| Literature DB >> 31142883 |
Duraiyarassu Uthaman1, Suman Lata Gupta1, Sandeep Kumar Mishra1, Satyen Parida1, Prasanna Udupi Bidkar1, Muthupillai Senthilnathan1.
Abstract
BACKGROUND AND AIMS: Ambu® AuraGain™ laryngeal airway (AuraGain) is one of the newer supraglottic airway device introduced in 2014. Cervical spine stabilisation with hard cervical collar makes insertion of supraglottic airways and tracheal intubation difficult. This study was conducted to investigate whether the presence of a cervical collar affects the oropharyngeal sealing pressure (OSP) and fibreoptic view of the glottis (Brimacombe score) in airways secured with the AuraGain.Entities:
Keywords: Ambu AuraGain; hard cervical collar; laryngeal airway; oropharyngeal sealing pressure; supraglottic airway
Year: 2019 PMID: 31142883 PMCID: PMC6530280 DOI: 10.4103/ija.IJA_787_18
Source DB: PubMed Journal: Indian J Anaesth ISSN: 0019-5049
Demographics
| Parameter | Patients ( | ||
|---|---|---|---|
| Age (Mean and SD) (years) | 40.4±11.3 | ||
| Gender (%) | Male:16 (45.7%); Female:19 (54.3%) | ||
| Weight (kg) | 63.9±8 | ||
| Height (cm) | 1.5±0.06 | ||
| BMI (kg/cm2) | 24.9±2.9 | ||
| MMP Class 1/2/3 | 9/18/8 (Median - 2) | ||
| ASA Class I/II/III | 20/13/2 | ||
| Sequence of insertion A/B | 18/17 | ||
| Age (Mean and SD) years | 39.8±10.3 | 41±12.7 | 40.4±11.3 |
| Gender (%) | Male:7 Female:11 | Male:9 Female:8 | Male:16 (45.7%) Female:19 (54.3%) |
| Weight (kg) | 63.4±5.2 | 64.4±2.8 | 63.9±8 |
| Height (m) | 1.5 | 1.5 | 1.5 |
| BMI (kg/m2) | 24.6±3.2 | 25.2±2.4 | 24.9±2.9 |
| MMP Class 1/2/3 | 4/10/4 | 5/8/4 | 9/18/8 (Median - 2) |
| ASA Class 1/2/3 | 8/8/2 | 12/5/0 | 20/13/2 |
Comparison of OSP and Brimacombe score
| Variable | Without collar | With collar | |
|---|---|---|---|
| Mean OSP (Mean±SD) | 29.6±3.7 | 30.17±3.1 | 0.310 |
| Brimacombe score 1/2/3/4 | 0/0/4/31 | 0/1/8/26 | 0.083 |
Analysis of insertion parameters
| Parameter | Without collar | With collar | |
|---|---|---|---|
| No of attempts:1/2 | 32/3 | 31/4 | 0.705 |
| Ease of insertion of Ambu:1/2/3/4 | 25/8/0/2 | 14/12/1/8 | 0.001* |
| Ease of insertion of nasogastric tube 1/2 | 32/3 | 26/9 | 0.04* |
| Time taken to insert (seconds) | 21.3±9.6 | 26.1±11.7 | 0.025* |
*P<0.05