Mehmet Kola1, Mehmet Önal1, Adem Türk1, Hidayet Erdöl1.
Abstract
Objectives: To evaluate the repeatability and agreement of macular thickness measurements obtained with E-MM5 and MM6, two different scan modes, on the Optovue RTVue optic coherence tomography (OCT) device. Materials andEntities:
Keywords: Eye; optical coherence tomography; retina
Year: 2019 PMID: 31055892 PMCID: PMC6517858 DOI: 10.4274/tjo.galenos.2018.88972
Source DB: PubMed Journal: Turk J Ophthalmol ISSN: 2149-8709
Figure 1The early treatment diabetic retinopathy study grid divides optic coherence tomography measurements into nine anatomic zones: Central fovea (F), nasal inner (NI), temporal inner (TI), inferior inner (II), superior inner (SI), nasal outer (NO), temporal outer (TO), superior outer (SO), inferior outer (IO) macula. The central ring is 1 mm in diameter, the inner ring is 3 mm in diameter, and the diameter of the outer ring is 5 mm in E-MM5 scan mode and 6 mm in MM6 scan mode
The mean (± standard deviation) and intraclass correlation coefficient values of three consecutive measurements of the participants’ right eyes (n=30) using the E-MM5 scanning mode of the optical coherence tomography device
The mean (± standard deviation) and intraclass correlation coefficient values of three consecutive measurements of the participants’ right eyes (n=30) using the MM6 scanning mode of the optical coherence tomography device
Coefficients of variation for pairwise comparisons of three consecutive measurements of the participants’ right eyes (n=30) using the E-MM5 scanning mode of the optical coherence tomography device
Coefficients of variation for pairwise comparisons of three consecutive measurements of the participants’ right eyes (n=30) using the MM6 scanning mode of the optical coherence tomography device
Comparisons of mean measurements obtained from the participants’ right eyes (n=30) using the E-MM5 and MM6 scanning modes of the optical coherence tomography device