Continuum Robot With Follow-the-Leader Motion for Endoscopic Third Ventriculostomy and Tumor Biopsy.

Yuanqian Gao, Kiyoshi Takagi, Takahisa Kato, Naoyuki Shono, Nobuhiko Hata.   

Abstract

BACKGROUND: In a combined endoscopic third ventriculostomy (ETV) and endoscopic tumor biopsy (ETB) procedure, an optimal tool trajectory is mandatory to minimize trauma to surrounding cerebral tissue.
OBJECTIVE: This paper presents wire-driven multi-section robot with push-pull wire. The robot is tested to attain follow-the-leader (FTL) motion to place surgical instruments through narrow passages while minimizing the trauma to tissues.
METHODS: A wire-driven continuum robot with six sub-sections was developed and its kinematic model was proposed to achieve FTL motion. An accuracy test to assess the robot's ability to attain FTL motion along a set of elementary curved trajectory was performed. We also used hydrocephalus ventricular model created from human subject data to generate five ETV/ETB trajectories and conducted a study assessing the accuracy of the FTL motion along these clinically desirable trajectories.
RESULTS: In the test with elementary curved paths, the maximal deviation of the robot was increased from 0.47 mm at 30 ° turn to 1.78 mm at 180 ° in a simple C-shaped curve. S-shaped FTL motion had lesser deviation ranging from 0.16 to 0.18 mm. In the phantom study, the greatest tip deviation was 1.45 mm, and the greatest path deviation was 1.23 mm.
CONCLUSION: We present the application of a continuum robot with FTL motion to perform a combined ETV/ETB procedure. The validation study using human subject data indicated that the accuracy of FTL motion is relatively high. The study indicated that FTL motion may be useful tool for combined ETV and ETB.

Entities:  

Mesh:

Year:  2019        PMID: 31034405      PMCID: PMC7098325          DOI: 10.1109/TBME.2019.2913752

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  19 in total

1.  Intraventricular and skull base neuroendoscopy in 2012: a global survey of usage patterns and the role of intraoperative neuronavigation.

Authors:  Felice Esposito; Federico Di Rocco; Gabriel Zada; Giuseppe Cinalli; Henry W S Schroeder; Conor Mallucci; Luigi M Cavallo; Philippe Decq; Carmela Chiaramonte; Paolo Cappabianca
Journal:  World Neurosurg       Date:  2013-07-10       Impact factor: 2.104

2.  Single burr hole endoscopic biopsy with third ventriculostomy-measurements and computer-assisted planning.

Authors:  Hannah Knaus; Schultz Matthias; Arend Koch; Ulrich-W Thomale
Journal:  Childs Nerv Syst       Date:  2011-02-16       Impact factor: 1.475

3.  Single burr hole rigid endoscopic third ventriculostomy and endoscopic tumor biopsy: what is the safe displacement range for the foramen of Monro?

Authors:  Xian Lun Zhu; Rong Gao; George Kwok Chu Wong; Hoi Tung Wong; Rebecca Yuen Ting Ng; Yong Yu; Rosanna Kit Man Wong; Wai Sang Poon
Journal:  Asian J Surg       Date:  2013-02-21       Impact factor: 2.767

4.  Tendon-driven continuum robot for neuroendoscopy: validation of extended kinematic mapping for hysteresis operation.

Authors:  Takahisa Kato; Ichiro Okumura; Hidekazu Kose; Kiyoshi Takagi; Nobuhiko Hata
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-10-17       Impact factor: 2.924

5.  Endoscopic resection of solid intraventricular brain tumors.

Authors:  Mark M Souweidane; Neal Luther
Journal:  J Neurosurg       Date:  2006-08       Impact factor: 5.115

6.  Three-Dimensional Simulation of Collision-Free Paths for Combined Endoscopic Third Ventriculostomy and Pineal Region Tumor Biopsy: Implications for the Design Specifications of Future Flexible Endoscopic Instruments.

Authors:  Kyle W Eastwood; Vivek P Bodani; James M Drake
Journal:  Oper Neurosurg (Hagerstown)       Date:  2016-09-01       Impact factor: 2.703

7.  Neuroendoscopic management of posterior third ventricle and pineal region tumors: technique, limitation, and possible complication avoidance.

Authors:  S Chibbaro; F Di Rocco; O Makiese; A Reiss; P Poczos; G Mirone; F Servadei; B George; P Crafa; M Polivka; A Romano
Journal:  Neurosurg Rev       Date:  2012-01-19       Impact factor: 3.042

8.  Neuroendoscopic aspiration of tumors in the posterior third ventricle and aqueduct lumen: a technical update.

Authors:  Alberto Feletti; Elisabetta Marton; Alessandro Fiorindi; Pierluigi Longatti
Journal:  Acta Neurochir (Wien)       Date:  2013-05-25       Impact factor: 2.216

9.  Surgical pathway and management of pineal region tumours in children.

Authors:  Benedetta Ludovica Pettorini; Rafid Al-Mahfoud; Michael D Jenkinson; Shivaram Avula; Barry Pizer; Conor Mallucci
Journal:  Childs Nerv Syst       Date:  2012-11-06       Impact factor: 1.475

10.  Optimizing Tube Precurvature to Enhance Elastic Stability of Concentric Tube Robots.

Authors:  Junhyoung Ha; Frank C Park; Pierre E Dupont
Journal:  IEEE Trans Robot       Date:  2016-11-22       Impact factor: 5.567

View more
  3 in total

1.  Transbronchial biopsy catheter enhanced by a multisection continuum robot with follow-the-leader motion.

Authors:  Lenny Dupourqué; Fumitaro Masaki; Yolonda L Colson; Takahisa Kato; Nobuhiko Hata
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-07-09       Impact factor: 2.924

2.  Towards FBG-Based Shape Sensing for Micro-scale and Meso-Scale Continuum Robots with Large Deflection.

Authors:  Yash Chitalia; Nancy Joanna Deaton; Seokhwan Jeong; Nahian Rahman; Jaydev P Desai
Journal:  IEEE Robot Autom Lett       Date:  2020-01-28

3.  FAS-A Fully Actuated Segment for Tendon-Driven Continuum Robots.

Authors:  Reinhard M Grassmann; Priyanka Rao; Quentin Peyron; Jessica Burgner-Kahrs
Journal:  Front Robot AI       Date:  2022-04-26