Christian Werner1,2, Georgia Chalvatzaki3, Xanthi S Papageorgiou3, Costas S Tzafestas3, Jürgen M Bauer1,2, Klaus Hauer2. 1. Centre for Geriatric Medicine, Heidelberg University, Heidelberg, Germany. 2. Agaplesion Bethanien Hospital Heidelberg, Geriatric Centre at the Heidelberg University, Heidelberg, Germany. 3. School of Electrical and Computer Engineering, National Technical University of Athens, Athens, Greece.
Abstract
OBJECTIVE: To assess the concurrent validity of a smart walker-integrated gait analysis system with the GAITRite® system for measuring spatiotemporal gait parameters in potential users of the smart walker. DESIGN: Criterion standard validation study. SETTING: Research laboratory in a geriatric hospital. PARTICIPANTS: Twenty-five older adults (⩾65 years) with gait impairments (habitual rollator use and/or gait speed <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination ⩾17). MAIN MEASURES: Stride, swing and stance time; stride length; and gait speed were simultaneously recorded using the smart walker-integrated gait analysis system and the GAITRite system while participants walked along a 7.8-m walkway with the smart walker. Concurrent criterion-related validity was assessed using the Bland-Altman method, percentage errors (acceptable if <30%), and intraclass correlation coefficients for consistency (ICC3,1) and absolute agreement (ICC2,1). RESULTS: Bias for stride, swing and stance time ranged from -0.04 to 0.04 seconds, with acceptable percentage errors (8.7%-23.0%). Stride length and gait speed showed higher bias (meanbias (SD) = 0.20 (0.11) m; 0.19 (0.13) m/s) and not acceptable percentage errors (31.3%-42.3%). Limits of agreement were considerably narrower for temporal than for spatial-related gait parameters. All gait parameters showed good-to-excellent consistency (ICC3,1 = 0.72-0.97). Absolute agreement was good-to-excellent for temporal (ICC2,1 = 0.72-0.97) but only poor-to-fair for spatial-related gait parameters (ICC2,1 = 0.37-0.52). CONCLUSION: The smart walker-integrated gait analysis system has good concurrent validity with the GAITRite system for measuring temporal but not spatial-related gait parameters in potential end-users of the smart walker. Stride length and gait speed can be measured with good consistency, but with only limited absolute accuracy.
OBJECTIVE: To assess the concurrent validity of a smart walker-integrated gait analysis system with the GAITRite® system for measuring spatiotemporal gait parameters in potential users of the smart walker. DESIGN: Criterion standard validation study. SETTING: Research laboratory in a geriatric hospital. PARTICIPANTS: Twenty-five older adults (⩾65 years) with gait impairments (habitual rollator use and/or gait speed <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination ⩾17). MAIN MEASURES: Stride, swing and stance time; stride length; and gait speed were simultaneously recorded using the smart walker-integrated gait analysis system and the GAITRite system while participants walked along a 7.8-m walkway with the smart walker. Concurrent criterion-related validity was assessed using the Bland-Altman method, percentage errors (acceptable if <30%), and intraclass correlation coefficients for consistency (ICC3,1) and absolute agreement (ICC2,1). RESULTS: Bias for stride, swing and stance time ranged from -0.04 to 0.04 seconds, with acceptable percentage errors (8.7%-23.0%). Stride length and gait speed showed higher bias (meanbias (SD) = 0.20 (0.11) m; 0.19 (0.13) m/s) and not acceptable percentage errors (31.3%-42.3%). Limits of agreement were considerably narrower for temporal than for spatial-related gait parameters. All gait parameters showed good-to-excellent consistency (ICC3,1 = 0.72-0.97). Absolute agreement was good-to-excellent for temporal (ICC2,1 = 0.72-0.97) but only poor-to-fair for spatial-related gait parameters (ICC2,1 = 0.37-0.52). CONCLUSION: The smart walker-integrated gait analysis system has good concurrent validity with the GAITRite system for measuring temporal but not spatial-related gait parameters in potential end-users of the smart walker. Stride length and gait speed can be measured with good consistency, but with only limited absolute accuracy.
Recent technological developments in gait analysis focus on ambulatory solutions that
allow for unobtrusive and continuous gait monitoring in real-life environments
outside the laboratory such as wearable sensors.[1] However, these sensors require an individual’s willingness to wear them and
may cause discomfort and adherence issues. In addition, to our knowledge, the
validity of body-worn sensors for measuring spatiotemporal gait parameters in older
adults with walking aids is still unknown. Considering that walkers or rollators are
prescribed routinely to patients during geriatric rehabilitation and that many older
adults with mobility limitations have to use them for ambulation, there is the need
for valid gait analysis systems to unobtrusively and continuously capture
spatiotemporal gait parameters also in these walking aid users.Technological advances have led to the development of ‘smart walkers’ with various
high-tech functionalities such as monitoring a user’s gait.[2] Different sensor types (e.g. vision-based sensors,[3] inertial measurement units,[4] force sensors[5]) have been used to implement gait analysis on a smart walker. Independent of
the technical implementations, to our knowledge, previous validation studies of
smart walker–integrated gait analysis systems suffered from methodological
shortcomings such as small sample sizes, participants not representative of
potential users, no criterion standard comparisons and/or no statistical
analyses.[3-5]The study aim was to assess the concurrent validity of a smart walker–integrated gait
analysis for measuring spatiotemporal gait parameters with a criterion standard
(GAITRite® system) in a reasonable number of potential smart walker
users.
Methods
The study was conducted between 1 November and 5 December 2014, with approval of the
ethics committee of the Medical Faculty of the Heidelberg University (S-358/2013)
and in accordance with the Declaration of Helsinki. All participants gave written
informed consent.Participants were recruited from rehabilitation wards of a geriatric hospital, from
nursing homes and from a hospital-associated sports club for geriatric outpatient
rehabilitation. According to the defined users of our smart walker,[6] inclusion criteria were age ⩾65 years, moderate gait impairments (rollator
use in daily life and/or 4 m usual gait speed[7] <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination[8] score ⩾17 points).The GAITRite system (CIR Systems Inc., Havertown, PA, USA) is an electronic walkway
with embedded pressure sensors, representing a well-established and validated method
for automated gait analysis in clinical settings.[9] The GAITRite system used in this study was 5.79 m long and 0.89 m wide
(active area: 4.88 m × 0.61 m; sampling rate 120 Hz).The smart walker integrates innovative functionalities such as sit-to-stand
assistance, obstacle avoidance, navigation assistance and gait monitoring. A
detailed description of all its functionalities has been provided
previously.[6,10] For this study, only the gait analysis system of the smart
walker was activated and all other innovative functionalities were deactivated. The
smart walker–integrated gait analysis system is based on a standard laser range
finder (UBG-04LX-F01; Hokuyo Automatic Co., Ltd, Osaka, Japan; sampling period
28 ms/scan) mounted at the rear side of the four-wheeled smart walker at a fixed
height of 35 cm from the ground with a viewing direction towards the user’s legs to
record their motion at a horizontal plane below the knee level. Gait parameters were
extracted by pre-processing the laser data using a Probabilistic Data Association
Particle Filtering system and subsequent modelling of the user’s walking pattern
based on a Hidden Markov Model approach, as previously described.[11] The overall goal of this gait analysis system is not only to validly measure
gait parameters continuously during smart walker use but also to serve as a basis
for future development of a context-aware smart walker that generates and provides
real-time assistive actions (e.g. distance/velocity adjustments) according to the
user‘s current walking pattern.After a familiarization phase, in which participants freely moved around with the
smart walker for approximately 2–5 minutes, they were instructed to walk along a
GAITRite instrumented walkway with the smart walker at self-selected maximum gait
speed. Each walk was initiated and terminated 1 m before and after the walkway
(total length = 7.79 m) to account for acceleration and deceleration. No practice
trials were performed on the instrumented walkway. After data recording, each walk
was checked to ensure that the same steps, and the same number of steps, were used
to calculate mean values for spatiotemporal gait parameters (stride, swing and
stance time; stride length; and gait speed) by both processing methods. Mean values
were used because average gait parameters are usually of clinical interest.Between-method differences (bias) and 95% limits of agreement
(meanbias ± 1.96 × SDbias) were determined using the
Bland–Altman method.[12] Percentage errors, calculated as
100 × (1.96 × SDbias)/((meansmart
walker + meanGAITRite)/2), were considered to be clinically
acceptable if <30%.[13] Intraclass correlation coefficients (ICCs) with 95% confidence intervals were
calculated to determine the consistency (ICC3,1) and absolute agreement
(ICC2,1) between the mean gait parameters measured by the two
methods. ICCs were interpreted as poor (<0.40), fair to good (0.40–0.75) and
excellent (>0.75).[14] The sample size for this study was estimated to be ⩾23 participants, based on
an acceptable ICC of 0.70 and an expected ICC of 0.90 for two measurements (smart
walker and GAITRite), a significance level (α) of 0.05 and a statistical power (1–β)
of 0.80.[15] A two-sided P-value of <0.05 indicated statistical
significance. Statistical analysis was performed using IBM SPSS Statistics for
Windows, Version 25.0 (IBM Corp., Armonk, NY, USA).
Results
The sample included 25 older adults with a mean (SD) age of 84.1 (5.4) years,
moderate gait impairments (usual gait speed = 0.48 (0.15) m/s) and no severe
cognitive impairment (Mini-Mental State Examination score = 24.5 (4.1) points).
Sixteen (64%) participants were geriatric rehabilitation patients, seven (28%) were
members of the sports club for geriatric outpatient rehabilitation, and two (8%)
were nursing home residents.Mean bias for the stride, swing and stance time ranged from –0.04 to 0.04 seconds,
with clinically acceptable percentage errors (8.7%–23.0%) (Table 1). Stride length and gait speed
showed both a substantially higher bias (meanbias (SD) = 0.20 (0.11) m;
0.19 (0.13) m/s) and a clinically not acceptable percentage error (31.3%–42.3%).
Limits of agreement were considerably narrower for the stride (–0.10 to
0.11 seconds), swing (–0.07 to 0.16 seconds) and stance time (–0.12 to 0.20 seconds)
than for the stride length (–0.07 to 0.44 m) and gait speed (–0.02 to 0.42 m/s)
(Figure 1). Consistency
between both methods was good to excellent for all gait parameters
(ICC3,1 = 0.72–0.97). Absolute agreement was also good to excellent
for the stride, swing and stance time (ICC2,1 = 0.72–0.97), but only poor
to fair for the stride length (ICC2,1 = 0.37) and gait speed
(ICC2,1 = 0.52).
Table 1.
Mean values (±SD), mean difference scores (bias ± SD), limits of agreement,
mean percentage errors and intraclass correlation coefficients for
consistency and absolute agreement for each gait parameter.
All ICCs for consistency and absolute agreement were significant at
P < 0.001.
Figure 1.
Bland–Altman plots for comparison between the GAITRite® system and
the smart walker–integrated gait analysis system. Dotted lines indicate bias
and dashed lines indicate upper and lower 95% limits of agreement (±1.96 SD
of the bias).
Mean values (±SD), mean difference scores (bias ± SD), limits of agreement,
mean percentage errors and intraclass correlation coefficients for
consistency and absolute agreement for each gait parameter.LOA: limits of agreement; PE: percentage error; ICC: intraclass
correlation coefficient; CI: confidence interval.All ICCs for consistency and absolute agreement were significant at
P < 0.001.Bland–Altman plots for comparison between the GAITRite® system and
the smart walker–integrated gait analysis system. Dotted lines indicate bias
and dashed lines indicate upper and lower 95% limits of agreement (±1.96 SD
of the bias).
Discussion
This initial validation study showed that the smart walker–integrated gait analysis
system provides comparable data to the GAITRite system for the temporal gait
parameters of stride, swing and stance time in potential smart walker users.
Although also a good consistency for the stride length and gait speed was found
between these two systems, they cannot be used interchangeably when absolute values
of these spatial-related gait parameters are required (e.g. for comparison with
normative values).The low absolute agreement for spatial-related gait parameters can be explained by
the fact that the GAITRite system refers to the distance between heel contacts on
the electronic walkway for measuring the stride length, while the smart
walker–integrated gait analysis system refers to the distance between leg placements
recorded by the laser range finder 35 cm above the walkway. The reference points of
the laser range finder are closer to the pivot of the lower legs (i.e. knee joint)
and thus travel a shorter distance during the gait cycle, resulting in the shorter
stride length and also the lower gait speed. While absolute agreement for these
spatial-related parameters seems lacking, the extent to which they agree with the
GAITRite system on the relative values (i.e. consistency) was good to excellent,
suggesting that the stride length and gait speed of the smart walker–integrated gait
analysis system may be themselves reliable and as good as those of the GAITRite
system in determining meaningful changes in a user’s walking pattern.Compared to previous validation studies of smart walker–integrated gait analysis
systems, the strengths of this study are that a reasonable number of participants
representative of potential smart walker users were recruited, a well-established,
validated gait analysis system was used as criterion standard for comparison, and
the data obtained were analysed by adequate statistical methods. However, this study
also has some limitations. Only short straight walking in a controlled laboratory
environment was evaluated, as limited by our criterion standard. Future studies
should assess the validity of the smart walker-gait analysis system in less
constrained movement situations. Our participants were predominantly females,
limiting the generalizability of the results to males. However, we did not expect
gender to affect the concurrent validity between the two systems.The smart walker–integrated gait analysis system can provide clinicians and
researchers the ability to unobtrusively capture gait parameters of smart walker
users, without any sensors being attached to the user’s body. Our study represents a
first step towards a continuous gait analysis of smart walker users in natural
environments. The applicability of the system for such long-term gait monitoring
needs to be confirmed in future studies.The smart walker–integrated gait analysis system has good concurrent
validity with the GAITRite® system for measuring temporal
gait parameters in potential smart walker users.Stride length and gait speed can also be measured consistently; however,
modifications are recommended to improve the absolute measurement
accuracy for these spatial-related gait parameters.
Authors: J M Guralnik; L Ferrucci; C F Pieper; S G Leveille; K S Markides; G V Ostir; S Studenski; L F Berkman; R B Wallace Journal: J Gerontol A Biol Sci Med Sci Date: 2000-04 Impact factor: 6.053