Literature DB >> 31045256

Transvenous extraction of very old (over 20-year-old) pacemaker leads using mechanical systems: Effectiveness and safety.

Andrzej Ząbek1, Krzysztof Boczar1, Maciej Dębski1, Paweł T Matusik1,2, Roman Pfitzner2,3, Mateusz Ulman1, Robert Musiał4, Jacek Lelakowski1,2, Barbara Małecka1,2.   

Abstract

OBJECTIVE: To analyze and compare the effectiveness and safety of transvenous lead extraction (TLE) with mechanical systems of pacing leads older than 20 years (group A) versus younger leads (group B).
METHODS: We performed TLE of 591 pacing leads in 377 patients. Fifty (8.5%) leads in 43 (11.4%) patients were implanted for equal to or more than 20 years. The mean dwell time of all extracted leads was 8.9 years (range, 0.1-36.0). Infection related to cardiovascular implantable electronic device was an indication for TLE in 18.3% of patients.
RESULTS: Complete lead removal and complete procedural success rates were similar between both groups (94.7% in group A vs 97.1% in group B, P = 0.445, and 90.7% in group A vs 95.8% in group B, P = 0.329, respectively). Incomplete lead removal in group A was observed only in leads older than 20 years. Removal of leads in group A was associated with significantly longer fluoroscopy time compared with group B (4.6 vs 1.9 minutes, P < 0.001). We did not find a significant difference in major and minor complication rates between groups (2.3% in group A vs 0.9% in group B and 2.3% in group A vs 2.2% in group B, P = 0.687, respectively). There were no deaths associated with the TLE procedure within 30 days after the procedure in either group.
CONCLUSION: This study shows that TLE of leads older than 20 years conducted at an experienced center seems to be comparably safe and effective as extraction of younger leads but requires longer fluoroscopy time.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  cardiovascular implantable electronic devices; effectiveness; safety; transvenous lead extraction

Mesh:

Year:  2019        PMID: 31045256     DOI: 10.1111/pace.13714

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Safety and efficacy of transvenous lead extraction of very old leads.

Authors:  Simon Pecha; Tibor Ziegelhoeffer; Yalin Yildirim; Yeong-Hoon Choi; Stephan Willems; Hermann Reichenspurner; Heiko Burger; Samer Hakmi
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-04-08

2.  Effectiveness and safety of transvenous extraction of single- versus dual-coil implantable cardioverter-defibrillator leads at single-center experience.

Authors:  Andrzej Ząbek; Krzysztof Boczar; Maciej Dębski; Mateusz Ulman; Roman Pfitzner; Robert Musiał; Jacek Lelakowski; Barbara Małecka
Journal:  Medicine (Baltimore)       Date:  2019-07       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.