Chih-Wen Chi1, Eddie Hsiang-Hua Lai1, Chao-Yuan Liu2, Chun-Pin Lin1,3, Chow-Shing Shin4.
Abstract
BACKGROUND/Entities:
Keywords: NiTi endodontic instrument; cutting efficiency; cyclic fatigue life; heat treatment
Year: 2016 PMID: 30895019 PMCID: PMC6395258 DOI: 10.1016/j.jds.2016.06.001
Source DB: PubMed Journal: J Dent Sci ISSN: 1991-7902 Impact factor: 2.080
Figure 1Component of cutting efficiency test device. (A) Illustration of the device. There were two counter rotated acrylic blocks that contacted the file with a force of 10 N. The file was operated by motor. (B) Photograph of the device.
Figure 2DSC results. (A) Control group: normal behavior of ProTaper Universal F2; (B) HT400 group: there was a single sharper and higher peak over the heating and cooling cycles; and (C) HT600 group: there was one peak combined with one plateau over the test. DSC = differential scanning calorimetry; HT400 = heat treatment at 400°C; HT600 = heat treatment at 600°C.
Cyclic fatigue lives of the files in the control and two experimental groups with files testing in the four different simulated canals.
| R10mmθ20° | R7.5mmθ40° | R7.5mmθ60° | R5mmθ60° | |||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Mean | SD | Mean | SD | |
| Control | 9953a | 2860.9 | 1325.3a | 215.2 | 513.6a | 180.1 | 562a | 129.1 |
| HT400 | 27668.8b | 566.1 | 3323.8b | 262.3 | 1233.4a | 284.7 | 1186.6b | 117.8 |
| HT600 | 38406.8c | 8439.0 | 2855.3b | 315.6 | 2680.4b | 382.2 | 2142.4c | 208.1 |
Different superscript letters indicate a significant difference among groups in each simulated canal.
HT400 = heat treatment at 400°C; HT600 = heat treatment at 600°C; SD = standard deviation.
Weight loss (mg) at the four cutting portions of the files in the control and two different experimental groups.
| 3 mm | 6 mm | 9 mm | 12 mm | |||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Mean | SD | Mean | SD | |
| Control | 18.09a | 0.23 | 25.82a | 1.05 | 34.27a | 0.28 | 38.66a | 0.30 |
| HT400 | 18.63b | 0.06 | 26.13a | 0.21 | 31.68b | 0.24 | 35.30b | 0.19 |
| HT600 | 16.28c | 0.11 | 21.07b | 0.18 | 25.23c | 0.22 | 28.21c | 0.08 |
Different superscript letters indicate a significant difference among groups in each simulated canal.
HT400 = heat treatment at 400°C; HT600 = heat treatment at 600°C; SD = standard deviation.