Literature DB >> 31072714

Comparison of Clinical and Radiologic Outcomes and Second-Look Arthroscopic Findings After Anterior Cruciate Ligament Reconstruction Using Fixed and Adjustable Loop Cortical Suspension Devices.

Hyeon Wook Ahn1, Jong Keun Seon2, Eun Kyoo Song1, Chan Jin Park1, Hong An Lim1.   

Abstract

PURPOSE: To compare clinical and radiological outcomes, including tunnel widening, and to evaluate graft status by second-look arthroscopy after anterior cruciate ligament (ACL) reconstruction using 2 different femoral cortical suspension devices (fixed and adjustable loop).
METHODS: Seventy-nine patients were included for this study. The patients were divided into 2 groups, 41 patients of fixed loop group (EndoButton) and 38 patients of adjustable loop group (TightRope). The Tegner activity, Lysholm knee score, Lachman test, and pivot-shift test were compared between the 2 groups. Anterior stress radiographs and femoral tunnel widening on plain radiographs were also compared. Twenty-seven patients (66%) in the EndoButton group and 21 patients (55%) in the TightRope group underwent the second-look arthroscopy.
RESULTS: At the final follow-up, the mean Tegner activity, Lysholm, and International Knee Documentation Committee scores were improved in both groups and there were no statistically significant differences (5.1 ± 1.6 and 5.5 ± 2.1, P = .312; 90.4 ± 7.8 and 91.0 ± 6.5, P = .525; and 87.4 ± 6.7 and 88.7 ± 5.3, P = .127, respectively). There were no statistical significances in the both groups in terms of Lachman test and pivot-shift tests (P = .392, .559) as well as anterior stress radiographs (mean 3.2 ± 1.4 mm and 2.9 ± 1.1 mm, P = .343). Moreover, radiologic measurements comparing femoral tunnel widening at proximal and distal half also showed no significant differences (P = .540 and .412 on anteroposterior view; P = .254 and .437 on lateral view). In the second-look arthroscopy findings for graft tear and synovial coverage, there were no significant differences (P = .784 and .897).
CONCLUSIONS: Both fixed loop and adjustable loop devices in ACL reconstruction provided good clinical and radiological outcomes. In patients with both devices, femoral tunnel widening at proximal and distal portion have no significant differences after ACL reconstruction. Moreover, second-look arthroscopy revealed no significant differences in terms of synovial coverage and rupture of the graft. LEVEL OF EVIDENCE: Level II, prospective comparative study.
Copyright © 2019 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 31072714     DOI: 10.1016/j.arthro.2019.01.051

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  8 in total

1.  Adjustable-loop implants are non-inferior to fixed-loop implants for femoral fixation in anterior cruciate ligament reconstruction.

Authors:  Simone B Elmholt; Torsten G Nielsen; Martin Lind
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2022-06-23       Impact factor: 4.342

2.  [Lengths of the fixed loop and the adjustable loop in the coarse bone tunnel were compared to influence the widening of the femoral bone tunnel and the function of the knee joint].

Authors:  Y Yin; Y Mei; Z G Wang; S Y Song; P F Liu; P F He; W J Wu; X Xie
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2021-10-18

Review 3.  Fixed-loop vs. adjustable-loop cortical button devices for femoral fixation in ACL reconstruction - a systematic review and meta-analysis.

Authors:  Simone Birkebæk Elmholt; Torsten Grønbech Nielsen; Martin Lind
Journal:  J Exp Orthop       Date:  2022-10-21

4.  No Difference in Outcomes Between Suspensory (Fixed-Loop Cortical Button) Versus Expandable Anteromedial Femoral Fixation in Anterior Cruciate Ligament Reconstruction With Autologous Hamstring Tendons: A Prospective, Randomized, Controlled Study in Male Patients.

Authors:  Andreas Panagopoulos; Dimitris Mylonas; Antonis Kouzelis; Petros Zampakis; Pantelis Kraniotis; John Lakoumentas; John Gliatis
Journal:  Arthrosc Sports Med Rehabil       Date:  2021-06-24

5.  Laterally shifted tibial tunnel can be the risk of residual knee laxity for double-bundle anterior cruciate ligament reconstruction.

Authors:  Daisuke Chiba; Yuji Yamamoto; Yuka Kimura; Shizuka Sasaki; Eiji Sasaki; Shohei Yamauchi; Eiichi Tsuda; Yasuyuki Ishibashi
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-04-30       Impact factor: 4.342

6.  Confirming Proper Button Deployment of Suspensory Fixation During ACL Reconstruction.

Authors:  Daniel F O'Brien; Lilah Fones; Victoria Stoj; Cory Edgar; Katherine Coyner; Robert A Arciero
Journal:  Orthop J Sports Med       Date:  2021-01-19

7.  In Vitro Testing of 2 Adjustable-Loop Cortical Suspensory Fixation Systems Versus Interference Screw for Anterior Cruciate Ligament Reconstruction.

Authors:  Gerardo L Garcés; Oscar Martel; Alejandro Yánez; Ignacio Manchado-Herrera; Luci M Motta
Journal:  Orthop J Sports Med       Date:  2021-09-28

8.  Adjustable Cortical Fixation Device for Quadriceps Tendon Repair: A Cadaveric Biomechanical Study.

Authors:  Heath P Gould; William R Rate; Pooyan Abbasi; Katherine L Mistretta; Jason W Hammond
Journal:  Orthop J Sports Med       Date:  2021-01-28
  8 in total