Literature DB >> 29506110

Three-Dimensional Simulation of Collision-Free Paths for Combined Endoscopic Third Ventriculostomy and Pineal Region Tumor Biopsy: Implications for the Design Specifications of Future Flexible Endoscopic Instruments.

Kyle W Eastwood1,2, Vivek P Bodani1,2, James M Drake1,2,3.   

Abstract

BACKGROUND: Recent innovations to expand the scope of intraventricular neuroendoscopy have focused on transitioning multiple-incision procedures into single-corridor approaches. However, the successful adoption of these combined procedures requires minimizing the unwanted torques applied to surrounding healthy structures.
OBJECTIVE: To define the geometry of relevant anatomical structures in endoscopic third ventriculostomy (ETV) and pineal region tumor biopsy (ETB). Second, to determine the optimal instrument shaft path required for collision-free single burr hole combined ETV/ETB.
METHODS: Magnetic resonance and computed tomography data from 15 pediatric patients who underwent both ETV and ETB procedures between 2006 and 2014 was segmented by using the 3DSlicer software package to create virtual 3-D patient models. Anatomical regions of interest were measured including the foramen of Monro, the massa intermedia, the floor of the third ventricle, and the tumor margin. Utilizing the MATLAB software package, virtual dexterous instruments were inserted into the models and optimal dimensions were calculated.
RESULTS: The diameters of the foramen of Monro, massa intermedia (anterior-posterior, superior-inferior), anterior third ventricle, and tumor margin are 6.85, 4.01, 5.05, 14.2, and 28.5 mm, respectively. The average optimal burr placement was determined to be 22.5 mm anterior to the coronal and 30 mm lateral to the sagittal sutures. Optimal flexible instrument geometries for novel instruments were calculated.
CONCLUSION: We have established a platform for estimating the shape of novel curved dexterous instruments for collision-free targeting of multiple intraventricular points, which is both patient and tool specific and can be integrated with image guidance. These data will aid in developing novel dexterous instruments.

Entities:  

Year:  2016        PMID: 29506110     DOI: 10.1227/NEU.0000000000001177

Source DB:  PubMed          Journal:  Oper Neurosurg (Hagerstown)        ISSN: 2332-4252            Impact factor:   2.703


  2 in total

1.  A multiport MR-compatible neuroendoscope: spanning the gap between rigid and flexible scopes.

Authors:  Sunil Manjila; Margherita Mencattelli; Benoit Rosa; Karl Price; Georgios Fagogenis; Pierre E Dupont
Journal:  Neurosurg Focus       Date:  2016-09       Impact factor: 4.047

2.  Continuum Robot With Follow-the-Leader Motion for Endoscopic Third Ventriculostomy and Tumor Biopsy.

Authors:  Yuanqian Gao; Kiyoshi Takagi; Takahisa Kato; Naoyuki Shono; Nobuhiko Hata
Journal:  IEEE Trans Biomed Eng       Date:  2019-04-29       Impact factor: 4.538

  2 in total