Comparison of compression screws used for mandible head fracture treatment-experimental study.

Marcin Kozakiewicz1.   

Abstract

OBJECTIVES: Compression screws that can fix a fracture of the condylar head of the mandible are available. Although typical titanium lag screws are generally used, a series of headless screws have been designed for this purpose. The "headless" advantage derived from the complete submergence of the screw in the bone after fixation is the lack of irritation of the lateral ligament of the temporo-mandibular joint by the screw head. The aim of this study is to compare the compression force of typically used lag screws with that of newly developed headless compression screws.
MATERIALS AND METHODS: This study tested five types of screws with a 14-mm length: 10 2.0-mm lag screws and 40 headless screws (four groups of equal size: 1.5-mm H screws, 1.8-mm H screws, 2.0-mm H screws, and 1.8-mm W screws). Two cuboid blocks of standardized polyurethane foam were fixed by one screw. The maximal compression force (F in N) and torque at maximal compression force (T in Ncm) were tested according to the guidelines of the American Society for Testing and Materials (ASTM F1839-08).
RESULTS: The test results were as follows: 2.0-mm lag screw: F = 211.5 ± 20.7 and T = 21.1 ± 1.6; 1.5-mm H screw: F = 47.2 ± 5.6 and T = 18.2 ± 1.1; 1.8-mm H screw: F = 95.5 ± 9.8 and T = 18.2 ± 1.1; 2.0-mm H screw: F = 145.1 ± 18.4 and T = 25.5 ± 1.5; and 1.8-mm W screw: F = 180.5 ± 8.7 and T = 26.4 ± 2.8. Each type of screw significantly differed from one another as far as compression forces (F) were considered (p < 0.05): the best is the lag screw contrary to the thinnest H screw which is the worst one. For all tested screws, the compression force depended on the torque (S curve relationship was revealed, p < 0.05) and screw diameter (a larger diameter yielded greater compression, p < 0.05).
CONCLUSIONS: The 8-mm headless screws featured only a small loss of compression ability compared to the titanium lag screws. CLINICAL RELEVANCE: Headless screws are mechanically attractive fixation materials for condylar head fracture treatment.

Entities:  

Keywords:  Head fracture; Headless compression screw; Mandible condyle; Surgical treatment

Mesh:

Year:  2019        PMID: 30891645     DOI: 10.1007/s00784-019-02842-y

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  7 in total

1.  Mechanical Evaluation of Titanium Plates for Osteoesynthesis High Neck Condylar Fracture of Mandible.

Authors:  Rafał Zieliński; Marcin Kozakiewicz; Bartłomiej Konieczny; Michał Krasowski; Jakub Okulski
Journal:  Materials (Basel)       Date:  2020-01-27       Impact factor: 3.623

2.  Forces Causing One-Millimeter Displacement of Bone Fragments of Condylar Base Fractures of the Mandible after Fixation by All Available Plate Designs.

Authors:  Marcin Kozakiewicz; Rafal Zieliński; Michał Krasowski; Jakub Okulski
Journal:  Materials (Basel)       Date:  2019-09-25       Impact factor: 3.623

3.  Clinical Evaluation of Magnesium Alloy Osteosynthesis in the Mandibular Head.

Authors:  Marcin Kozakiewicz; Izabela Gabryelczak; Bartosz Bielecki-Kowalski
Journal:  Materials (Basel)       Date:  2022-01-18       Impact factor: 3.623

4.  The Osteosynthesis of the Mandibular Head, Does the Way the Screws Are Positioned Matter?

Authors:  Marcin Kozakiewicz; Izabela Gabryelczak
Journal:  J Clin Med       Date:  2022-04-05       Impact factor: 4.241

5.  Open Rigid Internal Fixation of Low-Neck Condylar Fractures of the Mandible: Mechanical Comparison of 16 Plate Designs.

Authors:  Marcin Kozakiewicz; Rafał Zieliński; Bartłomiej Konieczny; Michał Krasowski; Jakub Okulski
Journal:  Materials (Basel)       Date:  2020-04-22       Impact factor: 3.623

6.  Are Magnesium Screws Proper for Mandibular Condyle Head Osteosynthesis?

Authors:  Marcin Kozakiewicz
Journal:  Materials (Basel)       Date:  2020-06-10       Impact factor: 3.623

7.  Bone Union Quality after Fracture Fixation of Mandibular Head with Compression Magnesium Screws.

Authors:  Marcin Kozakiewicz; Izabela Gabryelczak
Journal:  Materials (Basel)       Date:  2022-03-17       Impact factor: 3.623

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.