| Literature DB >> 30895003 |
Igor Križnar1, Katja Seme2, Aleš Fidler1.
Abstract
BACKGROUND/Entities:
Keywords: adhesive system; bacterial microleakage; temporary filling materials
Year: 2016 PMID: 30895003 PMCID: PMC6395255 DOI: 10.1016/j.jds.2016.06.004
Source DB: PubMed Journal: J Dent Sci ISSN: 1991-7902 Impact factor: 2.080
Application mode of the materials used in this study.
| Material | Application mode | |
|---|---|---|
| Group 1 | Cavit | Material was applied in bulk technique |
| Group 2 | Fuji II LC | Dentin conditioner was actively applied on all cavity walls using a microbrush and rinsed off after 10 s. The cavity was carefully dried, not to overdry the dentin, using a cotton pellet. Afterwards Fuji II LC was applied in 2 mm increments; each individually cured for 20 s with a LED curing light (Bluephase; IvoClar-Vivadent, Schaan, Liechtenstein; intensity 1200 mW/cm |
| Group 3 | Fuji IX | The access cavity was prepared in the same way as in Group Fuji II LC. Fuji IX was applied in a bulk technique and left to cure. The use of a coating agent was omitted so it would not penetrate into potential gaps. |
| Group 4 | Voco Clip | Material was applied in a bulk technique and cured for 40 s. |
| Group 5 | Tetric EvoCeram & AdheSE | AdheSE Primer was actively applied to all cavity walls for 30 s using a microbrush, and thoroughly dried afterwards. AdheSE Bond was applied for 15 s using a microbrush, light drying and curing with a LED curing light for 15 s followed. Tetric EvoCeram was applied in 2-mm increments; each was individually cured for 40 s. |
| Group 6 | Tetric EvoCeram & Excite | 37% phosphoric acid was applied to dentin walls for 15 s, rinsed off and lightly dried. An adhesive system Excite was thoroughly rubbed into dentin using a microbrush, lightly dried and cured for 20 s. Tetric EvoCeram was applied in 2-mm increments; each was individually cured for 40 s. |
The manufacturers and chemical composition of the materials used in this study.
| Material | Manufacturer | Composition (% weight) |
|---|---|---|
| Cavit | 3M ESPE, St. Paul, MN, USA | Zinc Oxide (30–50%) |
| Fuji II LC | GC EUROPE, Belgium | (Fuloro) Alumino silicate glass (100%) |
| Fuji IX | GC EUROPE, Belgium | Alumino silicate glass (95%) |
| Voco Clip | VOCO GmbH | 2-hydroxyethyl methacrylate (5–10%) |
| Tetric EvoCeram | Ivoclar-Vivadent, | Urethane dimethacrylate (2.5–< 10%) |
| AdheSE | Ivoclar-Vivadent, Schaan, Liechtenstein | Primer Phosphonic acid acrylate (25–50%) |
| Excite | Ivoclar-Vivadent, Schaan, Liechtenstein | SiO2 (silicon dioxide) |
Figure 1(A) Sectioning of the crowns filled with the dental filling materials used in this study; (B) obtaining 5.5-mm-high disks.
Figure 2Checking for turbidity. (A) Nonleaked samples; (B) leaked sample's turbidity in the bottom chamber.
Number of leaked samples and the estimated day of leakage.
| Material | Leaked samples (%) | Time range (d) | Mean day of leakage (d) |
|---|---|---|---|
| Tetric & AdheSE | 31.3 | 3–48 | 67.8 ± 35.1 |
| Cavit | 33.3 | 4–79 | 69.5 ± 34.6 |
| Tetric & Excite | 35.3 | 1–38 | 64.6 ± 36.3 |
| Fuji II LC | 66.7 | 1–55 | 42.9 ± 37.4 |
| Voco Clip | 83.3 | 5–86 | 44.2 ± 36.1 |
| Fuji IX | 88.2 | 1–74 | 21.1 ± 33.8 |
Figure 3Kaplan–Meier survival curve.