Literature DB >> 35411194

Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT.

Corey A Kemper1, Casper Mihl2,3, Bibi Martens2,3, Michael C McDermott2,4, Babs M F Hendriks2,3.   

Abstract

Purpose: To compare an investigational device (MEDRAD® Centargo CT Injection System, "Centargo") to the currently available MEDRAD® Stellant CT Injection System ("Stellant"), in terms of efficiency, injector performance, and user satisfaction. Patients and
Methods: A total of 425 patients at two sites were enrolled; 198 patients in phase one, a randomized study (98 Stellant and 100 Centargo). The second observational phase included 227 patients who were injected with Centargo. Phase one recorded times for setup, disassembly, and patient changeovers. Demographic data, subjective image quality, and injection parameters were collected. Phase two assessed usability via a questionnaire provided to all end-users of both systems (radiographers).
Results: Patient changeover times were statistically significantly faster with Centargo (15.4s ± 8.7s vs 53.7s ± 19.6s, p < 0.001). Centargo day-setup times were similar to Stellant (138.1s ± 92s vs 151.8s ± 30.6s, p = 0.33) and end-of-day-disassembly times were significantly slower (60.6s ± 27s vs 17.1s ± 12.9s, p < 0.001). Based on four different scenarios modelling patient throughput, the projected time savings with Centargo over Stellant was 40-63%, with the highest efficiency improvements for higher throughputs and the use of larger contrast medium bottles. Both Centargo and Stellant usability averaged between "Very Easy" and "Easy" in all responses to the questionnaire. There were no instances of interrupted injections due to communication loss or detected air and no insufficient images due to injector performance. No safety issues were identified.
Conclusion: Centargo was able to demonstrate improved efficiency as compared to Stellant while maintaining injector performance and high usability scores.
© 2022 Kemper et al.

Entities:  

Keywords:  Centargo; efficiency; power injection; time savings; usability

Year:  2022        PMID: 35411194      PMCID: PMC8994651          DOI: 10.2147/MDER.S353221

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


  11 in total

Review 1.  Intravenous contrast medium administration and scan timing at CT: considerations and approaches.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

Review 2.  Contrast agents in X-ray computed tomography and its applications in oncology.

Authors:  Annemarieke Rutten; Mathias Prokop
Journal:  Anticancer Agents Med Chem       Date:  2007-05       Impact factor: 2.505

3.  Impact of iodine delivery rate with varying flow rates on image quality in dual-energy CT of patients with suspected pulmonary embolism.

Authors:  Jan Hansmann; Christian Fink; Gregor Jost; Hubertus Pietsch; Mathias Meyer; John W Nance; Uwe J Schoepf; Ulrike I Attenberger; Stefan O Schoenberg; Thomas Henzler
Journal:  Acad Radiol       Date:  2013-06-06       Impact factor: 3.173

Review 4.  Evolution in Computed Tomography: The Battle for Speed and Dose.

Authors:  Michael M Lell; Joachim E Wildberger; Hatem Alkadhi; John Damilakis; Marc Kachelriess
Journal:  Invest Radiol       Date:  2015-09       Impact factor: 6.016

5.  Piston-Based vs Peristaltic Pump-Based CT Injector Systems.

Authors:  Amy Chaya; Gregor Jost; Jan Endrikat
Journal:  Radiol Technol       Date:  2019-03

6.  Comparison of dual-syringe and syringeless power injectors in outpatient MDCT practice: impact on the operator's performance, CT workflow, and operation cost.

Authors:  Xiaozhou Ma; Anand Singh; Joseph Fay; Giles Boland; Dushyant V Sahani
Journal:  J Am Coll Radiol       Date:  2012-08       Impact factor: 5.532

7.  Pressure injectors for radiologists: A review and what is new.

Authors:  Inna K Indrajit; Rajeev Sivasankar; John D'Souza; Rochan Pant; Raj S Negi; Samresh Sahu; Pi Hashim
Journal:  Indian J Radiol Imaging       Date:  2015 Jan-Mar

Review 8.  Contrast opacification on thoracic CT angiography: challenges and solutions.

Authors:  Abhishek Chaturvedi; Daniel Oppenheimer; Prabhakar Rajiah; Katherine A Kaproth-Joslin; Apeksha Chaturvedi
Journal:  Insights Imaging       Date:  2016-11-17

9.  Impact of CT Injector Technology and Contrast Media Viscosity on Vascular Enhancement: Evaluation in a Circulation Phantom.

Authors:  Michael McDermott; Corey Kemper; William Barone; Gregor Jost; Jan Endrikat
Journal:  Br J Radiol       Date:  2020-02-20       Impact factor: 3.039

10.  Syringeless power injector versus dual-syringe power injector: economic evaluation of user performance, the impact on contrast enhanced computed tomography (CECT) workflow exams, and hospital costs.

Authors:  Giorgio L Colombo; Ivo A Bergamo Andreis; Sergio Di Matteo; Giacomo M Bruno; Claudio Mondellini
Journal:  Med Devices (Auckl)       Date:  2013-11-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.