Does deep inspiration breath hold reduce plan complexity? Multicentric experience of left breast cancer radiotherapy with volumetric modulated arc therapy.

Serenella Russo1, Marco Esposito2, Victor Hernandez3, Jordi Saez4, Francesca Rossi5, Lisa Paoletti5, Silvia Pini2, Paolo Bastiani5, Giacomo Reggiori6, Giorgia Nicolini7, Eugenio Vanetti7, Stefano Tomatis6, Marta Scorsetti8, Pietro Mancosu6.   

Abstract

PURPOSE: Volumetric modulated arc therapy (VMAT) for left breast treatments allows heart sparing without compromising PTV coverage. However, this technique may require highly complex plans. Deep Inspiration Breath Hold (DIBH) procedure increases the heart-to-breast distance, facilitating the dose sparing of the heart. The aim of the present work was to investigate if the cardiac-sparing benefits of the DIBH technique were achieved with lower plan modulation and complexity than Free Breathing (FB) treatments. METHODS AND MATERIALS: Ten left side breast cases were considered by two centers with different treatment planning systems (TPS) and Linacs. VMAT plans were elaborated in FB and DIBH according to the same protocol. Plan complexity was evaluated by scoring several complexity indices. A new global score index accounting for both plan quality and dosimetric parameters was defined. Pre-treatment QA was performed for all VMAT plans using EPID and Epiqa software.
RESULTS: DIBH-VMAT plans were associated with significant PTV coverage improvement and mean heart dose reduction (p < 0.003), increasing the resulting global score index. All the evaluated complexity indices showed lower plan complexity for DIBH plans than FB ones, but only in few cases the results were statistically significant. All plans passed the gamma analysis with the selected criteria.
CONCLUSIONS: The DIBH technique is superior to the FB technique when the heart needs further sparing, allowing a reduction of the doses to OARs with a slightly lower degree of plan complexity and without compromising plan deliverability. These benefits were achieved regardless of the technological scenarios adopted.
Copyright © 2019. Published by Elsevier Ltd.

Entities:  

Keywords:  Complexity index; DIBH; Left breast; VMAT

Mesh:

Year:  2019        PMID: 30928069     DOI: 10.1016/j.ejmp.2019.02.018

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

Review 1.  Complexity metrics for IMRT and VMAT plans: a review of current literature and applications.

Authors:  Sophie Chiavassa; Igor Bessieres; Magali Edouard; Michel Mathot; Alexandra Moignier
Journal:  Br J Radiol       Date:  2019-07-24       Impact factor: 3.039

2.  Limiting treatment plan complexity by applying a novel commercial tool.

Authors:  Alessandro Scaggion; Marco Fusella; Giancarmelo Agnello; Andrea Bettinelli; Nicola Pivato; Antonella Roggio; Marco A Rossato; Matteo Sepulcri; Marta Paiusco
Journal:  J Appl Clin Med Phys       Date:  2020-05-21       Impact factor: 2.102

3.  Split-VMAT technique to control the deep inspiration breath hold time for breast cancer radiotherapy.

Authors:  Sara Poeta; Younes Jourani; Alex De Caluwé; Robbe Van den Begin; Dirk Van Gestel; Nick Reynaert
Journal:  Radiat Oncol       Date:  2021-04-20       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.