Alexander Pott1, Michael Baumhardt1, Karolina Weinmann1, Manuel Rattka1, Tilman Stephan1, Mirjam Keßler1, Carlo Bothner1, Kerstin Petscher1, Wolfgang Rottbauer1, Tillman Dahme1.
Abstract
PURPOSE: Real-time observation of pulmonary vein (PV) potentials with a spiral mapping catheter has emerged as a key electrogram-based procedural parameter to estimate lesion quality and titrate cryoenergy application during PV isolation (PVI) with the cryoballoon. Whether correct PV electrogram interpretation and thus PVI real-time observation rate depends on atrial rhythm during cryoballoon PVI is unknown. We compared observation rates of time-to PV isolation (TTI) during sinus rhythm (SR group) and during atrial fibrillation (AFib group) in cryoballoon PVI.Entities:
Keywords: Ablation; Atrial fibrillation; Cryoballoon; Pulmonary vein signals, time-to-isolation
Year: 2019 PMID: 31065585 PMCID: PMC6495074 DOI: 10.1016/j.ijcha.2019.100353
Source DB: PubMed Journal: Int J Cardiol Heart Vasc ISSN: 2352-9067
Fig. 1Surface ECG (□), PV signals recorded by 20 mm spiral mapping catheter () and intracardiac ECG of an 8-polar coronary sinus catheter (). (A) PV electrograms in SR with typical three-component signal consisting of LA far-field (#), sharp PV near-field signal (*) and LV far-field (&). (B) Pulmonary vein isolation using cryoballoon leads to progressive LA to PV conduction delay before complete disappearance of PV signals when the PV is isolated (§). (C) Atrial fibrillation is characterized by chaotic electrical activation of the atrium equalizing typical ECG pattern of near- and far-field potentials (D) complicating real-time observation of pulmonary vein isolation.
Baseline clinical parameters.
| Baseline characteristics | |
|---|---|
| Patients [n] | 157 |
| Gender, female [n (%)] | 62 (39.5) |
| Age (years) | 66.7 ± 10.9 |
| Paroxysmal AFib [n (%)] | 92 (58.6) |
| LA diameter (mm) | 45.4 ± 6.3 |
| Heart failure [n (%)] | 37 (23.6) |
| Arterial hypertension [n (%)] | 121 (77.1) |
| Diabetes mellitus [n (%)] | 26 (16.6) |
| CHA2DS2-VASc-score | 3.0 ± 1.8 |
| Stroke [n (%)] | 15 (9.6) |
Fig. 2Overall and per vein incidence of real-time PV observation during sinus rhythm and during atrial fibrillation. The significant higher overall TTI observation rate is mainly driven by distinct differences in LSPV signal interpretation between the SR group and the AFib group.
Procedural parameters.
| Procedural data | SR | AFib | p-Value |
|---|---|---|---|
| Pulmonary veins [n] | 365 | 233 | |
| PV isolation with 1. Freeze [n (%)] | 314 (86.0) | 194 (83.2) | 0.421 |
| LSPV [n (%)] | 82 (89.1) | 37 (68.5) | |
| LIPV [n (%)] | 73 (83.0) | 52 (89.1) | 0.129 |
| RSPV [n (%)] | 82 (86.3) | 52 (89.7) | 0.621 |
| RIPV [n (%)] | 77 (82.2) | 53 (84.1) | 0.829 |
| Mean nadir balloon temperature [°C, SD] | −45.0 ± 6.0 | −45.7 ± 6.2 | 0.195 |
| LSPV [°C, SD] | −46.4 ± 5.7 | −47.0 ± 5.4 | 0.329 |
| LIPV [°C, SD] | −42.5 ± 6.0 | −42.3 ± 4.7 | 0.689 |
| RSPV [°C, SD] | −46.3 ± 6.0 | −47.3 ± 6.0 | 0.399 |
| RIPV [°C, SD] | −44.6 ± 5.6 | - 46.1 ± 7.2 | 0.298 |
| Mean TTI all PV [sec., SD] | 43.4 ± 23.1 | 40.8 ± 20.3 | 0.276 |
| Mean TTI LSPV [n (%)] | 51.8 ± 26.2 | 44.8 ± 22.2 | 0.245 |
| Mean TTI LIPV [n (%)] | 36.6 ± 17.1 | 36.1 ± 1865 | 0.730 |
| Mean TTI RSPV [n (%)] | 38.1 ± 20.2 | 39.5 ± 20.1 | 0.740 |
| Mean TTI RIPV [n (%)] | 46.1 ± 24.4 | 43.1 ± 22.1 | 0.505 |
Bold values signifies if p <0.05.
Uni- and multivariate logistic regression for TTI detection.
| Variable | Univariate logistic regression | Multivariate logistic regression | ||
|---|---|---|---|---|
| OR (95%-CI) | p-Value | OR (95%-CI) | p-Value | |
| Left atrial diameter | 0.94 (0.88–1.01) | 0.95 (0.88–1.03) | 0.25 | |
| Age | 1.00 (0.96–1.04) | 0.88 | ||
| CHA2DS2-VASc-score | 0.90 (0.70–1.15) | 0.39 | ||
| Persistent AF | 1.44 (0.57–3.61) | 0.44 | ||
| Sex | 0.91 (0.37–2.21) | 0.83 | ||
| Heart failure | 0.73 (0.26–2.05) | 0.56 | ||
| Hypertension | 0.81 (0.28–2.35) | 0.69 | ||
| Diabetes | 0.39 (0.14–1.09) | 0.49 (0.16–1.50) | 0.21 | |
| Coronary vessel disease | 0.45 (0.18–1.10) | 0.62 (0.23–1.62) | 0.33 | |
| Sinus rhythm during PVI | 5.58 (2.13–14.59) | < | 4.98 (1.86–13.34) | |
Parameters with a p-value of 0.2 or lower in univariate logistic regression were included in multivariate logistic regression. Parameters with a p-value <0.05 were considered significant in the multivariate logistic regression.