| Literature DB >> 30929995 |
Glynis Frans1, Kurt Beuselinck2, Bart Peeters3, Marc Van Ranst4, Veroniek Saegeman5, Stefanie Desmet6, Katrien Lagrou7.
Abstract
The goal of the study was to adapt our Middle East respiratory syndrome coronavirus (MERS-CoV) lab-developed test (LDT) to 3 "Sample to Result" (S2R) systems: BD MAX (BD), ELITe InGenius (ELITechGroup), and ARIES (Luminex). The BD MAX and InGenius system allowed use of lab-developed primers and TaqMan probes, while ARIES required conversion to MultiCode primers for melting curve analysis. Each device required ≤1 day of training and assay optimization. No discordant results were noted after analysis of 32 External Quality Control (EQC) samples. On a 10-fold dilution series of a MERS-CoV-positive EQC sample, InGenius obtained the highest detection rate. Laboratory technicians rated the ARIES as the user-friendliest. It also required the least hands-on time. BD MAX had the lowest turnaround time and highest throughput. While each device had distinguishing system properties with associated (dis)advantages, the 3 S2R systems were comparable in terms of assay development and validation.Entities:
Keywords: ARIES; BD MAX; ELITe InGenius; Middle East respiratory syndrome coronavirus; Sample to result systems
Mesh:
Substances:
Year: 2019 PMID: 30929995 PMCID: PMC7127711 DOI: 10.1016/j.diagmicrobio.2019.02.006
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Required extraction/PCR optimization for the described MERS-CoV LDT.
| ARIES | BD MAX | InGenius | |
|---|---|---|---|
| Required PCR optimization (time) | ≤1 day | ≤1 day | ≤1 day |
| Sample volume (μL, range) | 200 (200–400) | 200 (100–750) | 200 (200 or 1000 |
| Extract volume (μL, range) | 150 (not applicable) | 12.5 (not applicable) | 50 (50, 100, or 200) |
| PCR reaction volume (μL, range) | 56 (not applicable) | 4 (not applicable) | 20 (20 to 50) |
| Extract volume used for PCR (μL) | 50 | 2 | 5 |
| Sample equivalent analyzed in PCR (μL) | 66,6 | 32 | 20 |
| Analysis method | Melting curve analysis | Taqman hydrolysis | Taqman hydrolysis |
| Internal extraction and PCR control | ARIES-specific, in cassette | Lab-developed PDV, 10 μL, pipetted manually (+ SPC | Lab-developed PDV, 10 μL, pipetted automatically |
| Primers | Unmodified lab-developed, mixed with mastermix, 0.5 μM each | Unmodified lab-developed, mixed with mastermix, 0.5 μM each | |
| Probes | Not applicable | Unmodified lab-developed, mixed with mastermix, 0.2 μM each | Unmodified lab-developed, mixed with mastermix, 0.2 μM each |
| Mastermix | 1× Taqman Fast Virus 1-Step, 12.5 μL, pipetted manually in snap-in tube | 1× Taqman Fast Virus 1-Step, 15 μL, pipetted automatically | |
| PCR optimization | - Generic ARIES PCR | Slightly modified lab-developed PCR with adjusted annealing and extension times | Identical to lab-developed PCR |
PDV = Phocine Distemper Virus; SPC = Specimen Processing Control; TNA = Total nucleic acid.
InGenius extraction kits for 1000-μL sample volumes are under development at the time of writing.
ARIES Exo + Ready Mix supporting TaqMan probes is under development at the time of writing.
The Sample Buffer Tube of the BD MAX extraction kit also contained a Specimen Processing Control (SPC), which can be used as an internal control but was not utilized in our study.
MERS-CoV LDT validation.
| QuantStudio Dx | ARIES | BD MAX | InGenius | |
|---|---|---|---|---|
| Accuracy | No discordance | No discordance | No discordance | No discordance |
| Specificity | No false positives | No false positives | No false positives | No false positives |
| Cross-reaction | None | None | None | None |
| Cross-contamination | None | None | None | None |
| MERS-CoV | Detected: 5/5 dilutions | Detected: 3/5 dilutions | Detected: 4/5 dilutions | Detected: 5/5 dilutions |
| MERS-CoV | Detected: 5/5 dilutions | Detected: 4/5 dilutions | Detected: 4/5 dilutions | Detected: 5/5 dilutions |
Analysis of 32 EQC samples: 16 MERS-CoV positive, 8 OC43-CoV positive, 1 NL63-CoV positive, and 7 CoV negative.
Determined by analysis of a negative sample adjacent to a positive sample.
MultiCode Ct values are typically higher than TaqMan Ct values (Voermans et al., 2016).
Determined on a 10-fold dilution series (5 dilutions) of a strong positive MERS-CoV EQC sample (Ct value Quantstudio ≈ 22).
Hands-on system evaluation.
| ARIES | BD MAX | InGenius | |
|---|---|---|---|
| Required on-site training | 1 day | 1 day | 1 day |
| Possibility to load primary samples | No, manual pipetting required | No, manual pipetting required | Yes |
| Storage of TNA extract | Not available | Not available | Available |
| Ease of use | Very easy | Easy | Easy/moderate |
| Most challenging step | Manual pipetting of primers | Manual pipetting of mastermix | Loading of disposables |
| Total hands-on time for 4 samples | 6 min 5 s | 12 min 25 s | 12 min 0 s |
| - Sample preparation | 4 min 25 s | 4 min 35 s | 1 min 20 s |
| - Device loading | 1 min 10 s | 5 min 40 s | 6 min 50 s |
| - Device clean-up | 0 min 30 s | 2 min 10 s | 3 min 50 s |
| On-board extraction and PCR time | 1 h 54 min | 1 h 22 min | 2 h 12 min |
| Total analysis time | 2 h 0 min 5 s | 1 h 34 min 25 s | 2 h 24 min 0 s |
| Reagent cost per sample | Contact company | 25.7 € | 12.1 € |
| Maximum throughput in a 1-day shift | - 2 × 6 samples/run | - 2 × 12 samples/run | - 12 samples/run |
| Different assays in the same run | Yes, same PCR profile | Yes, same extraction kit | Yes, same extract elution volume |
| Number of optical channels for multiplexing | 5 optical channels + melting curve | 5 optical channels + melting curve | 6 optical channels + melting curve |
| Software package and location | SYNCT: separate PC and device | BD MAX: PC next to device | InGenius: device |
| Automatic result interpretation | Programmable interpretation. | Programmable interpretation. | Programmable interpretation. |
| Onboard reagent traceability | Extraction cassette | Sample Buffer Tube | Extraction cassette |
| LIS communication | 2-way connection | 2-way connection | 2-way connection |
If available, based on the list price (VAT excluded) at the time of writing.
Was not evaluated during this study.
Fig. 1Graphical representation of the sample preparation, device loading, TNA extraction, PCR, and clean-up for the BD MAX, ELITe InGenius, and Luminex ARIES.
Copyright statement:Fig. 1contains several images derived from device product manuals. Permission was obtained from BD, ELITech, and Luminex to include these images in our manuscript.