Abstract
BACKGROUND: Few studies have reported on the utility of ultrathin bronchoscopes (UTBs) for performing radial probe endobronchial ultrasound (EBUS). Herein, we describe our experience with UTB and conventional bronchoscope (CB) for performing radial EBUS.Entities:
Keywords: Bronchoalveolar lavage; brush cytology; endobronchial ultrasound; pneumothorax; transbronchial lung biopsy; transbronchial needle aspiration
Year: 2019 PMID: 30829242 PMCID: PMC6410583 DOI: 10.4103/lungindia.lungindia_115_18
Source DB: PubMed Journal: Lung India ISSN: 0970-2113
Demographic profile and other parameters of the study population
| Parameter | Ultrathin scope ( | Conventional scope ( | |
|---|---|---|---|
| Age (years) | 55.3±14.6 | 55.1±14.9 | 0.991 |
| Male gender, | 25 (73.5) | 59 (67.8) | 0.662 |
| Size of lesion on CT (mm) | 16.4±9.1 | 24.7±14.6 | 0.006 |
| Location of lesion on CT, | |||
| Right upper lobe | 13 (38.2) | 28 (32.2) | 0.449 |
| Right middle lobe | 7 (20.6) | 11 (12.6) | |
| Right lower lobe | 3 (8.8) | 19 (21.8) | |
| Left upper lobe | 6 (17.6) | 17 (19.5) | |
| Lingula | 2 (5.9) | 2 (2.3) | |
| Left lower lobe | 3 (8.8) | 10 (11.5) | |
| CT findings | |||
| Consolidation | 16 (47.1%) | 27 (31.0%) | 0.10 |
| Solitary pulmonary nodule | 6 (17.6%) | 0 | 0.0001 |
| Multiple nodules | 11 (32.4%) | 29 (33.3%) | 0.92 |
| Location of probe in relation to the lesion, | |||
| Adjacent to the lesion | 3 (15.8) | 16 (32.7) | 0.164 |
| Within the lesion | 16 (84.2) | 33 (67.3) | |
| Bronchoscopy procedures, | |||
| Bronchial washings | 24 (70.6) | 82 (94.3) | <0.001 |
| Brush cytology | 26 (76.5) | 86 (98.9) | <0.001 |
| TBLB | 19 (55.9) | 63 (72.4) | 0.080 |
Chi-square test or Fisher’s exact test was used to analyze differences between categorical variables while the Mann–Whitney U-test was used for continuous variables. CT: Computed tomography, TBLB: Transbronchial lung biopsy, USG: Ultrasonography
Diagnostic yield of procedures performed using radial endobronchial ultrasound
| Parameter | Ultrathin scope ( | Conventional scope ( | |
|---|---|---|---|
| Bronchial washing | |||
| Nondiagnostic | 16 (66.7) | 61 (74.4) | 0.461 |
| Malignancy | 6 (25) | 13 (15.7) | |
| Tuberculosis | 2 (8.3) | 4 (4.8) | |
| Fungal pneumonia | 0 (0) | 4 (4.9) | |
| Brush cytology | |||
| Nondiagnostic | 21 (80.8) | 65 (75.6) | 0.639 |
| Malignancy | 4 (15.4) | 11 (12.8) | |
| Tuberculosis | 1 (3.8) | 6 (7) | |
| Fungal pneumonia | 0 (0) | 4 (4.7) | |
| TBLB | |||
| Nondiagnostic | 12 (63.2) | 35 (55.6) | 0.971 |
| Malignancy | 3 (15.8) | 12 (19) | |
| Tuberculosis | 2 (10.5) | 9 (14.3) | |
| Fungal pneumonia | 1 (5.3) | 4 (6.5) | |
| Lymphoma | 0 (0) | 1 (1.6) | |
| Sarcoidosis | 1 (5.3) | 2 (3.1) | |
| Final diagnosis | |||
| Malignancy | 14 (41.2) | 24 (27.6) | 0.369 |
| Lymphoma | 0 (0) | 1 (1.1) | |
| Tuberculosis | 2 (5.9) | 12 (13.8) | |
| Fungal pneumonia | 1 (2.9) | 6 (6.9) | |
| Sarcoidosis | 1 (2.9) | 2 (2.3) | |
| Foreign body | 1 (2.9) | 0 (0) | |
All values are represented as n (%). Chi-square test or Fisher’s exact test was used to analyze differences between categorical variables while the Mann–Whitney U-test was used for continuous variables. TBLB: Transbronchial lung biopsy
Outcome parameters between two bronchoscopes type
| Parameter | Ultrathin scope ( | Conventional scope ( | |
|---|---|---|---|
| Radial USG outcomes | |||
| Lesion visible on USG | 23 (67.6) | 55 (63.2) | <0.001 |
| Lesion not visible on USG | 3 (8.8) | 32 (36.8) | |
| Endobronchial lesion | 8 (23.5) | 0 (0) | |
| Duration of procedure in minutes, mean±SD | 20.1±8.7 | 22.6±9.2 | 0.018 |
| Overall yield | 19 (55.9) | 45 (51.7) | 0.680 |
| Complications | |||
| Hypoxia | 0 (0) | 2 (2.3) | 0.629 |
| Mild bleeding | 1 (2.9) | 1 (1.1) | |
| Pneumothorax | 1 (2.9) | 1 (1.1) | |
All values are represented as n (%) unless otherwise stated. Chi-square test or Fisher’s exact test was used to analyze differences between categorical variables, while the Mann–Whitney U-test was used for continuous variables. SD: Standard deviation, USG: Ultrasonography
Multivariate regression analysis of variables predicting successful yield of radial endobronchial ultrasound
| Adjusted OR (95% CI) | ||
|---|---|---|
| Type of scope (ultrathin vs. conventional) | 0.6 (0.2-2.2) | 0.45 |
| Size of lesion on CT thorax (mm) | 1.0 (0.9-1.1) | 0.86 |
| Upper lobe lesion on CT thorax | 2.2 (0.7-7.3) | 0.19 |
| Position of radial probe within the lesion (within vs. adjacent) | 2.6 (0.7-10.2) | 0.17 |
CI: Confidence interval, CT: Computed tomography, OR: Odds ratio