Intra-operative temperature monitoring with cutaneous zero-heat- flux-thermometry in comparison with oesophageal temperature: A prospective study in the paediatric population.

Hugo Carvalho1, Nadia Najafi1, Jan Poelaert1.   

Abstract

BACKGROUND: Maintenance of thermal homeostasis is of crucial importance in the anesthetized pediatric patient. Gold-standard methods for central core temperature measurement are however inappropriately invasive and impractical in daily practice. The SpotOn sensor uses zero-heat-flux thermometry technology and claims to bypass the invasiveness of classical methods and still accurately display central core temperatures. Up to date no formal analysis of this method in children has taken place. AIMS: The primary objective was to assess the accuracy in comparison with esophageal temperature; the secondary objective concerned the safety of the SpotOn sensor in the pediatric patients.
METHODS: Fifty-four children aged 1-12 years with an American Society of Anesthesiology classification I or II scheduled to undergo elective surgical procedures under general anesthesia for a minimum of 30 minutes were included. Exclusion criteria included: fragile forehead skin, procedures impeding proper SpotOn placement, thoracoscopic or gastroesophageal procedures, coagulopathy, hemodynamic instability, or vasoactive medication use. After sevoflurane induction, an esophageal temperature probe was placed in the lower third of the esophagus, and a SpotOn sensor on the lateral forehead. Temperatures were recorded in pairs per 1 minute intervals. Temperatures were subjected to bias analysis with 0.5°C as the a priori established clinical significance cutoff.
RESULTS: Bland-Altman analysis revealed the two methods differed on average 0.14°C (95% limits of agreement: -0.39 to 0.66), with 89.5% of the differences being under 0.5°C. No significant differences could be found between the two methods for the established 0.5°C cutoff. Linear regression analysis determined the following linear regression equation: 0.837x + 5.86 (R2  = 0.738). Lin's concordance correlation coefficient of 0.83 (95% CI: 0.81-0.84). No complications were observed with the use of the SpotOn sensor.
CONCLUSION: SpotOn revealed itself as accurate as an esophageal temperature probe when estimating central core temperatures under ideal conditions and over a narrow range of temperatures. No adverse effects were observed with the use of the SpotOn sensor.
© 2019 John Wiley & Sons Ltd.

Entities:  

Keywords:  anesthesia; children; equipment and supplies; equipment safety; general; pediatrics; spoton thermometry; temperature; thermometry; validation

Year:  2019        PMID: 31034706     DOI: 10.1111/pan.13653

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  5 in total

1.  Zero-heat-flux core temperature monitoring system: an observational secondary analysis to evaluate agreement with naso-/oropharyngeal probe during anesthesia.

Authors:  Nicholas West; Erin Cooke; Dan Morse; Richard N Merchant; Matthias Görges
Journal:  J Clin Monit Comput       Date:  2019-11-06       Impact factor: 2.502

2.  Development of local guidelines to prevent perioperative hypothermia in children: a prospective observational cohort study.

Authors:  Rehena Sultana; John C Allen; Yew Nam Siow; Choon Looi Bong; Shu Ying Lee
Journal:  Can J Anaesth       Date:  2022-09-15       Impact factor: 6.713

3.  Intraoperative zero-heat-flux thermometry overestimates esophageal temperature by 0.26 °C: an observational study in 100 infants and young children.

Authors:  Marcus Nemeth; Marijana Lovric; Thomas Asendorf; Anselm Bräuer; Clemens Miller
Journal:  J Clin Monit Comput       Date:  2020-10-31       Impact factor: 1.977

4.  Are Non-Contact Thermometers an Option in Anaesthesia? A Narrative Review on Thermometry for Perioperative Medicine.

Authors:  Andre van Zundert; Tonchanok Intaprasert; Floris Wiepking; Victoria Eley
Journal:  Healthcare (Basel)       Date:  2022-01-24

5.  Prospective comparative analysis of noninvasive body temperature monitoring using zero heat flux technology (SpotOn sensor) compared with esophageal temperature monitoring during pediatric surgery.

Authors:  Bo-Hyun Sang; Changjin Lee; Da Yeong Lee
Journal:  PLoS One       Date:  2022-08-08       Impact factor: 3.752

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.