| Literature DB >> 30863314 |
Abstract
The minimum stent area (MSA) has been clinically established as a significant predictor of restenosis, thrombosis, and ischemia using intra-vascular ultrasound (IVUS). Unfortunately, IVUS measurements are far from routine because of significant cost of IVUS, the training required, the subjectivity of image interpretation and the time added to the procedure. The objective of this study is to verify the accuracy of a conductance catheter for stent sizing. Here, we introduce an easy and entirely objective device and method for real time determination of MSA. A 10 kHz, 35 μA rms current is passed through the external electrodes of an intravascular catheter while the conductance is measured across a separate set of electrodes. Both phantom and ex vivo validations of metal stent sizing in five porcine carotid arteries were confirmed. The accuracy of the measurements were found to be excellent in phantoms (root mean square, rms, of 3.4% of actual value) and in ex-vivo vessels (rms = 3.2% of measured value). An offset of conductance occurs when a conductive metal stent (e.g., bare metal stent) is deployed in the vessel, while the slope remains the same. This offset is absent in the case of drug eluting stent where the metal is coated (i.e., insulated) or non-metal bioresorbable stent. The present device makes easy, accurate and reproducible measurements of the size of stented blood vessels within 3.2% rms error. This device provides an alternative method to sizing of stent (i.e., MSA) in real-time without subjective interpretation and with less cost than IVUS.Entities:
Keywords: conductance catheter; diameter; drug eluting stents; lumen sizing; minimum stent area
Year: 2019 PMID: 30863314 PMCID: PMC6399122 DOI: 10.3389/fphys.2019.00120
Source DB: PubMed Journal: Front Physiol ISSN: 1664-042X Impact factor: 4.566
FIGURE 1An illustration of an impedance catheter where the four electrodes are spaced at the tip (two inner and two outer electrodes) in the top panel; a zoom of the embedded portion of the electrode arrangement is shown in the middle panel; and a further zoom of the wire tunneling is shown in the lower panel.
FIGURE 2A schematic of an impedance catheter in the lumen of a stented vessel.
FIGURE 3Saline and stent calibrations of 0.45 and 0.9% NaCl solutions. Graph (A) shows conductance values for each corresponding CSA in acrylic tubes filled with saline only. Graph (B) shows the values with the stent embedded in each of the tubes. Graph (C) shows both saline only and stent values with normalized offsets. Lines of best fit are shown in Graphs (A) and (B).
FIGURE 4Phantom diameter measurements determined using impedance catheters versus a caliper. Four catheters were tested in five different tubing sizes. (A) A line of identity shown between caliper diameter and impedance diameter measurements. (B) A Bland-Altman plot of mean of diameter measurements versus percent difference in measurements.
FIGURE 5Inner diameters of five pig carotid arteries measured by impedance catheters versus diameters measured optically. Each artery was pressurized from 20 to 120 mmHg in increments of 20 mmHg during measurements. (A) The line of identity between optical diameter and impedance diameter measurements. (B) A Bland-Altman plot of mean of diameter measurements versus percent difference in measurements.
The calibration slope and offset of the impedance catheter in NaCl and NaCl plus wire coil (stainless steel and tungsten).
| Stainless Steel (SS) | ||||
|---|---|---|---|---|
| 0.45% NaCl | 0.45%NaCl w/ SS | 0.9% NaCl | 0.9%NaCl w/ SS | |
| Slope | 1.1 ± 0.17 | 1.1 ± 0.11 | 1.9 ± 0.31 | 1.9 ± 0.25 |
| Offset | -0.47 ± 0.18 | 3.8 ± 1.5 | -1.1 ± 0.71 | 3.1 ± 1.8 |
| 0.45% NaCl | 0.45%NaCl w/T | 0.9% NaCl | 0.9%NaCl w/T | |
| Slope | 1.2 ± 0.28 | 1.1 ± 0.26 | 1.8 ± 0.41 | 1.9 ± 0.35 |
| Offset | -0.78 ± 0.17 | 45 ± 5.6 | -0.84 ± 1.6 | 46 ± 6.4 |