Alexandre Bani Sadr, Nathalie Testart, Perrine Tylski1, Christian Scheiber.
Abstract
We evaluated the reproducibility of I-ioflupane (I-FP-CIT) SPECT with shorter scan times using a CZT camera. One hundred ninety-nine I-FP-CIT SPECT scans obtained with standard scan time (30 minutes) were truncated to provide 24-, 18-, and 15-minute study simulations. Striatal binding ratios were automatically calculated and remained stable for all series. At 15 minutes, only 10 of 398 striata (2.5%) showed statistically significant different striatal binding ratios compared with reference series. These series were reviewed by 2 operators, and a perfect agreement was found for each patient. Therefore, CZT camera allows a 2-fold scan time reduction in I-FP-CIT SPECT.Entities: