Piston-Based vs Peristaltic Pump-Based CT Injector Systems.

Amy Chaya, Gregor Jost, Jan Endrikat.   

Abstract

PURPOSE: To compare the fluid delivery performance of 2 different technologies for administration of iodinated contrast media in computed tomography (CT).
METHODS: The maximum achievable flow rate and the steady-state flow rate variance of a piston-based contrast media injector system (Stellant MP, MEDRAD) was compared with peristaltic pump-based injector systems (CT motion, Ulrich Medical; CT Exprès, Bracco). The contrast media iopromide (Ultravist) and iopamidol (Isovue) were used at 2 concentrations each (300 and 370 mg I/mL) and 3 catheter sizes (18, 20, and 22 G) to test the injector performance.
RESULTS: Average maximum achievable flow rates for room temperature iopromide (370 mg I/mL) using a 20 G catheter were 7.6, 7.1, and 4.8 mL per second for the Stellant MP injector, CT motion injector, and CT Exprès injector, respectively. The Stellant MP injector achieved significantly higher flow rates compared to the CT Exprès injector for all catheter sizes tested (P<.001). Higher flow rates also were observed for the Stellant MP injector compared to the CT motion injector, with 20 G and 22 G catheters (P<.001). The Stellant MP injector featured a constant steady-state flow rate (variance<0.04 mL/s), whereas the other systems injected in a pulsatile fashion, with significantly greater variance (P<.001). DISCUSSION: To the authors' knowledge, this is the first reported laboratory study providing preliminary evidence of differences between the fluid delivery performance of CT injection systems. Additional investigations using a dedicated flow phantom simulating human physiological flow parameters should be conducted, and depending on the results, a clinical study could assess the effect on image quality.
CONCLUSION: The piston-based injector demonstrated higher maximum achievable flow rates and more consistent steady-state flow when compared to peristaltic pump-based injectors.

Entities:  

Keywords:  CT injector technology; contrast media delivery; fluid delivery parameters; iodinated contrast media; peristaltic pump; piston-based; power injection

Mesh:

Substances:

Year:  2019        PMID: 30886031

Source DB:  PubMed          Journal:  Radiol Technol        ISSN: 0033-8397


  3 in total

1.  Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT.

Authors:  Corey A Kemper; Casper Mihl; Bibi Martens; Michael C McDermott; Babs M F Hendriks
Journal:  Med Devices (Auckl)       Date:  2022-04-05

2.  Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom.

Authors:  Michael McDermott; Corey Kemper; William Barone; Gregor Jost; Jan Endrikat
Journal:  Br J Radiol       Date:  2020-02-20       Impact factor: 3.039

3.  Disparity of time-contrast curves generated by various types of power injectors used in magnetic resonance imaging.

Authors:  Marcus Doppler; Ewald Moser; Uros Klickovic; Christian Nasel
Journal:  Sci Rep       Date:  2020-11-11       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.