Literature DB >> 31023305

Fully automated quantification of biventricular volumes and function in cardiovascular magnetic resonance: applicability to clinical routine settings.

Sören J Backhaus1,2, Wieland Staab2,3, Michael Steinmetz2,4, Christian O Ritter2,3, Joachim Lotz2,3, Gerd Hasenfuß1,2, Andreas Schuster1,2,5, Johannes T Kowallick6,7.   

Abstract

BACKGROUND: Cardiovascular magnetic resonance (CMR) represents the clinical gold standard for the assessment of biventricular morphology and function. Since manual post-processing is time-consuming and prone to observer variability, efforts have been directed towards automated volumetric quantification. In this study, we sought to validate the accuracy of a novel approach providing fully automated quantification of biventricular volumes and function in a "real-world" clinical setting.
METHODS: Three-hundred CMR examinations were randomly selected from the local data base. Fully automated quantification of left ventricular (LV) mass, LV and right ventricular (RV) end-diastolic and end-systolic volumes (EDV/ESV), stroke volume (SV) and ejection fraction (EF) were performed overnight using commercially available software (suiteHEART®, Neosoft, Pewaukee, Wisconsin, USA). Parameters were compared to manual assessments (QMass®, Medis Medical Imaging Systems, Leiden, Netherlands). Sub-group analyses were further performed according to image quality, scanner field strength, the presence of implanted aortic valves and repaired Tetralogy of Fallot (ToF).
RESULTS: Biventricular automated segmentation was feasible in all 300 cases. Overall agreement between fully automated and manually derived LV parameters was good (LV-EF: intra-class correlation coefficient [ICC] 0.95; bias - 2.5% [SD 5.9%]), whilst RV agreement was lower (RV-EF: ICC 0.72; bias 5.8% [SD 9.6%]). Lowest agreement was observed in case of severely altered anatomy, e.g. marked RV dilation but normal LV dimensions in repaired ToF (LV parameters ICC 0.73-0.91; RV parameters ICC 0.41-0.94) and/or reduced image quality (LV parameters ICC 0.86-0.95; RV parameters ICC 0.56-0.91), which was more common on 3.0 T than on 1.5 T.
CONCLUSIONS: Fully automated assessments of biventricular morphology and function is robust and accurate in a clinical routine setting with good image quality and can be performed without any user interaction. However, in case of demanding anatomy (e.g. repaired ToF, severe LV hypertrophy) or reduced image quality, quality check and manual re-contouring are still required.

Entities:  

Mesh:

Year:  2019        PMID: 31023305      PMCID: PMC8059518          DOI: 10.1186/s12968-019-0532-9

Source DB:  PubMed          Journal:  J Cardiovasc Magn Reson        ISSN: 1097-6647            Impact factor:   5.364


Introduction

Cardiovascular magnetic resonance (CMR) imaging is the gold standard for the assessment of cardiac function and morphology [1, 2]. Left ventricular (LV) ejection fraction (EF) is the most established parameter for cardiac functional assessments in clinical routine and is used for the evaluation of disease severity, treatment follow-up and risk assessment for adverse events [3, 4]. To extract clinically relevant information such as LV mass, LV and right ventricular (RV) end-diastolic and end-systolic volume (EDV/ESV), stroke volume (SV) and EF, accurate post-processing of the cine CMR images is required. In daily clinical routine, post-processing is typically performed manually by delineating endocardial and epicardial LV borders as well as endocardial RV borders in all short-axis (SAX) slices covering the ventricles from atrioventricular ring to apex, in both end-diastolic and end-systolic phases. This task is time-consuming, tedious and subject to observer-variability [2, 5–7]. Emerging post-processing software based on deep-learning algorithms using convolutional neural networks now offer a fully automated approach for LV and RV volume assessments and have recently become commercially available [8]. Initial evaluations of these automated approaches are promising [9], however often based on pre-selected cases with excellent image quality or ‘cropped’ data [9, 10], i.e. SAX stacks are manually triaged to only include end-diastolic and end-systolic images effectively covering the ventricles before applying the automatic algorithm. Importantly, most of the observer-variability results from discrepancies in defining the most apical and basal short-axis SAX slices [11], which is whitewashed by previous manual ‘cropping’ of SAX stacks. At the current time, it remains unknown whether fully automated quantification of biventricular morphology and function is feasible and accurate in clinical CMR routine. Accordingly, the aim of the present study was to evaluate the feasibility and accuracy of fully automated biventricular assessment of morphology and function in a variety of CMR data (neither pre-selected nor pre-processed) taken from a real-world data base of a tertiary care CMR unit.

Methods

Study design

The study population consisted of 300 randomly selected patients referred to CMR within clinical routine care between 2016 and 2018. The CMR imaging protocol was employed on clinical 1.5 or 3 Tesla (Magnetom Symphony or Magnetom Skyra, Siemens Healthineers, Erlangen, Germany) CMR scanners. Protocols were employed as appropriate for clinical routine, all of which including electrocardiogram (ECG)-gated balanced steady-state free precession (bSSFP) cine sequences for a SAX stack. Typical imaging parameters were as follows: 25 frames/cardiac cycle, pixel spacing 0.8 mm × 0.8 mm, 8 mm slice thickness as well as inter-slice gap, TE 1.5 ms, TR 3 ms. The study was approved by the Ethics Committee of the University Hospital Goettingen and complied with the Declaration of Helsinki. The Ethics Committee gave permission to waive informed consent for this retrospective analysis. Furthermore, agreement was assessed between the fully automated algorithm and expert consensus contours based on the Society for Cardiovascular Magnetic Resonance (SCMR) consensus data consisting of 15 cases with different pathologies [12].

CMR analyses

Volumetric analyses were performed manually in short-axis orientations by an experienced investigator according to standardized recommendations [11] using commercially available post-processing software (QMass®, Version 3.1.16.0, Medis Medical Imaging Systems, Leiden, Netherlands). Fully automated segmentation was performed employing dedicated commercially available software (suiteHEART®, Version 4.0.6, Neosoft, Pewaukee, Wisconsin, USA). The papillary muscles were included within the myocardium, trabecular tissue was excluded from the myocardial mass using both, the manual (QMass®) and the automated (suiteHEART®) software. Manual segmentation was performed by simple delineation of the LV endocardial- and epicardial borders and the RV endocardial border with Bézier smoothing at end-diastole and end-systole. No thresholding or edge detection was applied. Cross-referencing from long-axis locations was used to adjust for systolic atrioventricular ring descent. Fully automated segmentation was done overnight without any user-interaction neither by pre-processing the acquired short axis stack nor by post-processing automatically traced borders. Analyses included LV mass, and biventricular EDV, ESV, SV and EF. Agreement was tested between manual and fully automated analyses. Reproducibility was tested by reapplying the fully automated tracking algorithm on 20 randomly selected patients and by manual volumetric analyses by two experienced investigators including intra- and inter-observer reproducibility. All operators were blinded to each other’s results. Furthermore, the analysis time needed to perform manual segmentations was measured in the subset of 20 patients. The presence and relevance of artefacts impacting image quality was graded adopting the criteria proposed by Klinke et al. [13] taking wrap around, respiratory ghost, cardiac ghost, image blurring, metal and shimming artefacts into account (Table 1). One point was given if the artefact impeded the visualization of > 1/3 of the ventricular endocardial border at end-systole and/or end-diastole on a single SAX slice. If such artefact involved 2 slices or ≥ 3 slices, 2 and 3 points were given, respectively. Furthermore, accurate short-axis orientation was evaluated, resulting in an image quality score between 0 (= excellent quality) and 6 (= poor quality). Image quality scores were separately assessed for the LV and RV myocardium.
Table 1

Quality assessment of cine short-axis (SAX) images. The image quality score corresponds to the sum of qualitative scoring based on 6 criteria (range of score: 0–5). One point was given if an artefact impeded the visualization of > 1/3 of the ventricular endocardial border at end-systole and/or end-diastole on a single SAX slice. If such artefact involved 2 slices or ≥ 3 slices, 2 and 3 points were given, respectively. Incorrect short-axis orientation was graded with 2 points

0123Maximum Score
1. Wrap aroundNo1 slice2 slices≥3 slices3
2. Respiratory ghostNo1 slice2 slices≥3 slices
3. Cardiac ghostNo1 slice2 slices≥3 slices
4. Image blurring / mis-triggeringNo1 slice2 slices≥3 slices
5. Metallic artefactsNo1 slice2 slices≥3 slices
6. Orientation of stackCorrectIncorrect2
Quality assessment of cine short-axis (SAX) images. The image quality score corresponds to the sum of qualitative scoring based on 6 criteria (range of score: 0–5). One point was given if an artefact impeded the visualization of > 1/3 of the ventricular endocardial border at end-systole and/or end-diastole on a single SAX slice. If such artefact involved 2 slices or ≥ 3 slices, 2 and 3 points were given, respectively. Incorrect short-axis orientation was graded with 2 points For the SCMR consensus data, only LV parameters were compared between automated analyses and manual expert consensus parameters, since RV parameters were not provided. According to the method described by Suinesiaputra et al. [12], papillary muscles and trabecular tissue were excluded from the myocardial mass.

Statistics

Continuous variables were checked for normal distribution using the Shapiro-Wilk test and are presented as median with interquartile range (IQR). Biventricular volumes and LV mass were indexed to body surface area. Dependent variables were tested using the Wilcoxon signed-rank test. Agreement of manual and automated analyses as well as intra- and inter-observer variability was assessed first using Bland-Altman analysis [mean difference between measurements with 95% confidence interval (CI)] [14], second intra-class correlation coefficients (ICC) based on a model of absolute agreement, considered excellent if ICC > 0.74, good between 0.60 and 0.74, fair between 0.4 and 0.59 and poor below 0.4 [15], and third the coefficient of variation (CoV, = standard deviation [SD] of the differences divided by the mean) [16, 17]. P-values provided are two-sided, an alpha level below 0.05 was considered statistically significant. Statistical analyses were performed using IBM SPSS Statistic Software Version 24 (International Business Machines, Armonk, New York, USA) and Microsoft Excel (Microsoft, Redmond, Washington, USA).

Results

Study population

Patient characteristics and cardiac volumes for both manual and automated assessments are presented in Table 2. Biventricular automatic segmentation was feasible in all 300 cases. In comparison with manual evaluations, automatic assessments depicted higher LV volumes, lower LVEF, higher LV mass as well as higher RV EDV, RV SV and RVEF (p < 0.001 for all). The study population consisted of 100 referrals to evaluate ischemic heart disease, 120 patients with myocardial disease, 70 patients with congenital heart disease and 10 others. Table 3 provides an overview of clinical indications. There were 31 patients imaged after aortic valve replacement (AVR) of whom 18 received transcatheter aortic valve replacement (Edwards SAPIEN 3™, Edwards Lifesciences, Irvine, California, USA), 7 patients after open-surgery AVR using a bioprosthesis (Carpentier-Edwards Perimount™, Edwards Lifesciences) and 6 patients after open-surgery AVR with a mechanical aortic valve (SJM Regent™, St. Jude Medical Inc., St Paul, Minnesota, USA).
Table 2

Demographics and biventricular volumes

ParameterStudy population
Gender (f/m)118/182
Age52 (30, 71)
BSA1.92 (1.72, 2.05)
AutomatedManualp
LV Mass, g/m259.8 (48.8, 74.1)58.0 (46.0, 73.0) < 0.001
LV EDV, ml/m288.5 (77.4, 105.8)83.0 (71.0, 100.0) < 0.001
LV ESV, ml/m237.7 (29.4, 51.1)33.0 (25.0, 47.0) < 0.001
LV SV, ml/m248.8 (41.2, 55.8)48.0 (40.0, 56.0)0.133
LV EF58.0 (48.0, 63.0)60.0 (51.0, 66.0) < 0.001
RV EDV, ml/m287.0 (71.1, 107.4)79.0 (65.0, 97.0) < 0.001
RV ESV, ml/m238.4 (26.9, 48.0)39.0 (28.0, 51.0)0.125
RV SV, ml/m247.8 (40.3, 55.9)40.5 (32.0, 48.0) < 0.001
RV EF56.0 (50.0, 63.0)51.0 (44.0, 58.0) < 0.001

Continuous variables are expressed as median and interquartile range (IQR) and were tested using the Wilcoxon signed-rank test. LV/RV left/right ventricle, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction

Numbers in bold type indicate a significant difference

Table 3

Clinical CMR indications

Number of patients
Ischemic Heart Disease100
 Coronary Heart Disease97
 Acute Myocardial Infarction3
Myocardial disease120
 Myocarditis64
 Arrhythmogenic RV Cardiomyopathy19
 Dilated Cardiomyopathy14
 Hypertrophic Cardiomyopathy11
 Sarcoidosis7
 Iron Overload Cardiomyopathy3
 Non-Compaction Cardiomyopathy1
 Anderson Fabry Disease1
Congenital Heart Disease70
 Repaired Tetralogy of Fallot47
 Aortic Coarctation12
 Atrial Septal Defect6
 Aortic Dilatation in Bicuspid Aortic Valve5
Others10
 Cardiac Mass5
 Rheumatic Disease2
 Pericarditis Constrictiva1
 Pericardial Effusion1
 Pulmonary Arterial Hypertension1

RV right ventricular

Demographics and biventricular volumes Continuous variables are expressed as median and interquartile range (IQR) and were tested using the Wilcoxon signed-rank test. LV/RV left/right ventricle, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction Numbers in bold type indicate a significant difference Clinical CMR indications RV right ventricular

Image quality and post-processing

LV-image quality was graded with 1.0 (SD 1.3) (Score 0 n = 168, Score 1 n = 19, Score 2 n = 46, and Score 3 n = 67 points. RV-image quality was graded with 1.1 (SD 1.3) (Score 0 n = 151, Score 1 n = 37, Score 2 n = 39, and Score 3 n = 73 points. Appropriate short-axis orientation was fulfilled in 298 case, the highest image quality score assigned was 3. Manual post-processing took on average 11.3 ± 1.5 min as opposed to automated pre-processing with < 1 min/SAX stack. Representative examples of high and low segmentation accuracy are given in Fig. 1. Corresponding videos including automatic segmentation of all phases and SAX slices can be found in Additional file 1.
Fig. 1

Fully automated biventricular segmentation (upper panel) and manual segmentation (lower panel) of 6 representative cases. The figure comprises examples with good automated segmentation results (a-b) and limited automated segmentation results (c-f). Segmentation results of all phases and all short-axis slices can be found in the supplementary material. a End-diastolic phases of a patient with suspected myocarditis (1.5 T) and excellent biventricular segmentation. b End-systolic phases of a patient after transcatheter aortic valve replacement (TAVR) imaged at 3 T showing good segmentation results, except for insufficient segmentation of papillary muscles. c End-systolic phases of a patient with severe hypertrophic cardiomyopathy (HCM) imaged at 1.5 T with low accuracy of biventricular segmentations. d End-diastolic phases of a patient with dilated cardiomyopathy (DCM) imaged at 1.5 T with underestimation of right ventricular (RV) volume at the basal level. Also note misinterpretation of two apical thrombi as papillary muscles. e End-systolic phases of a patient with repaired Tetralogy of Fallot (ToF) with underestimation of RV volume due to severe metallic artefacts caused by sternal wires. f End-systolic phases of a patient with pulmonary arterial hypertension (PAH) with underestimation of RV volume most likely due to RV hypertrabeculation

Fully automated biventricular segmentation (upper panel) and manual segmentation (lower panel) of 6 representative cases. The figure comprises examples with good automated segmentation results (a-b) and limited automated segmentation results (c-f). Segmentation results of all phases and all short-axis slices can be found in the supplementary material. a End-diastolic phases of a patient with suspected myocarditis (1.5 T) and excellent biventricular segmentation. b End-systolic phases of a patient after transcatheter aortic valve replacement (TAVR) imaged at 3 T showing good segmentation results, except for insufficient segmentation of papillary muscles. c End-systolic phases of a patient with severe hypertrophic cardiomyopathy (HCM) imaged at 1.5 T with low accuracy of biventricular segmentations. d End-diastolic phases of a patient with dilated cardiomyopathy (DCM) imaged at 1.5 T with underestimation of right ventricular (RV) volume at the basal level. Also note misinterpretation of two apical thrombi as papillary muscles. e End-systolic phases of a patient with repaired Tetralogy of Fallot (ToF) with underestimation of RV volume due to severe metallic artefacts caused by sternal wires. f End-systolic phases of a patient with pulmonary arterial hypertension (PAH) with underestimation of RV volume most likely due to RV hypertrabeculation

Agreement of manual and automated analyses

Results comparing automated and manual volume assessments including mean differences with corresponding SD, ICC and CoV are presented in Tables 4 and 5. Corresponding Bland-Altman plots are presented in Figs. 2, 3 and 4. Agreement of manual and automated assessments in the overall cohort of 300 patients was excellent for all LV parameters (ICC ≥0.91), best for EDV (ICC 0.98) closely followed by ESV (ICC 0.96) as well as mass and EF (both ICC 0.95). The automated algorithm slightly overestimated LV mass, EDV and ESV while underestimating LV EF (mean difference − 2.5%, limits of agreement [LOA] -14.6 to 9.1%), p < 0.001). Agreement for RV volumes was excellent for RV EDV and ESV (both ICC 0.92) and good for RV SV (ICC 0.73) and EF (ICC 0.72). Similar to LV measurement, the automatic algorithm overestimated RV EDV, and also RV EF (mean difference 5.8%, LOA -13.0 to 24.6%, p < 0.001). Higher field strength (3 vs 1.5 Tesla) was associated with reduced agreement in biventricular volumes, though it was also associated with a decrease in image quality (1.5 T: LV image quality score 0.8 (SD 1.2), RV image quality score 0.7 (SD 1.1); 3.0 T: LV image quality score 1.4 (SD 1.3), RV image quality score 1.1 (SD 1.3); p < 0.001 for all). Similarly, aortic valve replacement resulted in lower agreement but was also accompanied by lower image quality (LV image quality score 1.9 (SD 1.2); RV image quality score 2.0 (SD 1.2). Repaired ToF was associated with decreased RV image quality (RV image quality score 1.8 [SD 1.1]) but preserved LV image quality (LV image quality score 0.62 [0.99]). Despite preserved LV image quality, agreement was reduced for both LV and RV volumes (Table 3.).
Table 4

Agreement between manual and automated segmentations. Agreement was analysed in the entire study group (n = 300) as well as in subgroups according to field strength, aortic valve replacement and repaired Tetralogy of Fallot

ParameterMean Difference (SD of the Diff.)ICC (95% CI)CoV (%)
AllLV Mass2.4 (9.3)0.95 (0.94–0.97)14.6
(n = 300)LV EDV5.0 (7.9)0.98 (0.94–0.99)8.5
LV ESV4.4 (11.1)0.96 (0.94–0.97)25.0
LV SV0.3 (7.3)0.91 (0.89–0.93)15.1
LV EF−2.5 (5.9)0.95 (0.92–0.97)10.6
RV EDV7.4 (12.0)0.92 (0.81–0.96)14.0
RV ESV−1.6 (9.8)0.92 (0.89–0.93)24.0
RV SV9.0 (10.6)0.73 (0.26–0.87)23.5
RV EF5.8 (9.6)0.72 (0.47–0.83)17.8
1.5 TLV Mass4.0 (7.2)0.96 (0.91–0.98)11.7
(n = 132)LV EDV4.4 (8.4)0.98 (0.96–0.99)8.8
LV ESV2.9 (6.1)0.99 (0.98–1.00)12.2
LV SV1.1 (6.4)0.95 (0.92–0.96)13.7
LV EF−1.5 (4.9)0.97 (0.96–0.98)9.3
RV EDV10.6 (9.6)0.90 (0.33–0.97)11.7
RV ESV2.1 (8.0)0.94 (0.91–0.96)20.5
RV SV8.5 (8.8)0.72 (0.11–0.88)20.6
RV EF3.9 (8.8)0.77 (0.62–0.86)16.5
3 TLV Mass0.1 (9.7)0.97 (0.95–0.98)13.2
(n = 90)LV EDV5.5 (7.3)0.98 (0.91–0.99)7.8
LV ESV6.9 (17.7)0.88 (0.79–0.92)40.5
LV SV−1.4 (8.5)0.84 (0.75–0.89)17.1
LV EF−3.8 (6.9)0.92 (0.83–0.96)12.3
RV EDV5.8 (12.2)0.89 (0.79–0.94)15.0
RV ESV−0.7 (7.9)0.94 (0.92–0.96)20.7
RV SV6.2 (11.0)0.64 (0.35–0.79)25.3
RV EF4.1 (8.4)0.83 (0.68–0.90)15.3
Aortic Valve replacementLV Mass1.3 (11.5)0.89 (0.76–0.94)16.7
(n = 31)LV EDV5.0 (7.7)0.97 (0.91–0.99)8.7
LV ESV5.5 (5.6)0.97 (0.76–0.99)14.8
LV SV−0.4 (7.6)0.93 (0.85–0.96)15.2
LV EF−4.4 (6.4)0.92 (0.73–0.97)10.8
RV EDV6.0 (13.9)0.90 (0.78–0.95)16.6
RV ESV−6.9 (10.6)0.86 (0.60–0.94)28.9
RV SV12.8 (15.0)0.54 (0.00–0.79)31.6
RV EF10.3 (13.7)0.54 (0.00–0.79)23.7
Tetralogy of FallotLV Mass2.9 (11.3)0.75 (0.55–0.86)23.0
(n = 47)LV EDV5.6 (7.8)0.91 (0.69–0.96)9.4
LV ESV3.1 (6.4)0.83 (0.66–0.91)18.7
LV SV1.9 (6.5)0.87 (0.76–0.93)13.1
LV EF−1.5 (5.3)0.73 (0.52–0.85)8.9
RV EDV2.6 (14.3)0.94 (0.90–0.97)13.5
RV ESV− 10.1 (10.5)0.82 (0.25–0.93)20.2
RV SV13.0 (9.4)0.81 (0.00–0.94)17.4
RV EF11.3 (7.5)0.41 (0.00–0.73)14.8

Biventricular volumes and LV mass were indexed to body surface area. T: Tesla. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction

Table 5

Agreement between manual and automated analyses according to image quality

ParameterMean Difference (SD of the Diff.)ICC (95% CI)CoV (%)
Good image quality (Score ≤ 1)LV (n = 187)LV Mass3.0 (7.9)0.96 (0.94–0.98)12.8
LV EDV3.4 (6.2)0.99 (0.97–0.99)6.5
LV ESV1.7 (4.1)0.99 (0.99–1.00)9.0
LV SV1.5 (5.2)0.95 (0.93–0.96)10.5
LV EF−0.6 (3.5)0.98 (0.98–0.99)6.2
RV (n = 188)RV EDV7.8 (10.5)0.93 (0.75–0.97)12.2
RV ESV1.0 (6.9)0.96 (0.95–0.97)16.9
RV SV6.7 (8.5)0.79 (0.40–0.90)18.9
RV EF3.0 (6.1)0.88 (0.78–0.93)11.5
Reduced image quality (Score ≥ 2)LV (n = 113)LV Mass1.3 (11.2)0.94 (0.91–0.96)16.7
LV EDV7.5 (9.7)0.95 (0.82–0.98)10.8
LV ESV8.9 (16.3)0.87 (0.74–0.93)37.9
LV SV−1.6 (9.6)0.86 (0.80–0.90)20.5
LV EF−5.6 (7.5)0.90 (0.67–0.96)13.8
RV (n = 112)RV EDV6.7 (14.3)0.91 (0.83–0.95)16.8
RV ESV−5.9 (12.1)0.84 (0.70–0.90)30.2
RV SV12.7 (12.6)0.67 (0.02–0.86)27.9
RV EF10.6 (12.1)0.56 (0.03–0.77)22.5

Biventricular volumes and LV mass were indexed to body surface area. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction

Fig. 2

Agreement of automatically and manually derived biventricular morphology and function. Bland Altman plots (automatic – manual) are shown for the entire study collective (n = 300). LV/RV: left/right ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference

Fig. 3

Agreement of automatically and manually derived left ventricular parameters according to image quality. Bland Altman plots (automatic – manual) are shown for studies with good image quality (score ≤ 1, n = 187) and for studies with reduced image quality (score ≥ 2, n = 113). LV: left ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference

Fig. 4

Agreement of automatically and manually derived right ventricular parameters according to image quality. Bland Altman plots (automatic – manual) are shown for studies with good image quality (score ≤ 1, n = 188) and for studies with reduced image quality (score ≥ 2, n = 112). RV: right ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference

Agreement between manual and automated segmentations. Agreement was analysed in the entire study group (n = 300) as well as in subgroups according to field strength, aortic valve replacement and repaired Tetralogy of Fallot Biventricular volumes and LV mass were indexed to body surface area. T: Tesla. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction Agreement between manual and automated analyses according to image quality Biventricular volumes and LV mass were indexed to body surface area. SD standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, LV left ventricular, RV right ventricular, EDV/ESV end-diastolic/systolic volume, SV stroke volume, EF ejection fraction Agreement of automatically and manually derived biventricular morphology and function. Bland Altman plots (automatic – manual) are shown for the entire study collective (n = 300). LV/RV: left/right ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference Agreement of automatically and manually derived left ventricular parameters according to image quality. Bland Altman plots (automatic – manual) are shown for studies with good image quality (score ≤ 1, n = 187) and for studies with reduced image quality (score ≥ 2, n = 113). LV: left ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference Agreement of automatically and manually derived right ventricular parameters according to image quality. Bland Altman plots (automatic – manual) are shown for studies with good image quality (score ≤ 1, n = 188) and for studies with reduced image quality (score ≥ 2, n = 112). RV: right ventricle, EDV/ESV: end-diastolic/systolic volume, SV: stroke volume, EF: ejection fraction, Δ: difference If classified according to image quality score, 0 to 1 point was associated with considerable better agreement than 2 to 3 points, both for LV and RV automated analyses. Considering an image quality score of ≤1, both LV and RV agreements were excellent for all variables with a bias of − 0.6% (LOA -7.6 to 6.4%) and 3.0% (LOA -9.2 to 15.2%) for LV EF and RV EF, respectively. However, large differences were observed in case of reduced image quality (score ≥ 2) with a bias of − 5.6% (LOA -20.6 to 9.4%) and 10.6% (LOA -13.6 to 34.8%) for LV EF and RV EF, respectively. LV and RV stroke volumes were very consistent in automated analyses, LV 48.1 ml/m2 compared to RV 47.6 ml/m2 in median, p = 0.435. Results from the comparison between automatically and manually derived expert consensus LV parameters based on the SCMR consensus data are provided in the Additional file 1. In accordance with the study’s results, agreement was excellent in the majority of cases (ICC ≥ 0.95 for all LV parameters) (Additional file 2: Tables S1 and S2). There was one patient with a 20% difference in LV EF between automatic and manual expert results (Case # 15), which was a patient with severe LV hypertrophy (Additional file 2: Table S1, Figure S1), similar to the case shown in Fig. 1c.

Reproducibility

Reproducibility for manual segmentations was better for LV than for RV measurements. The automated algorithm yielded exactly the same results when being reapplied. Table 6 shows ICC, CoV and mean differences (SD) within and between observers.
Table 6

Reproducibility of manual and automated analyses

ParameterMean Difference (SD of the Diff.)ICC (95% CI)CoV (%)
ManualIntra-observerLV Mass−0.3 (3.4)0.99 (0.97–1.00)5.3
LV EDV−3.3 (3.4)0.99 (0.91–1.00)4.0
LV ESV−2.6 (3.0)0.99 (0.95–1.00)8.4
LV SV−0.6 (3.9)0.97 (0.93–0.99)8.0
LV EF1.7 (3.7)0.98 (0.94–0.99)6.3
RV EDV−1.5 (6.7)0.96 (0.90–0.98)9.6
RV ESV1.8 (4.3)0.95 (0.87–0.98)14.2
RV SV−3.5 (6.3)0.81 (0.49–0.93)15.6
RV EF−2.9 (5.1)0.82 (0.49–0.93)8.9
Inter-observerLV Mass−0.7 (3.3)0.99 (0.97–1.00)5.6
LV EDV−5.6 (5.8)0.97 (0.74–0.99)6.7
LV ESV−6.3 (4.8)0.97 (0.46–0.99)12.5
LV SV0.8 (3.7)0.97 (0.93–0.99)7.7
LV EF5.2 (57.9)0.94 (0.44–0.98)8.0
RV EDV−15.5 (7.8)0.82 (0.00–0.96)10.0
RV ESV−7.3 (5.7)0.84 (0.00–0.96)16.4
RV SV−8.1 (6.9)0.75 (0.00–0.93)16.2
RV EF1.3 (7.2)0.62 (0.04–0.85)12.9
AutomaticIntra-observerLV Mass0.0 (0.0)1.000.0
LV EDV0.0 (0.0)1.000.0
LV ESV0.0 (0.0)1.000.0
LV SV0.0 (0.0)1.000.0
LV EF0.0 (0.0)1.000.0
RV EDV0.0 (0.0)1.000.0
RV ESV0.0 (0.0)1.000.0
RV SV0.0 (0.0)1.000.0
RV EF0.0 (0.0)1.000.0
Inter-observerLV Mass0.0 (0.0)1.000.0
LV EDV0.0 (0.0)1.000.0
LV ESV0.0 (0.0)1.000.0
LV SV0.0 (0.0)1.000.0
LV EF0.0 (0.0)1.000.0
RV EDV0.0 (0.0)1.000.0
RV ESV0.0 (0.0)1.000.0
RV SV0.0 (0.0)1.000.0
RV EF0.0 (0.0)1.000.0

Biventricular volumes and LV mass were indexed to body surface area. SD: standard deviation. ICC: intraclass correlation coefficient. CoV: coefficient of variation. LV: left ventricular. RV: right ventricular. EDV/ESV end-diastolic/systolic volume. SV: stroke volume. EF: ejection fraction

Reproducibility of manual and automated analyses Biventricular volumes and LV mass were indexed to body surface area. SD: standard deviation. ICC: intraclass correlation coefficient. CoV: coefficient of variation. LV: left ventricular. RV: right ventricular. EDV/ESV end-diastolic/systolic volume. SV: stroke volume. EF: ejection fraction

Discussion

The present study demonstrates the feasibility of fully automated quantification of biventricular morphology and function and reveals its current pitfalls and limitations in a ‘real-world’ clinical setting. Several notable findings should be considered. First, automatically and manually derived volumes agree well in case of good image quality; however, severe differences occur in case of reduced image quality. Second, agreement is better for LV than for RV volumes. Third, demanding anatomical circumstances (e.g. in patients with repaired ToF) result in lower agreement. Forth, different field strengths or the presence of valve replacements do not impede automated assessments as long as image quality is preserved.

Agreement of automated and manual assessment

CMR represents the reference standard for cardiac volume assessment [1] with incremental accuracy and reproducibility as compared to echocardiography [18]. However, CMR acquisition time is long and further requires time-consuming post-processing to extract clinically relevant information. Thus, efforts have been directed towards automated post-processing analyses based on deep-learning algorithms within the last decade [8, 19, 20]. The current literature demonstrates excellent agreement for automated and manual LV volume assessments [6, 10]; however, studies concerned with automatic RV segmentation are scarce [21]. Noteworthy, the study by Queirόs et al. [10] applied an automatic algorithm on cropped data, that is after manually defining the most basal and apical slices with subsequent cropping the SAX stack to include images effectively covering the LV before applying the automatic algorithm. Furthermore, ED and ES time points were manually pre-selected. However, the correct definition of the most basal slice is amongst the most challenging steps in SAX volume assessments and one of the most important source of observer variability [22], therefore representing a clear bias in testing the reliability of an automated algorithm. In the present study, we sought to simulate a real-world clinical scenario by randomly selecting patients from clinical routine imaging. We applied a commercially available automatic algorithm on clinically acquired SAX stacks – occasionally comprising both atria and ventricles – without any manual pre- or post-processing. The final data was acquired on 1.5 and 3.0 T scanners. Our data elaborates on the excellent agreement between automatically and manually derived volumes in case of good image quality, with overall better agreement for LV than for RV measurements. Indeed, quantification of RV volumes is generally more challenging as opposed to LV volumes due to the complex RV anatomy [23, 24]. Nevertheless, LV and RV stroke volumes were consistent in automated analyses in this patient group without intra- or extracardiac shunt. Manual post-processing time took on average more than 11 min as compared to fully automated assessments with < 1 min. Importantly, automatic analyses of several CMR examinations (in this case 300 scans) run completely user-independent and were performed overnight. Furthermore, automated analyses promise to overcome limitations in observer variability, since the algorithm yields exactly the same measures when being reapplied by different users. Thus, the automated frame-work provides a highly reproducible approach and is able to extremely shorten post-processing times of CMR examinations with subsequent potential to improve cost-effectiveness [25]. Furthermore, the framework may provide ‘on-the-fly’ post-processing parallel to finishing the CMR scan (e.g. during late gadolinium enhancement acquisitions).

Impact of image quality

Our data demonstrate that image quality is the leading determinant of accuracy for fully automatic volume assessment. In case of good image quality (image quality score of ≤1 adopted to the criteria proposed by Klinke et al. [13], Table 1), the bias of both LV and RV function was within acceptable limits. However, in case of reduced image quality (image quality score ≥ 2), a large bias of > 5% was observed for both LV and RV EF with wide LOA, particularly for RV EF. Importantly, the relevance of RV function and volumes is increasingly recognized in various diseases [26]. For example, the diagnosis of arrhythmogenic right ventricular cardiomyopathy is challenging and heavily relies on the assessment of RV EDV and RV EF [27]. If considered for clinical use and decision making, a precise volume assessment is of utmost importance, and cannot be achieved with the proposed fully automatic algorithm in case of impaired image quality yet.

Technical and anatomical considerations

To further elucidate limitations of the commercially available software, we compared the agreement of automated and manually derived volumes for subgroups according to field strengths, the presence of aortic valve replacements as well as repaired ToF. Agreement was better at 1.5 T compared to 3 T scans; however, at 3 T considerably more artefacts (mainly due to inadequate breath-holding and shimming) were present. Reduced agreement at 3 T is therefore more likely a result of lower image quality. Due to the growing number of percutaneously implanted aortic valves [28] and increasing indications for CMR imaging [29] including aortic valve stenosis [30], the presence of valve replacement in CMR studies is likely to grow. As long as image quality was preserved in these patients, agreement of LV volumes remained acceptable, enabling the use of automated algorithms in this group of patients. In contrast, in patients with repaired ToF, both RV and LV agreement were considerably decreased, despite low image quality solely affecting the RV (metal artefacts resulting from sternal wires). Since LV image quality was good, reduced agreement is most likely due to the more demanding anatomy in these patients (distinctly larger RV than LV volumes), which points out the current limitations of fully automated analysis. Here, it remains to be investigated whether or not the proposed automatic deep-learning frame-work is able to further learn from these cases with subsequent improvement of accuracy.

Limitations

The study’s conclusions are derived from the comparison of 300 automatically and manually quantified clinical CMR examinations from a single centre. Although manual contouring was performed by experienced observers, intra- and inter-observer variability may limit its use as a reference standard. Details of the automatic algorithm are not disclosed by the software vendor and therefore cannot be reported. RV mass was not measured, since the automatic algorithm does not provide RV mass quantification.

Conclusion

Fully automated quantification of biventricular morphology and function is feasible and accurate in the majority of cases in a clinical routine setting and has the potential to extremely accelerate post-processing times and to improve reproducibility. However, in case of limited image quality or in patients with demanding anatomy (e.g. in patients with repaired ToF) the proposed fully-automatic frame-work does not yet provide satisfying results and still requires manual re-contouring. A. Myocarditis. B. TAVR. C. HCM. D. DCM. E. Repaired ToF. F. PAH. (ZIP 23844 kb) Table S1. Individual comparison between manual consensus and fully automated LV parameters based on the SCMR consensus data. Table S2. Agreement between expert manual and automated segmentation based on the SCMR consensus data. Figure S1. Agreement of automatically and manually derived consensus LV parameters based on the SCMR consensus data. (DOCX 144 kb)
  29 in total

1.  Multi-centre validation of an automatic algorithm for fast 4D myocardial segmentation in cine CMR datasets.

Authors:  Sandro Queirós; Daniel Barbosa; Jan Engvall; Tino Ebbers; Eike Nagel; Sebastian I Sarvari; Piet Claus; Jaime C Fonseca; João L Vilaça; Jan D'hooge
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2015-10-22       Impact factor: 6.875

Review 2.  Cardiovascular magnetic resonance.

Authors:  Dudley J Pennell
Journal:  Circulation       Date:  2010-02-09       Impact factor: 29.690

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Comprehensive prognosis assessment by CMR imaging after ST-segment elevation myocardial infarction.

Authors:  Ingo Eitel; Suzanne de Waha; Jochen Wöhrle; Georg Fuernau; Phillipp Lurz; Matthias Pauschinger; Steffen Desch; Gerhard Schuler; Holger Thiele
Journal:  J Am Coll Cardiol       Date:  2014-09-23       Impact factor: 24.094

5.  Inter-study reproducibility of left ventricular torsion and torsion rate quantification using MR myocardial feature tracking.

Authors:  Johannes T Kowallick; Geraint Morton; Pablo Lamata; Roy Jogiya; Shelby Kutty; Joachim Lotz; Gerd Hasenfuß; Eike Nagel; Amedeo Chiribiri; Andreas Schuster
Journal:  J Magn Reson Imaging       Date:  2015-06-26       Impact factor: 4.813

6.  Left ventricular end-systolic volume as the major determinant of survival after recovery from myocardial infarction.

Authors:  H D White; R M Norris; M A Brown; P W Brandt; R M Whitlock; C J Wild
Journal:  Circulation       Date:  1987-07       Impact factor: 29.690

7.  Geometry-independent inclusion of basal myocardium yields improved cardiac magnetic resonance agreement with echocardiography and necropsy quantified left-ventricular mass.

Authors:  Lauren A Simprini; Parag Goyal; Noel Codella; David S Fieno; Anika Afroz; Jamie Mullally; Mitchell Cooper; Yi Wang; John Paul Finn; Richard B Devereux; Jonathan W Weinsaft
Journal:  J Hypertens       Date:  2013-10       Impact factor: 4.844

Review 8.  Worldwide TAVI registries: what have we learned?

Authors:  Stephan Haussig; Gerhard Schuler; Axel Linke
Journal:  Clin Res Cardiol       Date:  2014-03-20       Impact factor: 5.460

9.  Inter-vendor reproducibility of left and right ventricular cardiovascular magnetic resonance myocardial feature-tracking.

Authors:  Roman Johannes Gertz; Torben Lange; Johannes Tammo Kowallick; Sören Jan Backhaus; Michael Steinmetz; Wieland Staab; Shelby Kutty; Gerd Hasenfuß; Joachim Lotz; Andreas Schuster
Journal:  PLoS One       Date:  2018-03-14       Impact factor: 3.240

10.  Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing.

Authors:  Jeanette Schulz-Menger; David A Bluemke; Jens Bremerich; Scott D Flamm; Mark A Fogel; Matthias G Friedrich; Raymond J Kim; Florian von Knobelsdorff-Brenkenhoff; Christopher M Kramer; Dudley J Pennell; Sven Plein; Eike Nagel
Journal:  J Cardiovasc Magn Reson       Date:  2013-05-01       Impact factor: 5.364

View more
  12 in total

1.  Artificial intelligence and automation in valvular heart diseases.

Authors:  Qiang Long; Xiaofeng Ye; Qiang Zhao
Journal:  Cardiol J       Date:  2020-06-22       Impact factor: 2.737

Review 2.  Reference ranges ("normal values") for cardiovascular magnetic resonance (CMR) in adults and children: 2020 update.

Authors:  Nadine Kawel-Boehm; Scott J Hetzel; Bharath Ambale-Venkatesh; Gabriella Captur; Christopher J Francois; Michael Jerosch-Herold; Michael Salerno; Shawn D Teague; Emanuela Valsangiacomo-Buechel; Rob J van der Geest; David A Bluemke
Journal:  J Cardiovasc Magn Reson       Date:  2020-12-14       Impact factor: 5.364

3.  AI Based CMR Assessment of Biventricular Function: Clinical Significance of Intervendor Variability and Measurement Errors.

Authors:  Shuo Wang; Hena Patel; Tamari Miller; Keith Ameyaw; Akhil Narang; Daksh Chauhan; Simran Anand; Emeka Anyanwu; Stephanie A Besser; Keigo Kawaji; Xing-Peng Liu; Roberto M Lang; Victor Mor-Avi; Amit R Patel
Journal:  JACC Cardiovasc Imaging       Date:  2021-10-13

4.  Fully‑automated deep‑learning segmentation of pediatric cardiovascular magnetic resonance of patients with complex congenital heart diseases.

Authors:  Saeed Karimi-Bidhendi; Arghavan Arafati; Andrew L Cheng; Yilei Wu; Arash Kheradvar; Hamid Jafarkhani
Journal:  J Cardiovasc Magn Reson       Date:  2020-11-30       Impact factor: 5.364

Review 5.  The Role of AI in Characterizing the DCM Phenotype.

Authors:  Clint Asher; Esther Puyol-Antón; Maleeha Rizvi; Bram Ruijsink; Amedeo Chiribiri; Reza Razavi; Gerry Carr-White
Journal:  Front Cardiovasc Med       Date:  2021-12-21

Review 6.  Quantification of Myocardial Deformation Applying CMR-Feature-Tracking-All About the Left Ventricle?

Authors:  Torben Lange; Andreas Schuster
Journal:  Curr Heart Fail Rep       Date:  2021-05-01

7.  Assessment of right ventricular size and function from cardiovascular magnetic resonance images using artificial intelligence.

Authors:  Shuo Wang; Daksh Chauhan; Hena Patel; Alborz Amir-Khalili; Isabel Ferreira da Silva; Alireza Sojoudi; Silke Friedrich; Amita Singh; Luis Landeras; Tamari Miller; Keith Ameyaw; Akhil Narang; Keigo Kawaji; Qiang Tang; Victor Mor-Avi; Amit R Patel
Journal:  J Cardiovasc Magn Reson       Date:  2022-04-11       Impact factor: 6.903

8.  Training and clinical testing of artificial intelligence derived right atrial cardiovascular magnetic resonance measurements.

Authors:  Rob Van Der Geest; Andrew J Swift; Faisal Alandejani; Samer Alabed; Pankaj Garg; Ze Ming Goh; Kavita Karunasaagarar; Michael Sharkey; Mahan Salehi; Ziad Aldabbagh; Krit Dwivedi; Michail Mamalakis; Pete Metherall; Johanna Uthoff; Chris Johns; Alexander Rothman; Robin Condliffe; Abdul Hameed; Athanasios Charalampoplous; Haiping Lu; Sven Plein; John P Greenwood; Allan Lawrie; Jim M Wild; Patrick J H de Koning; David G Kiely
Journal:  J Cardiovasc Magn Reson       Date:  2022-04-07       Impact factor: 6.903

9.  Impact of fully automated assessment on interstudy reproducibility of biventricular volumes and function in cardiac magnetic resonance imaging.

Authors:  Sören J Backhaus; Andreas Schuster; Sebastian Kelle; Johannes T Kowallick; Torben Lange; Christian Stehning; Marcus Billing; Joachim Lotz; Burkert Pieske; Gerd Hasenfuß
Journal:  Sci Rep       Date:  2021-06-02       Impact factor: 4.379

10.  Fully Automated Cardiac Assessment for Diagnostic and Prognostic Stratification Following Myocardial Infarction.

Authors:  Andreas Schuster; Torben Lange; Sören J Backhaus; Carolin Strohmeyer; Patricia C Boom; Jonas Matz; Johannes T Kowallick; Joachim Lotz; Michael Steinmetz; Shelby Kutty; Boris Bigalke; Matthias Gutberlet; Suzanne de Waha-Thiele; Steffen Desch; Gerd Hasenfuß; Holger Thiele; Thomas Stiermaier; Ingo Eitel
Journal:  J Am Heart Assoc       Date:  2020-09-02       Impact factor: 5.501

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.