Literature DB >> 31076331

Evaluation of Pseudoreader Study Designs to Estimate Observer Performance Results as an Alternative to Fully Crossed, Multireader, Multicase Studies.

Rickey E Carter1, David R Holmes2, Joel G Fletcher3, Cynthia H McCollough3.   

Abstract

RATIONALE AND
OBJECTIVES: To examine the ability of a pseudoreader study design to estimate the observer performance obtained using a traditional fully crossed, multireader, multicase (MRMC) study.
MATERIALS AND METHODS: A 10-reader MRMC study with 20 computed tomography datasets was designed to measure observer performance on four novel noise reduction methods. This study served as the foundation for the empirical evaluation of three different pseudoreader designs, each of which used a similar bootstrap approach for generating 2000 realizations from the fully crossed study. Our three approaches to generating a pseudoreader varied in the degree to which reader performance was matched and integrated into the pseudoreader design. One randomly selected simulation was selected as a "mock study" to represent a hypothetical, prospective implementation of the design.
RESULTS: Using the traditional fully crossed design, figures of merit) (95% CIs) for the four noise reductions methods were 68.2 (55.5-81.0), 69.6 (58.4-80.8), 70.8 (60.2-81.4), and 70.9 (60.4-81.3), respectively. When radiologists' performances on the fourth noise reduction method were used to pair readers during the mock study, there was strong agreement in the estimated figures of merits with estimates using the pseudoreader design being within ±3% of the fully crossed design.
CONCLUSION: Fully crossed MRMC studies require significant investment in resources and time, often resulting in delayed implementation or minimal human testing before dissemination. The pseudoreader approach accelerates study conduct by combining readers judiciously and was found to provide comparable results to the traditional fully crossed design by making strong assumptions about exchangeability of the readers.
Copyright © 2019 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Multi-case study; Multi-reader; Observer performance; Study design

Mesh:

Year:  2019        PMID: 31076331      PMCID: PMC6834885          DOI: 10.1016/j.acra.2019.03.014

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  11 in total

1.  Receiver operating characteristic rating analysis. Generalization to the population of readers and patients with the jackknife method.

Authors:  D D Dorfman; K S Berbaum; C E Metz
Journal:  Invest Radiol       Date:  1992-09       Impact factor: 6.016

2.  A comparison of denominator degrees of freedom methods for multiple observer ROC analysis.

Authors:  Stephen L Hillis
Journal:  Stat Med       Date:  2007-02-10       Impact factor: 2.373

3.  Paired split-plot designs of multireader multicase studies.

Authors:  Weijie Chen; Qi Gong; Brandon D Gallas
Journal:  J Med Imaging (Bellingham)       Date:  2018-05-17

Review 4.  The value of observer performance studies in dose optimization: a focus on free-response receiver operating characteristic methods.

Authors:  John D Thompson; David J Manning; Peter Hogg
Journal:  J Nucl Med Technol       Date:  2013-04-26

5.  The measurement of observer agreement for categorical data.

Authors:  J R Landis; G G Koch
Journal:  Biometrics       Date:  1977-03       Impact factor: 2.571

6.  Recent developments in imaging system assessment methodology, FROC analysis and the search model.

Authors:  Dev P Chakraborty
Journal:  Nucl Instrum Methods Phys Res A       Date:  2011-08-21       Impact factor: 1.455

7.  Can radiologist training and testing ensure high performance in CT colonography? Lessons From the National CT Colonography Trial.

Authors:  Joel G Fletcher; Mei-Hsiu Chen; Benjamin A Herman; C Daniel Johnson; Alicia Toledano; Abraham H Dachman; Amy K Hara; Jeff L Fidler; Christine O Menias; Kevin J Coakley; Mark Kuo; Karen M Horton; Jugesh Cheema; Revathy Iyer; Bettina Siewert; Judy Yee; Richard Obregon; Peter Zimmerman; Robert Halvorsen; Giovanna Casola; Martina Morrin
Journal:  AJR Am J Roentgenol       Date:  2010-07       Impact factor: 3.959

8.  Multi-reader ROC studies with split-plot designs: a comparison of statistical methods.

Authors:  Nancy A Obuchowski; Brandon D Gallas; Stephen L Hillis
Journal:  Acad Radiol       Date:  2012-12       Impact factor: 3.173

9.  Low-dose CT for the detection and classification of metastatic liver lesions: Results of the 2016 Low Dose CT Grand Challenge.

Authors:  Cynthia H McCollough; Adam C Bartley; Rickey E Carter; Baiyu Chen; Tammy A Drees; Phillip Edwards; David R Holmes; Alice E Huang; Farhana Khan; Shuai Leng; Kyle L McMillan; Gregory J Michalak; Kristina M Nunez; Lifeng Yu; Joel G Fletcher
Journal:  Med Phys       Date:  2017-10       Impact factor: 4.071

10.  Observer Performance with Varying Radiation Dose and Reconstruction Methods for Detection of Hepatic Metastases.

Authors:  Joel G Fletcher; Jeff L Fidler; Sudhakar K Venkatesh; David M Hough; Naoki Takahashi; Lifeng Yu; Matthew Johnson; Shuai Leng; David R Holmes; Rickey Carter; Cynthia H McCollough
Journal:  Radiology       Date:  2018-09-11       Impact factor: 11.105

View more
  1 in total

1.  Low-dose CT image and projection dataset.

Authors:  Taylor R Moen; Baiyu Chen; David R Holmes; Xinhui Duan; Zhicong Yu; Lifeng Yu; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2020-12-16       Impact factor: 4.071

  1 in total