Chebyshev notch channel exhibits a flat magnitude
response in the passband and approximates the magnitude response of the
Bessel notch channel. This is attributed to the fact that it was designed as
a 0.5 dB Chebyshev filter and thus the amount of passband ripple is
limited.
Figure 5
Measured Bode magnitude plot with the gain of both channels set at 60 dB (a) and
input-referred noise (b) of the Chebyshev notch channel (red line) and the
Bessel notch channel (blue line). (a) Both channels provide a passband between
0.5 and 500 Hz. The roll-off of the high- and the low-pass filters equals +20
dB/decade and −40 dB/decade, respectively, for both topologies. However,
the Chebyshev notch channel provides a sharper transition between the passband
and the stopband and stronger attenuation at the central frequency of the notch
(=140 Hz), compared to the Bessel notch channel. Besides, the Chebyshev notch
channel exhibits a flat magnitude response in the passband and approximates the
magnitude response of the Bessel notch channel. (b) Based on the input-referred
noise graph, it is concluded that both channels are low-noise with the Chebyshev
notch channel presenting a slightly better noise performance. Noise power
spectral density estimates in the passband for the Chebyshev and the Bessel
notch channels are 4 nV and 4.4 nV respectively, with the residual
1/f corner estimated at roughly 10 Hz for both
channels.
An input-referred noise voltage graph presents the input noise
voltage of a system versus frequency. It is widely used to evaluate the
flicker (or 1/f) and the thermal noise of a system, as well
as the noise corner frequency, which is the point in the frequency spectrum
where 1/f noise and thermal (or white) noise are equal
(Wu ).
The input-referred noise was measured by connecting both inputs of the
front-end INA to the ground of the PCB, recording the output voltage of the
channel and then dividing it by the gain, which was equal to 60 dB. Since
there is no passive filtering network before the front-end AD8429 INA chip
and the gain of the first stage is sufficiently high (equal to 40 dB), which
allows the effective noise factor to be the noise factor of the first stage
without an impact on the subsequent stages (Northrop 2012, Poshala
), the input-referred noise of the
designed AFEs should approximate the measured input-referred noise reported
in the datasheet of the AD8429 chip. The integrated noise of the Chebyshev
and Bessel notch channels over the frequency range 0.5–500 Hz was
measured and found to be equal to 96 nV rms and 121 nV rms, respectively.
Figure 5(b) shows that both
channels are low-noise with the Chebyshev notch channel characterised by a
slightly better noise performance. Noise power spectral density estimates in
the passband for the Chebyshev and the Bessel notch channels are 4 nV
and 4.4 nV respectively, with the residual
1/f corner estimated at roughly 10 Hz for both
channels. Indeed, these measured results are in agreement with the noise
measurements reported in the datasheet of the front-end AD8429 INA chip.
Measured results versus specifications
Taking into consideration the previously presented Bode amplitude
plot and noise performance of the two channels, the recording capabilities
of the Chebyshev notch channel satisfy all of the requirements shown in
table 1. Regarding the Bessel
notch channel, it satisfies all of the requirements except for the one
related to the integrated noise of the channel. The integrated noise of the
Bessel notch channel over the frequency range 0.5–500 Hz was measured
and found to be equal to 121 nV rms which is higher than the imposed limit
of 100 nV rms (table 1). Since the
recording capabilities of the Bessel notch channel have not satisfied all of
the imposed specifications, measured results only from the Chebyshev notch
channel are presented in sections
3.1.4, 3.1.5, 3.2–3.4.
Total harmonic distortion (THD) and intermodulation distortion
(IMD)
Due to nonlinearities of electronic components, distortion is
generated. Two of the most common ways to assess the linearity of an
amplifying system is to specify its THD and its IMD levels.THD is the ratio of the root-sum-square value of all the harmonics
(2×, 3×, 4×, etc) to the rms signal level (Lai 2009). Generally speaking, only the
first five or six harmonics are significant in the THD measurement (Lai 2009). In other words, THD measures
the nonlinearity of a system, while applying a single sinusoidal signal as
its input. The THD of the Chebyshev notch channel for a gain of 60 dB is
shown in figure 6(a). After examining
the available dynamic range of the channel (from 1
μV peak to 2.3 mV peak), it is clear that the
achieved THD is less than 0.2%. Only input sinusoidal voltages with peak
amplitudes approaching the highest input voltage (=2.5 mV peak) that can be
handled by the rail-to-rail output INA located at the last stage of the AFE
present higher THD values.
Figure 6
THD (a) and third order IMD (the reference impedance equals 50 Ω) (b)
measured with the gain of the Chebyshev notch channel set at 60 dB. (a) After
examining the available dynamic range of the channel (from 1
μV peak to 2.3 mV peak), it is clear that the
achieved THD is less than 0.2%. (b) The two tones applied to the Chebyshev notch
channel were f1 = 4.9 Hz and
f2 = 5.1 Hz. The output power of a single
fundamental tone (in dBm—red line in the graph) and the relative
amplitude of the third order IMD3 products referenced to a single
tone (blue circles in figure (b)) are plotted as a function of the applied input
power. The third order intercept line (dashed blue line) is extended to
intersect the extension of the fundamental output signal line (dashed red line).
This intersection is termed the third order intercept point IP3. The
calculated IP3 is characterized by a relatively high value, which is
a positive result since the higher the IP3 values the better the
linearity of the amplifier and the weaker the output intermodulation products
that will be generated at the amplifier’s output.
In general, when a spectrally pure sinusoidal signal passes through
an amplifier, various harmonic distortion products are produced depending on
the nature and the severity of the non-linearity (Lai 2009). However, simply measuring harmonic
distortion levels produced by single tone sinusoidal signals of various
frequencies does not convey all the information required to assess the
amplifier’s potential performance in a clinical setting, where
reliable recording of neural signals is required. Hence, it is often
required that an amplifier be evaluated in terms of the IMD product levels
produced by two or more specified tones applied at the input of the
amplifier (Lai 2009).Thus, our aim was to not only examine the THD of the Chebyshev notch
channel, which is produced by a single tone sinusoidal input, but also to
investigate the distortion products produced by two input tones. When two
tones of frequencies, f1 and
f2, are applied to the input of a nonlinear
system, they produce second and third order products. The second order
products are located at frequencies f2 +
f1 and f2
− f1. The third order products located at
frequencies 2f1 + f2
and 2f2 + f1 can
often be filtered out. However, the third order products located at
2f1 − f2
and 2f2 - f1 are
situated close to the main tones f1 and
f2 and thus it is difficult to be rejected
by filtering (Lai 2009). It can be
shown that second order IMD levels increase by 2 dB for every 1 dB of input
signal increase while the third order IMD amplitudes increase by 3 dB for
every 1 dB of input signal increase (Lai
2009).Third order IMD performance is often specified in terms of the third
order intercept point (IP3). Two spectrally pure tones are
applied to the system. The two tones applied to the Chebyshev notch channel
were f1 = 4.9 Hz and
f2 = 5.1 Hz. In figure 6(b), the output power of a single fundamental
tone (in dBm—red line in the graph) and the power of the third order
products (blue circles in figure 6(b),
defined as IMD3) are plotted as a function of input power. It is
clear that the fundamental line is characterized by a slope that is equal to
1.The third order intercept line (dashed blue line) is extended to
intersect the extension of the fundamental output signal line (dashed red
line). This intersection is termed the third order intercept point
(IP3) and is a figure of merit for comparing
amplifiers’ linearity. The higher the IP3 values the more
linear the amplifier and the weaker the distortion products at its output.
As shown in figure 6(b), the
IP3 of the Chebyshev notch AFE is characterized by a high
value. It should be stressed that this high IP3 value of the
proposed AFE is a very desirable feature: the non-linearity of the AFE
should indeed be very low to avoid artefact coupling into the physiological
measurements through intermodulation.
Key properties of the Chebyshev notch AFE
Taking into account all the measured results acquired from the
Chebyshev notch channel, the key properties of this channel are summarized
in table 2.
Table 2
Key properties of the Chebyshev notch AFE.
Property
Value
Units/comments
Supply voltage
±5, ±2.5
Volts
Gain
60, 80
dB (programmable)
Integrated noise
26
nV rms (0.5–40 Hz)
33
nV rms (0.5–100 Hz)
96
nV rms (0.5–500 Hz)
CMRR
130
dB (DC to 60 Hz)
Maximum tolerable differential DC offset
±32/85[a]
mV
Dynamic range
±2.3
mV (peak), gain = 1000
±230
μV (peak), gain = 10000
SNR
30
dB (minimum)
Nonlinearity
<0.2%
THD
High-pass corner
0.5
Hz
Low-pass corner
500
Hz
Total current consumption
32
mA
Hours of continuous operation
28
h (900 mAh battery)
Measured differential DC offset rejection of 85 mV is achieved when
a 1st order analog 0.5 Hz high-pass filter is cascaded after the front-end
INA (stage 1 of the current AFE).
Performance evaluation of the AFE based on comparisons with commercial
biopotential acquisition devices
To compare our system with available devices, we introduced identical,
extremely weak single tones to the Chebyshev notch channel and to a
state-of-the-art, commercial biological amplifier (Bioamplifier included in
Powerlab 26T, ADInstruments), which is optimized for measuring a wide variety of
biological signals such as ECG, EMG and EEG. In the first case, the signals were
recorded by the AFE of the Chebyshev notch channel and were digitized by the
16-bit ADC of the Powerlab 16/35 system at 1 kSPS, whereas in the second case
the signals were recorded and digitized by the Powerlab 26T (bioamplifier and
16-bit ADC) system at 1 kSPS.More specifically, a weak sinusoidal single tone (100 nV peak, 25 Hz)
was presented to the inputs of the two systems. This weak sinusoidal single tone
was provided by the Agilent 33220A waveform generator. However, since the
weakest signal that can be injected by the specific generator is a 10 mV peak
sinewave, we made use of ohmic attenuators (Cinch Connectivity Solutions) that
provided 100 dB attenuation to the signals injected by the waveform generator. A
digital low-pass filter at 30 Hz was applied on the recordings of both systems
in order to ensure that the noise coming from the front-ends of both systems
(integrated noise) stays at levels lower than 100 nV peak (so that the 100 nV
peak signal dominates the noise), and be able to compare them on an equal
footing. Moreover, the mean values of the signals recorded by the two systems
were removed in order to facilitate a more direct comparison between the two
AFEs in terms of signal quality.Regarding the first system (Chebyshev notch channel), its gain was set
at 80 dB (or 10 000 V/V) and the range of the Powerlab ADC at ± 10 V
(maximum available). The reason behind the choice of applying a gain of 80 dB
lies with the fact that a gain of 60 dB would not allow for the amplified
signals to overcome the smallest input increment the specific ADC
can resolve (20/65 536 = 305 μV). Regarding the second
system (Powerlab 26T bioamplifier), its recording range was set at ± 100
μV (lowest available), which means that a gain of
100 dB (or 100 000 V/V) was applied upon the input signals by the
bioamplifier’s AFE.Figure 7 shows that the Chebyshev
notch channel (figure 7(a)) is less
vulnerable to DC offsets that exist in the weak sinusoidal signal and can thus
provide more stable signal recordings compared to the commercial bioamplifier
(figure 7(b)). Next, a second
sinusoidal single tone with the same amplitude but lower frequency (=5 Hz) was
injected to the inputs of the two AFEs and the amplitude spectrum of the overall
recorded signal was calculated (figure
7(c)). It is clear that the spectrums of both systems include visible
spectral peaks at the two test frequencies (5 and 25 Hz). Based on the graph,
the amplitude (the reference voltage equals 1 V) of each of these two spectral
peaks is approximately equal to −143 dB, which is in accordance to the
expected theoretical value of
Figure 7
(a) Output voltage (after removing the gain of 80 dB) recorded from the Chebyshev
notch channel when a sinusoidal single tone (25 Hz, amplitude 100 nV peak) was
injected to the input of the channel. (b) Output voltage recorded from the
Powerlab 26T bioamplifier when a sinusoidal single tone (25 Hz, amplitude 100 nV
peak) was injected to the input of the system. (c) Amplitude spectrum calculated
when two sinusoidal tones, one low-frequency (=5 Hz) and one higher-frequency
(=25 Hz) are sequentially injected to the inputs of the two AFEs. The amplitude
spectrums of both systems present two spectral peaks at 5 and 25 Hz, which are
characterized by the same amplitude. (d) Output voltage (after removing the gain
of 80 dB) recorded from the Chebyshev notch channel when a sinusoidal single
tone (25 Hz, amplitude 30 nV peak) was injected to the input of the channel. (e)
Output voltage recorded from the Powerlab 26T bioamplifier when a sinusoidal
single tone (25 Hz, amplitude 30 nV peak) was injected to the input of the
system. (f) Amplitude spectrum calculated when two sinusoidal tones, one
low-frequency (=5 Hz) and one higher-frequency (=25 Hz), are sequentially
injected to the inputs of the two AFEs. The amplitude spectrums of both systems
present two spectral peaks at 5 and 25 Hz, which are characterized by the same
amplitude.
To push the limits of the Chebyshev notch channel’s recording
capabilities towards the noise floor of the system, an extremely weak sinusoidal
single tone (30 nV peak, 25 Hz) was presented to the inputs of the two systems.
Again, this sinusoidal single tone was provided by the Agilent 33220A waveform
generator in combination with attenuators that provided 110 dB attenuation to
the signals injected by the waveform generator. A digital low-pass filter at 30
Hz was applied on the recordings of both systems. The gain and digitization
settings were left the same with the ones used in the experiment where 100 nV
peak test tones were applied.As anticipated based on the results acquired by the injection of the 100
nV peak test tone, the Chebyshev notch channel (figure 7(d)) provides more stable signal recordings compared to the
commercial bioamplifier (figure 7(e)).
Next, a second sinusoidal single tone with the same amplitude but lower
frequency (=5 Hz) was injected to the inputs of the two AFEs and the amplitude
spectrum of the overall recorded signal was calculated (figure 7(f)). It is clear that the spectrums of both systems
include visible spectral peaks at the two test frequencies (5 and 25 Hz). It is
important to note here that the noise floor of the Chebyshev notch channel
(spectrum in red) is lower than the noise floor of the biological amplifier
(spectrum in blue). Based on the graph, the amplitude (the reference voltage
equals 1 V) of each of these two spectral peaks is approximately equal to
−153 dB, which is in accordance to the expected theoretical value ofFinally, a 50 s segment of LFP signal recorded (low-pass filtered by a
high-order digital low-pass filter at 553 Hz) from the subthalamic nucleus in a
patient with PD withdrawn from levodopa was injected by means of a waveform
generator to the input of the Chebyshev notch channel. Moreover, in order to
ensure that no phase distortion or ringing oscillations are introduced by the
analog Chebyshev notch filter when LFP recordings are obtained, the same LFP
signal was injected to a commercial high-performance differential amplifier that
does not include any analog notch filtering stage in its front-end electronics.
The commercial amplifier used in this series of experiments is the DP-301 model
(ADInstruments), which has been designed for amplifying weak signals such as
extracellular action potentials, and weaker EEG and ECG signals.However, since the waveform generator is not able to inject signals that
are weaker that 10 mV peak, the injected LFP signal at the generator’s
output (which was in mV range) had to be attenuated before entering the input of
the Chebyshev notch channel. More specifically, four attenuators (Cinch
Connectivity Solutions) that provided 80 dB attenuation were used in order to
bring the amplitude of the LFP signal injected by the waveform generator down to
the level that characterizes the original LPF signal, which is approximately
equal to 0.32 μV rms. The spectrum of this signal
contains a peak in the beta frequency band (13–30 Hz) and another peak at
80 Hz. The gain of the DP-301 amplifier was set at 80 dB (maximum available) in
order to ensure that this instrument will provide a reliable recording of the
weak LFP signal. The gain of the Chebyshev notch channel was also set at 80 dB
to compare the two systems on an equal footing. Finally, the analog outputs of
the two systems were sampled by the ADC of the Powerlab 16/35 system at 1 kSPS
(the range was set at ± 2 V so the smallest resolvable input increment of
the ADC and the smallest detectable signal by the two systems (Chebyshev notch
channel and commercial amplifier) were equal to 61 μV
and 6.1 nV, respectively). The analog high-pass filter included in the DP-301
amplifier (cut-off frequency at 1 Hz) was activated so that its temporal
response can be compared with the temporal response of the Chebyshev notch
channel’s ac-coupled AFE (cut-off frequency at 0.5 Hz) on an equal
footing.Figure 8(a) depicts the LFP signal
recorded by the high-order Chebyshev notch channel (red line) and the DP-301
amplifier (blue line). The two signals approximate each other which shows that
the Chebyshev notch channel is able to record the LFP signal without introducing
any phase distortion or ringing oscillations (figure 8(c)). Moreover, since the amplitude spectrum of the LFP
signal recorded by our channel (figure
8(b)) contains both the beta peak and the peak at 80 Hz, it can be
concluded that the proposed AFE architecture can record, save for the stopband
frequencies, both the low and high frequencies of the original LFP signal. It is
important to note here that the peak (red line) existing in the stopband
(125–155 Hz) is introduced by the notch operation (bear in mind figure 5(b)). As is shown in figure 8(b), the noise added by the Chebyshev
notch filter does not significantly affect the frequencies below and above the
stopband of this filter. However, physiological information should not be sought
after in the stopband of the notch (pink region in figure 8(b)).
Figure 8
A weak LFP signal is injected into the inputs of the Chebyshev notch channel and
the DP-301 commercial differential amplifier (ADInstruments). (a) Comparison of
the Chebyshev notch channel’s output (red line—after removing the
gain of 80 dB) with the DP-301 amplifier’s output (blue line—after
removing the gain of 80 dB) in the time domain. (b) The amplitude spectrum of
the Chebyshev notch channel’s output (red line) approximates the
amplitude spectrum of the original LFP signal (pink line). (c) The LFP
recordings acquired by the Chebyshev notch channel and the DP-301 amplifier
approximate each other. This shows that the proposed AFE architecture is capable
of recording weak LFP signals without introducing any phase distortion or
ringing oscillations. (d) The proposed AFE architecture provides more accurate
recording of the high frequencies (f > 350 Hz) included
in the original LFP signal in comparison to the DP-301 amplifier.
On the other hand, as is shown in figure
8(a), the stopband noise does not seem to significantly affect the
time-domain recording of the Chebyshev notch channel. The normalised root mean
square error, or RMSE, between the time-domain LFP signals recorded by the two
systems (Chebyshev notch channel and DP-301 amplifier) was measured and found to
be equal to 4.6%. This error, which can be considered tolerable taking into
account the extremely low amplitude of the specific LFP signal, can be
attributed to (1) the fact that the DP-301 amplifier cannot accurately record
frequencies of the LFP signal that are higher than 350 Hz (figure 8(d)), (2) the fact that small DC offsets existing in
the extremely weak LFP signal are not completely rejected by the two systems,
and (3) the noise in the stopband coming from the Chebyshev notch operation.
Evaluation of the artefact suppression capabilities of the AFE by in
vitro DBS tests
To examine the capability of the proposed Chebyshev notch channel to
suppress stimulation artefacts and thus allow artefact-free LFP recording during
stimulation, we prepared two in vitro setups, one for testing
unipolar DBS and one for testing bipolar DBS. The details of these two setups
have been given in figure 3 (section 2.2). More specifically, our aim was
to investigate whether or not the proposed Chebyshev notch channel could extend
the available bandwidth of LFP recording during stimulation, and to further
compare its performance with the Bessel notch channel’s performance but
with the focus to be on their stimulation suppression capabilities rather than
their recording capabilities, which have already been tested (sections 3.1.1 and 3.1.2).The strategy followed for the tests was to gradually increase the
available bandwidth and thus allow more artefacts to affect the recorded
signals. At each bandwidth setting, the quality of the recorded signals was
assessed. The shortening of the available bandwidth was achieved by the
application of a real-time and high-order digital low-pass filter. The first
step towards increasing the available bandwidth for recording during stimulation
setups, was to define a passband between 0.5 and 140 Hz, with 140 Hz being the
stimulation frequency and the central frequency of the notch filters. The next
step was to define a passband between 0.5 and 250 Hz to examine the impact of
the artefacts coming from the stimulation harmonic at 280 Hz on the recorded
signals.The first test (for both bandwidths) was to inject a weak sinusoidal
single tone (1 μV peak, 15 Hz) into tyrode solution and
examine the recording capabilities of the Chebyshev notch channel in and without
the presence of bipolar stimulation (140 Hz, 3 V peak, 100
μs). Given 0.5–140 Hz bandwidth, the
Chebyshev notch channel was able to record the weak sinusoidal single tone
without (figures 9(a) and (e)) and in
(figures 9(b) and (f)) the presence of
bipolar stimulation. Finally, when the bandwidth was set from 0.5 to 250 Hz, the
Chebyshev notch channel was again able to record the weak sinusoidal single tone
without (figures 9(c) and (g)) and in
(figures 9(d) and (h)) the presence of
bipolar stimulation.
Figure 9
Time and frequency responses of the Chebyshev notch channel, in and without the
presence of bipolar stimulation (140 Hz, 3 V peak, 100
μs). The test signal was a sinusoidal single tone with
an amplitude of approximately 1 μV peak and a frequency
of 15 Hz. (a) Time-domain recording without the presence of stimulation for a
passband set from 0.5 to 140 Hz. (b) Time-domain recording in the presence of
stimulation for a passband set from 0.5 to 140 Hz. (c) Time-domain recording
without the presence of stimulation for a passband ranging from 0.5 to 250 Hz.
(d) Time-domain recording in the presence of stimulation for a passband ranging
from 0.5 to 250 Hz. (e) Amplitude spectrum of the signals presented in figure
(a). (f) Amplitude spectrum of the signals presented in figure (b). (g)
Amplitude spectrum of the signals presented in figure (c). (h) Amplitude
spectrum of the signals presented in figure (d).
Having ensured that the designed Chebyshev notch channel was able to
record a weak sinusoidal single tone during stimulation without facing
saturation issues, the next step was to compare its artefact reduction
capabilities with the capabilities of the Bessel notch channel. In these tests,
‘played back’ LFP signals (repetitions of an LFP segment lasting
for 10 s, obtained from (Oostenveld )) with two visible spectral peaks at
approximately 167 Hz and 221 Hz were injected in tyrode solution from a waveform
generator, as described in section 2.2
(figure 3). The goal of these
experiments was to test all possible circumstances (in/without the presence of
bipolar/unipolar stimulation) and prove that 1) the designed Chebyshev AFE can
indeed provide a bandwidth that extends beyond the stimulation frequency of 140
Hz and 2) the application of the Chebyshev notch filter does not prevent the
successful recording of frequencies that are close to the stop band (for
instance the 167 Hz spectral peak of the LFP signal used in this series of
experiments).Figures 10 and 11 illustrate detailed views of the
time-domain LFP recordings taken from the Chebyshev (blue line) and the Bessel
(red line) notch channels, with (solid line) and without (dash-dot line)
unipolar (figure 10) and bipolar (figure 11) stimulation. In both figures,
some stimulation artefacts appear in the LFP recordings collected by means of
the Bessel notch channel. This observation leads to the conclusion that the
Chebyshev notch channel provides more stable and reliable recordings of the LFP
signals during stimulation in comparison to the Bessel notch channel. This is
mainly attributed to the fact that the Chebyshev notch filter provides a
stronger attenuation at the notch frequency than the Bessel notch filter (figure 5(a)). Moreover, a close inspection of
the signals recorded by the Chebyshev notch channel in and without the presence
of stimulation (figures 10 and 11) shows that the quality of the recovered
LFP signals is not affected by ringing that could be introduced by the notch
filter. This may be attributed to the fact that the ringing oscillations
introduced by the Chebyshev notch filter as a response to DBS are low in
amplitude and short in duration (bear in mind figure 4(d) where DBS pulses of the same duration were presented to
the input of the high-order notch channel).
Figure 10
Detailed view of the time-domain LFP recordings taken from the Chebyshev (blue
line corresponding to the left y-axis) and the Bessel (red line
corresponding to the right y-axis) notch channels, with (solid
line) and without (dash-dot line) unipolar stimulation. (a) The passband of both
channels is between 0.5 Hz and 140 Hz. (b) The passband of both channels is
between 0.5 Hz and 250 Hz.
Figure 11
Detailed view of the time-domain LFP recordings taken from the Chebyshev (blue
line corresponding to the left y-axis) and the Bessel (red line
corresponding to the right y-axis) notch channels, with (solid
line) and without (dash-dot line) bipolar stimulation. (a) The passband of both
channels is between 0.5 Hz and 140 Hz. (b) The passband of both channels is
between 0.5 Hz and 250 Hz.
Furthermore, a graphical representation of the amplitude spectrum of the
contaminating signals entering the positive (red colour) and negative (green)
input of the front-end INA of the designed Chebyshev notch channel, along with
the amplitude spectrum of the channel’s output in (black) and without
(pink) the presence of stimulation, after digitally removing the 280 Hz harmonic
from the recorded LFP signals during stimulation (spectrum in black), is
depicted in figure 12. Figures 12(a) and (b) correspond to the
unipolar and bipolar stimulation setting, respectively. It is clear that
aliasing artefacts located at various frequencies that are not harmonic
repetitions of the stimulation frequency (=140 Hz) exist in the spectrum of the
contaminating signals. This finding is in accordance with results measured from
existing DBS devices (Stanslaski , Pinnell
). However, figure 12 shows that the amplitude spectrum of the
channel’s output in the presence of either unipolar or bipolar
stimulation is free from these artefacts and thus approximates the spectrum of
the signals recorded without stimulation. This important finding could be
attributed to the fact that the proposed AFE does not include passive filtering
before the front-end INA. Front-end passive filters can lead to the degradation
of the combined (passive filter plus INA) apparent CMRR of the front-end due to
component mismatches (Casas ). The absence of such a passive filter network
enhances the ability of the proposed AFE to reject common-mode disturbances
stemming from the electrode-solution interface, thus offering a smooth spectrum
at the output of the AFE and an artefact-free LFP recording in both unipolar and
bipolar stimulation setups.
Figure 12
Amplitude spectrum (recorded from the Chebyshev notch channel) of the (1) signals
entering the negative (green) and positive (red) inputs of the front-end INA
during stimulation, (2) the AFE output voltage during stimulation (black), and
(3) the AFE output voltage without the presence of stimulation (pink). (a)
Unipolar stimulation setting, and (b) bipolar stimulation setting.
To quantify the differences between the recorded signals during
stimulation and the ones recorded without stimulation, we used the normalised
RMSE calculation. The normalisation for the RMSE calculation was performed over
the range of the reference signal, which is the signal recorded without the
presence of stimulation. Figure 13 shows
the normalised RMSE values that represent the differences between the recorded
signals in and without the presence of unipolar (figure 13(a)) and bipolar (figure
13(b)) stimulation for the Chebyshev (dashed blue line) and Bessel
(dotted blue line) notch channels. The vertical red lines shown in the graph
represent the amplitude Bode plots in each recording case. In other words, they
describe the available passband, set by the application of a very steep
real-time digital low-pass filter. To examine the benefits gained by the use of
an analog notch filter for artefact suppression, we introduced a third AFE
(solid blue line in figure 13), which does
not include any analog notch filtering circuitry and suppresses the stimulation
artefacts in the digital domain (by applying a very steep real-time digital
low-pass filter). More specifically, this AFE includes a passive, 1st order
low-pass filter at 8 kHz (to suppress high-frequency noise components), followed
by the INA chip AD8420 from Analog Devices set to provide a gain of 20 dB.
Figure 13
Normalised root mean square errors (RMSE) between the signals recorded in and
without the presence of unipolar (a) and bipolar (b) stimulation. The green
vectors show the main stimulation frequency component (=140 Hz) and the
stimulation harmonic that is closer to the available passband (=280 Hz). The red
lines (correspond to the right y-axis) show the available
bandwidth (BW) for each recording trial and the blue lines (correspond to the
left y-axis) depict the calculated RMSE values.
Referring to figure 13(a), in the
first bandwidth setting (0.5–50 Hz), the Chebyshev and Bessel notch
channels present similar RMSE values, whereas the -channel without notch
filter’ already shows a bigger error. In the next three bandwidth
settings (100 Hz, 140 Hz and 250 Hz), the Chebyshev notch channel presents the
lowest error, whereas the ‘channel without notch filter’ shows
unacceptably high errors, which is attributed to the fact that: (a) aliasing
artefacts exist in the frequency range 50–100 Hz, which is a finding that
is in agreement with the literature (Pinnell
), and (b) the stimulation frequency
(=140 Hz) is in the passband (when the digital low-pass filter is set at 140 or
250 Hz). The same conclusions are drawn from figure 13(b), where the Chebyshev notch channel presents the lowest
error. As in the case of unipolar stimulation, the ‘channel without notch
filter’ is characterised by unacceptably high errors at bandwidths
greater than 50 Hz.Another important observation is that the RMSE errors produced by the
Chebyshev and Bessel notch channels decrease when the bandwidth increases from
50 Hz to 140 Hz and then slightly increase when the bandwidth is set at 250 Hz.
This is attributed to a small intrinsic error that mostly comes from the dc
offset voltage which is generated by the electrodes and is not completely
rejected by the system. Hence, this small error in voltage is more apparent in
smaller bandwidths where the recorded LFP signals are weaker due to filtering
(0.5–50 Hz), decreases when the available bandwidth (and thus recorded
LFP signal strength) increases (0.5–140 Hz) and, finally, slightly
increases when the available bandwidth increases even more (0.5–250 Hz)
since more interference leaks into the wide passband (the 280 Hz stimulation
harmonic is getting closer to the passband).
Evaluation of the artefact suppression capabilities of the AFE by in
vivo DBS tests
To provide a proof-of-function in vivo, we recorded
LFPs from the thalamus of a non-human primate, at the end of a non-recovery
procedure that was performed for the primary purpose of another ongoing study.
The experiments were approved by the local ethics committee at Newcastle
University and performed under appropriate UK Home Office licenses in accordance
with the Animals (Scientific Procedures) Act 1986. A female rhesus macaque was
anesthetised with a ketamine/midazolam/alfentanil infusion and a segmented DBS
electrode (electrode A, model DB-2201, Boston Scientific Neuromodulation) was
implanted into the thalamus as shown in figure
14. The monophasic stimulation pulses (6 V peak-to-peak amplitude,
142 Hz frequency and 100 μs pulse width) were delivered
by means of a commercial stimulator (Grass, Astromed, Inc., USA). Unipolar
stimulation was applied to contact 2 of electrode A (illustrated as A2 in figure 14) and LFP signals were
differentially recorded through contacts 1 and 3 of electrode A (illustrated as
A1 and A3 in figure 14, respectively). The
stimulation ground was introduced into the neural tissue through contact 1
(illustrated as B1 in figure 14) of
electrode B (model 401261, St. Jude Medical), which was placed over the frontal
cortex. A commercial isolator (SIU5 stimulus isolation point, Grass, Astromed,
Inc., USA) was used to electrically isolate the stimulation ground from the
mains. The non-human primate was under anaesthesia during the entire experiment
with the head held in a primate stereotactic frame, which was connected to the
ground of our recording system. The LFP signals recorded by the proposed AFE
were digitized at a sampling frequency of 20 kSPS and depicted on a computer by
the Powerlab data acquisition system (ADInstruments).
Figure 14
Experimental setup for evaluating the artefact suppression capabilities of the
proposed Chebyshev AFE channel architecture in vivo. A DBS
electrode (electrode A, model DB-2201, Boston Scientific Neuromodulation) was
implanted into the thalamus of an anaesthetised non-human primate. The
monophasic stimulation pulses (6 V peak-to-peak amplitude, 142 Hz frequency and
100 μs pulse width) were delivered by means of a
commercial stimulator (Grass, Astromed, Inc., USA). Unipolar stimulation was
applied to contact A2 and LFP signals were differentially recorded through
contacts A1 and A3. The stimulation ground was introduced into the brain tissue
through a second electrode (contact B1, model 401261, St. Jude Medical) that was
placed over the frontal cortex. The stimulation ground was electrically isolated
from the mains using a commercial isolator (SIU5 stimulus isolation point,
Grass, Astromed, Inc., USA).The non-human primate was under anaesthesia with the
head held in a primate stereotactic frame, which was connected to the ground of
our recording system. The LFP signals recorded by the proposed AFE were
digitized at a sampling frequency of 20 kSPS (samples per second) and depicted
on a computer by the Powerlab data acquisition system (ADInstruments).
As shown in figures 15(a) and (b),
the Chebyshev notch channel can provide artefact-free LFP recordings during DBS.
Moreover, after examining the detailed views of the LFP recordings acquired
without and in the presence of DBS (figures
15(c)–(f)), we conclude that: (1) the stimulation artefacts
(at 142 Hz and 284 Hz) induced by DBS have been significantly suppressed (blue
line in figure 15(f)), (2) the amplitude
spectrum of the LFP signals recorded during DBS (figure 15(f)) is free from aliasing artefacts, which is in full
agreement with the in vitro experimental results shown in figure 12, and (3) the contaminating 142 Hz
DBS pulses are successfully suppressed by 68 dBs (amplitude spectrums in red and
black in figure 15(f)) thanks to the
combined notch filtering action and the front-end INA’s (high) CMRR.
Figure 15
The proposed Chebyshev AFE architecture for artefact-free LFP recordings during
unipolar DBS in vivo. LFP signals were recorded from the
thalamus of an anaesthetised non-human primate in and without the presence of
DBS with the experimental setup illustrated in figure 14. (a) Bipolar (differential) LFP recordings without DBS.
(b) Bipolar (differential) LFP recordings during DBS. (c) Detailed view of the
LFP recordings acquired without DBS. (d) Detailed view of the LFP recordings
acquired during DBS. (e) Amplitude spectrum of the LFP signal recorded without
DBS. (f) Amplitude spectrum of: (1) the LFP signal recorded during DBS (blue
line), and (2) the stimulation pulses presented at the positive (red line) and
negative (black line) inputs of the front-end instrumentation amplifier. It is
clear that the proposed artefact suppression strategy (analog notch filtering at
140 Hz and digital low-pass filtering at 250 Hz) allows for artefact-free LFP
recordings during DBS (observe the 142 Hz stimulation fundamental frequency,
which has been strongly attenuated by the high-order notch filtering
action).
Discussion
Methodological significance
The analog filtering strategy used in the proposed AFE architecture is
an effective approach to adequately suppress stimulation artefacts. In our
application, the stimulation artefact and the signal of interest are highly
overlapping both in the time and frequency domain. More specifically, the
stimulation frequency (=140 Hz) and its first two harmonics (280 and 420 Hz) are
located within the desired system bandwidth (0.5–500 Hz). Therefore, the
stimulation artefact and its harmonics could not be separated from the neural
signals of interest using an analog high-order low-pass filter, which was the
strategy employed by Rossi ).An alternative approach that has been extensively adopted is to provide
a switching circuit that disconnects the front-end leads of the amplifier during
stimulation (Rossi ). This strategy is effective for applications, such as
transcranial magnetic stimulation (Paus
, Fuggetta , Van Der Werf and Paus 2006, Van Der Werf ) and evoked potentials
(Knaflitz ), because in these setups the stimulation artefact and the signal of
interest are well separated in the time domain but highly overlapping in the
frequency domain (Rossi ). Since our aim was to provide neural recordings during
stimulation, we avoided employing this technique.Moreover, in a typical stimulation therapy (for PD or Dystonia) the
neural signals of interest, which typically are on the order of 1–10
μV when measured from DBS electrodes, are up to six
orders of magnitude weaker than the stimulation pulses. As a result, a high gain
is required to make the neural signals detectable by the front-end electronics
and the subsequent ADC blocks. However, this high amplification is applied on
both neural signals and stimulation artefact, which often leads to the
saturation of the front-end amplifier. Hence, the approach to completely shift
the stimulation suppression to the digital domain by using FIR filtering or
template subtraction techniques increases the risk of saturation (Rossi ).Besides that, template subtraction techniques suffer from varying
artefact morphology stemming from undersampling the artefact shape and
misalignment between stimulation and sample timing (Qian , Zhou ). Similarly, adaptive
filtering methods filter the stimulation pulse (Mendrela ) or the artefact recorded on a
neighbouring channel (Basir-Kazeruni ) in order to estimate and subtract the artefact
while filter coefficients are adapted. According to Zhou ), these subtraction
methods can be implemented with low latency, but require artefact detection,
template building and on-board memory for template storage. To avoid signal
distortion, templates must be regularly updated to track any changes in artefact
shape or stimulation waveform. Furthermore, estimated templates often take time
to converge, resulting in varying levels of cancellation over time. Finally,
high dynamic range front-ends are required for subtraction and component
decomposition techniques, since the undistorted artefact waveform has to be
recorded.Reconstruction methods remove samples contaminated with artefacts and
replace them with interpolated values. More specifically, sample-and-hold
methods hold over the last known good sample for the duration of each artefact
(Montgomery , Hartmann ). This procedure requires only a single sample of memory, but
may cause significant distortion (Zhou
). To reduce distortion, samples may
be replaced by linear interpolation between the nearest clean samples (Heffer and Fallon 2008, Zhou ), an
estimation from a learned Gaussian probability density (Hoffmann ) for data segments, or
a reconstruction using cubic spline interpolation (Waddell ). Although simple to
implement, reconstruction methods require artefact detection. This can be done
using blind detection algorithms (Montgomery
, Heffer and Fallon 2008, Hoffmann
), or using timing indicators from the
stimulator (Zhou ). These methods lose information during the artefact, degrading
the achieved SNR (Zhou ).Taking into consideration the above discussion of the digital artefact
suppression approaches proposed so far, it can be argued that an alternative
strategy for recording in real time artefact-free LFP signals during stimulation
could be based on the efficient application of a combination of analog and
digital filters. The preservation of real-time operation is particularly
important in neuromodulation because the stimulation must change in real time
based on the measured state of the neural network (Stanslaski ). The strategy of
removing all of the artefacts by employing (usually high-order) digital
filtering techniques could lead to significant delays in data processing and
challenge the practicality of a real-time, closed-loop system that would employ
a digital only artefact removal strategy; thus this strategy was also put
aside.The finally adopted approach was to introduce a Bainter analog notch
filter to increase the available bandwidth by suppressing the artefact
originating from the stimulation frequency (=140 Hz) and apply high-order
low-pass filtering to reject the higher-frequency harmonics. For convenience and
testing purposes the low-pass filtering was realised in the digital domain to
facilitate the experimental study of our approach. Conceivably, however, a
high-order analog low-pass filter could also be used to reject high-order
harmonics albeit at the expense of size for the externalised device and limited
flexibility during the experimental testing of our approach. The Bainter notch
filter topology was selected because its Q is dependent on the gain of the
amplifiers as opposed to component matching. Consequently, the notch depth is
not sensitive to temperature drift or aging (Baker 2015). The analog notch
filter introduces a negligible delay in the signal processing chain. The digital
low-pass filtering block was provided by the Powerlab 16/35 system and
introduced a processing delay of 75 ms. This delay is in full agreement with the
delays introduced by wearable recording systems that apply real-time digital
signal processing (Salehizadeh ).Another reason for introducing an analog notch filter in the signal
chain is to suppress the stimulation interference produced by the
electrode/tissue impedance mismatch. This mismatch exists even in a symmetric
sensing and stimulation setup and is hard to be controlled within a biological
environment (Stanslaski ). Therefore, an AFE that includes an eight-pole Bainter notch
filter with Chebyshev response was designed, developed and tested in
vitro and in vivo. Besides that, a comparison, in
terms of recording quality and artefact suppression capability, between the
designed AFE, an identical AFE employing an 8th order Bessel notch filter, and
an AFE that does not include any analog band-stop filtering and rejects all the
artefacts digitally, was drawn and measured results were presented.Since the next harmonic after the main stimulation frequency is at 280
Hz, we decided to gradually increase the available bandwidth from 140 Hz to 250
Hz and calculate the normalised RMSE between the recorded signals in and without
the presence of stimulation. In both unipolar and bipolar stimulation, when the
bandwidth is restricted between 0.5 and 50 Hz, the RMSE values of the three
channels are kept low. However, the channel lacking notch filtering presents
unacceptably high RMSE values when the bandwidth is extended beyond 50 Hz. This
finding, which stresses the necessity of using an analog notch filter for
artefact suppression, is in accordance with results measured from existing DBS
systems that offer a passband reaching 100 Hz. Those results showed that
prominent stimulation artefacts existed on the raw LFP trace, consisting of both
harmonic repetitions of the stimulation frequency and aliasing artefacts (Pinnell ).
Regarding the Chebyshev and Bessel notch channels, no significant change in the
calculated RMSE was introduced by the extension of the recording bandwidth to
250 Hz, which leads to the conclusion that the artefacts introduced by the 280
Hz harmonic (and higher ones) do not significantly affect the quality of the
recorded signals when adequately suppressed by a high-order low-pass filter at
250 Hz.A thorough examination of the measured results presented in this paper
leads to the conclusion that the sensing and stimulation artefact suppression
capabilities of the Chebyshev notch channel outperform the capabilities of the
Bessel notch channel. Another important finding is that the proposed AFE
architecture provides LFP recordings that are not affected by aliasing artefacts
located at frequencies that are not harmonic repetitions of the stimulation
frequency. Hence, bearing in mind the desire for wide bandwidth LFP recording
with DBS artefacts suppressed, the approach studied here suggests that cascading
two analog high-order notch filters at 140 and 280 Hz with a steep, high-order,
also analog, low-pass filter of a cut-off frequency ~400 Hz should enable
practical, low delay, high-quality and higher bandwidth (~400 Hz) LFP
recording with DBS artefacts strongly suppressed; at the expense of somewhat
increased size and power consumption of the externalised AFE. The full digital
realisation in the Lab (e.g. via Powerlab) of the same architecture (two notch
filters and a low-pass one) introduces a total approximate delay higher than 0.6
s [2 × 270 ms (notch filters) + 75 ms (low-pass)] which challenges the
practicality of a fast, closed-loop neuromodulation system.One of the important merits of the proposed approach is that biosignal
blanking during stimulation is avoided. Moreover, the proposed artefact
suppression strategy allows for artefact-free LFP recordings during monophasic
DBS, which can be perceived as the worst case scenario, since in biphasic DBS
the artefact and the electrochemical DC offsets (Zhou ) produced are weaker (figures 4(d) and (e)). Indeed, it should be
stressed that in both in vitro and in vivo
experimental setups, LFP signals were successfully recorded during stimulation
even in the presence of inherent asymmetries introduced by
the use of segmented DBS electrodes that allowed differential-mode interference
to enter the signal chain. Furthermore, the design decision to directly connect
the front-end amplifier to the DBS electrodes allowed us to avoid the
introduction of a passive high-pass filter network at the first stage of the AFE
and the subsequent noise deterioration and CMRR reduction this strategy would
entail. The input bias current of the front-end INA (150 nA maximum) is lower
than the limit imposed by the IEC 60601-1 standard (maximum allowable patient
auxiliary current for Type B, normally connected applied parts equals 10
μA). Thus, the designed AFE complies with safety
requirements.Finally, when pairs of electrodes are used for measuring differential
voltages from the human body (invasively or non-invasively), it is recommended
to use the same material for each of the electrodes because, in such a case,
their half-cell potentials are approximately equal. According to Webster (2010),
this strategy: (1) ensures that the net DC potential seen at the input of the
amplifier connected to the electrodes is relatively small, and (2) minimizes
possible saturation effects in the case of high-gain direct-coupled amplifiers.
Hence, since our proposed high-gain front-end INA (with high-pass
characteristics) differentially records LFP signals from two same material
(platinum-iridium) contacts of a DBS electrode, the differential DC offset that
is produced at the electrode-tissue interface is in the order of tens of
millivolts (Denison ), and it is thus removed by our system (the maximum dc offset
rejection that can be accomplished by our system in its current format is
±32 mV, which is in full agreement with the rejection offered by recent
bidirectional neural interface systems (Greenwald ), as verified by the
in vitro and in vivo measured experimental
results presented here. Based on the datasheet of the front-end INA (AD8429)
chip, the maximum electrochemical DC offset that can be rejected by our system
is approximately equal to 2.4 V, so long as it is presented as a common-mode
signal at both inputs of our Chebyshev notch AFE.
Limitations and further improvements
All of the practical merits described above, are achieved at the expense
of signal loss in the frequency band ranging from 125 to 155 Hz and the
appearance of noise in this frequency band when extremely weak (0.32
μV rms) LFP signals, which are close to the
Chebyshev notch channel’s noise floor (~0.1
μV rms), are recorded (figure 8(b)). Crucially, the impulse response of the analog notch
filter when stimulated by a DBS pulse is characterized by low-amplitude and
short-duration transient ringing (figures 4(d) and
(e)), which, as the measured results in figures 8(c), 10 and 11 show, does not affect the quality of the
recorded LFP signals. However, even this short and weak ringing can be avoided
by adding an analog multiplexer which would allow the user to introduce, through
software, the analog notch filter in the signal chain just before the onset of
DBS and exclude it from the signal chain before the termination of stimulation,
e.g. 15 ms before the last stimulation pulse. In this way, the weak ringing
effect that may be introduced by the Chebyshev notch filter after the last DBS
pulse it senses, would not appear after the (actual) last stimulation pulse, and
thus it would not interfere with evoked resonant neural activity, which is a
physiological signal of interest that appears 4 ms after the (actual) last DBS
pulse and lasts for ~20 ms (Sinclair
). Furthermore, in a future version of
this AFE (comprised, for example, of two high-order analog notch filters and one
high-order analog low-pass filter, as explained above) the introduction of
digitally selectable/tunable notch and cut-off frequencies would provide the
researchers and clinicians with the ability to suppress more than one
stimulation rate and preserve wide bandwidth recording.Finally, the existence of small intrinsic errors produced by the
Chebyshev and Bessel notch channels (figure
13) mainly results from the dc offset voltage, which is produced by
the electrodes and is not completely rejected by the system. However, further
rejection of this differential dc offset (up to 85 mV of differential dc offset
voltage rejection can be achieved, as reported in table 2) could be accomplished by introducing a passive 1st
order high-pass filter between the front-end INA and notch filtering stages (in
other words between stages 1 and 2 of the current AFE design shown in figure 1). The addition of this high-pass
filter block would introduce a small delay in the transient response of the
system (Henderson and Kautz 1958), however, this delay may be considered
tolerable taking into account the enhancement this strategy could offer in the
recording capabilities of the Chebyshev notch channel during stimulation.
Pathophysiological significance
The key aim for this system is to remove existing constraints for
clinical neuroscience discovery with bi-directional neural interfaces. The major
attributes of our system are its wide pass band and low noise floor relative to
other devices allowing the recording of signals during stimulation at the same
site with greater resolution, especially at higher frequencies. An alternative
approach is to shift attention from LFPs in subcortical nuclei to
electrocorticographic recordings as tractable feedback biomarkers for
closed-loop stimulation (De Hemptinne , Herron
, Swann , 2018). In some cases, this is necessary as electrocorticographic
signals have better signal-to-noise ratios and are spatially separated from DBS
sites, so they are less corrupted by stimulation-induced artefact.However, subcortical recordings have their own merits. These include the
inherent convergence in basal ganglia targets, so that extensive cortical
regions can be modulated, and clearer pathological correlates (Oswal ).
Moreover, the recording of signals from the same electrodes as used for DBS
limits instrumentation of the brain and thereby the incremental morbidity and
expense. The currently described system facilitates consideration of LFP
activity as a source of feedback control. Specifically, it widens the potential
feature space beyond the beta band (Little
, Priori , Arlotti ) to lower amplitude, higher
frequency activities. These include finely tuned gamma activity centred around
70 Hz and associated with dyskinesias (Fogelson
), and high frequency oscillations of
over 200 Hz in frequency, together with related phase amplitude coupling which
have both been linked to bradykinesia and rigidity (Meidahl ). In addition,
stimulation evoked subcortical potentials have high frequency components that
may carry information about targeting and motor impairment in Parkinson’s
disease (Sinclair ). A richer feature space also improves the capability of machine
learning approaches to identify control signals (Shah , Yao ). In its current embodiment, the
large power dissipation would not allow for placement of the circuit in an
implantable system. The intention is to achieve a more complete sampling of the
physiomarker space, and based on what is discovered, define a bespoke
application specific integrated circuit (ASIC) that would provide the resolution
required within the power constraints of the implant.To sum up, from the pathophysiologic point of view, being able to extend
the available bandwidth for LFP recording from the target stimulation site
constitutes pivotal progress for developing novel closed-loop neurostimulation
systems that use low- and higher-frequency LFPs as control signals. The
in vivo experimental results presented in this study show
that our recording system: 1) does not saturate during DBS, and 2) is able to
provide artefact-free LFP recordings during DBS. The next step towards
developing the aforementioned closed-loop DBS technology would be to use the
proposed AFE architecture in a clinical study to continuously record LFPs during
DBS from the STN of parkinsonian patients and, more specifically, from the
neural tissue surrounding the stimulating electrode. This study would allow the
identification of changes in LFP rhythms induced by DBS and the determination of
features extracted from the LFP signals that could be used to regulate and
optimise ongoing DBS.
Conclusion
The novel and versatile Chebyshev notch channel designed, realised and
tested in vitro and in vivo, allows for real-time,
low-noise and artefact-free LFP recordings during DBS for a bandwidth of
0.5–250 Hz. Proof of the proposed channel’s recording and artefact
suppression capabilities has been provided and its performance has been evaluated
quantitatively by means of a series of in vitro and in
vivo experiments and comparisons with commercial high-performance
biopotential acquisition systems. It has been proven that the designed AFE is able
to reliably record weak LFP signals (1–10 μV peak in
figures 10, 11 and 15), in and without the
presence of either unipolar or bipolar DBS, which renders it a functional and
practical AFE architecture to be utilised in a wide range of applications and
environments. This work is the first step towards developing a closed-loop
neurostimulation system that uses low and higher-frequency LFPs as control
signals.
Authors: Nicole C Swann; Coralie de Hemptinne; Svjetlana Miocinovic; Salman Qasim; Sarah S Wang; Nathan Ziman; Jill L Ostrem; Marta San Luciano; Nicholas B Galifianakis; Philip A Starr Journal: J Neurosci Date: 2016-06-15 Impact factor: 6.167
Authors: Hayriye Cagnan; David Pedrosa; Simon Little; Alek Pogosyan; Binith Cheeran; Tipu Aziz; Alexander Green; James Fitzgerald; Thomas Foltynie; Patricia Limousin; Ludvic Zrinzo; Marwan Hariz; Karl J Friston; Timothy Denison; Peter Brown Journal: Brain Date: 2016-12-22 Impact factor: 13.501
Authors: Mattia Arlotti; Matteo Colombo; Andrea Bonfanti; Tomasz Mandat; Michele Maria Lanotte; Elena Pirola; Linda Borellini; Paolo Rampini; Roberto Eleopra; Sara Rinaldo; Luigi Romito; Marcus L F Janssen; Alberto Priori; Sara Marceglia Journal: Front Neurosci Date: 2021-12-07 Impact factor: 4.677
Authors: Jean Debarros; Lea Gaignon; Shenghong He; Alek Pogosyan; Moaad Benjaber; Timothy Denison; Peter Brown; Huiling Tan Journal: Annu Int Conf IEEE Eng Med Biol Soc Date: 2020-07