The knowledge of in vitro and in vivo stability of polymeric nanoparticles is vital for the development of clinical formulations for drug delivery and cell labeling applications. Förster resonance energy transfer (FRET)-based fluorescence labeling approaches are promising tools to study nanoparticle stability under different physiological conditions. Here, we present the FRET-based stability assessment of poly(lactic-co-glycolic acid) (PLGA) nanoparticles encapsulating BODIPY-FL12 and Nile Red as the donor and acceptor, respectively. The stability of PLGA nanoparticles is studied via monitoring the variations of fluorescence emission characteristics along with colloidal characterization. Accordingly, PLGA nanoparticles are colloidally stable for more than 2 weeks when incubated in aqueous buffers in situ, whereas in vitro particle degradation starts in between 24 and 48 h, reaching a complete loss of FRET at 72 h as shown with fluorescence microscopy imaging and flow cytometry analysis. PLGA nanoparticles systemically administered to mice predominantly accumulate in the liver, in which FRET no longer takes place at time points as early as 24 h postadministration as determined by ex vivo organ imaging and flow cytometry analysis. The results of this study expand our knowledge on drug release and degradation behavior of PLGA nanoparticles under different physiological conditions, which will prove useful for the rational design of PLGA-based formulations for various applications that can be translated into clinical practice.
The knowledge of in vitro and in vivo stability of polymeric nanoparticles is vital for the development of clinical formulations for drug delivery and cell labeling applications. Förster resonance energy transfer (FRET)-based fluorescence labeling approaches are promising tools to study nanoparticle stability under different physiological conditions. Here, we present the FRET-based stability assessment of poly(lactic-co-glycolic acid) (PLGA) nanoparticles encapsulating BODIPY-FL12 and Nile Red as the donor and acceptor, respectively. The stability of PLGA nanoparticles is studied via monitoring the variations of fluorescence emission characteristics along with colloidal characterization. Accordingly, PLGA nanoparticles are colloidally stable for more than 2 weeks when incubated in aqueous buffers in situ, whereas in vitro particle degradation starts in between 24 and 48 h, reaching a complete loss of FRET at 72 h as shown with fluorescence microscopy imaging and flow cytometry analysis. PLGA nanoparticles systemically administered to mice predominantly accumulate in the liver, in which FRET no longer takes place at time points as early as 24 h postadministration as determined by ex vivo organ imaging and flow cytometry analysis. The results of this study expand our knowledge on drug release and degradation behavior of PLGA nanoparticles under different physiological conditions, which will prove useful for the rational design of PLGA-based formulations for various applications that can be translated into clinical practice.
Drug
delivery systems utilizing polymeric nanocarriers have emerged
as versatile tools with tremendous potential in clinical applications.[1−4] Entrapment of drugs within polymeric nanoparticles (NPs) offers
the advantages of reduced toxicity and systemic side effects, as well
as improved stability and targeted delivery of the drug.[5] Furthermore, colloidal (e.g., size, composition,
loading efficiency) and functional (e.g., release profile) properties
of nanoparticles can be adjusted to achieve the desired therapeutic
effects.[6] An optimally designed drug delivery
system should ensure the availability of the drug at a specific location
for a required period of time.[7] Despite
recent advances in the development of drug delivery nanoparticles
with precisely controlled colloidal and functional properties, their
translation to clinical use remains challenging. This is largely because
the nanoparticles developed and tested in vitro do not necessarily
reflect the in vivo performance, especially when drug release is also
dependent on polymer degradation.[8−10] Furthermore, complex
biological environments, including blood components such as serum
proteins, lipids, enzymes, or cells, can interfere with the colloidal
and functional stability of nanoparticles, causing drug leakage or
premature release.[11−13] Therefore, the rational design of drug delivery nanoparticles
with an improved therapeutic efficacy requires the development of
robust techniques to monitor the functional stability of nanoparticles
in different physiological conditions.Fluorescence labeling
strategies are commonly employed to evaluate
drug retention and cellular uptake of nanoparticles.[14] Typically, nanoparticles that encapsulate a single type
of fluorescent label are tracked using fluorescence microscopy or
flow cytometry analysis.[15] However, the
colloidal stability of nanoparticles cannot be assessed with this
approach because the fluorescence signals originating from the released
and retained dyes cannot be distinguished.[16] Alternatively, two-color labeled nanoparticles can be designed using
fluorescent dyes that exhibit a Förster resonance energy transfer
(FRET).[17] FRET is a nonradiative transfer
of energy from an excited donor molecule (D) to a nearby acceptor
(A), displaying spectral overlap via dipole–dipole coupling.[18] The strong dependence of FRET on the D–A
separation distance can provide useful information on the release[13] and degradation[19] profile of nanoparticles encapsulating D–A dyes. When dyes
are in close proximity within the intact nanoparticle, the energy
is transferred from the excited D to A, resulting in D quenching and
A sensitization. Upon dye release and nanoparticle degradation, the
D–A distance becomes too large for an efficient energy transfer.
Therefore, release and degradation profiles of nanoparticles can be
determined by monitoring the recovery of D emission and loss of A
sensitization using fluorescence spectroscopy and imaging.[20,21]FRET-based assessment of dye release and particle degradation
have
been reported for various types of polymeric carriers formulated as
self-assembled micelles[11,22−24] and nanoparticles.[10,25−28] Among the polymeric carriers,
poly(lactic-co-glycolic acid) (PLGA) has received
particular attention due to its excellent biocompatibility, tunable
degradation characteristics, and long clinical history.[29] Drug delivery systems based on PLGA particles
are already available on the market (e.g., Lupron Depot by Abbott
Laboratories, USA, and Trelstar by Watson Pharmaceuticals, USA), and
new formulations are under continuous development for various applications.
Although the development of PLGA particles of varying size,[30,31] surface functionality,[32−34] and encapsulated cargo[35−37] have been shown in several studies, reports on the colloidal and
functional stability of these particles in vivo are scarce.[38] We have previously shown that PLGA nanoparticles
of ∼200 nm diameter are suitable for cell loading, do not affect
cell viability, and can specifically deliver their cargo to desired
cell populations[39−42] with a relatively higher intracellular uptake compared to micron-sized
particles.[43]In this study, we present
a detailed stability assessment of PLGA
nanoparticles of ∼200 nm diameter in situ, in vitro, and in
vivo, exploiting a FRET-based fluorescence labeling approach. We study
both the functional (i.e., release profile) and colloidal stability
(i.e., degradation) of the PLGA nanoparticles using FRET. PLGA nanoparticles
were loaded with BODIPY-FL and Nile Red dyes that acted as D and A,
respectively. After optimization of the dye loading ratio, we studied
the dye release and particle degradation profile in different buffers
in situ. In vitro assessment of particle uptake and stability was
done using human-derived peripheral blood mononuclear cells and mouse-derived
bone marrow dendritic cells. Finally, we tracked nanoparticles systemically
administered to mouse by extracting and analyzing organs for the evaluation
of in vivo particle stability. Monitoring the recovery of quenched
D fluorescence upon dye release and particle degradation revealed
striking differences under different physiological conditions. The
results demonstrated here will prove useful for rational design of
PLGA nanoparticles for applications, such as drug delivery or cell
loading for therapy monitoring with an improved therapeutic efficacy.
Materials and Methods
Materials
PLGA (Resomer RG 502H),
with a 50:50 ratio of lactic acid–glycolic acid and MW 7000–17 000
Da, was obtained from Evonik Nutrition & Care GmbH (Germany).
Poly(vinyl alcohol) (PVA, 9000–10 000 Mw, 80%, hydrolyzed)
and Nile Red were obtained from Sigma-Aldrich (USA). CholEsteryl BODIPY-FL
C12 was purchased from Thermo Fisher Scientific (USA) and acetonitrile
was from VWR (The Netherlands). Ultrapure Milli-Q water (18.2 MΩ
cm) was used where necessary (Merck, USA). RPMI-1640 medium, Anti-Anti
(AA), and ß-mercaptoethanol were obtained from Gibco. X-Vivo
medium and ultraglutamine were from Lonza. Fetal bovine serum (FBS)
was purchased from Hyclone (GE Healthcare, USA).
Nanoparticle Formulation
Dye-loaded
PLGA nanoparticles were prepared via a nanoprecipitation method. Briefly,
the organic phase containing 4 mg of PLGA in 72 μL of acetonitrile
was mixed with the appropriate amounts of fluorescent dyes followed
by dropwise addition to 1 mL of 2% aqueous PVA solution under constant
stirring at 450 rpm. Particles were washed three times with ultrapure
water by centrifugation at 15 000 rpm for 35 min after evaporation
of the organic solvent overnight at 4 °C. Finally, particles
were lyophilized. For the determination of optimal dye concentration
without inducing dye quenching, BODIPY-FL green was loaded within
PLGA nanoparticles at concentrations of 0.1, 0.5, 1.0, 1.3, 1.5, 1.6,
and 2.0% (w/w). For the Nile Red encapsulation, the dye concentration
was varied as 0.1, 0.3, 0.7, 1.0, and 1.3% (w/w). Dye loading ratios
determined in the previous optimization steps were used for coencapsulation
of donor (1.0%) and acceptor (0.2%) dyes to prepare PLGA nanoparticles
that exhibit FRET.
Colloidal and Optical Characterization
of
Nanoparticles
The size distribution of nanoparticles was
measured via dynamic light scattering (DLS) (Nanotrac Flex, Microtrac).
Briefly, 50 μL of the particle suspension was diluted to 1 mL
using filtered Milli-Q water to avoid multiple scattering events.
The measurements were obtained at room temperature (25 °C ±
1 °C), at which the corresponding water viscosity and refractive
index were 0.872 and 1.330 cP, respectively. An average of three measurements
was used to report the Z-average values for each sample. The ζ
potential of nanoparticles was measured using a Zetasizer Nano ZS
(Malvern Instruments). Particles were dispersed in 5 mM NaCl solution
for the measurements, and an average of three measurements was used
to report the ζ potential. Atomic force microscopy (AFM) images
of nanoparticles were obtained with a Catalyst BioScope (Bruker) coupled
to a confocal microscope (TCS SP5II, Leica). 100 μL of the particle
suspension was dried on clean glass substrates, and particles were
imaged in peak-force tapping mode using silicon nitride cantilevers
with nominal spring constants of 0.4 N/m (Bruker). AFM images were
analyzed using NanoScope analysis software (Bruker).Optical
characterization of fluorescent nanoparticles was done by measuring
the steady-state fluorescence of diluted samples (1 mg/mL) using Eppendorf
semimicro Vis Cuvettes on an LS 55 Fluorescence spectrometer (PerkinElmer).
A xenon lamp was used as the excitation source. Samples were excited
at 488 nm, and the emission was recorded between 500 and 800 nm. The
width of excitation and detection slits and other data collection
parameters were kept the same for each measurement.
Stability Assessment of PLGA Nanoparticles
in Situ
PLGA nanoparticles encapsulating D and A dyes (will
be referred to as “FRET NP”) were studied in situ to
determine dye release and particle integrity. A suspension of FRET
NP in Milli-Q water was incubated at 37 °C for a period of 2
weeks. Emission spectra and size distribution of particles were measured
at different time points. PLGA nanoparticles encapsulating the only
donor BODIPY-FL green dye (will be referred to as “green NP”)
and acceptor Nile Red dye (will be referred to as “red NP”)
were studied in parallel as a control. The recovery of FRET-quenched
D emission was monitored using fluorescence spectroscopy and was compared
to variations of the green NP emission; thus, the influence of direct
Nile Red excitation was excluded from the analysis. This process was
repeated for nanoparticles incubated in PBS buffers at pH 7.4 and
at pH 5.8 for a period of 1 week.
Cell
Culture
Peripheral blood mononuclear
cells (PBMCs) were isolated from buffy coats of healthy individuals
after informed consent, using ficoll density centrifugation (Lymphoprep,
STEMCELL Technologies, Vancouver, Canada). Adherent monocytes were
cultured in X-VIVO 15 media supplemented with 2% human serum and in
the presence of interleukin-4 (300 U/mL) and granulocyte-monocyte
colony stimulating factor (GM-CSF, 450 U/mL) to obtain immature dendritic
cells (DCs).Bone marrow-derived DCs (BMDCs) were isolated from
the femur and tibia of donor mice. Isolated cells were then cultured
in RPMI-1640 medium (supplemented with 10% FBS, 1% AA, 1% Ultraglutamine,
and β-mercaptoethanol) in the presence of murine GM-CSF (20
ng/mL).
Cell Labeling
Uptake and intracellular
trafficking of nanoparticles were studied on a monocyte-derived DC
(moDC) culture. Day-3 immature moDCs were collected from the Costar
flasks and counted using Trypan Blue. The cell suspension was diluted
to the desired concentration (0.1 × 106/mL). Then,
0.2 × 105 cells were plated on sterile coverslips
in 24-well plates and were labeled with green NP at a concentration
1 mg of nanoparticles per million cells. The cells were then incubated
for 6 and 24 h at 37 °C. At each time point, the medium was removed
and the coverslips were washed carefully with PBS. The cells were
then fixed by adding 300 μL of 2% paraformaldehyde (PFA) and
permeabilized with PBA + 0.1% Saponin. LAMP1 (Sigma-Aldrich)- or EEA1
(BD Bioscience)-specific primary antibodies (Ab) were incubated with
cells followed by staining with isotype-specific 568AlexaFluor-conjugated
secondary Ab. The cell nuclei were stained with DAPI using 4 μL
droplets of a Mowiol–DAPI mixture. Coverslips were transferred
to the slides, kept overnight in the dark, and then examined with
an Olympus FV1000 confocal laser scanning microscope. An argon laser
(488 nm) was used as the excitation source, and images were collected
with an UPLSAPO 60× objective at a 502–538 nm range. Images
were then processed with ImageJ software.
Stability
Assessment of PLGA NP in Vitro
Day-3 human moDCs and day-14
murine CD103+ BMDCs were
labeled with the particles at a concentration of 1 mg of nanoparticles
per million cells and were incubated for different time points (0.5,
1, 3, 6, 48, and/or 72 h). At each time point, the cells were analyzed
with flow cytometry (BD FACS Verse, BD Biosciences). FlowJo analysis
software was used to determine the mean fluorescence intensity of
cells. BMDCs were also imaged with a Leica DMI6000 epi-fluorescence
microscope equipped with a 63× 1.4 NA oil immersion objective,
a metal halide EL6000 lamp for excitation, a DFC365FX CCD camera,
and GFP filter sets (all from Leica) to measure the variations of
intracellular intensity at the green emission channel for green NP
and FRET NP at 24 and 72 h of incubation.
Stability
Assessment of PLGA NP in Vivo
Albino C57BL/6J female mice
(B6 (Cg)-Tyrc-2J) were obtained from
Jax Laboratory and maintained under specific pathogen-free conditions
at the Central Animal Laboratory in Nijmegen, The Netherlands. Experiments
were performed according to the guidelines for animal care of the
Nijmegen Animal Experiments Committee (DEC 2016-0045). Mice were injected
intravenously (i.v.) with 200 μL of either FRET NP or green
NP at a concentration of 10 mg/mL. The organs (liver, spleen, kidneys,
lungs) were isolated at 2 and 24 h postinjection after sacrificing
the mice and were imaged with an IVIS LUMINA in vivo imaging system
with the excitation set to 465 nm, and the emission was detected at
GFP channel (5 s exposure time). Finally, the cells isolated from
the liver were analyzed with flow cytometry.
Results and Discussion
Formulation and Characterization
of Dye-Loaded
PLGA Nanoparticles
PLGA nanoparticles (PLGA NPs) encapsulating
fluorescent dyes were prepared via a nanoprecipitation method.[6] A water-miscible organic solvent (acetonitrile)
containing PLGA and fluorescent dyes was mixed with the aqueous phase
containing 2% PVA solution under constant stirring. The hydrophobic
nature of both dyes allowed for their encapsulation within PLGA NPs
during the particle formation. PLGA NPs encapsulating only donor (green
nanoparticles), only acceptor (red nanoparticles), and a combination
of donor/acceptor dyes (FRET nanoparticles) were prepared (Figure A). Particles with
an average diameter of ∼200 nm (PDI < 0.2) and a slightly
negative ζ potential (−2.74 ± 0.377 mV for FRET
nanoparticles) were obtained using this preparation method. A representative
atomic force microscopy image of FRET nanoparticles is shown in Figure B. BODIPY-FL and
Nile Red were selected as the D–A pair, respectively, due to
their excellent spectral overlap[44] (Figure C).
Figure 1
(A) Schematic illustration
for the preparation of PLGA NPs encapsulating
only donor (green NPs), only acceptor (red NPs), and coencapsulating
donor and acceptor dyes (FRET NPs). (B) A representative atomic force
microscopy height image of FRET NPs coencapsulating 1% donor and 0.2%
acceptor (w/w) dyes. Scan size: 750 nm × 750 nm. Scale bar: Z-axis. (C) Normalized excitation (dashed curves) and emission
(solid curves) spectra of BODIPY-FL (green) and Nile Red (red) donor–acceptor
pair.
(A) Schematic illustration
for the preparation of PLGA NPs encapsulating
only donor (green NPs), only acceptor (red NPs), and coencapsulating
donor and acceptor dyes (FRET NPs). (B) A representative atomic force
microscopy height image of FRET NPs coencapsulating 1% donor and 0.2%
acceptor (w/w) dyes. Scan size: 750 nm × 750 nm. Scale bar: Z-axis. (C) Normalized excitation (dashed curves) and emission
(solid curves) spectra of BODIPY-FL (green) and Nile Red (red) donor–acceptor
pair.The large spectral overlap integral,
determined as J = 2.25 × 1015 nm4 M–1 cm–1, ensured an
efficient energy transfer between
the BODIPY-FL and Nile Red. The D–A separation distance, at
which the FRET efficiency is 50% (i.e., the Förster distance, R0) was calculated using eq .The FRET orientation factor
κ2 = 2/3, the refractive index of PLGA n = 1.46, and the BODIPY-FL quantum yield ΦD = 0.9
led to a Förster distance of R0 = 5.5 nm. Therefore, an efficient FRET distance of approximately
2.75–11 nm that corresponded to ca. 0.5 to 2-fold R0 was obtained using this FRET pair.An ideal FRET
probe to assess particle stability should be robust
and efficient, which largely depends on the photophysical properties
of the fluorescent dyes used as FRET pairs. The robustness ensures
that observed changes in the emission characteristics are solely due
to dye release and particle degradation rather than modified emission
properties of fluorescent dyes due to, e.g., temperature. In this
respect, BODIPY dyes with an excellent stability are good candidates
to be utilized as FRET probes. However, previously reported FRET pairs
constructed with BODIPY dyes as both donors and acceptors displayed
smaller Förster distances.[20] The
Förster distance is a measure of the donor–acceptor
separation distances that can be probed efficiently using FRET. The
ability to detect FRET at larger molecular separation distances results
in improved precision even at larger length scales. By pairing the
BODIPY donor with a Nile Red acceptor, we drastically improved the
Förster distance as 5.5 nm, which enabled an efficient energy
transfer between donor and acceptor molecules at separation distances
as large as 11 nm (2-fold of Förster distance). Therefore,
compared to previously reported studies, the FRET system we developed
displayed both a high robustness and a more efficient energy transfer
at larger distances.Prior to preparation of FRET NPs, loading
ratios of both dyes were
optimized. A series of dye concentrations in the 0.1–2.0% (w/w)
range were used for encapsulation within PLGA NPs. Self-quenching
of dyes was observed at concentrations above 1.3% and 0.3% (w/w) for
the donor (Figure A) and acceptor dyes (Figure B), respectively. Therefore, loading ratios of 1.3% D and
0.3% A were used for coencapsulation to prepare FRET NPs, for which
an efficient quenching of D emission was observed. Lowering the dye
fractions to 1.0% D and 0.2% A was observed to increase the efficiency
of D quenching further. Figure C shows the emission spectra of FRET NPs with different D–A
fractions normalized at the acceptor emission wavelength (∼615
nm). The size distribution analysis of FRET NPs showed a mean diameter
of ∼200 nm with a small PDI (<0.2) for all particles (Figure D). Other tested
loading ratios displayed either a low FRET efficiency (Figure S1A) or larger average particle size (Figure S1B). Therefore, the formulation containing
1.0% D and 0.2% A was selected for further studies. For this loading
ratio, the FRET efficiency (E) was calculated as
∼70% using eq .IDA and ID are the total donor fluorescence
intensities in the presence and absence of the acceptor, respectively.
Figure 2
(A) Fluorescence
intensity of green NPs prepared with various concentrations
of BODIPY-FL (from 0.1% to 2.0%). (B) Fluorescence intensity of red
NPs prepared with various concentrations of Nile Red (from 0.1% to
1.3%). (C) Emission spectra of FRET NPs loaded with 1.0% D–0.2%
A (black), 1.0% D–0.3% A (red), and 1.3% D–0.3% A (blue)
normalized at the acceptor emission wavelength (λexcitation 488 nm). (D) Size distribution of FRET NPs with the depicted D–A
loading ratios. (E) Emission spectra of green NPs loaded with 1.0%
BODIPY-FL (green), red NPs loaded with 0.2% Nile Red (red), and FRET
NPs coencapsulating 1.0% BODIPY-FL and 0.2% Nile Red (orange). (F)
Emission spectra of intact FRET NPs dispersed in water (orange) and
disassembled FRET NPs dispersed in acetonitrile (green).
(A) Fluorescence
intensity of green NPs prepared with various concentrations
of BODIPY-FL (from 0.1% to 2.0%). (B) Fluorescence intensity of red
NPs prepared with various concentrations of Nile Red (from 0.1% to
1.3%). (C) Emission spectra of FRET NPs loaded with 1.0% D–0.2%
A (black), 1.0% D–0.3% A (red), and 1.3% D–0.3% A (blue)
normalized at the acceptor emission wavelength (λexcitation 488 nm). (D) Size distribution of FRET NPs with the depicted D–A
loading ratios. (E) Emission spectra of green NPs loaded with 1.0%
BODIPY-FL (green), red NPs loaded with 0.2% Nile Red (red), and FRET
NPs coencapsulating 1.0% BODIPY-FL and 0.2% Nile Red (orange). (F)
Emission spectra of intact FRET NPs dispersed in water (orange) and
disassembled FRET NPs dispersed in acetonitrile (green).Steady-state fluorescence spectroscopy analysis
revealed quenching
of D emission accompanied by a sensitized A emission for FRET NPs
compared to the emission spectra of green NPs containing only 1.0%
BODIPY-FL and red NPs containing only 0.2% of Nile Red (Figure E). Disassembly of FRET NPs
upon resuspension in acetonitrile resulted in the recovery of D emission
and loss of A sensitization (Figure F), indicating the relevance of emission characteristics
to structural integrity. Relying on these observations, the spectral
changes of FRET NPs were monitored under different physiological conditions
to obtain information on the structural changes of NPs that result
in an increase in the D–A separation distance, such as dye
release and particle degradation.
Assessment
of Dye Release and Particle Stability
in Situ
The optical and colloidal analysis of FRET NPs incubated
in an aqueous solution at 37 °C enabled the particle stability
assessment in situ. Fluorescence emission spectra and size distribution
of NPs were measured at different time points up to 2 weeks. Figure A shows the emission
spectra of FRET NPs normalized at the emission peak of the acceptor dye.
A gradual recovery of D emission was observed at longer incubation
times, which indicated a lower FRET efficiency at these time points.
The FRET ratios were calculated by dividing the emission intensity
of A by the emission intensity of D at each measurement point (Figure B). As a control,
green NPs containing only 1.0% D were studied in parallel. Calculation
of FRET ratios is a useful way to analyze FRET as it involves the
contribution of both D and A emission intensities simultaneously.
A decrease in FRET ratio would indicate a less efficient energy transfer
between the D/A pair due to dye release or particle degradation. Indeed,
a sharp decrease of FRET ratio was observed at day 1, which was due
to an initial burst release of dyes. The biphasic release pattern
that involves this initial burst followed by a more sustained release
is typically observed for PLGA NPs.[45] The
gradual decrease of FRET ratio at the following time points corresponded
to the sustained release phase.
Figure 3
Stability assessment of PLGA NPs in situ.
(A) Fluorescence emission
spectra of FRET NPs normalized at the acceptor emission peak measured
at 4 h (black), day 1 (red), day 2 (blue), day 3 (magenta), day 6
(green), day 8 (navy), and day 14 (purple). Excitation wavelength:
488 nm. (B) FRET ratio plot for FRET NPs (orange) and green NPs (green)
at different measurement time points. (C) Average particle size of
FRET NPs measured at different incubation times. (D) Normalized emission
spectra of FRET NPs (black), green NPs (blue), and disassembled FRET
NPs (red) on day 14.
Stability assessment of PLGA NPs in situ.
(A) Fluorescence emission
spectra of FRET NPs normalized at the acceptor emission peak measured
at 4 h (black), day 1 (red), day 2 (blue), day 3 (magenta), day 6
(green), day 8 (navy), and day 14 (purple). Excitation wavelength:
488 nm. (B) FRET ratio plot for FRET NPs (orange) and green NPs (green)
at different measurement time points. (C) Average particle size of
FRET NPs measured at different incubation times. (D) Normalized emission
spectra of FRET NPs (black), green NPs (blue), and disassembled FRET
NPs (red) on day 14.Monitoring the green NPs in parallel ensured that the observed
changes in the FRET ratio were not due to the influence of other factors
(such as temperature) on the photophysical properties of the BODIPY-FL.
As shown in Figure B, the variations in the FRET ratio values of green NPs were negligible
compared to FRET NPs due to the remarkable stability of BODIPY dyes.[46]Here, the decrease of FRET efficiency
was mainly the result of
dye release, particle degradation, or a combination of both. A nearly
complete degradation of PLGA NPs with an average size of 230 nm was
previously reported to take place within 10 weeks of incubation.[47] Interestingly, these nanoparticles maintained
their size until they were totally degraded, despite the decrease
in their molecular weight. In our study, the particle size measurements
showed that the average size of nanoparticles remained within the
200 nm range during the incubation period as well (Figure C). Although a certain degree
of particle degradation possibly took place during the 2 weeks of
incubation,[48] it can be concluded that
dye release was the major mechanism that resulted in a decreased FRET
efficiency.On the last day of incubation, FRET NPs were redispersed
in acetonitrile
in order to induce particle disassembly. These “broken”
FRET NPs served as “no-FRET controls”. The emission
spectra of FRET NPs dispersed in water and acetonitrile in comparison
to green NPs are shown in Figure D. The comparison of the emission spectra of FRET NPs
to their “broken” counterparts on day 14 revealed the
presence of FRET still on week two, which was evident by the A sensitization
peak for the FRET NPs dispersed in water. The small shoulder observed
around 620 nm for the “broken” FRET particles was due
to the direct excitation of acceptor dye.The release profile
of FRET NPs was also studied in neutral and
slightly acidic conditions as pH can influence cargo release mechanisms
and particle degradation characteristics.[49] FRET NPs dispersed in PBS buffer with pH 5.8 (Figure A,B) or pH 7.4 (Figure C,D) were incubated at 37 °C for a period
of 7 days. Fluorescence emission spectra and particle size distribution
(Figure S2) were recorded at different
time points. While a higher initial burst release was observed at
pH 7.4 (Figure D),
the release was faster at pH 5.8 in the following release period (Figure B). The burst release
takes place during the initial water uptake and swelling of the lyophilized
nanoparticles. The PLGA used for the preparation of nanoparticles
was terminated with carboxylic acid groups. Therefore, it is very
likely that a higher degree of interactions took place between the
cationic species present in the buffer and the carboxylic acid groups
on the polymer, which resulted in the facilitated diffusion and a
higher burst release of dyes at pH 7.4. A higher burst release at
pH 7.4 compared to more acidic conditions was reported for PLGA NPs
also in previous studies.[45]
Figure 4
Normalized fluorescence
emission spectra of FRET NPs incubated
at pH 5.8 (A) and pH 7.4 (C) at 0.5 h (black), 4 h (red), day 1 (blue),
day 3 (magenta), and day 7 (green). Excitation wavelength: 488 nm.
FRET ratio plot for FRET NPs (orange) and green NPs (green) dispersed
in PBS buffer with pH 5.8 (B) and pH 7.4 (D).
Normalized fluorescence
emission spectra of FRET NPs incubated
at pH 5.8 (A) and pH 7.4 (C) at 0.5 h (black), 4 h (red), day 1 (blue),
day 3 (magenta), and day 7 (green). Excitation wavelength: 488 nm.
FRET ratio plot for FRET NPs (orange) and green NPs (green) dispersed
in PBS buffer with pH 5.8 (B) and pH 7.4 (D).In addition to burst release, the pH has been shown to influence
the degradation mechanism of PLGA NPs as well.[50] However, an incubation period of 7 days is probably too
short to observe the influence of polymer degradation on the release
profile at different pH values. In one of our previous studies, we
also have shown that the release of a hydrophobic dye was faster in
a more acidic environment.[39] Thus, in situ
release behavior of the formulations that we reported previously is
in good agreement with the results presented in this work.
Assessment of Cellular Uptake and Particle
Stability in Vitro
The intracellular fate of nanoparticles
has a strong influence on their therapeutic efficacy particularly
when the encapsulated cargo is unstable at specific intracellular
locations due to, e.g., organelle’s pH.[51] Therefore, it is necessary to study the cellular uptake,
intracellular trafficking, and in vitro stability of polymeric drug
delivery nanoparticles.Monocyte-derived DCs (moDCs) obtained
from healthy donors, known to have a high phagocytic capacity, were
used for the analysis of uptake and intracellular trafficking of NPs.
Green NPs containing only 1.0% D were used to examine the intracellular
trafficking. MoDCs incubated with green NPs were fixed at 6 and 24
h of incubation for confocal microscopy (Figure A). In order to monitor the colocalization
of nanoparticles with early endosomes and lysosomes, EEA1 and LAMP1
staining were performed, respectively. The intracellular BODIPY-FL
signal was observed to increase at 24 h compared to 6 h of incubation.
In addition, while the NPs colocalized with the early endosomes at
6 h, only a partial overlap with the early endosomes and lysosomes
was observed at 24 h. This is in accordance with the capability of
PLGA nanoparticles to escape endolysosomal pathways as has been shown
also in previous studies.[43,52,53] The mechanism behind the nanoparticle escape is likely due to reversal
of the surface charge of nanoparticles from anionic to cationic in
acidic endolysosomal pH. This reversal leads to the interaction of
nanoparticles with endolysosomal membrane and escape to cytosol.[53] Thus, the fact that we observe only partial
colocalization with an endolysosomal compartment could be explained
by this escape mechanism.
Figure 5
(A) Confocal microscopy images of moDCs incubated
with green NPs
for 6 and 24 h. Cells were stained with DAPI for nuclei (blue) and
either with EEA1 for early endosomes or with LAMP1 for lysosomes (red).
The fluorescence emission of green NPs is presented in green. The
overlap between green NPs and EEA1 or LAMP1 is shown in yellow. (B)
Mean fluorescence intensity values obtained with flow cytometry at
4 and 24 h for moDCs incubated at 4 °C (light gray) and 37 °C
(dark gray) with green NPs. (C) Mean fluorescence intensity plot for
moDCs incubated with green NPs (green) and FRET NPs (orange) for various
time points up to 48 h.
(A) Confocal microscopy images of moDCs incubated
with green NPs
for 6 and 24 h. Cells were stained with DAPI for nuclei (blue) and
either with EEA1 for early endosomes or with LAMP1 for lysosomes (red).
The fluorescence emission of green NPs is presented in green. The
overlap between green NPs and EEA1 or LAMP1 is shown in yellow. (B)
Mean fluorescence intensity values obtained with flow cytometry at
4 and 24 h for moDCs incubated at 4 °C (light gray) and 37 °C
(dark gray) with green NPs. (C) Mean fluorescence intensity plot for
moDCs incubated with green NPs (green) and FRET NPs (orange) for various
time points up to 48 h.In order to evaluate the energy dependence of NP uptake,
cells
were incubated with green NPs at 4 °C and mean fluorescence intensities
were measured at 4 and 24 h using flow cytometry. The intracellular
BODIPY-FL signal was negligible at 4 °C compared to 37 °C
at both incubation times (Figure B), which demonstrated that NPs were mainly taken up
via energy-dependent endocytic pathways. The low level of intracellular
fluorescence at 4 °C can be attributed to the passive entry of
leaked dye molecules via diffusion[39] or
adsorption of nanoparticles to the cell surface.[54] Dye leakage is indeed a common problem for polymeric nanoparticles,
which can result in misinterpretation of their cellular uptake and
intracellular distribution.[16]Cellular
uptake and in vitro stability of nanoparticles was evaluated
by flow cytometry analysis of moDCs incubated with green NPs and FRET
NPs for a period of 48 h (Figure C). The intracellular fluorescence intensity increased
rapidly up to ca. 6 h, and then reached a plateau at 24 h for both
particle types. A slight decrease of the mean fluorescence intensity
was observed for green NPs at 48 h, which can be an indication of
dye release as free dyes show lower emission intensities in an aqueous
environment than in the NPs.[20] Nevertheless,
the emission intensity difference between green NPs and FRET NPs revealed
the presence of FRET still at 48 h.In vitro particle stability
assessment was also performed on murine
CD103+ BMDCs. Cells incubated with FRET NPs and green NPs
were analyzed with fluorescence microscopy imaging and flow cytometry
at 24 and 72 h of incubation (Figure ). Fluorescence microscopy images were obtained using
excitation and emission filters suitable for the detection of BODIPY-FL.
An efficient donor quenching was observed at 24 h of incubation as
indicated by the dimmer intracellular fluorescence intensity for FRET
NPs compared to green NPs (Figure A). On the other hand, similar intracellular intensities
of donor emission observed at 72 h revealed that there was no longer
an efficient energy transfer at this time point between donor and
acceptor dyes. These findings were confirmed by the flow cytometry
analysis as well (Figure B). The mean fluorescence intensity values of FRET NPs measured
at the green emission channel was comparable to that of green NPs
at 72 h, which confirmed the recovery of donor emission at this time
of incubation.
Figure 6
(A) Fluorescence microscopy images of murine BMDCs incubated
with
FRET NPs and green NPs for 24 and 72 h collected at the green emission
channel. Cell nuclei are stained with DAPI (blue). (B) Mean fluorescence
intensity value BMDCs incubated with green NPs (light gray) and FRET
NPs (dark gray) measured at the green emission channel using flow
cytometry. A complete recovery of donor quenching is observed at 72
h.
(A) Fluorescence microscopy images of murine BMDCs incubated
with
FRET NPs and green NPs for 24 and 72 h collected at the green emission
channel. Cell nuclei are stained with DAPI (blue). (B) Mean fluorescence
intensity value BMDCs incubated with green NPs (light gray) and FRET
NPs (dark gray) measured at the green emission channel using flow
cytometry. A complete recovery of donor quenching is observed at 72
h.Overall, fluorescence microscopy
imaging and flow cytometry analyses
revealed a significantly faster donor recovery in vitro, reaching
a complete loss of FRET at 72 h probably due to both dye release and
particle degradation. It should be noted that cell culture media can
alter the colloidal and chemical properties of NPs due to the presence
of proteins and high ion content.[55] Furthermore,
during their uptake through the endocytic pathway, nanoparticles are
exposed to varying physiological conditions such as acidic pH of late
endosomes[55] and high levels of hydrolytic
enzymes in lysosomes.[56] All of these factors
affect nanoparticle stability and degradation, resulting in a faster
release of the encapsulated cargo in vitro when compared to in situ
settings.
Assessment of Particle Stability in Vivo
In vivo stability of particles systemically administered to mice
was studied via ex vivo organ imaging and flow cytometry (Figure ). Green NPs and
FRET NPs were injected intravenously (i.v.) for liver imaging, and
cells were isolated at 2 and 24 h after injection. In accordance with
the biodistribution studies reported for PLGA NPs after i.v. administration,[8,57,58] NPs were detected mainly in liver
as confirmed by flow cytometry analysis and ex vivo organ imaging
(Figure S3). Therefore, cells isolated
from the liver were used in the further analyses. A higher fluorescence
emission intensity was detected in the liver of mice injected with
green NPs (Figure A) compared to those injected with FRET NPs (Figure B). For both experimental groups, a significant
decrease of fluorescence emission was observed at the 24 h time point.
The decrease in emission probably resulted from the particle degradation
or enhanced release, accompanied by fast clearance of the dyes from
the liver. A similar observation was also reported by Simon et al.,
who showed a significant decrease of nanoparticle concentration in
various tissues over 24 h period.[57] Therefore,
the observed decrease in emission could be also attributed to a rapid
clearance of the nanoparticles from the liver as well.
Figure 7
Ex vivo imaging of the
liver 2 and 24 h after i.v. injection of
(A) Green NPs and (B) FRET NPs. (Excitation wavelength = 465 nm, emission
filter = GFP, exposure time = 5 s.) (C) Flow cytometry histograms
of cells isolated from the liver at 2 and 24 h postinjection. (Excitation
wavelength = 488 nm, detection filter = FITC). Red: unlabeled control.
Green: Green NPs. Orange: FRET NPs.
Ex vivo imaging of the
liver 2 and 24 h after i.v. injection of
(A) Green NPs and (B) FRET NPs. (Excitation wavelength = 465 nm, emission
filter = GFP, exposure time = 5 s.) (C) Flow cytometry histograms
of cells isolated from the liver at 2 and 24 h postinjection. (Excitation
wavelength = 488 nm, detection filter = FITC). Red: unlabeled control.
Green: Green NPs. Orange: FRET NPs.Next, we isolated the cells from this organ and analyzed
them with
flow cytometry. FRET was still detectable at 2 h after administration
as depicted by a lower mean fluorescence intensity at the donor emission
channel compared to the cells obtained from the liver of a mouse injected
with green NPs (Figure C). On the other hand, a reduced mean fluorescence intensity was
observed for the cells isolated at the 24 h time point for both FRET
NPs and green NPs (Figure C). Furthermore, intensity peaks for both conditions showed
an overlap, indicating the loss of FRET at this time point. Additional
analysis of the cells shown in Figure C revealed an almost complete loss of FRET in vivo
within 24 h after injection (Figure S4).A previous study by Mohammad et al. showed degradation of PLGA
nanoparticles with a slight reduction in molecular weight already
within the first 24 h after i.v. administration and predominant accumulation
in the liver.[8] In that study, the degradation
profile of 200 nm PLGA nanoparticles showed differences between in
vitro and in vivo conditions, as well as between various types of
tissues. In our study, a faster recovery of the donor emission was
observed in vivo. However, here we used a lower molecular weight polymer
(7000–17 000 Da), which can lead to faster polymer degradation
and a more rapid drug release.[59] The impact
of enzymes on the PLGA degradation is unclear; nevertheless, due to
the fact that liver has a high concentration of esterases,[8] we cannot exclude a possible influence of these
enzymes on the faster release and degradation observed in our study.
The presence of proteins in blood and variations of pH in different
tissue types could also affect the degradation profile of the nanoparticles.
Therefore, our in vivo study suggests the involvement of additional
parameters, e.g., enzymes, on release and nanoparticle degradation.
Conclusion
It is
of major importance to monitor the stability of polymeric
nanoparticles under different physiological conditions for the development
of clinically applicable formulations. In this study, we demonstrated
that FRET is a powerful tool to assess the stability of PLGA nanoparticles
by encapsulating BODIPY-FL green and Nile Red dyes as the donor–acceptor
pair, respectively. These FRET NPs of ∼200 nm in size showed
∼70% FRET efficiency. Optical and colloidal analysis of nanoparticles
revealed that PLGA nanoparticles were stable for at least 2 weeks
in situ, while in vitro particle degradation started in between 24
and 48 h, reaching a complete recovery of donor emission at 72 h.
The nanoparticles were readily taken up by human-derived moDCs mainly
via energy-dependent endocytic pathways, but could escape the endolysosomal
routes after uptake. In vivo studies showed the accumulation of systemically
administered PLGA nanoparticles in the liver and particle degradation
within 24 h. The results of our study will have a major impact on
research focusing on the applications such as drug delivery and cell
therapy monitoring using PLGA nanoparticles, as they provide a better
understanding of the degradation processes and an attractive way to
monitor the uptake of the nanoparticles and release of encapsulated
cargo.
Authors: Andre E Nel; Lutz Mädler; Darrell Velegol; Tian Xia; Eric M V Hoek; Ponisseril Somasundaran; Fred Klaessig; Vince Castranova; Mike Thompson Journal: Nat Mater Date: 2009-06-14 Impact factor: 43.841
Authors: Luis J Cruz; Paul J Tacken; Remco Fokkink; Ben Joosten; Martien Cohen Stuart; Fernando Albericio; Ruurd Torensma; Carl G Figdor Journal: J Control Release Date: 2010-02-13 Impact factor: 9.776
Authors: Malgorzata S Cartiera; Katherine M Johnson; Vanathy Rajendran; Michael J Caplan; W Mark Saltzman Journal: Biomaterials Date: 2009-02-20 Impact factor: 12.479
Authors: Maria Camilla Operti; Yusuf Dölen; Jibbe Keulen; Eric A W van Dinther; Carl G Figdor; Oya Tagit Journal: Pharmaceutics Date: 2019-11-08 Impact factor: 6.321