Avinav Bharati1, Susama Rani Mandal2, Arun Kumar Gupta3, Amlesh Seth4, Raju Sharma3, Ashu S Bhalla3, Chandan J Das3, S Chatterjee5, Pratik Kumar6.
Abstract
AIM: This study aims to develop a method using dual-energy computed tomography (DECT) to determine the effective atomic number and electron density of substances.Entities:
Keywords: Dual-energy computed tomography; effective atomic number; electron density
Year: 2019 PMID: 30983771 PMCID: PMC6438052 DOI: 10.4103/jmp.JMP_125_18
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Calibration phantom consisted of two parts the outer sphere and the inner cylindrical tubes of various sizes to accommodate different size of pellets. The cylindrical tubes could be fitted to outer sphere with the help of a zig. The pellets of calibrators and validators were put in cylindrical tube which was inserted into outer sphere containing water. The inner cylindrical tube also contained a suitable liquid
Ten calibrators, which were used to do calibration and five validators used for checking the validity of inversion algorithm
| Sample type | Sample code | Sample name | Exponent X |
|---|---|---|---|
| Calibration samples (calibrators) | S1 | Sodium acetate anhydrous | 2.426 |
| S2 | Sodium | 2.586 | |
| S3 | Sucrose | 2.238 | |
| S4 | Sodium bisulfite | 2.626 | |
| S5 | Magnesium | 2.627 | |
| S6 | Aluminum | 2.659 | |
| S7 | Silicon | 2.720 | |
| S8 | Calcium sulfate dihydrate | 2.655 | |
| S9 | Ammonium chloride | 2.680 | |
| S10 | Phosphorus | 1.766 | |
| Samples used for verification of calibration (validators) | S11 | Ethylenediamine acetate | 2.346 |
| S12 | Potassium hydrogen carbonate | 2.656 | |
| S13 | Potassium sulfate | 2.701 | |
| S14 | Calcium acetate | 2.614 | |
| S15 | Ammonium sulfate | 2.553 |
Exponent Y was found to be ~3.07 for all the above samples
Scanning details
| Parameters | Details |
|---|---|
| Protocol | Abdominal protocol |
| Reconstruction algorithm | D26f convolution kernel |
| Slice thickness | 0.3 mm |
| Collimation | 64×0.6 |
| Temporal resolution (ms) | 83 |
| Pitch | 0.25 |
| Rotation time (ms) | 330 |
| Matrix size | 512×512 |
Data for the calibrators like their Hounsfield unit values for dual-energy computed tomography scan, Zxeff, Zeff, ρ, ρe
| Sample number | HU 100 | HU 140 Sn | Zxeff | Zeff | Density ρ (g/cm3) | ρe×1023/cm3 | |
|---|---|---|---|---|---|---|---|
| S1 | 442.72 | 375.18 | 0.0185 | 191.31 | 8.71 | 1.38 | 4.32 |
| S2 | 98.14 | −30.68 | 0.0816 | 494.28 | 11.00 | 0.97 | 2.69 |
| S3 | 530.00 | 480.00 | 0.0166 | 79.523 | 7.07 | 1.51 | 4.81 |
| S4 | 1564.72 | 1155.36 | 0.0867 | 821.15 | 12.88 | 2.08 | 6.23 |
| S5 | 2052.00 | 1599.2 | 0.0964 | 794.99 | 12.00 | 2.42 | 7.35 |
| S6 | 2470.91 | 1789.01 | 0.1089 | 916.83 | 13.00 | 2.70 | 7.79 |
| S7 | 1052.86 | 701.51 | 0.0829 | 1311.60 | 14.00 | 1.71 | 5.10 |
| S8 | 1842.44 | 1197.06 | 0.1280 | 1326.48 | 14.99 | 2.06 | 6.34 |
| S9 | 1425.31 | 807.33 | 0.1460 | 1432.26 | 15.05 | 1.47 | 4.74 |
| S10 | 2257.30 | 1324.20 | 0.1671 | 1897.54 | 16.00 | 1.88 | 5.56 |
The HU values at 100 and 140 Sn kVp were obtained after DECT scan and other function like ρe, S, Zeff X, and Zeff was obtained using equation 5,8, 11 and 15. HU: Hounsfield unit, DECT: Dual-energy computed tomography
Effective atomic number obtained experimentally (Zeff (DECT)) as well as by using formulae (Zeff (actual)) given in equation 1
| Sample code | HU 100 | HU 140 Sn | Zeff (DECT) | Zeff (Actual) | Percentage error (R1) | ρe (DECT) ×1023/cm3 | ρe (actual) ×1023/cm3 | Percentage error (R2) |
|---|---|---|---|---|---|---|---|---|
| S11 | 600 | 540 | 7.51 | 7.91 | 5.0 | 4.90 | 5.02 | 2.3 |
| S12 | 2154.02 | 1401.84 | 14.69 | 15.08 | 2.7 | 6.39 | 6.31 | 1.2 |
| S13 | 3098.5 | 1789.94 | 16.13 | 16.34 | 1.2 | 6.83 | 7.01 | 2.5 |
| S14 | 836.62 | 469.58 | 13.89 | 14.23 | 2.3 | 4.06 | 4.12 | 1.4 |
| S15 | 998.1 | 763.9 | 11.71 | 11.47 | 2.0 | 5.25 | 5.35 | 1.8 |
Also electron density obtained experimentally (ρe (obtained)) as well as by using formulae (ρe (actual)) given in equation 5. Percentage error R1 and R2 are the ratio of actual to DECT obtained values of effective atomic number and electron density respectively. HU: Hounsfield unit, DECT: Dual-energy computed tomography
Figure 2Least square fit graph between Log Zxeff and Log Zeff which was used to determine correlation coefficients a1 and b1 of equation 8. 95% confidence interval was shown by lines drawn above and below the central line. Value of a1 = 0.3491 and b1 = 0.2593 (linear regression coefficient r = 0.9949 with P < 0.0001 that is statistically highly significant result) were found by calibrators
Figure 4Linear regression fit of equation 19 that is [Log (ρe)-LogG(ν)] versus Zxeff 95% confidence interval was shown by lines drawn above and below the central line. (a) The value of correlation coefficient a4 and b4 were found to be −0.501 and 1.366 × 10−4, respectively (linear regression coefficient r = 0.9707 with P < 0.0001 that is statistically highly significant result) for voltage ν =140 kVp. (b) Similarly, at ν =100 kVp, a4 and b4 were found to be −0.509 and 5.932 × 10−5, respectively (linear regression coefficient r = 0.8696 with P < 0.01 that is statistically significant result)