Literature DB >> 30849014

The Influence of Distance on Radiant Exposure and Degree of Conversion Using Different Light-Emitting-Diode Curing Units.

A O Al-Zain, G J Eckert, J A Platt.   

Abstract

OBJECTIVES: To investigate the influence of curing distance on the degree of conversion (DC) of a resin-based composite (RBC) when similar radiant exposure was achieved using six different light-curing units (LCUs) and to explore the correlation among irradiance, radiant exposure, and DC. METHODS AND MATERIALS: A managing accurate resin curing-resin calibrator system was used to collect irradiance data for both top and bottom specimen surfaces with a curing distance of 2 mm and 8 mm while targeting a consistent top surface radiant exposure. Square nanohybrid-dual-photoinitiator RBC specimens (5 × 5 × 2 mm) were cured at each distance (n=6/LCU/distance). Irradiance and DC (micro-Raman spectroscopy) were determined for the top and bottom surfaces. The effect of distance and LCU on irradiance, radiant exposure, and DC as well as their linear associations were analyzed using analysis of variance and Pearson correlation coefficients, respectively (α=0.05).
RESULTS: While maintaining a similar radiant exposure, each LCU exhibited distinctive patterns in decreased irradiance and increased curing time. No significant differences in DC values (63.21%-70.28%) were observed between the 2- and 8-mm distances, except for a multiple-emission peak LCU. Significant differences in DC were detected among the LCUs. As expected, irradiance and radiant exposure were significantly lower on the bottom surfaces. However, a strong correlation between irradiance and radiant exposure did not necessarily result in a strong correlation with DC.
CONCLUSIONS: The RBC exhibited DC values >63% when the top surface radiant exposure was maintained, although the same values were not reached for all lights. A moderate-strong correlation existed among irradiance, radiant exposure, and DC.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30849014     DOI: 10.2341/18-004-L

Source DB:  PubMed          Journal:  Oper Dent        ISSN: 0361-7734            Impact factor:   2.440


  5 in total

1.  Influence of irradiation distance on the mechanical performances of resin composites polymerized with high-irradiance light curing units.

Authors:  Soram Oh; Hyun Ju Kim; Hyun-Jung Kim; Sibel A Antonson; Sun-Young Kim
Journal:  Biomater Res       Date:  2022-05-20

2.  Effectiveness of Using an Instructional Video in Teaching Light-Curing Technique.

Authors:  Afnan O Al-Zain; Tasneem M Al-Osaimi
Journal:  Adv Med Educ Pract       Date:  2021-03-29

3.  Errors in light-emitting diodes positioning when curing bulk fill and incremental composites: impact on properties after aging.

Authors:  Abdulrahman A Balhaddad; Isadora M Garcia; Haifa Maktabi; Maria Salem Ibrahim; Qoot Alkhubaizi; Howard Strassler; Fabrício M Collares; Mary Anne S Melo
Journal:  Restor Dent Endod       Date:  2021-09-24

4.  Influence of Tip Diameter and Light Spectrum of Curing Units on the Properties of Bulk-Fill Resin Composites.

Authors:  Igor Oliveiros Cardoso; Alexandre Coelho Machado; Luísa de Oliveira Fernandes; Paulo Vinícius Soares; Luís Henrique Araújo Raposo
Journal:  Eur J Dent       Date:  2021-12-14

5.  Effects of aging and light-curing unit type on the volume and internal porosity of bulk-fill resin composite restoration.

Authors:  Afnan O Al-Zain; Elaf A Alboloshi; Walaa A Amir; Maryam A Alghilan; Eliseu A Münchow
Journal:  Saudi Dent J       Date:  2022-01-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.