Failure Probability, Stress Distribution and Fracture Analysis of Experimental Screw for Micro Conical Abutment.

Antonio Braulino de Melo Filho1, João Paulo Mendes Tribst2, Nathália de Carvalho Ramos2, Julio Nogueira Luz2, Maria Aparecida Neves Jardini3, Alexandre Luiz Souto Borges2, Mauro Pedrine Santamaria3, Renata Marques de Melo2.   

Abstract

The aim of this study was to evaluate the failure probability of two types of abutment screws after compressive load and to analyze the stress distribution with finite element method. Sixty (60) single-tooth implant restorations were assembled on titanium implants (e-fix, A.S. Technology - Titanium Fix). The groups were divided into Conventional screw (Screw neck 1.5 ø mm) and Experimental screw (Screw neck constricted with 1.2 ø mm). Specimens were subjected to single load to failure with compressive test according ISO 14801. The fractured specimens were subjected to stereomicroscopy for measurement of remaining screws inside the implant and characterization of fracture origin. Representative specimens were analyzed by scanning electronic microscopy. For finite element method (FEM), an identical 3D model of the two in vitro test groups were used with similar conditions (30º, 100 N load). The stress in the abutment screw was analyzed by von-Mises criteria. The results of strength means were 4132.5 ± 76 MPa and 4528.2 ± 127.2 for conventional and experimental groups, respectively. During microscopy, the mean (mm) of the remaining screw piece inside the implants were 0.97 ± 0.23 and 1.32 ± 0.12 for conventional and experimental groups, respectively. In FEM, the conventional group showed stress concentered in an unfavorable region (peak of 39.23 MPa), while the experimental group showed more stress areas but less concentration than the conventional group (36.6 MPa). In using the tested experimental geometry, the abutment screw can have its strength improved, and the origin of failure can be more favorable to clinical resolution.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30970059     DOI: 10.1590/0103-6440201902401

Source DB:  PubMed          Journal:  Braz Dent J        ISSN: 0103-6440


  2 in total

1.  Effect of framework type on survival probability of implant-supported temporary crowns: An in vitro study.

Authors:  Vinícius-Anéas Rodrigues; Amanda-Maria-de Oliveira Dal Piva; Claudio-Akira Yamaguchi; Alexandre-Luiz-Souto Borges; Marcio-Katsuyoshi Mukai; João-Paulo-Mendes Tribst
Journal:  J Clin Exp Dent       Date:  2020-05-01

2.  Mechanical behavior of implant assisted removable partial denture for Kennedy class II.

Authors:  João-Paulo-Mendes Tribst; Rodrigo-Máximo de Araújo; Naiara-Pires Ramanzine; Natália-Ribeiro Santos; Amanda-Maria-de Oliveira Dal Piva; Alexandre-Luiz-Souto Borges; João-Mauricio-Ferraz da Silva
Journal:  J Clin Exp Dent       Date:  2020-01-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.