| Literature DB >> 3106413 |
D W Lander, D L Jungkind, L L Stone.
Abstract
Thirty-five modifications of Abbott MS-2 growth medium were evaluated in an attempt to reduce the high frequency of false-susceptible errors produced by the MS-2 system when testing resistant Pseudomonas aeruginosa against cefotaxime. A hypotonic modification with 0.1% KNO3 reduced the errors from 41 to 20% for 23 resistant isolates. All hypertonic modifications proved unsatisfactory. No modification reduced false-susceptible errors to acceptable clinical levels.Entities:
Mesh:
Substances:
Year: 1987 PMID: 3106413 PMCID: PMC266070 DOI: 10.1128/jcm.25.4.732-733.1987
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948