Clinical value of (dedicated) 3 Tesla and 7 Tesla MRI for cT1 glottic carcinoma: A feasibility study.

Sylvia L van Egmond1,2, Bernard M D Vonck2, Johanna J Bluemink3, Frank A Pameijer4, Jan Willem Dankbaar4, Inge Stegeman2,5, Marielle E P Philippens3, Cornelis A T van den Berg3, Luuk M Janssen6, Chris H Terhaard3.   

Abstract

OBJECTIVE: To assess the feasibility of the clinical use of 3 Tesla and 7 Tesla Magnetic Resonance Imaging for early (cT1) glottic carcinoma, including structural assessment of technical image quality and visibility of the tumor; and if feasible, to correlate MRI findings to routine diagnostics.
METHODS: Prospective feasibility study. Twenty patients with primary clinical T1 glottic carcinoma underwent both routine clinical staging and CT. In addition, a 3 T and 7 T MRI protocol, developed for small laryngeal lesions, was performed in a 4-point immobilization mask, using dedicated surface coils. Afterwards, routine endoscopic direct suspension laryngoscopy under general anaesthesia was performed.
RESULTS: Only 2 of 7 (29%) of 7 T MRI scans were rated as moderate to good technical image quality. After exclusion of three patients with only mild to moderate dysplasia at the time of MRI, 13 of 17 (76%) of 3 T MRIs were of adequate technical image quality. Tumor visualization was adequate in 8 of 13 (62%) of patients with invasive squamous cell carcinomas. With exclusion of the four MRIs with motion artefacts, the tumor and its boundaries could be adequately seen in 8 of 9 (89%) patients with squamous cell carcinoma versus only one in four (25%) of patients with carcinoma in situ lesions.
CONCLUSIONS: 7 Tesla MRI was considered not feasible. 3 Tesla MRI, with adequate patient selection, namely clinical exclusion of patients with a history of claustrophobia and inclusion of only histologically proven invasive squamous cell carcinoma, can be feasible. Especially with further improvement of MR image quality. LEVEL OF EVIDENCE: 2B, prospective diagnostic study.

Entities:  

Keywords:  3 Tesla; 7 Tesla; MRI; early glottic cancer; magnetic resonance imaging

Year:  2019        PMID: 30828625      PMCID: PMC6383319          DOI: 10.1002/lio2.246

Source DB:  PubMed          Journal:  Laryngoscope Investig Otolaryngol        ISSN: 2378-8038


INTRODUCTION

Early stage squamous cell carcinoma (SCC) of the glottic larynx, treated by radiotherapy or endoscopic laser surgery, is characterized by a low tumor volume and rare incidence of regional or distant metastatic spread.1, 2 Reported local control rates for both treatment modalities are equally high (71–95%), depending on invasion of the anterior commissure (AC).3 Therefore, the aim of treatment is to achieve local control with the best possible functional outcome. Topic of controversy remains which treatment modality results in tumor eradication with preservation of functions, especially when the cancer involves the AC, involves the vocal muscle, or extends laterally in the laryngeal ventricle. The main factor that prohibits the systematic evaluation and pooling of data is the uncertainty of tumor comparability, due to an absence of a standardized method that accurately measures tumor extent and depth. Both flexible laryngoscopy, videolaryngostroboscopy (VLS), and direct suspension laryngoscopy with an operating microscope can evaluate AC invasion, but no claims can be made about the extent of the tumor in depth, as well as submucosal intralaryngeal growth. Rates reported of correct staging of AC invasion vary beteen 40% to 72%.4, 5 Visualization of early glottic tumors by computed tomography (CT) imaging can also be difficult due to small tumor size and volume and the poor soft tissue detail. Invasion of the AC is more difficult to detect. Diagnostic value of CT varies widely in literature with a reported accuracy of 83% to 86%,6, 7 a sensitivity of 43% and a specificity of 83% when compared to histopathological results5. However, these number account for all T1–T4 laryngeal tumors. No subanalysis was done for cT1 tumors only. False‐positive results may be caused by difficulty to discriminate reactive inflammation and edema from tumor.8. A recent systematic review evaluating the diagnostic value of MRI for pre‐therapeutic staging of early glottic cancer, showed that MRI data of T1 and T2 glottic carcinoma is limited, heterogenous and based on sub analysis9. Both slice thickness and the use of neck receive coils varied and was suboptimal for small laryngeal lesions. 7 T ultra‐high field MRI has the potential to depict smaller lesions.10, 11 In the head and neck region, the inner ear and the parotid gland and duct have been visualized with varying results.12, 13 Up to now, no studies have been performed to assess the diagnostic value of 7 T for glottic carcinomas. The primary goal of this study was to assess the feasibility of the clinical use of both an dedicated 3 T MRI protocol for small laryngeal lesions, and 7 T MRI of clinical T1 glottic carcinoma; Both (technical) image quality and tumor visibility were evaluated. Secondary goals were comparison of MRI data to routine diagnostics, including suspension laryngoscopy and computed tomography. And, in case of laser excision, verification of MRI with histopathological data.

MATERIALS AND METHODS

Ethical Considerations

This prospective feasibility study was approved by the local ethics committee and written informed consent was obtained from all patients included in the study.

Methods

Following flexible laryngoscopy and videolaryngostroboscopy at the Out Patients Department, patients were staged by two independent head and neck surgeons according to the International TNM classification for malignant glottic tumors.14 All patients with clinically staged T1 glottic carcinomas, referred to a tertiary referral center (Utrecht University Medical Center) between December 2014 and March 2017, were considered for this study. All patients had a histologically proven carcinoma in situ or invasive SCC of the vocal cord(s). We excluded patients with a history of previous glottic surgery, patients with previous radiation therapy for a head and neck carcinoma, patients with recurrence of glottic disease, patients with legal incapability or insufficient command of the Dutch language, or patients who did not pass MRI screening criteria. Videolaryngoscopy was recorded and saved and glottal closure, regularity, mucosal wave and symmetry were scored using a 4‐point (0–3) grading scale by two ENT surgeons, specialized in head and neck oncology. Patients subsequently underwent CT, 3 Tesla MRI, and 7 Tesla MRI.

CT Imaging Protocol

All contrast enhanced CT studies were performed on a 128 detector‐row scanner (Philips iCT, Cleveland, OH, USA). For the acquisition, first an injection of 90 ml of nonionic contrast material (Ultravist (Iopromide) 300 mg/ml, Bayer Healthcare, Whippany, NJ), at a rate of 4 ml/s followed by a 40‐ml saline flush was given. Forty seconds after the first injection a second injection of 30 ml of contrast material and a 30‐ml saline flush at a rate of 4 ml/s was given. A bolus triggered scan was started simultaneous with the second bolus with a locator in the descending aorta at a Hounsfield Units (HU) threshold of 170 with no scan delay. Acquisition was performed in axial plane during quiet breathing. Scanning parameters were: 128 x 0.625 mm collimation, 120 kVp, 143 mAs, a rotation time of 0.4 s, pitch of 0.914, 250 mm Field of View, and 512 x 512 matrix. Also, 3‐mm axial, coronal and sagital sections were reconstructed. The axial images were reconstructed parallel to the plane of the true vocal cords.

MR Imaging Protocol

To minimize movements of the head and neck region, 3 Tesla and 7 Tesla MRI were both performed using a 4‐point immobilization mask for the head and neck region with individual head support, made prior to scanning at the Department of Radiotherapy. This mask is an adaption of the 5‐point immobilization mask used for head and neck radiotherapy, covering the region form the chin to the shoulders. Its use is standard practice for MRI performed for patients who are scheduled for radiotherapy of the head and neck region at our center.15 Both MRI exams were performed with administration of 0.1 ml/kg single‐dose intravenous contrast agent gadolinium (Gadovist) and 30‐cc saline flush. Scans were angulated along the C3–C4 or C4–C5 disk space (ie, parallel to the plane of the true vocal cords). The 3 T and 7 T protocol, optimized after a pilot study with healthy volunteers, consisted of planning scans, preparation scans to optimize the radiofrequency (RF) transmit field and for 7 T the magnetic field homogeneity in the region of interest and T1 and T2 weighted scans optimized in volunteers. The 3 T and 7 T sequences (following the scans for planning and optimization of the RF and magnetic field homogeneity) are listed in Table 1. For the 7 Tesla MRI exam, a dedicated transmit neck coil was used in the form of a dielectric waveguide.16 The weight of the dielectric bag filled with deuterated water was not sensed by the patient due to the stiffness of the mask material. Local receive coils were used to maximize the signal intensity obtained from the glottic area.17 The receive coils were placed on the mask, covering the region of the larynx. Prior to the study, a risk analysis was performed in cooperation with the Department of Medical Technology.
Table 1

3 T and 7 T Scan Protocol Sequences.

3 Tesla MRI7 Tesla MRI
Subsequent scan protocols1234123
Sequence3D T2W SE3D T2W SEMS T1W SE + and ‐ Gado3D T1W SE GadoMS T2W SEMS T1W SE + and – GadoMS T1W SE + Gado
OrientationTransverseTransverseTransverseSagittalTransverseTransverseSagittal
TE/TR (ms)157/2225235/210020/50025/40068/350037/80037/700
TSE factor801453201188
Flip angle (degrees)9090909090100100
Refocusing angle (degrees)40‐12025‐120 * 40809090
Field of view (mm3)160x160x34160x160x39160x160x44160x160x44140x140x35140x140x35140x140x50
No transverse slices34392088321665 ‐ C
Acquisition pixel size (m3)0.5x0.5x11x1x10.6x0.61x1x0.50.5x0.5x10.4x0.4x20.7x0.7x0.7
Imaging matrix320x318160x158272x270160x160280x275348x346200x200
NSA2222222
Sense factor1.91.922122
Scan duration (minutes)624362.54

3 T = 3 Tesla; 7 T = 7 Tesla; C = coronal; MS = multislice; NSA = number of signal averages; TE/TR = echo time/repetition time; TSE = turbo spin echo.

No refocusing control

3 T and 7 T Scan Protocol Sequences. 3 T = 3 Tesla; 7 T = 7 Tesla; C = coronal; MS = multislice; NSA = number of signal averages; TE/TR = echo time/repetition time; TSE = turbo spin echo. No refocusing control

Diagnostic Evaluation

CT, 3 T MRI, and 7 T MRI scans were individually assessed for technical image quality and visibility of the tumor, scored with a 4‐point Likert scale, by three independent researchers (two experienced head and neck radiologists and one experienced head and neck radiation oncologist), blinded for patient data. In a second joint meeting, in case of visible tumor, the tumor was delineated and location, tumor size (in mm), tumor volume and depth were assessed. Following imaging, as part of standard clinical procedure, direct suspension laryngoscopy with a rigid endoscope was performed under general anaesthesia. Extent of the tumor was evaluated by visual examination through an operating microsope with a standard eye‐to‐object‐distance of 400 mm and by palpation with microforceps. Images were recorded using a measuring tape (Fig. 1). Lesions were staged according to the international TNM (for Tumor, regional lymph Nodes, distant Metastasis) Classification of Malignant Tumors in T1a, T1b, or T2. Furthermore, appearance (exophytic growth pattern), length (in mm), expected depth invasion of the tumor (after palpation) and possible paraglottic or subglottic extension as well as involvement of the anterior comissure were reported. The decision for direct endoscopic excision with CO2 laser or radiotherapy was made during direct suspension laryngoscopy by the head and neck surgeon, independent of MRI (and CT) results. In general, lesions with involvement of the anterior commissure, lesions with paraglottic extension or patients in which tumor visualization was inadequate, were not considered suitable for laser resection. When endoscopic laser resection was performed, pathologic specimens were fixed with needles onto a standardized full‐scale image of the larynx, to enable optimal structured histopathological evaluation of tumor margins. When considered unsuitable for endoscopic excision, Intensity Modulated Radiotherapy (IMRT) was commenced with a daily fraction of 2.4 Gy, 5 days/week, for 25 days, resulting in a total irradiation fraction dose of 60 Gy in 5 weeks.
Figure 1

Tumor was measured with a measuring tape during direct laryngoscopy

Tumor was measured with a measuring tape during direct laryngoscopy

Statistics

Since direct laryngoscopy cannot be used as a reference standard and a pathology specimen was not available in each patient, no diagnostic accuracy measures could be calculated. Results will be descriptive.

RESULTS

Number of Scans

A flow diagram of all included patients and performed scans is shown in Figure 2. After 10 patients, we evaluated the scan quality of both 3 T and 7 T MRI and considered the 7 T MRI not feasible. Technical image quality was moderate to good in only one or two out of seven patients. As a result, the tumor was not adequately visible in 72% to 86% of the patients (Tables 2 and 3). With these results, the clinical burden to undergo 7 T MRI for the remaining 10 patients to be included in the study could therefore not be justified. Therefore, the last 10 patients did not undergo 7 T scanning. In retrospect, two carcinoma in situ lesions and one SCC were excised during an initial biopsy at the referral hospital. These three patients only had mild to moderate dysplasia during time of scanning and were excluded for evaluation. Therefore, 17 3 T MRIs and seven 7 T MRIs were included in our analysis.
Figure 2

Flow diagram of included patients, number of scans, scan quality, and tumor visibility

Table 2

Assessment of Image Quality by Three Independent Observers, Blinded for Clinical Data.

CT3 Tesla MRI7 Tesla MRI
Image qualityR1R2R3R1R2R3R1R2R3
Bad 1 (6%)003 (18%)1(6%)1 (6%)3 (43%)1 (14%)4 (57%)
Poor 001 (6%)1 (6%)3 (18%)3 (18%)2 (29%)4 (57%)2 (29%)
Moderate 02 (12%)10 (59%)5 (29%)8 (47%)3 (18%)1 (14%)1 (14%)0
Good 16 (94%)15 (88%)6 (35%)8 (47%)5 (29%)10 (59%)1 (14%)1 (14%)1 (14%)
Total 171717171717777

CT = computed tomography; MRI = magnetic resonance imaging; R1 = radiologist 1 (FP); R2 = radiologist 2 (JD); R3 = radiation oncologist (CT).

Table 3

Assessment of Image Quality by Three Independent Observers, Blinded for Clinical Data.

CT3 Tesla MRI7 Tesla MRI
Tumor visibilityR1R2R3R1R2R3R1R2R3
Tumor not visible 4 (24%)4 (24%)7 (41%)5 (29%)5 (29%)5 (29%)5 (72%)2 (29%)5 (72%)
Tumor visible, but not the tumor boundaries 3 (18%)3 (18%)5 (29%)3 (18%)4 (24%)4 (24%)1 (14%)3 (43%)1 (14%)
Tumor boundaries visible, but not clear 6 (35%)2 (12%)2 (12%)6 (35%)5 (29%)3 (18%)01 (14%)0
Both tumor and boundaries clearly visible 4 (24%)8 (47%)3 (18%)3 (18%)3 (18%)5 (29%)1 (14%)1 (14%)1 (14%)
Total 171717171717777

CT = computed tomography; MRI = magnetic resonance imaging; R1 = radiologist 1 (FP); R2 = radiologist 2 (JD); R3 = radiation oncologist (CT).

Flow diagram of included patients, number of scans, scan quality, and tumor visibility Assessment of Image Quality by Three Independent Observers, Blinded for Clinical Data. CT = computed tomography; MRI = magnetic resonance imaging; R1 = radiologist 1 (FP); R2 = radiologist 2 (JD); R3 = radiation oncologist (CT). Assessment of Image Quality by Three Independent Observers, Blinded for Clinical Data. CT = computed tomography; MRI = magnetic resonance imaging; R1 = radiologist 1 (FP); R2 = radiologist 2 (JD); R3 = radiation oncologist (CT).

Inter Observer Agreement

Assessment of MRI data by two radiologists and one radiation oncologist appeared to be heterogeneous. Intra‐class correlation coefficient for image quality and tumor visibility for CT was 0.76 and 0.80. Intra‐class correlation coefficient for image quality and tumor visibility for 3 Tesla MRI was 0.86 and 0.91. When classified as “acceptable” (moderate or good) or “unacceptable” (bad or poor), a higher degree of inter observer agreement was reached.

Histopathological Results

Five out of 17 tumors were considered suitable for endoscopic laser resection. Four tumors were excised, one was evaporated. Tumor diameters could only be assessed in the four excision histopathological specimens. A total of 12 patients received radiotherapy.

Technical Image Quality of CT and MRI

Data following exclusion of three patients with only mild/moderate dysplasia are presented in Table 2. Ninety‐four percent to 100%, 76%, and 14% to 29% of CT, 3 T MRI, and 7 T MRI scans were rated as moderate to good technical image quality, respectively.

Tumor Visibility of CT and MRI

Data are presented in Table 3. In 30% to 59%, 47% to 53% and 14% to 29% of CT, 3 T MRI, and 7 T MRI, respectively, the tumor and boundaries could be identified. One of the patients with the best quality on 3 T MRI and 7 T MRI is shown in Figure 3. CT and a clinical image during suspension laryngoscopy shows a comparable lesion of the left anterior vocal cord. This tumor was clinically scored as a T1b tumor, with involvement of the anterior commissure. In this case, radiotherapy was performed. Findings could therefore not be related to histopathological results.
Figure 3

Axial images at the mid glottic level showing a lesion on the anterior half of the left true vocal cord with extension to the anterior commissure (arrows). On CT the lesion is isointense to the vocal cord and therefore difficult to see. 3 T MRI (a), 7 T MRI (b), contrast enhanced CT (c) and clinical photo during suspension laryngoscopy (d) of the same study patient.

3 T = 3 Tesla; 7 T = 7 Tesla; CT = computed tomography; MRI = magnetic resonance imaging.

Axial images at the mid glottic level showing a lesion on the anterior half of the left true vocal cord with extension to the anterior commissure (arrows). On CT the lesion is isointense to the vocal cord and therefore difficult to see. 3 T MRI (a), 7 T MRI (b), contrast enhanced CT (c) and clinical photo during suspension laryngoscopy (d) of the same study patient. 3 T = 3 Tesla; 7 T = 7 Tesla; CT = computed tomography; MRI = magnetic resonance imaging.

Relation Between Image Quality, Tumor Visibility, and Tumor Dimensions in Subgroups of Patients on 3 T MRI and CT

Patients with carcinoma in situ lesions

Four out of 17 patients had carcinoma in situ lesions. Although all 3 T MRI scans were of moderate to good (“acceptable”) image quality, in only one patient (25%), tumor boundaries could be visualized, although not clear. Concerning CT, the technical quality of all scans were good, but the tumor seen on MRI was not visible on CT. In another patient with Cis, visibility of tumor boundaries was slightly better on CT than on MR, leading to an “acceptable” tumor visibility on CT.

Patients with invasive squamous cell carcinoma

Thirteen out of 17 patients had an invasive scc on histopathological report. Eight patients had a cT1a lesion, four patients had a cT1b tumor, and one patient had an SCC staged as T2. Four MRI scans were of bad image quality due to severe motion artefacts. One of these patients suffered from claustrophobia. The remaining nine scans were of moderate to good (“acceptable”) image quality. In eight patients, the tumor and its boundaries could be adequately visualized. On CT, image quality was good in all patients, except for the one with claustrophobia. Tumor visibility was adequate in nine patients.

Comparison of tumor dimensions on CT, 3 T, 7 T and direct laryngoscopy

In eight patients with invasive SCC, the tumor and its boundaries could be adequately visualized on 3 T MRI and tumor dimensions could be measured (Table 4). In a consensus meeting, tumors were delineated on 3 T MRI and CT. Volumes variated from 0.036 to 2.38 cc and were slightly larger on MRI than on CT for all tumors. Tumor volume was not related to visibility of the tumor. In only one of these eight patients, laser excision was performed. Correlation of tumor diameters to pathological specimens was therefore not contributive.
Table 4

Tumor Dimensions Measured on Imaging and Direct Laryngoscopy.

Anterior‐posterior dimeter (mm)Lateral diameter (mm)
In mmCT3 T MRILaryngoscopyCT3 T MRILaryngoscopy
Pt 28.614.2173.78.95
Pt 38.69104.54.25
Pt 42027158.51010
Pt 95117553
Pt 121010123.43.43.5
Pt 1411.826.0155.09.08
Pt 165.26.473.83.55
Pt 2045.891.42.31

3 T = 3 Tesla; CT = computed tomography; MRI = magnetic resonance imaging.

Tumor Dimensions Measured on Imaging and Direct Laryngoscopy. 3 T = 3 Tesla; CT = computed tomography; MRI = magnetic resonance imaging.

Assessment of anatomical/glottic subsites

Presence or absence of tumor involvement of anatomical subsites was scored on both imaging and during suspension laryngoscopy. For the anterior commissure, concordance between CT, 3 T MRI and laryngoscopy was low for the whole study population (including both superficial carcinoma in situ lesions and invasive bulky disease). For the eight patients with invasive SCC with adequate tumor visualization, shown in Table 4, concordance was higher. For these patients, a consensus was reached by the observers, blinded for clinical data. Two out of eight CTs and 3 T MRIs (of the same patients) were false negative for tumor invasion of the AC, which was visible during direct laryngoscopy. This was due to superficial tumor extension of a more bulky midcord lesion into the AC. Four out of eight CTs and MRIs (of the same patients) were true positive. Two CT and MRIs were true negative. The vocal muscle was scored as involved in five patients on both CT and MRI, it was scored as not involved in two patients and was scored as involved on MRI but not involved on CT in the same patient. No verification with histopathology could be done. Other subsites were not, or rarely, involved.

DISCUSSION

Feasibility: Primary Endpoint

Twenty percent of patients refused 7 T scanning because of claustrophobia. In contrast to our expectations, based on pilot data with detailed laryngeal visualization in healthy subjects, in only one out of seven 7 T scans, the tumor and its boundaries could be clearly seen. We therefore considered 7 T MRI not feasible. Although the vast majority of 3 T scans was of moderate to good technical quality, this only resulted in adequate tumor visibility on 3 T MRI in nine out of the total group of 17 patients (53%) versus 10 out of 17 patients (59%) on CT. The tumor visibility was adequate in nine out of 13 (69%) 3 T MRIs with moderate to good image quality, versus 10 out of 16 patients (63%) with good CT image quality. This was probably due to inadequate visibility of superficial (carcinoma in situ) lesions. When subgroup analysis was done, tumor visualization was adequate in 62% (or even 89% after exclusion of four MRIs with bad technical quality) with SSC versus only 25% of patients with carcinoma in situ lesions. On CT, tumor visibility was adequate in 25% of Cis lesions and in 69% of SCC lesions (or 75% after exclusion of one CT with bad technical quality).

Study Limitations

This study has several limitations. First, histopathology could not be used as a reference standard to validate the diagnostic value of the scans. Random biopsy of anatomic subsites, only for study purposes, was considered unethical and harmful for the study patients. Second, data were very heterogeneous. The low concordance between laryngoscopy and CT does not allow for a combined reference standard to validate MRI findings.18 Third, the moderate intraclass correlation coefficient on a 4‐point scale between two experienced head and neck radiologists and the radiation oncologist reflects the difficulty of image interpretation. When scored as “adequate/acceptable” or “inadequate/unacceptable,” agreement was high. The intra‐observer and inter‐observer reproducibility in the delineation of head and neck tumors have a wide range, with a reported percentage measurement error for primary tumors of 0% to 53%.19, 20, 21 Between observers, the size of the largest gross tumor volume (GTV) can be more than eight times the size of the smallest volume. Last limitation of this study is that the majority of the glottic tumors were clinically superficial lesions. Four out of 17 were carcinoma in situ lesions. Another three lesions were biopsied in the referral hospital as carcinoma in situ lesions, but were invasive carcinomas after second biopsy at our tertiary hospital. Lesions that were cT2 on first visit were excluded from this study. This could have influenced the results. In larger glottic carcinomas MRI has been shown to be of additional value in delineating tumor boundaries and differentiating tumor from edema or fibrosis.22, 23

Comparison of MRI Data to Routine Diagnostics; Secondary Endpoint

Anterior‐posterior diameters measured during laryngoscopy compared to MRI and CT were comparable. However, clinical measurements can be difficult due to both invasive and exophytic tumor components with varying shapes of the vocal cord. The superficial component results in larger tumor diameters measured during laryngoscopy, which was not detected on imaging. Because of the lack of histopathological data, unfortunately, no relation between tumor visibility and depth invasion of the tumor could be addressed.

Challenges of 3 T and 7 T MRI Scanning for Small Laryngeal Lesions

For optimal detection of small laryngeal lesions and to avoid partial volume effect of scanning small lesions, we used a slice thickness of 0.5 to 1 mm. However, small slice thickness results in a diminished Signal to Noise Ratio (SNR). To optimize the signal in the area of interest, we used neck receive coils close to the larynx. At the surface the subcutaneous fat is located. This fat can cause large artefacts when patients move (for example due to breathing). These artefacts are more pronounced in the 7 T scans than in the 3 T MRI scans, due to the high sensitivity of the small surface coils in the 7 T setup close to the surface. To minimize motion artefacts, all patients were scanned in a 4‐point immobilization mask. A benefit of the stiff mask material was that the transmit neck coil used on 7 T, consisting of a dielectric bag filled with deuterated water, which can be quite heavy on the neck, was not sensed by the patient. Although the immobilization mask minimizes external movements of the head and neck, internal motion as a result of breathing and swallowing cannot be prevented. To diminish the susceptibility for motion artefacts, we performed four dynamic scans with one‐fourth scan duration each, but this was not beneficial. A solution for this problem appears to be “in‐vivo respiratory triggering,” a method in which data acquisition is only done during expiration, since internal breathing motion will reduce and swallowing cannot occur.24 At 7 T, the SNR increases linearly with field strength. However, the image homogeneity is generally less due to the smaller RF wavelength causing RF interferences.25, 26 For this reason, a body coil is very hard to manufacture at such high field strength. We used a dedicated neck transmit coil with 2 antenna's to homogenize the transmit field based on the principle of a dielectric waveguide16. Another challenge, especially at 7 T, is the energy absorbed by the human body (Specific Absorption Ratio of SAR in W/kg), which scales quadratically with the field strength. The higher SAR at 7 T has to be compensated by fewer RF pulses per time unit, resulting in a longer acquisition time. Also, the water‐fat shift is increased at 7 T compared to 3 T. This can be compensated for by a higher read‐out bandwidth, this however reduces the SNR. Further improvement of scan quality at 3 T might be accomplished by a neck receive coil consisting of multiple elements. At 7 T, we already used receive coils with multiple elements. At 3 T, we used a receive coil with two elements. Adding receive coils can lead to a higher acceleration factor with a shorter scan duration. This will especially be of benefit when data acquisition is limited to the expiration phase of breathing. With multiple elements, the signal remains high enough for optimal tumor visibility.

CONCLUSION

We conducted a feasibility study for 3 T and 7 T MRI for clinically staged Tis, T1a and T1b glottic carcinoma. 7 T MRI was considered not feasible. 3 T MRI, with clinical exclusion of patients with a history of claustrophobia, and with histologically proven invasive squamous cell carcinoma, can be feasible. Especially with further improvement of MRI quality. However, its superiority to CT could not be confirmed in this study. Superficial tumor extension remains difficult to detect on imaging.
  25 in total

1.  7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images.

Authors:  J T Vaughan; M Garwood; C M Collins; W Liu; L DelaBarre; G Adriany; P Andersen; H Merkle; R Goebel; M B Smith; K Ugurbil
Journal:  Magn Reson Med       Date:  2001-07       Impact factor: 4.668

Review 2.  The impact of gross tumor volume (GTV) and clinical target volume (CTV) definition on the total accuracy in radiotherapy theoretical aspects and practical experiences.

Authors:  Elisabeth Weiss; Clemens F Hess
Journal:  Strahlenther Onkol       Date:  2003-01       Impact factor: 3.621

3.  Subjective acceptance of 7 Tesla MRI for human imaging.

Authors:  Jens M Theysohn; Stefan Maderwald; Oliver Kraff; Christoph Moenninghoff; Mark E Ladd; Susanne C Ladd
Journal:  MAGMA       Date:  2007-12-07       Impact factor: 2.310

4.  Time-of-flight magnetic resonance angiography at 7 Tesla.

Authors:  Johannes T Heverhagen; Eric Bourekas; Steffen Sammet; Michael V Knopp; Petra Schmalbrock
Journal:  Invest Radiol       Date:  2008-08       Impact factor: 6.016

5.  Anterior vocal commissure invasion in laryngeal carcinoma diagnosis.

Authors:  M M Barbosa; V J F Araújo; E Boasquevisque; R Carvalho; S Romano; R A Lima; F L Dias; S K Salviano
Journal:  Laryngoscope       Date:  2005-04       Impact factor: 3.325

6.  Pretherapeutic staging of recurrent laryngeal carcinoma: clinical findings and imaging studies compared with histopathology.

Authors:  Peter Zbären; Andreas Christe; Marco D Caversaccio; Edouard Stauffer; Harriet C Thoeny
Journal:  Otolaryngol Head Neck Surg       Date:  2007-09       Impact factor: 3.497

7.  First clinical study on ultra-high-field MR imaging in patients with multiple sclerosis: comparison of 1.5T and 7T.

Authors:  K Kollia; S Maderwald; N Putzki; M Schlamann; J M Theysohn; O Kraff; M E Ladd; M Forsting; I Wanke
Journal:  AJNR Am J Neuroradiol       Date:  2009-01-15       Impact factor: 3.825

8.  Intraobserver variability in the MR determination of tumor volume in squamous cell carcinoma of the pharynx.

Authors:  Andrew R Gordon; Laurie A Loevner; Amita Shukla-Dave; Regina O Redfern; Adina I Sonners; Alex M Kilger; Mark A Elliott; Mitchell Machtay; Randal S Weber; Jerry D Glickson; David I Rosenthal
Journal:  AJNR Am J Neuroradiol       Date:  2004 Jun-Jul       Impact factor: 3.825

9.  Magnetic resonance imaging protocol optimization for delineation of gross tumor volume in hypopharyngeal and laryngeal tumors.

Authors:  Gerda M Verduijn; Lambertus W Bartels; Cornelis P J Raaijmakers; Chris H J Terhaard; Frank A Pameijer; Cornelis A T van den Berg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-06-01       Impact factor: 7.038

Review 10.  Imaging of the larynx and hypopharynx.

Authors:  Minerva Becker; Karim Burkhardt; Pavel Dulguerov; Abdelkarim Allal
Journal:  Eur J Radiol       Date:  2008-05-20       Impact factor: 3.528

View more