Assessing the concurrent validity of a gait analysis system integrated into a smart walker in older adults with gait impairments.

Christian Werner1,2, Georgia Chalvatzaki3, Xanthi S Papageorgiou3, Costas S Tzafestas3, Jürgen M Bauer1,2, Klaus Hauer2.   

Abstract

OBJECTIVE: To assess the concurrent validity of a smart walker-integrated gait analysis system with the GAITRite® system for measuring spatiotemporal gait parameters in potential users of the smart walker.
DESIGN: Criterion standard validation study.
SETTING: Research laboratory in a geriatric hospital. PARTICIPANTS: Twenty-five older adults (⩾65 years) with gait impairments (habitual rollator use and/or gait speed <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination ⩾17). MAIN MEASURES: Stride, swing and stance time; stride length; and gait speed were simultaneously recorded using the smart walker-integrated gait analysis system and the GAITRite system while participants walked along a 7.8-m walkway with the smart walker. Concurrent criterion-related validity was assessed using the Bland-Altman method, percentage errors (acceptable if <30%), and intraclass correlation coefficients for consistency (ICC3,1) and absolute agreement (ICC2,1).
RESULTS: Bias for stride, swing and stance time ranged from -0.04 to 0.04 seconds, with acceptable percentage errors (8.7%-23.0%). Stride length and gait speed showed higher bias (meanbias (SD) = 0.20 (0.11) m; 0.19 (0.13) m/s) and not acceptable percentage errors (31.3%-42.3%). Limits of agreement were considerably narrower for temporal than for spatial-related gait parameters. All gait parameters showed good-to-excellent consistency (ICC3,1 = 0.72-0.97). Absolute agreement was good-to-excellent for temporal (ICC2,1 = 0.72-0.97) but only poor-to-fair for spatial-related gait parameters (ICC2,1 = 0.37-0.52).
CONCLUSION: The smart walker-integrated gait analysis system has good concurrent validity with the GAITRite system for measuring temporal but not spatial-related gait parameters in potential end-users of the smart walker. Stride length and gait speed can be measured with good consistency, but with only limited absolute accuracy.

Entities:  

Keywords:  Smart walker; assistive devices; elderly; gait analysis; validity

Mesh:

Year:  2019        PMID: 31131630      PMCID: PMC6745611          DOI: 10.1177/0269215519852143

Source DB:  PubMed          Journal:  Clin Rehabil        ISSN: 0269-2155            Impact factor:   3.477


Introduction

Recent technological developments in gait analysis focus on ambulatory solutions that allow for unobtrusive and continuous gait monitoring in real-life environments outside the laboratory such as wearable sensors.[1] However, these sensors require an individual’s willingness to wear them and may cause discomfort and adherence issues. In addition, to our knowledge, the validity of body-worn sensors for measuring spatiotemporal gait parameters in older adults with walking aids is still unknown. Considering that walkers or rollators are prescribed routinely to patients during geriatric rehabilitation and that many older adults with mobility limitations have to use them for ambulation, there is the need for valid gait analysis systems to unobtrusively and continuously capture spatiotemporal gait parameters also in these walking aid users. Technological advances have led to the development of ‘smart walkers’ with various high-tech functionalities such as monitoring a user’s gait.[2] Different sensor types (e.g. vision-based sensors,[3] inertial measurement units,[4] force sensors[5]) have been used to implement gait analysis on a smart walker. Independent of the technical implementations, to our knowledge, previous validation studies of smart walker–integrated gait analysis systems suffered from methodological shortcomings such as small sample sizes, participants not representative of potential users, no criterion standard comparisons and/or no statistical analyses.[3-5] The study aim was to assess the concurrent validity of a smart walker–integrated gait analysis for measuring spatiotemporal gait parameters with a criterion standard (GAITRite® system) in a reasonable number of potential smart walker users.

Methods

The study was conducted between 1 November and 5 December 2014, with approval of the ethics committee of the Medical Faculty of the Heidelberg University (S-358/2013) and in accordance with the Declaration of Helsinki. All participants gave written informed consent. Participants were recruited from rehabilitation wards of a geriatric hospital, from nursing homes and from a hospital-associated sports club for geriatric outpatient rehabilitation. According to the defined users of our smart walker,[6] inclusion criteria were age ⩾65 years, moderate gait impairments (rollator use in daily life and/or 4 m usual gait speed[7] <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination[8] score ⩾17 points). The GAITRite system (CIR Systems Inc., Havertown, PA, USA) is an electronic walkway with embedded pressure sensors, representing a well-established and validated method for automated gait analysis in clinical settings.[9] The GAITRite system used in this study was 5.79 m long and 0.89 m wide (active area: 4.88 m × 0.61 m; sampling rate 120 Hz). The smart walker integrates innovative functionalities such as sit-to-stand assistance, obstacle avoidance, navigation assistance and gait monitoring. A detailed description of all its functionalities has been provided previously.[6,10] For this study, only the gait analysis system of the smart walker was activated and all other innovative functionalities were deactivated. The smart walker–integrated gait analysis system is based on a standard laser range finder (UBG-04LX-F01; Hokuyo Automatic Co., Ltd, Osaka, Japan; sampling period 28 ms/scan) mounted at the rear side of the four-wheeled smart walker at a fixed height of 35 cm from the ground with a viewing direction towards the user’s legs to record their motion at a horizontal plane below the knee level. Gait parameters were extracted by pre-processing the laser data using a Probabilistic Data Association Particle Filtering system and subsequent modelling of the user’s walking pattern based on a Hidden Markov Model approach, as previously described.[11] The overall goal of this gait analysis system is not only to validly measure gait parameters continuously during smart walker use but also to serve as a basis for future development of a context-aware smart walker that generates and provides real-time assistive actions (e.g. distance/velocity adjustments) according to the user‘s current walking pattern. After a familiarization phase, in which participants freely moved around with the smart walker for approximately 2–5 minutes, they were instructed to walk along a GAITRite instrumented walkway with the smart walker at self-selected maximum gait speed. Each walk was initiated and terminated 1 m before and after the walkway (total length = 7.79 m) to account for acceleration and deceleration. No practice trials were performed on the instrumented walkway. After data recording, each walk was checked to ensure that the same steps, and the same number of steps, were used to calculate mean values for spatiotemporal gait parameters (stride, swing and stance time; stride length; and gait speed) by both processing methods. Mean values were used because average gait parameters are usually of clinical interest. Between-method differences (bias) and 95% limits of agreement (meanbias ± 1.96 × SDbias) were determined using the Bland–Altman method.[12] Percentage errors, calculated as 100 × (1.96 × SDbias)/((meansmart walker + meanGAITRite)/2), were considered to be clinically acceptable if <30%.[13] Intraclass correlation coefficients (ICCs) with 95% confidence intervals were calculated to determine the consistency (ICC3,1) and absolute agreement (ICC2,1) between the mean gait parameters measured by the two methods. ICCs were interpreted as poor (<0.40), fair to good (0.40–0.75) and excellent (>0.75).[14] The sample size for this study was estimated to be ⩾23 participants, based on an acceptable ICC of 0.70 and an expected ICC of 0.90 for two measurements (smart walker and GAITRite), a significance level (α) of 0.05 and a statistical power (1–β) of 0.80.[15] A two-sided P-value of <0.05 indicated statistical significance. Statistical analysis was performed using IBM SPSS Statistics for Windows, Version 25.0 (IBM Corp., Armonk, NY, USA).

Results

The sample included 25 older adults with a mean (SD) age of 84.1 (5.4) years, moderate gait impairments (usual gait speed = 0.48 (0.15) m/s) and no severe cognitive impairment (Mini-Mental State Examination score = 24.5 (4.1) points). Sixteen (64%) participants were geriatric rehabilitation patients, seven (28%) were members of the sports club for geriatric outpatient rehabilitation, and two (8%) were nursing home residents. Mean bias for the stride, swing and stance time ranged from –0.04 to 0.04 seconds, with clinically acceptable percentage errors (8.7%–23.0%) (Table 1). Stride length and gait speed showed both a substantially higher bias (meanbias (SD) = 0.20 (0.11) m; 0.19 (0.13) m/s) and a clinically not acceptable percentage error (31.3%–42.3%). Limits of agreement were considerably narrower for the stride (–0.10 to 0.11 seconds), swing (–0.07 to 0.16 seconds) and stance time (–0.12 to 0.20 seconds) than for the stride length (–0.07 to 0.44 m) and gait speed (–0.02 to 0.42 m/s) (Figure 1). Consistency between both methods was good to excellent for all gait parameters (ICC3,1 = 0.72–0.97). Absolute agreement was also good to excellent for the stride, swing and stance time (ICC2,1 = 0.72–0.97), but only poor to fair for the stride length (ICC2,1 = 0.37) and gait speed (ICC2,1 = 0.52).
Table 1.

Mean values (±SD), mean difference scores (bias ± SD), limits of agreement, mean percentage errors and intraclass correlation coefficients for consistency and absolute agreement for each gait parameter.

Gait parameterGAITRite®Smart walkerBias95% LOAPEConsistencyICC3,1 (95% CI)Absolute agreementICC2,1 (95% CI)
Stride time (s)1.21 ± 0.241.21 ± 0.22–0.01 ± 0.05–0.10 to 0.118.70.97 (0.94 to 0.99)0.97 (0.94 to 0.98)
Swing time (s)0.48 ± 0.090.52 ± 0.10–0.04 ± 0.06–0.07 to 0.1622.80.80 (0.59 to 0.91)0.72 (0.27 to 0.89)
Stance time (s)0.73 ± 0.180.69 ± 0.130.04 ± 0.08–0.12 to 0.2023.00.86 (0.70 to 0.93)0.83 (0.62 to 0.93)
Stride length (m)0.80 ± 0.170.60 ± 0.130.20 ± 0.11–0.02 to 0.4231.30.72 (0.45 to 0.86)0.37 (–0.10 to 0.73)
Gait speed (m/s)0.70 ± 0.220.51 ± 0.150.19 ± 0.13–0.07 to 0.4442.30.76 (0.53 to 0.89)0.52 (–0.10 to 0.82)

LOA: limits of agreement; PE: percentage error; ICC: intraclass correlation coefficient; CI: confidence interval.

All ICCs for consistency and absolute agreement were significant at P < 0.001.

Figure 1.

Bland–Altman plots for comparison between the GAITRite® system and the smart walker–integrated gait analysis system. Dotted lines indicate bias and dashed lines indicate upper and lower 95% limits of agreement (±1.96 SD of the bias).

Mean values (±SD), mean difference scores (bias ± SD), limits of agreement, mean percentage errors and intraclass correlation coefficients for consistency and absolute agreement for each gait parameter. LOA: limits of agreement; PE: percentage error; ICC: intraclass correlation coefficient; CI: confidence interval. All ICCs for consistency and absolute agreement were significant at P < 0.001. Bland–Altman plots for comparison between the GAITRite® system and the smart walker–integrated gait analysis system. Dotted lines indicate bias and dashed lines indicate upper and lower 95% limits of agreement (±1.96 SD of the bias).

Discussion

This initial validation study showed that the smart walker–integrated gait analysis system provides comparable data to the GAITRite system for the temporal gait parameters of stride, swing and stance time in potential smart walker users. Although also a good consistency for the stride length and gait speed was found between these two systems, they cannot be used interchangeably when absolute values of these spatial-related gait parameters are required (e.g. for comparison with normative values). The low absolute agreement for spatial-related gait parameters can be explained by the fact that the GAITRite system refers to the distance between heel contacts on the electronic walkway for measuring the stride length, while the smart walker–integrated gait analysis system refers to the distance between leg placements recorded by the laser range finder 35 cm above the walkway. The reference points of the laser range finder are closer to the pivot of the lower legs (i.e. knee joint) and thus travel a shorter distance during the gait cycle, resulting in the shorter stride length and also the lower gait speed. While absolute agreement for these spatial-related parameters seems lacking, the extent to which they agree with the GAITRite system on the relative values (i.e. consistency) was good to excellent, suggesting that the stride length and gait speed of the smart walker–integrated gait analysis system may be themselves reliable and as good as those of the GAITRite system in determining meaningful changes in a user’s walking pattern. Compared to previous validation studies of smart walker–integrated gait analysis systems, the strengths of this study are that a reasonable number of participants representative of potential smart walker users were recruited, a well-established, validated gait analysis system was used as criterion standard for comparison, and the data obtained were analysed by adequate statistical methods. However, this study also has some limitations. Only short straight walking in a controlled laboratory environment was evaluated, as limited by our criterion standard. Future studies should assess the validity of the smart walker-gait analysis system in less constrained movement situations. Our participants were predominantly females, limiting the generalizability of the results to males. However, we did not expect gender to affect the concurrent validity between the two systems. The smart walker–integrated gait analysis system can provide clinicians and researchers the ability to unobtrusively capture gait parameters of smart walker users, without any sensors being attached to the user’s body. Our study represents a first step towards a continuous gait analysis of smart walker users in natural environments. The applicability of the system for such long-term gait monitoring needs to be confirmed in future studies. The smart walker–integrated gait analysis system has good concurrent validity with the GAITRite® system for measuring temporal gait parameters in potential smart walker users. Stride length and gait speed can also be measured consistently; however, modifications are recommended to improve the absolute measurement accuracy for these spatial-related gait parameters.
  10 in total

1.  Lower extremity function and subsequent disability: consistency across studies, predictive models, and value of gait speed alone compared with the short physical performance battery.

Authors:  J M Guralnik; L Ferrucci; C F Pieper; S G Leveille; K S Markides; G V Ostir; S Studenski; L F Berkman; R B Wallace
Journal:  J Gerontol A Biol Sci Med Sci       Date:  2000-04       Impact factor: 6.053

2.  "Mini-mental state". A practical method for grading the cognitive state of patients for the clinician.

Authors:  M F Folstein; S E Folstein; P R McHugh
Journal:  J Psychiatr Res       Date:  1975-11       Impact factor: 4.791

3.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

4.  Sample size formulas for estimating intraclass correlation coefficients with precision and assurance.

Authors:  G Y Zou
Journal:  Stat Med       Date:  2012-07-04       Impact factor: 2.373

Review 5.  A review of the functionalities of smart walkers.

Authors:  Maria Martins; Cristina Santos; Anselmo Frizera; Ramón Ceres
Journal:  Med Eng Phys       Date:  2015-08-22       Impact factor: 2.242

6.  Validity of the GAITRite walkway system for the measurement of averaged and individual step parameters of gait.

Authors:  Kate E Webster; Joanne E Wittwer; Julian A Feller
Journal:  Gait Posture       Date:  2004-12-10       Impact factor: 2.840

7.  User-Oriented Evaluation of a Robotic Rollator That Provides Navigation Assistance in Frail Older Adults with and without Cognitive Impairment.

Authors:  Christian Werner; George P Moustris; Costas S Tzafestas; Klaus Hauer
Journal:  Gerontology       Date:  2017-11-29       Impact factor: 5.140

Review 8.  Toward Pervasive Gait Analysis With Wearable Sensors: A Systematic Review.

Authors:  Shanshan Chen; John Lach; Benny Lo; Guang-Zhong Yang
Journal:  IEEE J Biomed Health Inform       Date:  2016-11       Impact factor: 5.772

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

10.  Automatic Assessment of a Rollator-User's Condition During Rehabilitation Using the i-Walker Platform.

Authors:  Joaquin Ballesteros; Cristina Urdiales; Antonio B Martinez; Marina Tirado
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2017-04-25       Impact factor: 3.802

  10 in total
  7 in total

1.  Smartphone-Based Body Location-Independent Functional Mobility Analysis in Patients with Parkinson's Disease: A Step towards Precise Medicine.

Authors:  Diogo Vila-Viçosa; Mariana Leitão; Raquel Bouça-Machado; Filipa Pona-Ferreira; Sara Alberto; Joaquim J Ferreira; Ricardo Matias
Journal:  J Pers Med       Date:  2022-05-19

2.  Quantitative analysis of the bilateral coordination and gait asymmetry using inertial measurement unit-based gait analysis.

Authors:  Seung Hwan Han; Chang Oh Kim; Kwang Joon Kim; Jeanhong Jeon; Hsienhao Chang; Eun Seo Kim; Hoon Park
Journal:  PLoS One       Date:  2019-10-01       Impact factor: 3.240

3.  Rehabilitation of older people with Parkinson's disease: an innovative protocol for RCT study to evaluate the potential of robotic-based technologies.

Authors:  Roberta Bevilacqua; Elvira Maranesi; Mirko Di Rosa; Riccardo Luzi; Elisa Casoni; Nadia Rinaldi; Renato Baldoni; Fabrizia Lattanzio; Valentina Di Donna; Giuseppe Pelliccioni; Giovanni Renato Riccardi
Journal:  BMC Neurol       Date:  2020-05-13       Impact factor: 2.474

4.  Concurrent Validity of GAITRite and the 10-m Walk Test to Measure Gait Speed in Adults with Chronic Ankle Instability.

Authors:  Ho Kim; Dongmin Kum; Insu Lee; Jongduk Choi
Journal:  Healthcare (Basel)       Date:  2022-08-09

5.  Algorithm based on one monocular video delivers highly valid and reliable gait parameters.

Authors:  Arash Azhand; Sophie Rabe; Swantje Müller; Igor Sattler; Anika Heimann-Steinert
Journal:  Sci Rep       Date:  2021-07-07       Impact factor: 4.379

6.  Randomised controlled trial assessing the effect of a technology-assisted gait and balance training on mobility in older people after hip fracture: study protocol.

Authors:  Elvira Maranesi; Giovanni Renato Riccardi; Fabrizia Lattanzio; Mirko Di Rosa; Riccardo Luzi; Elisa Casoni; Nadia Rinaldi; Renato Baldoni; Valentina Di Donna; Roberta Bevilacqua
Journal:  BMJ Open       Date:  2020-06-15       Impact factor: 2.692

Review 7.  Walking with rollator: a systematic review of gait parameters in older persons.

Authors:  Marion Mundt; Joao Pedro Batista; Bernd Markert; Cornelius Bollheimer; Thea Laurentius
Journal:  Eur Rev Aging Phys Act       Date:  2019-09-10       Impact factor: 3.878

  7 in total