Does Size Matter? Cryoballoon-Based Pulmonary Vein Isolation Using a Novel 25-mm Circular Mapping Catheter.

Bruno Reissmann1, Michael Schlüter2, Francesco Santoro1, Tilman Maurer1, Christian-Hendrik Heeger1, Christine Lemes1, Thomas Fink1, Johannes Riedl1, Andreas Rillig1, Shibu Mathew1, Feifan Ouyang1, Karl-Heinz Kuck1, Andreas Metzner1.   

Abstract

BACKGROUND: Real-time recording of pulmonary vein isolation (PVI) using a circular mapping catheter has become a key aspect of cryoballoon (CB) ablation. The aim of this study was to investigate the procedural safety, efficacy and rate of real-time pulmonary vein (PV) recording using a novel circular mapping catheter with a 25-mm loop size for CB-based PVI.Methods and 
Results: A total of 40 patients with symptomatic atrial fibrillation (AF) underwent PVI using a second-generation CB and a novel 25-mm circular mapping catheter. A total of 159 PV were identified and successfully isolated. Real-time PV recording was achieved in 80% of the PV. In 3 right inferior PV the circular mapping catheter had to be exchanged for a stiff guidewire due to insufficient mechanical support. Therefore, acute PVI using exclusively the circular mapping catheter was achieved in 156/159 PV (98%). Mean procedure and fluoroscopy times were 66±21 min and 15±6 min, respectively. Transient phrenic nerve palsy occurred in 1 patient as the only procedural complication.
CONCLUSIONS: The exclusive use of a novel 25-mm circular mapping catheter for CB ablation of AF results in a real-time PV recording rate of 80% and isolation of 98% of targeted PV.

Entities:  

Keywords:  Atrial fibrillation; Circular mapping catheter; Cryoballoon ablation; Pulmonary vein isolation; Real-time recording

Mesh:

Year:  2017        PMID: 29279459     DOI: 10.1253/circj.CJ-17-0956

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  3 in total

1.  A decapolar 25-mm-diameter circular mapping catheter with improved detection of pulmonary vein potential disappearance in cryoballoon ablation.

Authors:  Yasuhiro Matsuda; Masaharu Masuda; Mitsutoshi Asai; Osamu Iida; Shin Okamoto; Takayuki Ishihara; Kiyonori Nanto; Takashi Kanda; Takuya Tsujimura; Shota Okuno; Aki Tsuji; Toshiaki Mano
Journal:  J Arrhythm       Date:  2019-05-14

2.  Impact of atrial rhythm on pulmonary vein signals in cryoballoon ablation - Sinus rhythm predicts real-time observation of pulmonary vein isolation.

Authors:  Alexander Pott; Michael Baumhardt; Karolina Weinmann; Manuel Rattka; Tilman Stephan; Mirjam Keßler; Carlo Bothner; Kerstin Petscher; Wolfgang Rottbauer; Tillman Dahme
Journal:  Int J Cardiol Heart Vasc       Date:  2019-04-30

3.  The assessment of pulmonary vein potentials using the new achieve advance during cryoballoon ablation of atrial fibrillation.

Authors:  Giacomo Mugnai; Massimiliano Manfrin; Carlo de Asmundis; Erwin Ströker; Massimo Longobardi; Werner Rauhe; Cesare Storti; Pedro Brugada; Gian-Battista Chierchia
Journal:  Indian Pacing Electrophysiol J       Date:  2019-06-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.